Open Access 01-12-2011 | Research

Statistical analysis of IMRT dosimetry quality assurance measurements for local delivery guideline

Authors: Jin Beom Chung, Jae Sung Kim, Sung Whan Ha, Sung-Joon Ye

Published in: Radiation Oncology | Issue 1/2011

Login to get access

Abstract

Purpose

To establish our institutional guideline for IMRT delivery, we statistically evaluated the results of dosimetry quality assurance (DQA) measurements and derived local confidence limits using the concept confidence limit of |mean|+1.96σ.

Materials and methods

From June 2006 to March 2009, 206 patients with head and neck cancer, prostate cancer, liver cancer, or brain tumor were treated using LINAC-based IMRT technique. In order to determine site specific DQA tolerances at a later stage, a hybrid plan with the same fluence maps as in the treatment plan was generated on CT images of a cylindrical phantom of acryl. Points of measurement using a 0.125 cm3 ion-chamber were typically located in the region of high and uniform doses. The planar dose distributions perpendicular to the central axis were measured by using a diode array in solid water with all fields delivered, and assessed using gamma criteria of 3%/3 mm. The mean values and standard deviations were used to develop the local confidence and tolerance limits. The dose differences and gamma pass rates for the different treatment sites were also evaluated in terms of total monitor uints (MU), MU/cGy, and the number of PTV's pieces.

Results

The mean values and standard deviations of ion-chamber dosimetry differences between calculated and measured doses were -1.6 ± 1.2% for H&N cancer, -0.4 ± 1.2% for prostate and abdominal cancer, and -0.6 ± 1.5% for brain tumor. Most of measured doses (92.2%) agreed with the calculated doses within a tolerance limit of ±3% recommended in the literature. However, we found some systematic under-dosage for all treatment sites. The percentage of points passing the gamma criteria, averaged over all treatment sites was 97.3 ± 3.7%. The gamma pass rate and the agreement of ion-chamber dosimetry generally decreased with increasing the number of PTV's pieces, the degree of modulation (MU/cGy), and the total MU beyond 700. Our local confidence limits were comparable to those of AAPM TG 119 and ESTRO guidelines that were provided as a practical baseline for center-to-center commissioning comparison. Thus, our institutional confidence and action limits for IMRT delivery were set into the same levels of those guidelines.

Discussion and Conclusions

The systematic under-dosage were corrected by tuning up the MLC-related factors (dosimetric gap and transmission) in treatment planning system (TPS) and further by incorporating the tongue-and groove effect into TPS. Institutions that have performed IMRT DQA measurements over a certain period of time need to analyze their accrued DQA data. We confirmed the overall integrity of our IMRT system and established the IMRT delivery guideline during this procedure. Dosimetric corrections for the treatment plans outside of the action level can be suggested only with such rigorous DQA and statistical analysis.
Appendix
Available only for authorised users
Literature
1.
go back to reference Intensity Modulated Radiation Therapy Collaborative Working Group: Intensity-modulated radiotherapy: current status and issues of interest. Int J Radiat Oncol Biol Phys 2001, 51: 880-914. 10.1016/S0360-3016(01)01749-7CrossRef Intensity Modulated Radiation Therapy Collaborative Working Group: Intensity-modulated radiotherapy: current status and issues of interest. Int J Radiat Oncol Biol Phys 2001, 51: 880-914. 10.1016/S0360-3016(01)01749-7CrossRef
2.
go back to reference Kao J, Turian J, Meyers A, Hamilton RJ, Smith B, Vijayakumar S, Jani AB: Sparing of the penile bulb and proximal penile structures with intensity-modulated radiation therapy for prostate cancer. Br J Radiol 2004, 77: 129-136. 10.1259/bjr/37893924CrossRefPubMed Kao J, Turian J, Meyers A, Hamilton RJ, Smith B, Vijayakumar S, Jani AB: Sparing of the penile bulb and proximal penile structures with intensity-modulated radiation therapy for prostate cancer. Br J Radiol 2004, 77: 129-136. 10.1259/bjr/37893924CrossRefPubMed
3.
go back to reference Ahamad A, D'Souza W, Salehpour M, Iyer R, Tucker SL, Jhingran A, Eifel PJ: Intensity-modulated radiation therapy after hysterectomy: comparison with conventional treatment and sensitivity of the normal-tissue-sparing effect to margin size. Int J Radiat Oncol Biol Phys 2005, 62: 1117-1124. 10.1016/j.ijrobp.2004.12.029CrossRefPubMed Ahamad A, D'Souza W, Salehpour M, Iyer R, Tucker SL, Jhingran A, Eifel PJ: Intensity-modulated radiation therapy after hysterectomy: comparison with conventional treatment and sensitivity of the normal-tissue-sparing effect to margin size. Int J Radiat Oncol Biol Phys 2005, 62: 1117-1124. 10.1016/j.ijrobp.2004.12.029CrossRefPubMed
4.
go back to reference Sultanem K, Shu HK, Xia P, Akazawa C, Quivey JM, Verhey LJ, Fu KK: Three dimension intensity modulated radiotherapy in the treatment of nasopharyngeal carcinoma: The University of California - San Francisco experience. Int J Radiat Oncol Biol Phys 2000, 48: 711-722. 10.1016/S0360-3016(00)00702-1CrossRefPubMed Sultanem K, Shu HK, Xia P, Akazawa C, Quivey JM, Verhey LJ, Fu KK: Three dimension intensity modulated radiotherapy in the treatment of nasopharyngeal carcinoma: The University of California - San Francisco experience. Int J Radiat Oncol Biol Phys 2000, 48: 711-722. 10.1016/S0360-3016(00)00702-1CrossRefPubMed
5.
go back to reference Boyer LA, Geis P, Grant W, Carol M: Modulated beam conformal therapy for head and neck tumors. Int J Radiat Oncol Biol Phys 1997, 39: 227-236.CrossRefPubMed Boyer LA, Geis P, Grant W, Carol M: Modulated beam conformal therapy for head and neck tumors. Int J Radiat Oncol Biol Phys 1997, 39: 227-236.CrossRefPubMed
6.
go back to reference Li XA, Wang JZ, Jursinic PA, Lawton CA, Wang D: Dosimetric advantages of IMRT simultaneous integrated boost for high-risk prostate cancer. Int J Radiat Oncol Biol Phys 2005, 61: 1251-1257. 10.1016/j.ijrobp.2004.11.034CrossRefPubMed Li XA, Wang JZ, Jursinic PA, Lawton CA, Wang D: Dosimetric advantages of IMRT simultaneous integrated boost for high-risk prostate cancer. Int J Radiat Oncol Biol Phys 2005, 61: 1251-1257. 10.1016/j.ijrobp.2004.11.034CrossRefPubMed
7.
go back to reference Cilla S, Macchia G, Digesu C, Deodato F, Romanella M, Ferrandina G, Padula G, Picardi V, Scambia G, Piermattei A, Morganti AG: 3D-Conformal versus intensity- modulated postoperative radiotherapy of vaginal vault: a dosimetric comparison. Med Dosm 2010, 35: 135-142. 10.1016/j.meddos.2009.04.003CrossRef Cilla S, Macchia G, Digesu C, Deodato F, Romanella M, Ferrandina G, Padula G, Picardi V, Scambia G, Piermattei A, Morganti AG: 3D-Conformal versus intensity- modulated postoperative radiotherapy of vaginal vault: a dosimetric comparison. Med Dosm 2010, 35: 135-142. 10.1016/j.meddos.2009.04.003CrossRef
8.
go back to reference Palta JR, Liu C, Li JG: Quality assurance of intensity modulated radiation therapy. Int J Radiat Oncol Biol Phys 2008, 71: S108-S112. 10.1016/j.ijrobp.2007.05.092CrossRefPubMed Palta JR, Liu C, Li JG: Quality assurance of intensity modulated radiation therapy. Int J Radiat Oncol Biol Phys 2008, 71: S108-S112. 10.1016/j.ijrobp.2007.05.092CrossRefPubMed
9.
go back to reference Boehmer D, Bohsung J, Eichwurzel I, Moys A, Budach V: Clinical and physical quality assurance for intensity modulated radiotherapy of prostate cancer. Radiother Oncol 2004, 71: 319-325. 10.1016/j.radonc.2004.02.016CrossRefPubMed Boehmer D, Bohsung J, Eichwurzel I, Moys A, Budach V: Clinical and physical quality assurance for intensity modulated radiotherapy of prostate cancer. Radiother Oncol 2004, 71: 319-325. 10.1016/j.radonc.2004.02.016CrossRefPubMed
10.
go back to reference Ezzell GA, Galvin JM, Low D, Palta JR, Rosen I, Sharpe MB, Xia P, Xiao Y, Xing L, Yu CX: Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee. Med Phys 2003, 30: 2089-2115. 10.1118/1.1591194CrossRefPubMed Ezzell GA, Galvin JM, Low D, Palta JR, Rosen I, Sharpe MB, Xia P, Xiao Y, Xing L, Yu CX: Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee. Med Phys 2003, 30: 2089-2115. 10.1118/1.1591194CrossRefPubMed
11.
go back to reference Doblado FS, Capote R, Rosello JV, Leal A, Lagares JI, Arrans R, Hartmann GH: Micro ionization-chamber dosimetry in IMRT verification: Clinical implications of dosimetric errors in the PTV. Radiother Oncol 2005, 75: 342-348. 10.1016/j.radonc.2005.04.011CrossRef Doblado FS, Capote R, Rosello JV, Leal A, Lagares JI, Arrans R, Hartmann GH: Micro ionization-chamber dosimetry in IMRT verification: Clinical implications of dosimetric errors in the PTV. Radiother Oncol 2005, 75: 342-348. 10.1016/j.radonc.2005.04.011CrossRef
12.
go back to reference Mijheer B, Georg D: Guidelines for the verification of IMRT. Brussels, Belgium: ESTRO; 2008. Mijheer B, Georg D: Guidelines for the verification of IMRT. Brussels, Belgium: ESTRO; 2008.
13.
go back to reference Ezzell GA, Burmeister JW, Dogan N, LoSasso TJ, Mechalakos JG, Mihailidis D, Molineu A, Palta JR, Ramsey CR, Salter BJ, Shi J, Xia P, Yue NJ, Xiao Y: IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys 2009, 36: 5359-5373. 10.1118/1.3238104CrossRefPubMed Ezzell GA, Burmeister JW, Dogan N, LoSasso TJ, Mechalakos JG, Mihailidis D, Molineu A, Palta JR, Ramsey CR, Salter BJ, Shi J, Xia P, Yue NJ, Xiao Y: IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys 2009, 36: 5359-5373. 10.1118/1.3238104CrossRefPubMed
14.
go back to reference Agazaryan N, Solberg TD, Demarco JJ: Patient specific quality assurance for the delivery of intensity modulated radiotherapy. J Appl Clin Med Phys 2003, 4: 453-461. Agazaryan N, Solberg TD, Demarco JJ: Patient specific quality assurance for the delivery of intensity modulated radiotherapy. J Appl Clin Med Phys 2003, 4: 453-461.
15.
go back to reference Yan G, Liu C, Simon TA, Peng LC, Fox C, Li JG: On the sensitivity of patient-specific IMRT QA to MLC positioning errors. J A ppl Clin Med Phys 2009, 10: 121-128. Yan G, Liu C, Simon TA, Peng LC, Fox C, Li JG: On the sensitivity of patient-specific IMRT QA to MLC positioning errors. J A ppl Clin Med Phys 2009, 10: 121-128.
16.
go back to reference Tasi JS, Wazer DE, Ling MN, Wu K, Fagundes M, Dipetrillo T, Kramer B, Koistinen M, Engler MJ: Dosimetric verification of the dynamic intensity modulated radiation therapy of 92 patients. Int J Radiat Oncol Biol Phys 1998, 40: 1213-1230. 10.1016/S0360-3016(98)00009-1CrossRef Tasi JS, Wazer DE, Ling MN, Wu K, Fagundes M, Dipetrillo T, Kramer B, Koistinen M, Engler MJ: Dosimetric verification of the dynamic intensity modulated radiation therapy of 92 patients. Int J Radiat Oncol Biol Phys 1998, 40: 1213-1230. 10.1016/S0360-3016(98)00009-1CrossRef
17.
go back to reference Ting JY, Davis LW: Dose verification for patients undergoing IMRT. Med Dosim 2001, 26: 179-188. 10.1016/S0958-3947(01)00059-0CrossRef Ting JY, Davis LW: Dose verification for patients undergoing IMRT. Med Dosim 2001, 26: 179-188. 10.1016/S0958-3947(01)00059-0CrossRef
18.
go back to reference Dong L, Antolak J, Salehpour M, Forster K, O'Neil L, Kendall R, Rosen I: Patient-specific point dose measurement for IMRT monitor unit verification. Int J Radiat Oncol Biol Phys 2003, 26: 867-877.CrossRef Dong L, Antolak J, Salehpour M, Forster K, O'Neil L, Kendall R, Rosen I: Patient-specific point dose measurement for IMRT monitor unit verification. Int J Radiat Oncol Biol Phys 2003, 26: 867-877.CrossRef
19.
go back to reference Pawlicki T, Whitaker M, Boyer AL: Statical process control for radiotherapy quality assurance. Med Phys 2005, 32: 2777-2786. 10.1118/1.2001209CrossRefPubMed Pawlicki T, Whitaker M, Boyer AL: Statical process control for radiotherapy quality assurance. Med Phys 2005, 32: 2777-2786. 10.1118/1.2001209CrossRefPubMed
20.
go back to reference Pawlicki T, Mundt AJ: Quality in radiation oncology. Med Phys 2007, 34: 1529-1534. 10.1118/1.2727748CrossRefPubMed Pawlicki T, Mundt AJ: Quality in radiation oncology. Med Phys 2007, 34: 1529-1534. 10.1118/1.2727748CrossRefPubMed
21.
go back to reference Breen SL, Moseley DJ, Zhang B, Sharpe MB: Statistical process control for IMRT dosimetric verification. Med Phys 2008, 35: 4417-4425. 10.1118/1.2975144CrossRefPubMed Breen SL, Moseley DJ, Zhang B, Sharpe MB: Statistical process control for IMRT dosimetric verification. Med Phys 2008, 35: 4417-4425. 10.1118/1.2975144CrossRefPubMed
22.
go back to reference Low DA, Dempsey JF: Evaluation of the gamma dose distribution comparison method. Med Phys 2003, 30: 2455-2464. 10.1118/1.1598711CrossRefPubMed Low DA, Dempsey JF: Evaluation of the gamma dose distribution comparison method. Med Phys 2003, 30: 2455-2464. 10.1118/1.1598711CrossRefPubMed
23.
go back to reference Clark CH, Hansen VN, Chantler H, Edwards C, James HV, Webster G, Miles EA, Guerrero Urbano MT, Bhide SA, Bidmead AM, Nutting CM: Dosimetry audit for a multi-centre IMRT head and neck trial. Radiother Oncol 2009, 93: 102-108. 10.1016/j.radonc.2009.04.025CrossRefPubMed Clark CH, Hansen VN, Chantler H, Edwards C, James HV, Webster G, Miles EA, Guerrero Urbano MT, Bhide SA, Bidmead AM, Nutting CM: Dosimetry audit for a multi-centre IMRT head and neck trial. Radiother Oncol 2009, 93: 102-108. 10.1016/j.radonc.2009.04.025CrossRefPubMed
24.
go back to reference Ibbott GS, Followill DS, Molineu HA, Lowenstein JR, Alvarez PE, Roll JE: Challenges in credentialing institutions and participants in advanced technology multi-institutional clinical trials. Int J Radiat Oncol Biol Phys 2008, 71: S71-S75. 10.1016/j.ijrobp.2007.08.083PubMedCentralCrossRefPubMed Ibbott GS, Followill DS, Molineu HA, Lowenstein JR, Alvarez PE, Roll JE: Challenges in credentialing institutions and participants in advanced technology multi-institutional clinical trials. Int J Radiat Oncol Biol Phys 2008, 71: S71-S75. 10.1016/j.ijrobp.2007.08.083PubMedCentralCrossRefPubMed
25.
go back to reference Jeraj R, Keall PJ, Ostwald PM: Agreement criteria between excepted and measured field fluence in IMRT of head and neck cancer: The importance and use of the γ histograms statical analysis. Radiother Oncol 2007, 85: 399-406.CrossRef Jeraj R, Keall PJ, Ostwald PM: Agreement criteria between excepted and measured field fluence in IMRT of head and neck cancer: The importance and use of the γ histograms statical analysis. Radiother Oncol 2007, 85: 399-406.CrossRef
26.
go back to reference Spirou SV, Chui CS: A gradient inverse planning algorithm with dose volume constraints. Med Phys 1998, 25: 321-333. 10.1118/1.598202CrossRefPubMed Spirou SV, Chui CS: A gradient inverse planning algorithm with dose volume constraints. Med Phys 1998, 25: 321-333. 10.1118/1.598202CrossRefPubMed
27.
go back to reference Studer G, Huguenin PU, Davis JB, Kunz G, Lütolf UM, Glanzmann C: IMRT using simultaneously integrated boost (SIB) in head and neck cancer patients. Radiation Oncology 2006, 1: 7-21. 10.1186/1748-717X-1-7PubMedCentralCrossRefPubMed Studer G, Huguenin PU, Davis JB, Kunz G, Lütolf UM, Glanzmann C: IMRT using simultaneously integrated boost (SIB) in head and neck cancer patients. Radiation Oncology 2006, 1: 7-21. 10.1186/1748-717X-1-7PubMedCentralCrossRefPubMed
28.
go back to reference Giorgia N, Antonella F, Eugenio V, Alessandro C, Filippo A, Luca C: What is an acceptably smoothed fluence? Dosimetric and delivery considerations for dynamic sliding window IMRT. Radiation Oncology 2007, 2: 42-54. 10.1186/1748-717X-2-42PubMedCentralCrossRefPubMed Giorgia N, Antonella F, Eugenio V, Alessandro C, Filippo A, Luca C: What is an acceptably smoothed fluence? Dosimetric and delivery considerations for dynamic sliding window IMRT. Radiation Oncology 2007, 2: 42-54. 10.1186/1748-717X-2-42PubMedCentralCrossRefPubMed
29.
go back to reference Jang SY, Liu HH, Mohan R: Underestimation of low-dose radiation in treatment planning of intensity-modulated radiotherapy. Int J Radiat Oncol Biol Phys 2008, 71: 1537-1546. 10.1016/j.ijrobp.2008.04.014CrossRefPubMed Jang SY, Liu HH, Mohan R: Underestimation of low-dose radiation in treatment planning of intensity-modulated radiotherapy. Int J Radiat Oncol Biol Phys 2008, 71: 1537-1546. 10.1016/j.ijrobp.2008.04.014CrossRefPubMed
30.
go back to reference Das IJ, Ding GX, Ahnesjo A, Small fields: Non-equilibrium radiation dosimetry. Med Phys 2008, 35: 206-213. 10.1118/1.2815356CrossRefPubMed Das IJ, Ding GX, Ahnesjo A, Small fields: Non-equilibrium radiation dosimetry. Med Phys 2008, 35: 206-213. 10.1118/1.2815356CrossRefPubMed
31.
go back to reference Papatheodorou S, Rosenwald JC, Zefkili S, Murillo MC, Drouard J, Gaboriaud G: Dose calculation and verification of intensity modulation generated by dynamic multileaf collimators. Med Phys 2000, 27: 960-971. 10.1118/1.598960CrossRefPubMed Papatheodorou S, Rosenwald JC, Zefkili S, Murillo MC, Drouard J, Gaboriaud G: Dose calculation and verification of intensity modulation generated by dynamic multileaf collimators. Med Phys 2000, 27: 960-971. 10.1118/1.598960CrossRefPubMed
32.
go back to reference Xing L, Curran B, Hill R, Holmes T, Ma L, Forster KM, Boyer AL: Dosimetric verification of a commercial inverse treatment planning system. Phys Med Biol 1999, 44: 463-478. 10.1088/0031-9155/44/2/013CrossRefPubMed Xing L, Curran B, Hill R, Holmes T, Ma L, Forster KM, Boyer AL: Dosimetric verification of a commercial inverse treatment planning system. Phys Med Biol 1999, 44: 463-478. 10.1088/0031-9155/44/2/013CrossRefPubMed
33.
go back to reference Capote R, Sanchez-Doblado F, Leal A, Lagares JL, Arrans R, Hartman GH: An EGSnrc Monte Carlo study of the microionization-chamber for reference dosimetry of narrow irregular IMRT beamlets. Med Phys 2004, 31: 2416-2422. 10.1118/1.1767691CrossRefPubMed Capote R, Sanchez-Doblado F, Leal A, Lagares JL, Arrans R, Hartman GH: An EGSnrc Monte Carlo study of the microionization-chamber for reference dosimetry of narrow irregular IMRT beamlets. Med Phys 2004, 31: 2416-2422. 10.1118/1.1767691CrossRefPubMed
34.
go back to reference Deng J, Pawlicki T, Chen Y, Li J, Jiang SB, Ma CM: The MLC tongue-and-groove effect on IMRT dose distributions. Phys Med Biol 2001, 46: 1039-1060. 10.1088/0031-9155/46/4/310CrossRefPubMed Deng J, Pawlicki T, Chen Y, Li J, Jiang SB, Ma CM: The MLC tongue-and-groove effect on IMRT dose distributions. Phys Med Biol 2001, 46: 1039-1060. 10.1088/0031-9155/46/4/310CrossRefPubMed
35.
go back to reference Li JS, Lin T, Chen L, Price RA, Ma CM: Uncertainties in IMRT dosimetry. Med Phys 2010, 37: 2491-2500. 10.1118/1.3413997CrossRefPubMed Li JS, Lin T, Chen L, Price RA, Ma CM: Uncertainties in IMRT dosimetry. Med Phys 2010, 37: 2491-2500. 10.1118/1.3413997CrossRefPubMed
36.
go back to reference Alfonso R, Andreo P, Capote R, Saiful Huq M, Kilby W, Kjäll P, Mackie TR, Palmans H, Rosser K, Seuntjens J, Ullrich W, Vatnitsky S: A new formalism for reference dosimetry of small and nonstandard fields. Med Phys 2008, 35: 5179-5186. 10.1118/1.3005481CrossRefPubMed Alfonso R, Andreo P, Capote R, Saiful Huq M, Kilby W, Kjäll P, Mackie TR, Palmans H, Rosser K, Seuntjens J, Ullrich W, Vatnitsky S: A new formalism for reference dosimetry of small and nonstandard fields. Med Phys 2008, 35: 5179-5186. 10.1118/1.3005481CrossRefPubMed
37.
go back to reference Sánchez-Doblado F, Andreo P, Capote R, Leal A, Perucha M, Arráns R, Núñez L, Mainegra E, Lagares JI, Carrasco E: Ionization-chamber dosimetry of small photon fields: a Monte Carlo study on stopping power ratio for radiosurgery and IMRT beams. Phys Med Biol 2003, 48: 2081-2099.CrossRefPubMed Sánchez-Doblado F, Andreo P, Capote R, Leal A, Perucha M, Arráns R, Núñez L, Mainegra E, Lagares JI, Carrasco E: Ionization-chamber dosimetry of small photon fields: a Monte Carlo study on stopping power ratio for radiosurgery and IMRT beams. Phys Med Biol 2003, 48: 2081-2099.CrossRefPubMed
Metadata
Title
Statistical analysis of IMRT dosimetry quality assurance measurements for local delivery guideline
Authors
Jin Beom Chung
Jae Sung Kim
Sung Whan Ha
Sung-Joon Ye
Publication date
01-12-2011
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2011
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/1748-717X-6-27

Other articles of this Issue 1/2011

Radiation Oncology 1/2011 Go to the issue