Published in: BMC Ophthalmology 1/2018

Open Access 01-12-2018 | Research article

Shielding effect of the smoke plume by the ablation of excimer lasers

Authors: Csaba Szekrényesi, Huba Kiss, Tamás Filkorn, Zoltán Zsolt Nagy

Published in: BMC Ophthalmology | Issue 1/2018

Login to get access

Abstract

Background

Shielding and scattering effect of the smoke plume column ejected from the laser ablated material is a well-known phenomenon. Debris evacuation system of the excimer laser equipment removes these particles, but insufficient air flow can result in undesired refractive outcomes of the treatment. The aim of this study was to reveal the effect of the air flow speed on the actual ablation depth.

Methods

SCWIND AMARIS 500E flying spot excimer laser was tested in this study. A 150 μm phototherapeutic keratectomy (PTK) profile with 8 mm diameter was applied to the surface of polymethyl methacrylate (PMMA) plates. The velocity of the air flow was changed with adjustable air aspiration system. Ablation depth was measured with highly-precise contact micrometer.

Results

The prediction model was statistically significant, F(1,8) = 552.85, p < 0.001, and accounted for approximately 98.7% of variance of ablation (R2 = 0.987, R2adj = 0.986). Lower air flow speed resulted in a weaker ablation capability of the excimer laser.

Conclusion

Air flow generated by the aspiration equipment is a key factor for the predictable outcomes of refractive treatment. Therefore, manufacturer inbuilt debris removal system should be regularly checked and maintained to ensure proper clinical and predictable refractive results.
Literature
1.
go back to reference Szekrényesi C, Sándor GL, Gyenes A, Kiss H, Filkorn T, Nagy Z. Relationship between corneal surface temperature and air flow conditions during refractive laser eye surgery using three different excimer lasers. Orv Hetil. 2016;157(43):1717–21.CrossRef Szekrényesi C, Sándor GL, Gyenes A, Kiss H, Filkorn T, Nagy Z. Relationship between corneal surface temperature and air flow conditions during refractive laser eye surgery using three different excimer lasers. Orv Hetil. 2016;157(43):1717–21.CrossRef
2.
go back to reference Dantas PE, Martins CL, de Souza LB, Dantas MC. Do environmental factors influence excimer laser pulse fluence and efficacy? J Refract Surg. 2007;23(3):307–9.PubMed Dantas PE, Martins CL, de Souza LB, Dantas MC. Do environmental factors influence excimer laser pulse fluence and efficacy? J Refract Surg. 2007;23(3):307–9.PubMed
3.
go back to reference Maldonado-Codina C, Morgan PB, Efron N. Thermal consequences of photorefractive keratoectomy. Cornea. 2001;20:509–15.CrossRef Maldonado-Codina C, Morgan PB, Efron N. Thermal consequences of photorefractive keratoectomy. Cornea. 2001;20:509–15.CrossRef
4.
go back to reference Müller B, Boeck T, Hartmann C. Effect of excimer laser beam delivery and beam shaping on corneal sphericity in photorefractive keratectomy. J Cataract Refract Surg. 2004 Feb;30(2):464–70.CrossRef Müller B, Boeck T, Hartmann C. Effect of excimer laser beam delivery and beam shaping on corneal sphericity in photorefractive keratectomy. J Cataract Refract Surg. 2004 Feb;30(2):464–70.CrossRef
5.
go back to reference Bor Z, Hopp B, Rácz B, Szabó G, Ratkay I, Süveges I, Füst A, Mohay J. Plume emission, shock wave and surface wave formation during excimer laser ablation of the cornea. Refract Corneal Surg. 1993 Mar-Apr;9(2 Suppl):S111–5.PubMed Bor Z, Hopp B, Rácz B, Szabó G, Ratkay I, Süveges I, Füst A, Mohay J. Plume emission, shock wave and surface wave formation during excimer laser ablation of the cornea. Refract Corneal Surg. 1993 Mar-Apr;9(2 Suppl):S111–5.PubMed
6.
go back to reference Hahn DW, Ediger MN, Pettit GH. Dynamics of ablation plume particles generated during excimer laser corneal ablation. Lasers Surg Med. 1995;16(4):384–9.CrossRef Hahn DW, Ediger MN, Pettit GH. Dynamics of ablation plume particles generated during excimer laser corneal ablation. Lasers Surg Med. 1995;16(4):384–9.CrossRef
7.
go back to reference Noack J, Tönnies R, Hohla K, Birngruber R, Vogel A. Influence of ablation plume dynamics on the formation of central islands in excimer laser photorefractive keratectomy. Ophthalmology. 1997 May;104(5):823–30.CrossRef Noack J, Tönnies R, Hohla K, Birngruber R, Vogel A. Influence of ablation plume dynamics on the formation of central islands in excimer laser photorefractive keratectomy. Ophthalmology. 1997 May;104(5):823–30.CrossRef
8.
go back to reference Dorronsoro C, Siegel J, Remon L, Marcos S. Suitability of Filofocon A and PMMA for experimental models in excimer laser ablation refractive surgery. Opt Express. 2008;16(25):20955–67.CrossRef Dorronsoro C, Siegel J, Remon L, Marcos S. Suitability of Filofocon A and PMMA for experimental models in excimer laser ablation refractive surgery. Opt Express. 2008;16(25):20955–67.CrossRef
10.
go back to reference Gottsch JD, Rencs EV, Cambier JL. Excimer laser calibration system. J Refract Surg. 1996;12(3):401–11.PubMed Gottsch JD, Rencs EV, Cambier JL. Excimer laser calibration system. J Refract Surg. 1996;12(3):401–11.PubMed
11.
go back to reference Doga AV, Shpak AA, Sugrobov VA. Smoothness of ablation on polymethylmethacrylate plates with four scanning excimer lasers. J Refract Surg. 2004;20(5 suppl):730–3. Doga AV, Shpak AA, Sugrobov VA. Smoothness of ablation on polymethylmethacrylate plates with four scanning excimer lasers. J Refract Surg. 2004;20(5 suppl):730–3.
12.
go back to reference Naroo SA, Charman WN. Surface roughness after excimer laser ablation using a PMMA model: profilometry and effects on vision. J Refract Surg. 2005;21(3):260–8.PubMed Naroo SA, Charman WN. Surface roughness after excimer laser ablation using a PMMA model: profilometry and effects on vision. J Refract Surg. 2005;21(3):260–8.PubMed
13.
go back to reference Wernli J, Schumacher S, Wuellner C, et al. Initial surface temperature of PMMA plates used for daily laser calibration to control the predictability of corneal refractive surgery. J Refract Surg. 2012;28(9):639–44.CrossRef Wernli J, Schumacher S, Wuellner C, et al. Initial surface temperature of PMMA plates used for daily laser calibration to control the predictability of corneal refractive surgery. J Refract Surg. 2012;28(9):639–44.CrossRef
14.
go back to reference Arba-Mosquera S, Shraiki M. Analysis of the PMMA and cornea temperature rise during excimer laser ablation. J Mod Opt. 2010;57(5):400–7.CrossRef Arba-Mosquera S, Shraiki M. Analysis of the PMMA and cornea temperature rise during excimer laser ablation. J Mod Opt. 2010;57(5):400–7.CrossRef
17.
go back to reference Kim JM, Kim JC, Park WC, Seo JS, Chang HR. Effect of thermal preconditioning before excimer laser photoablation. J Korean Med Sci. 2004 Jun;19(3):437–46.CrossRef Kim JM, Kim JC, Park WC, Seo JS, Chang HR. Effect of thermal preconditioning before excimer laser photoablation. J Korean Med Sci. 2004 Jun;19(3):437–46.CrossRef
Metadata
Title
Shielding effect of the smoke plume by the ablation of excimer lasers
Authors
Csaba Szekrényesi
Huba Kiss
Tamás Filkorn
Zoltán Zsolt Nagy
Publication date
01-12-2018
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2018
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-018-0942-8

Other articles of this Issue 1/2018

BMC Ophthalmology 1/2018 Go to the issue