01-11-2017 | Technical Report
Test-retest reliability of echo intensity parameters in healthy Achilles tendons using a semi-automatic tracing procedure
Published in: Skeletal Radiology | Issue 11/2017
Login to get accessAbstract
Objective
To evaluate the test-retest reliability of the ultrasound echo intensity parameters on healthy Achilles tendon using a semi-automatic tracing procedure.
Materials and methods
Eighteen healthy volunteers participated. B-mode images were acquired in the transverse plane (mid-tendon; insertion) and used to analyze tendon echogenicity. Grayscale distribution of the pixels within the selected ROIs was represented as a histogram. Descriptive statistics of the grayscale distribution (mean, variance, skewness, kurtosis, and entropy) and parameters from the co-occurrence matrix (contrast, energy, and homogeneity) were calculated.
Results
Reliability of echo intensity parameters of the mid-Achilles tendon ranged from high to very high with an ICC2.k of 0.94 for echogenicity, 0.87 for variance, 0.80 for skewness, 0.72 for kurtosis, 0.89 for entropy, 0.90 for contrast, 0.91 for energy, and 0.93 for homogeneity, while for tendon insertion they ranged from moderate to high with an ICC2.k of 0.74 for echogenicity, 0.88 for variance, 0.75 for skewness, 0.55 for kurtosis, 0.87 for entropy, 0.70 for contrast, 0.77 for energy, and 0.56 for homogeneity.
Conclusion
Ultrasound echo intensity is a reliable technique to characterize the internal structure of the Achilles tendon.