Open Access 01-12-2016 | Research

Stereoscopic X-ray imaging, cone beam CT, and couch positioning in stereotactic radiotherapy of intracranial tumors: preliminary results from a cross-modality pilot installation

Authors: Barbara Zollner, Christian Heinz, Sabrina Pitzler, Farkhad Manapov, Steffi Kantz, Maya Christine Rottler, Maximilian Niyazi, Ute Ganswindt, Claus Belka, Hendrik Ballhausen

Published in: Radiation Oncology | Issue 1/2016

Login to get access

Abstract

Background

To assess the accuracy and precision of a fully integrated pilot installation of stereoscopic X-ray imaging and kV-CBCT for automatic couch positioning in stereotactic radiotherapy of intracranial tumors. Positioning errors as detected by stereoscopic X-ray imaging are compared to those by kV-CBCT (i.e. the accuracy of the new method is verified by the established method), and repeated X-ray images are compared (i.e. the precision of new method is determined intra-modally).

Methods

Preliminary results are reported from a study with 32 patients with intracranial tumors. Patients were treated with stereotactic radiotherapy guided by stereoscopic X-ray imaging and kV-CBCT. Patient positioning was automatically corrected by a robotic couch. Cross-modal discrepancies in position detection were measured (N = 42). Intra-modal improvements after correction and re-verification by stereoscopic X-ray imaging were measured (N = 70). The accuracy and precision of stereoscopic X-ray imaging and the accuracy and precision of CBCT were confirmed in phantom measurements (N = 12 shifts of a ball bearing phantom, N = 24 shifts of a head phantom).

Results

After correction based on stereoscopic X-ray imaging 95% of residual mean errors were below 0.4, 0.4, 0.5, and 0.7 mm (lateral, longitudinal, vertical, radial, respectively). Stereoscopic X-ray imaging and CBCT were in close agreement with an average discrepancy of 0.1, 0.5, 0.3 and 0.8 mm, respectively. 95% of discrepancies were below 0.8, 1.2, 1.0, and 1.4 mm, respectively. After correction and re-verification by stereoscopic X-ray imaging, the remaining intra-modal residual error was consistent with zero (p = 0.31, p = 0.48, p = 0.81 in lateral, longitudinal, and vertical direction; p-values from two-tailed t-test). The inherent technical accuracy and precision of stereoscopic X-ray imaging and the accuracy and precision of CBCT were found to be of the order of 0.1 mm in controlled phantom settings.

Conclusions

In a routine clinical setting, both stereoscopic X-ray imaging and CBCT were able to reduce positioning errors by an order of magnitude. The end-to-end precision of the system, measured from the discrepancy (mean) between ExacTrac and CBCT, in a clinical setting seems to be about 0.8 mm radially, including couch positioning. The precision (measured from repeatability of ExacTrac, intra-modal) was found to be about 0.7 mm radially in a clinical setting.
Literature
1.
go back to reference Badakhshi H, Kaul D, Wust P, Wiener E, Budach V, Graf R. Image-guided stereotactic radiosurgery for cranial lesions: large margins compensate for reduced image guidance frequency. Anticancer Res. 2013;33(10):4639–43.PubMed Badakhshi H, Kaul D, Wust P, Wiener E, Budach V, Graf R. Image-guided stereotactic radiosurgery for cranial lesions: large margins compensate for reduced image guidance frequency. Anticancer Res. 2013;33(10):4639–43.PubMed
2.
go back to reference Nieder C, Grosu AL, Gaspar LE. Stereotactic radiosurgery (SRS) for brain metastases: a systematic review. Radiat Oncol. 2014;9:155. doi:10.1186/1748-717X-9-155 Nieder C, Grosu AL, Gaspar LE. Stereotactic radiosurgery (SRS) for brain metastases: a systematic review. Radiat Oncol. 2014;9:155. doi:10.1186/1748-717X-9-155
3.
go back to reference Harrabi SB, Adeberg S, Welzel T, Rieken S, Habermehl D, Debus J, et al. Long term results after fractionated stereotactic radiotherapy (FSRT) in patients with craniopharyngioma: maximal tumor control with minimal side effects. Radiat Oncol. 2014;9:203. doi:10.1186/1748-717X-9-203.CrossRefPubMedPubMedCentral Harrabi SB, Adeberg S, Welzel T, Rieken S, Habermehl D, Debus J, et al. Long term results after fractionated stereotactic radiotherapy (FSRT) in patients with craniopharyngioma: maximal tumor control with minimal side effects. Radiat Oncol. 2014;9:203. doi:10.​1186/​1748-717X-9-203.CrossRefPubMedPubMedCentral
4.
go back to reference El Gantery MM, Abd El Baky HM, El Hossieny HA, Mahmoud M, Youssef O. Management of brain metastases with stereotactic radiosurgery alone versus whole brain irradiation alone versus both. Radiat Oncol. 2014;9:116. doi:10.1186/1748-717X-9-116 El Gantery MM, Abd El Baky HM, El Hossieny HA, Mahmoud M, Youssef O. Management of brain metastases with stereotactic radiosurgery alone versus whole brain irradiation alone versus both. Radiat Oncol. 2014;9:116. doi:10.1186/1748-717X-9-116
6.
go back to reference Ohtakara K, Hayashi S, Tanaka H, Hoshi H, Kitahara M, Matsuyama K, et al. Clinical comparison of positional accuracy and stability between dedicated versus conventional masks for immobilization in cranial stereotactic radiotherapy using 6-degree-of-freedom image guidance system-integrated platform. Radiother Oncol. 2012;102(2):198–205. doi:10.1016/j.radonc.2011.10.012.CrossRefPubMed Ohtakara K, Hayashi S, Tanaka H, Hoshi H, Kitahara M, Matsuyama K, et al. Clinical comparison of positional accuracy and stability between dedicated versus conventional masks for immobilization in cranial stereotactic radiotherapy using 6-degree-of-freedom image guidance system-integrated platform. Radiother Oncol. 2012;102(2):198–205. doi:10.​1016/​j.​radonc.​2011.​10.​012.CrossRefPubMed
8.
go back to reference Ma J, Chang Z, Wang Z, Jackie Wu Q, Kirkpatrick JP, Yin FF. ExacTrac X-ray 6 degree-of-freedom image-guidance for intracranial non-invasive stereotactic radiotherapy: comparison with kilo-voltage cone-beam CT. Radiother Oncol. 2009;93(3):602–8. doi:10.1016/j.radonc.2009.09.009.CrossRefPubMed Ma J, Chang Z, Wang Z, Jackie Wu Q, Kirkpatrick JP, Yin FF. ExacTrac X-ray 6 degree-of-freedom image-guidance for intracranial non-invasive stereotactic radiotherapy: comparison with kilo-voltage cone-beam CT. Radiother Oncol. 2009;93(3):602–8. doi:10.​1016/​j.​radonc.​2009.​09.​009.CrossRefPubMed
11.
go back to reference Jonsson M, Ceberg S, Nordstrom F, Thornberg C, Back SA. Technical evaluation of a laser-based optical surface scanning system for prospective and retrospective breathing adapted computed tomography. Acta Oncol. 2015;54(2):261–5. doi:10.3109/0284186X.2014.948059 Jonsson M, Ceberg S, Nordstrom F, Thornberg C, Back SA. Technical evaluation of a laser-based optical surface scanning system for prospective and retrospective breathing adapted computed tomography. Acta Oncol. 2015;54(2):261–5. doi:10.3109/0284186X.2014.948059
12.
go back to reference Placht S, Stancanello J, Schaller C, Balda M, Angelopoulou E. Fast time-of-flight camera based surface registration for radiotherapy patient positioning. Med Phys. 2012;39(1):4–17. doi:10.1118/1.3664006.CrossRefPubMed Placht S, Stancanello J, Schaller C, Balda M, Angelopoulou E. Fast time-of-flight camera based surface registration for radiotherapy patient positioning. Med Phys. 2012;39(1):4–17. doi:10.​1118/​1.​3664006.CrossRefPubMed
13.
go back to reference Gilles M, Fayad H, Miglierini P, Clement JF, Scheib S, Cozzi L, et al. Patient positioning in radiotherapy based on surface imaging using time of flight cameras. Med Phys. 2016;43(8):4833. doi:10.1118/1.4959536.CrossRefPubMed Gilles M, Fayad H, Miglierini P, Clement JF, Scheib S, Cozzi L, et al. Patient positioning in radiotherapy based on surface imaging using time of flight cameras. Med Phys. 2016;43(8):4833. doi:10.​1118/​1.​4959536.CrossRefPubMed
17.
go back to reference Kim J, Jin JY, Walls N, Nurushev T, Movsas B, Chetty IJ, et al. Image-guided localization accuracy of stereoscopic planar and volumetric imaging methods for stereotactic radiation surgery and stereotactic body radiation therapy: a phantom study. Int J Radiat Oncol Biol Phys. 2011;79(5):1588–96. doi:10.1016/j.ijrobp.2010.05.052.CrossRefPubMed Kim J, Jin JY, Walls N, Nurushev T, Movsas B, Chetty IJ, et al. Image-guided localization accuracy of stereoscopic planar and volumetric imaging methods for stereotactic radiation surgery and stereotactic body radiation therapy: a phantom study. Int J Radiat Oncol Biol Phys. 2011;79(5):1588–96. doi:10.​1016/​j.​ijrobp.​2010.​05.​052.CrossRefPubMed
18.
go back to reference Tominaga H, Araki F, Shimohigashi Y, Ishihara T, Kawasaki K, Kanetake N, et al. Accuracy of positioning and irradiation targeting for multiple targets in intracranial image-guided radiation therapy: a phantom study. Phys Med Biol. 2014;59(24):7753–66. doi:10.1088/0031-9155/59/24/7753.CrossRefPubMed Tominaga H, Araki F, Shimohigashi Y, Ishihara T, Kawasaki K, Kanetake N, et al. Accuracy of positioning and irradiation targeting for multiple targets in intracranial image-guided radiation therapy: a phantom study. Phys Med Biol. 2014;59(24):7753–66. doi:10.​1088/​0031-9155/​59/​24/​7753.CrossRefPubMed
19.
go back to reference Brainlab. Exactrac frameless radiosurgery, clinical white paper. 2014. Brainlab. Exactrac frameless radiosurgery, clinical white paper. 2014.
21.
go back to reference Infusino E, Trodella L, Ramella S, D’Angelillo RM, Greco C, Iurato A, et al. Estimation of patient setup uncertainty using BrainLAB Exatrac X-Ray 6D system in image-guided radiotherapy. J Appl Clin Med Phys. 2015;16(2):5102. doi:10.1120/jacmp.v16i2.5102.PubMed Infusino E, Trodella L, Ramella S, D’Angelillo RM, Greco C, Iurato A, et al. Estimation of patient setup uncertainty using BrainLAB Exatrac X-Ray 6D system in image-guided radiotherapy. J Appl Clin Med Phys. 2015;16(2):5102. doi:10.​1120/​jacmp.​v16i2.​5102.PubMed
24.
go back to reference Benedict SH, Yenice KM, Followill D, Galvin JM, Hinson W, Kavanagh B, et al. Stereotactic body radiation therapy: the report of AAPM Task Group 101. Med Phys. 2010;37(8):4078–101. doi:10.1118/1.3438081.CrossRefPubMed Benedict SH, Yenice KM, Followill D, Galvin JM, Hinson W, Kavanagh B, et al. Stereotactic body radiation therapy: the report of AAPM Task Group 101. Med Phys. 2010;37(8):4078–101. doi:10.​1118/​1.​3438081.CrossRefPubMed
Metadata
Title
Stereoscopic X-ray imaging, cone beam CT, and couch positioning in stereotactic radiotherapy of intracranial tumors: preliminary results from a cross-modality pilot installation
Authors
Barbara Zollner
Christian Heinz
Sabrina Pitzler
Farkhad Manapov
Steffi Kantz
Maya Christine Rottler
Maximilian Niyazi
Ute Ganswindt
Claus Belka
Hendrik Ballhausen
Publication date
01-12-2016
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2016
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/s13014-016-0735-2

Other articles of this Issue 1/2016

Radiation Oncology 1/2016 Go to the issue