Published in: Insights into Imaging 6/2012

Open Access 01-12-2012 | Original Article

Should less motion sensitive T2-weighted BLADE TSE replace Cartesian TSE for female pelvic MRI?

Authors: Johannes M. Froehlich, Thierry Metens, Bianka Chilla, Nik Hauser, Markus Klarhoefer, Rahel A. Kubik-Huch

Published in: Insights into Imaging | Issue 6/2012

Login to get access

Abstract

Objectives

To prospectively compare the diagnostic performance of a non-Cartesian k-space sampling T2-weighted TSE BLADE sequence with a conventional T2-weighted TSE sequence in female pelvic organs.

Methods

Forty-seven patients with sonographically indeterminate adnexal masses or uterine lesions underwent sagittal BLADE and conventional TSE at 1.5 T after glucagon administration. Two radiologists independently determined their preferred sequence by rating: overall image diagnostic quality, conspicuity of the zonal anatomy and delineation of pathologies of the uterus and cervix, presence of artefacts, and of fluid in the pouch of Douglas (Wilcoxon signed rank test). Signal-to noise ratios (SNRs) and contrast-to-noise ratios (CNRs) were measured for the myometrium versus the rectus abdominis muscle (Student’s t-test).

Results

BLADE significantly (p < 0.0001) reduced motion and ghosting artefacts and showed improved conspicuity (p = 0.3/0.24), but overall image quality did not differ significantly (inter-observer agreement BLADE κ = 0.89; TSE κ = 0.84). In the majority of cases (53.2 % vs 59.6 %, respectively, κ = 0.82) radiologists preferred conventional TSE due to better image contrast (p < 0.0001) and visibility of free pelvic fluid (p ≤ 0.0001). SNR (TSE 57.5 ± 37.7; BLADE 16.6 ± 12.2) and CNR (TSE 40.4 ± 33.5; BLADE 7.2 ± 8.8) were significantly higher on conventional TSE (p < 0.0001).

Conclusions

Although BLADE reduces motion artefacts and provides a clearer delineation of uterine zonal anatomy compared with conventional TSE, this comes at the expense of overall contrast.

Main Messages

Use of BLADE may reduce T2 contrast and thus visibility of free pelvic fluid or cystic structures
Non-Cartesian sampling of k-space such as BLADE is beneficial due to less motion sensitivity
BLADE provides clearer delineation and conspicuity of uterine zonal anatomy on pelvic MRIs
Literature
1.
go back to reference Spencer JA, Forstner R, Cunha TM, Kinkel K, ESUR Female Imaging Sub-Committee (2010) ESUR guidelines for MR imaging of the sonographically indeterminate adnexal mass: an algorithmic approach. Eur Radiol 20:25–35PubMedCrossRef Spencer JA, Forstner R, Cunha TM, Kinkel K, ESUR Female Imaging Sub-Committee (2010) ESUR guidelines for MR imaging of the sonographically indeterminate adnexal mass: an algorithmic approach. Eur Radiol 20:25–35PubMedCrossRef
2.
go back to reference Kinkel K, Forstner R, Danza FM et al (2009) Staging of endometrial cancer with MRI: guidelines of the European Society of Urogenital Imaging. Eur Radiol 19:1565–1574PubMedCrossRef Kinkel K, Forstner R, Danza FM et al (2009) Staging of endometrial cancer with MRI: guidelines of the European Society of Urogenital Imaging. Eur Radiol 19:1565–1574PubMedCrossRef
3.
go back to reference Hirokawa Y, Isoda H, Maetani YS et al (2008) MRI artifact reduction and quality improvement in the upper abdomen with PROPELLER and prospective acquisition correction (PACE) techniques. AJR 191:1154–1158PubMedCrossRef Hirokawa Y, Isoda H, Maetani YS et al (2008) MRI artifact reduction and quality improvement in the upper abdomen with PROPELLER and prospective acquisition correction (PACE) techniques. AJR 191:1154–1158PubMedCrossRef
4.
go back to reference Wintersperger BJ, Runge VM, Biswas J et al (2006) Brain magnetic resonance imaging at 3 Tesla using BLADE compared with standard rectilinear data sampling. Investigative Radiology 41:586–592PubMedCrossRef Wintersperger BJ, Runge VM, Biswas J et al (2006) Brain magnetic resonance imaging at 3 Tesla using BLADE compared with standard rectilinear data sampling. Investigative Radiology 41:586–592PubMedCrossRef
5.
go back to reference Pipe JG (1999) Motion correction with PROPELLER MRI: application to head motion and free-breathing cardiac imaging. Magn Reson Med 42:963–969PubMedCrossRef Pipe JG (1999) Motion correction with PROPELLER MRI: application to head motion and free-breathing cardiac imaging. Magn Reson Med 42:963–969PubMedCrossRef
6.
go back to reference Lane BF, Vandermeer FQ, Oz RC, Irwin EW, McMillan AB, Wong-You-Cheong JJ (2011) Comparison of sagittal T2-weighted BLADE and fast spin-echo MRI of the female pelvis for motion artifact and lesion detection. AJR 197:W307–W313PubMedCrossRef Lane BF, Vandermeer FQ, Oz RC, Irwin EW, McMillan AB, Wong-You-Cheong JJ (2011) Comparison of sagittal T2-weighted BLADE and fast spin-echo MRI of the female pelvis for motion artifact and lesion detection. AJR 197:W307–W313PubMedCrossRef
7.
go back to reference Koyama T, Fujimoto K, Togashi K (2007) The clinical advantages of T2-weighted MR imaging in the female pelvis with syngo BLADE. Magnetom Flash 3:60–62 Koyama T, Fujimoto K, Togashi K (2007) The clinical advantages of T2-weighted MR imaging in the female pelvis with syngo BLADE. Magnetom Flash 3:60–62
8.
go back to reference Fujimoto K, Koyama T, Tamai K et al (2011) BLADE acquisition method improves T2-weighted MR images of the female pelvis compared with a standard fast spin-echo sequence. Eur J Radiol 80:796–801PubMedCrossRef Fujimoto K, Koyama T, Tamai K et al (2011) BLADE acquisition method improves T2-weighted MR images of the female pelvis compared with a standard fast spin-echo sequence. Eur J Radiol 80:796–801PubMedCrossRef
9.
go back to reference Haneder S, Tinter D, Gutfleisch A, Schoenberg SO, Michaely HJ (2011) Image quality of T2w-TSE of the abdomen and pelvis with Cartesian or BLADE-type k-space sampling: a retrospective interindividual comparison study. Eur J Radiol 79:177–182PubMedCrossRef Haneder S, Tinter D, Gutfleisch A, Schoenberg SO, Michaely HJ (2011) Image quality of T2w-TSE of the abdomen and pelvis with Cartesian or BLADE-type k-space sampling: a retrospective interindividual comparison study. Eur J Radiol 79:177–182PubMedCrossRef
10.
go back to reference Chilla B, Hauser N, Singer G, Trippel M, Froehlich JM, Kubik-Huch RA (2011) Indeterminate adnexal masses at ultrasound: effect of MRI imaging findings on diagnostic thinking and therapeutic decisions. Eur Radiol 21:1301–1310PubMedCrossRef Chilla B, Hauser N, Singer G, Trippel M, Froehlich JM, Kubik-Huch RA (2011) Indeterminate adnexal masses at ultrasound: effect of MRI imaging findings on diagnostic thinking and therapeutic decisions. Eur Radiol 21:1301–1310PubMedCrossRef
11.
go back to reference Froehlich JM, Daenzer M, von Weymarn C et al (2009) Aperistaltic effect of hyoscine N-butylbromide versus glucagon on the small bowel assessed by magnetic resonance imaging. Eur Radiol 19:1387–1393PubMedCrossRef Froehlich JM, Daenzer M, von Weymarn C et al (2009) Aperistaltic effect of hyoscine N-butylbromide versus glucagon on the small bowel assessed by magnetic resonance imaging. Eur Radiol 19:1387–1393PubMedCrossRef
12.
go back to reference Michaely HJ, Gramer H, Weckbach S et al (2008) Renal T2-weighted turbo-spin-echo imaging with BLADE at 3.0 Tesla: initial experience. JMRI 27:148–153PubMedCrossRef Michaely HJ, Gramer H, Weckbach S et al (2008) Renal T2-weighted turbo-spin-echo imaging with BLADE at 3.0 Tesla: initial experience. JMRI 27:148–153PubMedCrossRef
13.
go back to reference Deng J, Larson AC (2009) Modified PROPELLER approach for T2-mapping of the abdomen. Magn Reson Med 61:1269–1278PubMedCrossRef Deng J, Larson AC (2009) Modified PROPELLER approach for T2-mapping of the abdomen. Magn Reson Med 61:1269–1278PubMedCrossRef
14.
go back to reference Pandit P, Qi Y, King KF, Johnson GA (2011) Reduction of artifacts in T2-weighted PROPELLER in high-field preclinical imaging. Magn Reson Med 65:538–543PubMedPubMedCentralCrossRef Pandit P, Qi Y, King KF, Johnson GA (2011) Reduction of artifacts in T2-weighted PROPELLER in high-field preclinical imaging. Magn Reson Med 65:538–543PubMedPubMedCentralCrossRef
15.
go back to reference Hussain SM, Outwater EK, Siegelmann ES (2000) MR imaging features of pelvic mucinous carcinomas. Eur Radiol 10:885–981PubMedCrossRef Hussain SM, Outwater EK, Siegelmann ES (2000) MR imaging features of pelvic mucinous carcinomas. Eur Radiol 10:885–981PubMedCrossRef
Metadata
Title
Should less motion sensitive T2-weighted BLADE TSE replace Cartesian TSE for female pelvic MRI?
Authors
Johannes M. Froehlich
Thierry Metens
Bianka Chilla
Nik Hauser
Markus Klarhoefer
Rahel A. Kubik-Huch
Publication date
01-12-2012
Publisher
Springer Berlin Heidelberg
Published in
Insights into Imaging / Issue 6/2012
Electronic ISSN: 1869-4101
DOI
https://doi.org/10.1007/s13244-012-0193-9

Other articles of this Issue 6/2012

Insights into Imaging 6/2012 Go to the issue