Published in: Radiation Oncology 1/2014

Open Access 01-12-2014 | Research

The effects of computed tomography image characteristics and knot spacing on the spatial accuracy of B-spline deformable image registration in the head and neck geometry

Authors: Charlotte L Brouwer, Roel GJ Kierkels, Aart A van ’t Veld, Nanna M Sijtsema, Harm Meertens

Published in: Radiation Oncology | Issue 1/2014

Login to get access

Abstract

Objectives

To explore the effects of computed tomography (CT) image characteristics and B-spline knot spacing (BKS) on the spatial accuracy of a B-spline deformable image registration (DIR) in the head-and-neck geometry.

Methods

The effect of image feature content, image contrast, noise, and BKS on the spatial accuracy of a B-spline DIR was studied. Phantom images were created with varying feature content and varying contrast-to-noise ratio (CNR), and deformed using a known smooth B-spline deformation. Subsequently, the deformed images were repeatedly registered with the original images using different BKSs. The quality of the DIR was expressed as the mean residual displacement (MRD) between the known imposed deformation and the result of the B-spline DIR.
Finally, for three patients, head-and-neck planning CT scans were deformed with a realistic deformation field derived from a rescan CT of the same patient, resulting in a simulated deformed image and an a-priori known deformation field. Hence, a B-spline DIR was performed between the simulated image and the planning CT at different BKSs. Similar to the phantom cases, the DIR accuracy was evaluated by means of MRD.

Results

In total, 162 phantom registrations were performed with varying CNR and BKSs. MRD-values < 1.0 mm were observed with a BKS between 10–20 mm for image contrast ≥ ± 250 HU and noise < ± 200 HU. Decreasing the image feature content resulted in increased MRD-values at all BKSs. Using BKS = 15 mm for the three clinical cases resulted in an average MRD < 1.0 mm.

Conclusions

For synthetically generated phantoms and three real CT cases the highest DIR accuracy was obtained for a BKS between 10–20 mm. The accuracy decreased with decreasing image feature content, decreasing image contrast, and higher noise levels. Our results indicate that DIR accuracy in clinical CT images (typical noise levels < ± 100 HU) will not be effected by the amount of image noise.
Appendix
Available only for authorised users
Literature
1.
go back to reference Schwartz DL: Current Progress in Adaptive Radiation Therapy for Head and Neck Cancer. Curr Oncol Rep 2012, 14: 139-147.CrossRefPubMed Schwartz DL: Current Progress in Adaptive Radiation Therapy for Head and Neck Cancer. Curr Oncol Rep 2012, 14: 139-147.CrossRefPubMed
2.
go back to reference Castadot P, Lee JA, Geets X, Grégoire V: Adaptive Radiotherapy of Head and Neck Cancer. Semin Oncol 2010, 20: 84-93.CrossRef Castadot P, Lee JA, Geets X, Grégoire V: Adaptive Radiotherapy of Head and Neck Cancer. Semin Oncol 2010, 20: 84-93.CrossRef
3.
go back to reference Castadot P, Lee J, Parraga A, Geets X, Macq B, Grégoire V: Comparison of 12 deformable registration strategies in adaptive radiation therapy for the treatment of head and neck tumors. Radiother Oncol 2008, 89: 1-12.CrossRefPubMed Castadot P, Lee J, Parraga A, Geets X, Macq B, Grégoire V: Comparison of 12 deformable registration strategies in adaptive radiation therapy for the treatment of head and neck tumors. Radiother Oncol 2008, 89: 1-12.CrossRefPubMed
4.
go back to reference Elstrøm UV, Wysocka BA, Muren LP, Petersen JBB, Grau C: Daily kV cone-beam CT and deformable image registration as a method for studying dosimetric consequences of anatomic changes in adaptive IMRT of head and neck cancer. Acta Oncol 2010, 49: 1101-1108.CrossRefPubMed Elstrøm UV, Wysocka BA, Muren LP, Petersen JBB, Grau C: Daily kV cone-beam CT and deformable image registration as a method for studying dosimetric consequences of anatomic changes in adaptive IMRT of head and neck cancer. Acta Oncol 2010, 49: 1101-1108.CrossRefPubMed
5.
go back to reference Hardcastle N, Tomé WA, Cannon DM, Brouwer CL, Wittendorp PWH, Dogan N, Guckenberger M, Allaire S, Mallya Y, Kumar P, Oechsner M, Richter A, Song S, Myers M, Polat B, Bzdusek K: A multi-institution evaluation of deformable image registration algorithms for automatic organ delineation in adaptive head and neck radiotherapy. Rad Oncol 2012, 7: 1.CrossRef Hardcastle N, Tomé WA, Cannon DM, Brouwer CL, Wittendorp PWH, Dogan N, Guckenberger M, Allaire S, Mallya Y, Kumar P, Oechsner M, Richter A, Song S, Myers M, Polat B, Bzdusek K: A multi-institution evaluation of deformable image registration algorithms for automatic organ delineation in adaptive head and neck radiotherapy. Rad Oncol 2012, 7: 1.CrossRef
6.
go back to reference Kaus MR, Brock KK, Pekar V, Dawson LA, Nichol AM, Jaffray DA: Assessment of a model-based deformable image registration approach for radiation therapy planning. Int J Radiat Oncol Biol Phys 2007, 68: 572-580.CrossRefPubMed Kaus MR, Brock KK, Pekar V, Dawson LA, Nichol AM, Jaffray DA: Assessment of a model-based deformable image registration approach for radiation therapy planning. Int J Radiat Oncol Biol Phys 2007, 68: 572-580.CrossRefPubMed
7.
go back to reference Wang H, Dong L, O’Daniel J, Mohan R, Garden AS, Ang KK, Kuban DA, Bonnen M, Chang JY, Cheung R: Validation of an accelerated “demons” algorithm for deformable image registration in radiation therapy. Phys Med Biol 2005, 50: 2887-2905.CrossRefPubMed Wang H, Dong L, O’Daniel J, Mohan R, Garden AS, Ang KK, Kuban DA, Bonnen M, Chang JY, Cheung R: Validation of an accelerated “demons” algorithm for deformable image registration in radiation therapy. Phys Med Biol 2005, 50: 2887-2905.CrossRefPubMed
8.
go back to reference Brock KK, Hawkins M, Eccles C, Moseley JL, Moseley DJ, Jaffray DA, Dawson LA: Improving image-guided target localization through deformable registration. Acta Oncol 2008, 47: 1279-1285.CrossRefPubMed Brock KK, Hawkins M, Eccles C, Moseley JL, Moseley DJ, Jaffray DA, Dawson LA: Improving image-guided target localization through deformable registration. Acta Oncol 2008, 47: 1279-1285.CrossRefPubMed
9.
go back to reference Davis MH, Khotanzad A, Flamig DP, Harms SE: A physics-based coordinate transformation for 3-D image matching. IEEE Trans Med Imaging 1997, 16: 317-328.CrossRefPubMed Davis MH, Khotanzad A, Flamig DP, Harms SE: A physics-based coordinate transformation for 3-D image matching. IEEE Trans Med Imaging 1997, 16: 317-328.CrossRefPubMed
10.
go back to reference Rohde GK, Aldroubi A, Dawant BM: The adaptive bases algorithm for intensity-based nonrigid image registration. IEEE Trans Med Imaging 2003, 22: 1470-1479.CrossRefPubMed Rohde GK, Aldroubi A, Dawant BM: The adaptive bases algorithm for intensity-based nonrigid image registration. IEEE Trans Med Imaging 2003, 22: 1470-1479.CrossRefPubMed
11.
go back to reference Rueckert D, Sonoda LI, Hayes C, Hill DLG, Leach MO, Hawkes DJ: Nonrigid Registration Using Free-Form Deformations: Application to Breast MR Images. IEEE Trans Med Imaging 1999, 18: 712-721.CrossRefPubMed Rueckert D, Sonoda LI, Hayes C, Hill DLG, Leach MO, Hawkes DJ: Nonrigid Registration Using Free-Form Deformations: Application to Breast MR Images. IEEE Trans Med Imaging 1999, 18: 712-721.CrossRefPubMed
12.
go back to reference Unser M: Splines: a perfect fit for signal and image processing. IEEE Signal Process Mag 1999, 16: 22-38.CrossRef Unser M: Splines: a perfect fit for signal and image processing. IEEE Signal Process Mag 1999, 16: 22-38.CrossRef
14.
go back to reference Murphy MJ, Wei Z, Fatyga M, Williamson J, Anscher M, Wallace T, Weiss E: How does CT image noise affect 3D deformable image registration for image-guided radiotherapy planning? Med Phys 2008, 35: 1145.CrossRefPubMed Murphy MJ, Wei Z, Fatyga M, Williamson J, Anscher M, Wallace T, Weiss E: How does CT image noise affect 3D deformable image registration for image-guided radiotherapy planning? Med Phys 2008, 35: 1145.CrossRefPubMed
16.
go back to reference Siewerdsen JH, Cunningham IA, Jaffray DA: A framework for noise-power spectrum analysis of multidimensional images. Med Phys 2002, 29: 2655.CrossRefPubMed Siewerdsen JH, Cunningham IA, Jaffray DA: A framework for noise-power spectrum analysis of multidimensional images. Med Phys 2002, 29: 2655.CrossRefPubMed
17.
go back to reference Klein S, Staring M, Murphy K, Viergever MA, Pluim JPW: Elastix: a Toolbox for Intensity-Based Medical Image Registration. IEEE Trans Med Imaging 2010, 29: 196-205.CrossRefPubMed Klein S, Staring M, Murphy K, Viergever MA, Pluim JPW: Elastix: a Toolbox for Intensity-Based Medical Image Registration. IEEE Trans Med Imaging 2010, 29: 196-205.CrossRefPubMed
19.
go back to reference Kybic J, Unser M: Fast parametric elastic image registration. IEEE Trans Image Process 2003, 12: 1427-1442.CrossRefPubMed Kybic J, Unser M: Fast parametric elastic image registration. IEEE Trans Image Process 2003, 12: 1427-1442.CrossRefPubMed
20.
go back to reference Klein S, Staring M, Pluim JPW: Evaluation of optimization methods for nonrigid medical image registration using mutual information and B-splines. IEEE Trans Image Process 2007, 16: 2879-2890.CrossRefPubMed Klein S, Staring M, Pluim JPW: Evaluation of optimization methods for nonrigid medical image registration using mutual information and B-splines. IEEE Trans Image Process 2007, 16: 2879-2890.CrossRefPubMed
21.
go back to reference Tanner C, Schnabel JA, Hill DLG, Hawkes DJ, Degenhard A, Leach MO, Hose DR, Hall-Craggs MA, Usiskin SI: Quantitative evaluation of free-form deformation registration for dynamic contrast-enhanced MR mammography. Med Phys 2007, 34: 1221.CrossRefPubMed Tanner C, Schnabel JA, Hill DLG, Hawkes DJ, Degenhard A, Leach MO, Hose DR, Hall-Craggs MA, Usiskin SI: Quantitative evaluation of free-form deformation registration for dynamic contrast-enhanced MR mammography. Med Phys 2007, 34: 1221.CrossRefPubMed
22.
go back to reference Varadhan R, Karangelis G, Krishnan KHS: A framework for deformable image registration vlaidation in radiotherapy clinical applications. J Appl Clin Med Phys 2013, 14: 763-784. Varadhan R, Karangelis G, Krishnan KHS: A framework for deformable image registration vlaidation in radiotherapy clinical applications. J Appl Clin Med Phys 2013, 14: 763-784.
23.
go back to reference Østergaard Noe K, De Senneville BD, ElstrØm UV, Tanderup K, Sangild T: Acceleration and validation of optical flow based deformable registration for image-guided radiotherapy. Acta Oncol 2008, 47: 1286-1293.CrossRefPubMed Østergaard Noe K, De Senneville BD, ElstrØm UV, Tanderup K, Sangild T: Acceleration and validation of optical flow based deformable registration for image-guided radiotherapy. Acta Oncol 2008, 47: 1286-1293.CrossRefPubMed
24.
go back to reference Christensen GE, Johnson HJ: Invertibility and transitivity analysis for nonrigid image registration. J Electron Imaging 2003, 12: 106.CrossRef Christensen GE, Johnson HJ: Invertibility and transitivity analysis for nonrigid image registration. J Electron Imaging 2003, 12: 106.CrossRef
25.
go back to reference Bender ET, Hardcastle N, Tomé WA: On the dosimetric effect and reduction of inverse consistency and transitivity errors in deformable image registration for dose accumulation. Med Phys 2012, 39: 272.CrossRefPubMed Bender ET, Hardcastle N, Tomé WA: On the dosimetric effect and reduction of inverse consistency and transitivity errors in deformable image registration for dose accumulation. Med Phys 2012, 39: 272.CrossRefPubMed
26.
go back to reference Staring M, Klein S, Pluim JPW: A rigidity penalty term for nonrigid registration. Med Phys 2007, 34: 4098-4108.CrossRefPubMed Staring M, Klein S, Pluim JPW: A rigidity penalty term for nonrigid registration. Med Phys 2007, 34: 4098-4108.CrossRefPubMed
Metadata
Title
The effects of computed tomography image characteristics and knot spacing on the spatial accuracy of B-spline deformable image registration in the head and neck geometry
Authors
Charlotte L Brouwer
Roel GJ Kierkels
Aart A van ’t Veld
Nanna M Sijtsema
Harm Meertens
Publication date
01-12-2014
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2014
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/1748-717X-9-169

Other articles of this Issue 1/2014

Radiation Oncology 1/2014 Go to the issue