The impact of blood pressure cuff location on the accuracy of noninvasive blood pressure measurements in obese patients: an observational study

Authors: Nicholas Anast, MD, Megan Olejniczak, MD, Jerry Ingrande, MD, John Brock-Utne, MD, PhD

Published in: Canadian Journal of Anesthesia/Journal canadien d'anesthésie | Issue 3/2016

Login to get access

Abstract

Purpose

Obesity presents many challenges to the anesthesiologist, including poorly fitting blood pressure (BP) cuffs due to the conical shape of the upper arm. The aim of this study was to determine the accuracy of noninvasive BP readings, obtained from a noninvasive BP cuff using various cuff locations and wrapping techniques, compared with invasive intra-arterial BP readings.

Methods

Thirty American Society of Anesthesiologists physical status I-III obese (body mass index > 30 kg·m−2) individuals undergoing non-cardiac surgery were enrolled in this observational study. Serial oscillometric noninvasive BP (NIBP) measurements were taken in the patients’ forearm and upper arm with two different wrapping formations (one following the contour of the upper arm, the other keeping cuff edges parallel). These NIBP measurements were compared with invasive arterial blood pressure (ABP) measurements taken from the ipsilateral radial artery. The precision and bias of the NIBP and ABP measurements were determined using Bland-Altman analysis. Analysis of variance and Welch’s t test were used to determine between-group differences in bias.

Results

There was poor agreement between the ABP measurements and all types of NIBP measurements. Each of our study participants had a least one NIBP parameter (mean arterial pressure, systolic BP, or diastolic BP) that was > 10 mmHg different than the corresponding ABP parameter. Upper arm BP measurements showed a statistically insignificant trend toward underestimating ABP. For all cuff positions and wrapping techniques, systolic BP offered the best agreement between NIBP and ABP measurements.

Conclusions

All the forms of NIBP cuff orientation studied had unacceptable precision and bias compared with invasive ABP measurements. When patient and/or surgical conditions necessitate accurate BP monitoring, direct arterial measurement should be considered over NIBP measurements in obese patients.
Literature
1.
go back to reference Ogden CL, Carroll MD, Kit BK, Flegal KM. Prevalence of childhood and adult obesity in the United States, 2011-2012. JAMA 2014; 311: 806-14.CrossRefPubMed Ogden CL, Carroll MD, Kit BK, Flegal KM. Prevalence of childhood and adult obesity in the United States, 2011-2012. JAMA 2014; 311: 806-14.CrossRefPubMed
2.
go back to reference American Society of Anesthesiologists. Standards for Basic Anesthetic Monitoring. Standards and Practice Parameters. 2011. Available from URL: www.asahq.org (accessed August 2015). American Society of Anesthesiologists. Standards for Basic Anesthetic Monitoring. Standards and Practice Parameters. 2011. Available from URL: www.​asahq.​org (accessed August 2015).
3.
go back to reference Merchant R, Chartrand D, Dain S, et al. Guidelines to the practice of anesthesia—revised edition 2015. Can J Anesth 2015; 62: 54-67.CrossRefPubMed Merchant R, Chartrand D, Dain S, et al. Guidelines to the practice of anesthesia—revised edition 2015. Can J Anesth 2015; 62: 54-67.CrossRefPubMed
4.
go back to reference White WB, Berson AS, Robbins C, et al. National standard for measurement of resting and ambulatory blood pressures with automated sphygmomanometers. Hypertension 1993; 21: 504-9.CrossRefPubMed White WB, Berson AS, Robbins C, et al. National standard for measurement of resting and ambulatory blood pressures with automated sphygmomanometers. Hypertension 1993; 21: 504-9.CrossRefPubMed
5.
go back to reference Wan Y, Heneghan C, Stevens R, et al. Determining which automatic digital blood pressure device performs adequately: a systematic review. J Hum Hypertens 2010; 24: 431-8.PubMedCentralCrossRefPubMed Wan Y, Heneghan C, Stevens R, et al. Determining which automatic digital blood pressure device performs adequately: a systematic review. J Hum Hypertens 2010; 24: 431-8.PubMedCentralCrossRefPubMed
7.
go back to reference Umana E, Ahmed W, Fraley MA, Alpert MA. Comparison of oscillometric and intraarterial systolid and diastolic blood pressures in lean, overweight and obese patients. Angiology 2006; 57: 41-5.CrossRefPubMed Umana E, Ahmed W, Fraley MA, Alpert MA. Comparison of oscillometric and intraarterial systolid and diastolic blood pressures in lean, overweight and obese patients. Angiology 2006; 57: 41-5.CrossRefPubMed
8.
go back to reference Durie M, Beckmann U, Gillies DM. Incidents relating to arterial cannulation as identified in 7,525 reports submitted to the Australian incident monitoring study (AIMS-ICU). Anaesth Intensive Care 2002; 30: 60-5.PubMed Durie M, Beckmann U, Gillies DM. Incidents relating to arterial cannulation as identified in 7,525 reports submitted to the Australian incident monitoring study (AIMS-ICU). Anaesth Intensive Care 2002; 30: 60-5.PubMed
9.
go back to reference Singer AJ, Kahn SR, Thode HC Jr, Hollander JE. Comparison of forearm and upper arm blood pressures. Prehosp Emerg Care 1999; 3: 123-6.CrossRefPubMed Singer AJ, Kahn SR, Thode HC Jr, Hollander JE. Comparison of forearm and upper arm blood pressures. Prehosp Emerg Care 1999; 3: 123-6.CrossRefPubMed
10.
go back to reference Pierin AMG, Alavarce DC, Gusmao JL, Halpern A, Mion D Jr. Blood pressure measurement in obese patients: comparison between upper arm and forearm measurements. Blood Press Monit 2004; 9: 101-5.CrossRefPubMed Pierin AMG, Alavarce DC, Gusmao JL, Halpern A, Mion D Jr. Blood pressure measurement in obese patients: comparison between upper arm and forearm measurements. Blood Press Monit 2004; 9: 101-5.CrossRefPubMed
11.
go back to reference Schell K, Bradley E, Bucher L, et al. Clinical comparison of automatic, noninvasive measurements of blood pressure in the forearm and upper arm. Am J Crit Care 2005; 14: 232-41.PubMed Schell K, Bradley E, Bucher L, et al. Clinical comparison of automatic, noninvasive measurements of blood pressure in the forearm and upper arm. Am J Crit Care 2005; 14: 232-41.PubMed
12.
go back to reference Fortune M, Jeselnik K, Johnson S, et al. A comparison of forearm and upper arm blood pressure measurements in a sample of healthy young adults. J Undergrad Nurs Scholarsh 2009; 11: 1-9. Fortune M, Jeselnik K, Johnson S, et al. A comparison of forearm and upper arm blood pressure measurements in a sample of healthy young adults. J Undergrad Nurs Scholarsh 2009; 11: 1-9.
13.
go back to reference Domiano KL, Hinck SM, Savinske DL, Hope KL. Comparison of upper arm and forearm blood pressure. Clin Nurs Res 2008; 17: 241-50.CrossRefPubMed Domiano KL, Hinck SM, Savinske DL, Hope KL. Comparison of upper arm and forearm blood pressure. Clin Nurs Res 2008; 17: 241-50.CrossRefPubMed
14.
go back to reference Pickering TG, Hall JE, Appel LJ, et al. Recommendations for blood pressure measurement in humans and experimental animals. Part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Hypertension 2005; 45: 142-61. Pickering TG, Hall JE, Appel LJ, et al. Recommendations for blood pressure measurement in humans and experimental animals. Part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Hypertension 2005; 45: 142-61.
15.
go back to reference American National Standard. Manual. Arlington, VA: Electronic or Automated Sphygmomanometers. Association for the Advancement of Medical Instrumentation; 1992 . American National Standard. Manual. Arlington, VA: Electronic or Automated Sphygmomanometers. Association for the Advancement of Medical Instrumentation; 1992 .
16.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 327: 307-10.CrossRef Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 327: 307-10.CrossRef
17.
go back to reference Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res 1999; 8: 135-60.CrossRefPubMed Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res 1999; 8: 135-60.CrossRefPubMed
18.
go back to reference R Development Core Team. A Language and Environment for Statistical Computing. The R Project for Statistical Computing. Vienna, Austria: R Foundation for Statistical Computing - ISBN 3-900051-07-0 - 2011. Available from URL: http://www.R-project.org (accessed August 2015). R Development Core Team. A Language and Environment for Statistical Computing. The R Project for Statistical Computing. Vienna, Austria: R Foundation for Statistical Computing - ISBN 3-900051-07-0 - 2011. Available from URL: http://​www.​R-project.​org (accessed August 2015).
19.
20.
go back to reference Leblanc ME, Croteau S, Ferland A, et al. Blood pressure assessment in severe obesity: validation of a forearm approach. Obesity 2013; 21: E533-41.CrossRefPubMed Leblanc ME, Croteau S, Ferland A, et al. Blood pressure assessment in severe obesity: validation of a forearm approach. Obesity 2013; 21: E533-41.CrossRefPubMed
21.
go back to reference Hersh L, Sesing JC, Luczyk WJ, Friedman BA, Zhou S, Batchelder PB. Validation of a conical cuff on the forearm for estimating radial artery blood pressure. Blood Press Monit 2014; 19: 38-45.CrossRefPubMed Hersh L, Sesing JC, Luczyk WJ, Friedman BA, Zhou S, Batchelder PB. Validation of a conical cuff on the forearm for estimating radial artery blood pressure. Blood Press Monit 2014; 19: 38-45.CrossRefPubMed
22.
go back to reference Mauck GW, Smith CR, Geddes LA, Bourland JD. The meaning of the point of maximum oscillations in cuff pressure in the indirect measurement of blood pressure—part ii. J Biomech Eng 1980; 102: 28-33.CrossRefPubMed Mauck GW, Smith CR, Geddes LA, Bourland JD. The meaning of the point of maximum oscillations in cuff pressure in the indirect measurement of blood pressure—part ii. J Biomech Eng 1980; 102: 28-33.CrossRefPubMed
23.
go back to reference Babbs CF. Oscillometric measurement of systolic and diastolic blood pressure validated in a physiologic mathematical model. Biomed Eng Online 2012; 11: 56.PubMedCentralCrossRefPubMed Babbs CF. Oscillometric measurement of systolic and diastolic blood pressure validated in a physiologic mathematical model. Biomed Eng Online 2012; 11: 56.PubMedCentralCrossRefPubMed
Metadata
Title
The impact of blood pressure cuff location on the accuracy of noninvasive blood pressure measurements in obese patients: an observational study
Authors
Nicholas Anast, MD
Megan Olejniczak, MD
Jerry Ingrande, MD
John Brock-Utne, MD, PhD
Publication date
01-03-2016
Publisher
Springer US
Published in
Canadian Journal of Anesthesia/Journal canadien d'anesthésie / Issue 3/2016
Print ISSN: 0832-610X
Electronic ISSN: 1496-8975
DOI
https://doi.org/10.1007/s12630-015-0509-6

Other articles of this Issue 3/2016

Canadian Journal of Anesthesia/Journal canadien d'anesthésie 3/2016 Go to the issue