Stability of two angular stable locking plates for open wedge high tibial osteotomy (HTO): TomoFix™ versus LOQTEQ® HTO plate
Authors:
Wolf Petersen, Alexander Wall, Thomas Paulin, Hi Un Park, Lydia Heymann
Published in:
Archives of Orthopaedic and Trauma Surgery
|
Issue 10/2014
Login to get access
Abstract
Introduction
The aim of this study was to test the mechanical stability of LOQTEQ® HTO plate in comparison to the TomoFix™ internal plate fixator.
Materials and methods
An unstable fracture gap model with two standardized polyoxymethylene (POM) cylinders was used for this study. With this model 5 LOQTEQ® HTO plates and 5 TomoFix™ were cyclically loaded. The start load was 100 N and increased by 50 N after each successful completed cycle. Each load step included 250,000 cycles. All tests were performed in accordance with the American Society for Testing and Materials (ASTM) Standard Specification and Test Method for Metallic Bone Plates F382–99(2003)e1.
Results
The mean fatigue strength of the LOQTEQ® HTO Plate was 155 % of the value gained by the TomoFix™ implant. This difference was statistically significant (t test <0.001). There was also a significant difference in the cycles but both implants survived. The LOQTEQ® HTO Plate system survived 1,520,336.8 (±121,687.7) cycles, whereas the TomoFix™ group passed on average 847,802,600 (±134,660.7) cycles. In both groups, the only failure mode was plate breakage in a shaft hole.
Conclusions
The results of the present study showed that fatigue strength of the LOQTEQ® HTO plate was significantly higher in comparison to TomoFix™. These results justify the clinical use of the LOQTEQ® HTO plate.