Semi-robotic 6 degree of freedom positioning for intracranial high precision radiotherapy; first phantom and clinical results

Authors: Jürgen Wilbert, Matthias Guckenberger, Bülent Polat, Otto Sauer, Michael Vogele, Michael Flentje, Reinhart A Sweeney

Published in: Radiation Oncology | Issue 1/2010

Login to get access

Abstract

Background

To introduce a novel method of patient positioning for high precision intracranial radiotherapy.

Methods

An infrared(IR)-array, reproducibly attached to the patient via a vacuum-mouthpiece(vMP) and connected to the table via a 6 degree-of-freedom(DoF) mechanical arm serves as positioning and fixation system. After IR-based manual prepositioning to rough treatment position and fixation of the mechanical arm, a cone-beam CT(CBCT) is performed. A robotic 6 DoF treatment couch (HexaPOD™) then automatically corrects all remaining translations and rotations. This absolute position of infrared markers at the first fraction acts as reference for the following fractions where patients are manually prepositioned to within ± 2 mm and ± 2° of this IR reference position prior to final HexaPOD-based correction; consequently CBCT imaging is only required once at the first treatment fraction.
The preclinical feasibility and attainable repositioning accuracy of this method was evaluated on a phantom and human volunteers as was the clinical efficacy on 7 pilot study patients.

Results

Phantom and volunteer manual IR-based prepositioning to within ± 2 mm and ± 2° in 6DoF was possible within a mean(± SD) of 90 ± 31 and 56 ± 22 seconds respectively. Mean phantom translational and rotational precision after 6 DoF corrections by the HexaPOD was 0.2 ± 0.2 mm and 0.7 ± 0.8° respectively. For the actual patient collective, the mean 3D vector for inter-treatment repositioning accuracy (n = 102) was 1.6 ± 0.8 mm while intra-fraction movement (n = 110) was 0.6 ± 0.4 mm.

Conclusions

This novel semi-automatic 6DoF IR-based system has been shown to compare favourably with existing non-invasive intracranial repeat fixation systems with respect to handling, reproducibility and, more importantly, intra-fraction rigidity. Some advantages are full cranial positioning flexibility for single and fractionated IGRT treatments and possibly increased patient comfort.
Appendix
Available only for authorised users
Literature
1.
go back to reference Groh BA, Siewerdsen JH, Drake DG, Wong JW, Jaffray DA: A performance comparison of flat-panel imager-based MV and kV cone-beam CT. Med Phys 2002, 29: 967-975. 10.1118/1.1477234CrossRefPubMed Groh BA, Siewerdsen JH, Drake DG, Wong JW, Jaffray DA: A performance comparison of flat-panel imager-based MV and kV cone-beam CT. Med Phys 2002, 29: 967-975. 10.1118/1.1477234CrossRefPubMed
2.
go back to reference Thilmann C, Nill S, Tucking T, Hoss A, Hesse B, Dietrich L, Bendl R, Rhein B, Haering P, Thieke C, et al.: Correction of patient positioning errors based on in-line cone beam CTs: clinical implementation and first experiences. Radiat Oncol 2006, 1: 16. 10.1186/1748-717X-1-16PubMedCentralCrossRefPubMed Thilmann C, Nill S, Tucking T, Hoss A, Hesse B, Dietrich L, Bendl R, Rhein B, Haering P, Thieke C, et al.: Correction of patient positioning errors based on in-line cone beam CTs: clinical implementation and first experiences. Radiat Oncol 2006, 1: 16. 10.1186/1748-717X-1-16PubMedCentralCrossRefPubMed
3.
go back to reference Polat B, Guenther I, Wilbert J, Goebel J, Sweeney RA, Flentje M, Guckenberger M: Intra-fractional uncertainties in image-guided intensity-modulated radiotherapy (IMRT) of prostate cancer. Strahlenther Onkol 2008, 184: 668-673. 10.1007/s00066-008-1875-6CrossRefPubMed Polat B, Guenther I, Wilbert J, Goebel J, Sweeney RA, Flentje M, Guckenberger M: Intra-fractional uncertainties in image-guided intensity-modulated radiotherapy (IMRT) of prostate cancer. Strahlenther Onkol 2008, 184: 668-673. 10.1007/s00066-008-1875-6CrossRefPubMed
4.
go back to reference Linthout N, Verellen D, Tournel K, Storme G: Six dimensional analysis with daily stereoscopic x-ray imaging of intrafraction patient motion in head and neck treatments using five points fixation masks. Med Phys 2006, 33: 504-513. 10.1118/1.2165417CrossRefPubMed Linthout N, Verellen D, Tournel K, Storme G: Six dimensional analysis with daily stereoscopic x-ray imaging of intrafraction patient motion in head and neck treatments using five points fixation masks. Med Phys 2006, 33: 504-513. 10.1118/1.2165417CrossRefPubMed
5.
go back to reference Yan H, Yin FF, Kim JH: A phantom study on the positioning accuracy of the Novalis Body system. Med Phys 2003, 30: 3052-3060. 10.1118/1.1626122CrossRefPubMed Yan H, Yin FF, Kim JH: A phantom study on the positioning accuracy of the Novalis Body system. Med Phys 2003, 30: 3052-3060. 10.1118/1.1626122CrossRefPubMed
6.
go back to reference van Santvoort J, Wiggenraad R, Bos P: Positioning accuracy in stereotactic radiotherapy using a mask system with added vacuum mouth piece and stereoscopic X-ray positioning. Int J Radiat Oncol Biol Phys 2008, 72: 261-267.CrossRefPubMed van Santvoort J, Wiggenraad R, Bos P: Positioning accuracy in stereotactic radiotherapy using a mask system with added vacuum mouth piece and stereoscopic X-ray positioning. Int J Radiat Oncol Biol Phys 2008, 72: 261-267.CrossRefPubMed
7.
go back to reference D'Souza WD, McAvoy TJ: An analysis of the treatment couch and control system dynamics for respiration-induced motion compensation. Med Phys 2006, 33: 4701-4079. 10.1118/1.2372218CrossRefPubMed D'Souza WD, McAvoy TJ: An analysis of the treatment couch and control system dynamics for respiration-induced motion compensation. Med Phys 2006, 33: 4701-4079. 10.1118/1.2372218CrossRefPubMed
8.
go back to reference Guckenberger M, Meyer J, Wilbert J, Baier K, Sauer O, Flentje M: Precision of image-guided radiotherapy (IGRT) in six degrees of freedom and limitations in clinical practice. Strahlenther Onkol 2007, 183: 307-313. 10.1007/s00066-007-1695-0CrossRefPubMed Guckenberger M, Meyer J, Wilbert J, Baier K, Sauer O, Flentje M: Precision of image-guided radiotherapy (IGRT) in six degrees of freedom and limitations in clinical practice. Strahlenther Onkol 2007, 183: 307-313. 10.1007/s00066-007-1695-0CrossRefPubMed
9.
go back to reference Wilbert J, Meyer J, Baier K, Guckenberger M, Herrmann C, Hess R, Janka C, Ma L, Mersebach T, Richter A, et al.: Tumor tracking and motion compensation with an adaptive tumor tracking system (ATTS): system description and prototype testing. Med Phys 2008, 35: 3911-3921. 10.1118/1.2964090CrossRefPubMed Wilbert J, Meyer J, Baier K, Guckenberger M, Herrmann C, Hess R, Janka C, Ma L, Mersebach T, Richter A, et al.: Tumor tracking and motion compensation with an adaptive tumor tracking system (ATTS): system description and prototype testing. Med Phys 2008, 35: 3911-3921. 10.1118/1.2964090CrossRefPubMed
10.
go back to reference Meyer J, Wilbert J, Baier K, Guckenberger M, Richter A, Sauer O, Flentje M: Positioning accuracy of cone-beam computed tomography in combination with a HexaPOD robot treatment table. Int J Radiat Oncol Biol Phys 2007, 67: 1220-1228.CrossRefPubMed Meyer J, Wilbert J, Baier K, Guckenberger M, Richter A, Sauer O, Flentje M: Positioning accuracy of cone-beam computed tomography in combination with a HexaPOD robot treatment table. Int J Radiat Oncol Biol Phys 2007, 67: 1220-1228.CrossRefPubMed
11.
go back to reference Breneman JC, Steinmetz R, Smith A, Lamba M, Warnick RE: Frameless image-guided intracranial stereotactic radiosurgery: clinical outcomes for brain metastases. Int J Radiat Oncol Biol Phys 2009, 74: 702-706.CrossRefPubMed Breneman JC, Steinmetz R, Smith A, Lamba M, Warnick RE: Frameless image-guided intracranial stereotactic radiosurgery: clinical outcomes for brain metastases. Int J Radiat Oncol Biol Phys 2009, 74: 702-706.CrossRefPubMed
12.
go back to reference Lamba M, Breneman JC, Warnick RE: Evaluation of image-guided positioning for frameless intracranial radiosurgery. Int J Radiat Oncol Biol Phys 2009, 74: 913-919.CrossRefPubMed Lamba M, Breneman JC, Warnick RE: Evaluation of image-guided positioning for frameless intracranial radiosurgery. Int J Radiat Oncol Biol Phys 2009, 74: 913-919.CrossRefPubMed
13.
go back to reference Boda-Heggemann J, Walter C, Rahn A, Wertz H, Loeb I, Lohr F, Wenz F: Repositioning accuracy of two different mask systems-3D revisited: comparison using true 3D/3D matching with cone-beam CT. Int J Radiat Oncol Biol Phys 2006, 66: 1568-1575.CrossRefPubMed Boda-Heggemann J, Walter C, Rahn A, Wertz H, Loeb I, Lohr F, Wenz F: Repositioning accuracy of two different mask systems-3D revisited: comparison using true 3D/3D matching with cone-beam CT. Int J Radiat Oncol Biol Phys 2006, 66: 1568-1575.CrossRefPubMed
14.
go back to reference Polat B, Wilbert J, Baier K, Flentje M, Guckenberger M: Nonrigid patient setup errors in the head-and-neck region. Strahlenther Onkol 2007, 183: 506-511. 10.1007/s00066-007-1747-5CrossRefPubMed Polat B, Wilbert J, Baier K, Flentje M, Guckenberger M: Nonrigid patient setup errors in the head-and-neck region. Strahlenther Onkol 2007, 183: 506-511. 10.1007/s00066-007-1747-5CrossRefPubMed
15.
go back to reference Guckenberger M, Meyer J, Wilbert J, Baier K, Sauer O, Flentje M: Precision of image-guided radiotherapy (IGRT) in six degrees of freedom and limitations in clinical practice. Strahlenther Onkol 2007, 183: 307-313. 10.1007/s00066-007-1695-0CrossRefPubMed Guckenberger M, Meyer J, Wilbert J, Baier K, Sauer O, Flentje M: Precision of image-guided radiotherapy (IGRT) in six degrees of freedom and limitations in clinical practice. Strahlenther Onkol 2007, 183: 307-313. 10.1007/s00066-007-1695-0CrossRefPubMed
16.
go back to reference Sterzing F, Welzel T, Sroka-Perez G, Schubert K, Debus J, Herfarth KK: Reirradiation of multiple brain metastases with helical tomotherapy. A multifocal simultaneous integrated boost for eight or more lesions. Strahlenther Onkol 2009, 185: 89-93. 10.1007/s00066-009-1971-2CrossRefPubMed Sterzing F, Welzel T, Sroka-Perez G, Schubert K, Debus J, Herfarth KK: Reirradiation of multiple brain metastases with helical tomotherapy. A multifocal simultaneous integrated boost for eight or more lesions. Strahlenther Onkol 2009, 185: 89-93. 10.1007/s00066-009-1971-2CrossRefPubMed
17.
go back to reference Murphy MJ, Chang SD, Gibbs IC, Le QT, Hai J, Kim D, Martin DP, Adler JR Jr: Patterns of patient movement during frameless image-guided radiosurgery. Int J Radiat Oncol Biol Phys 2003, 55: 1400-1408.CrossRefPubMed Murphy MJ, Chang SD, Gibbs IC, Le QT, Hai J, Kim D, Martin DP, Adler JR Jr: Patterns of patient movement during frameless image-guided radiosurgery. Int J Radiat Oncol Biol Phys 2003, 55: 1400-1408.CrossRefPubMed
18.
go back to reference Hoogeman MS, Nuyttens JJ, Levendag PC, Heijmen BJ: Time dependence of intrafraction patient motion assessed by repeat stereoscopic imaging. Int J Radiat Oncol Biol Phys 2008, 70: 609-618.CrossRefPubMed Hoogeman MS, Nuyttens JJ, Levendag PC, Heijmen BJ: Time dependence of intrafraction patient motion assessed by repeat stereoscopic imaging. Int J Radiat Oncol Biol Phys 2008, 70: 609-618.CrossRefPubMed
19.
go back to reference Sweeney RA, Bale R, Auberger T, Vogele M, Foerster S, Nevinny-Stickel M, Lukas P: A simple and non-invasive vacuum mouthpiece-based head fixation system for high precision radiotherapy. Strahlenther Onkol 2001, 177: 43-47. 10.1007/PL00002357CrossRefPubMed Sweeney RA, Bale R, Auberger T, Vogele M, Foerster S, Nevinny-Stickel M, Lukas P: A simple and non-invasive vacuum mouthpiece-based head fixation system for high precision radiotherapy. Strahlenther Onkol 2001, 177: 43-47. 10.1007/PL00002357CrossRefPubMed
20.
go back to reference Olch AJ, Lavey RS: Reproducibility and treatment planning advantages of a carbon fiber relocatable head fixation system. Radiother Oncol 2002, 65: 165-168. 10.1016/S0167-8140(02)00282-7CrossRefPubMed Olch AJ, Lavey RS: Reproducibility and treatment planning advantages of a carbon fiber relocatable head fixation system. Radiother Oncol 2002, 65: 165-168. 10.1016/S0167-8140(02)00282-7CrossRefPubMed
21.
go back to reference van Herk M: Errors and margins in radiotherapy. Semin Radiat Oncol 2004, 14: 52-64. 10.1053/j.semradonc.2003.10.003CrossRefPubMed van Herk M: Errors and margins in radiotherapy. Semin Radiat Oncol 2004, 14: 52-64. 10.1053/j.semradonc.2003.10.003CrossRefPubMed
22.
go back to reference Kooy HM, Dunbar SF, Tarbell NJ, Mannarino E, Ferarro N, Shusterman S, Bellerive M, Finn L, McDonough CV, Loeffler JS: Adaptation and verification of the relocatable Gill-Thomas-Cosman frame in stereotactic radiotherapy. Int J Radiat Oncol Biol Phys 1994, 30: 685-691.CrossRefPubMed Kooy HM, Dunbar SF, Tarbell NJ, Mannarino E, Ferarro N, Shusterman S, Bellerive M, Finn L, McDonough CV, Loeffler JS: Adaptation and verification of the relocatable Gill-Thomas-Cosman frame in stereotactic radiotherapy. Int J Radiat Oncol Biol Phys 1994, 30: 685-691.CrossRefPubMed
23.
go back to reference Burtscher J, Sweeney R, Bale R, Eisner W, Twerdy K: Neuroendoscopy based on computer assisted adjustment of the endoscope holder in the laboratory. Minim Invasive Neurosurg 2003, 46: 208-214. 10.1055/s-2003-42348CrossRefPubMed Burtscher J, Sweeney R, Bale R, Eisner W, Twerdy K: Neuroendoscopy based on computer assisted adjustment of the endoscope holder in the laboratory. Minim Invasive Neurosurg 2003, 46: 208-214. 10.1055/s-2003-42348CrossRefPubMed
24.
go back to reference Maciunas RJ, Galloway RL Jr, Latimer JW: The application accuracy of stereotactic frames. Neurosurgery 1994, 35: 682-694. 10.1227/00006123-199410000-00015CrossRefPubMed Maciunas RJ, Galloway RL Jr, Latimer JW: The application accuracy of stereotactic frames. Neurosurgery 1994, 35: 682-694. 10.1227/00006123-199410000-00015CrossRefPubMed
25.
go back to reference Jakel O, Debus J: Selection of beam angles for radiotherapy of skull base tumours using charged particles. Phys Med Biol 2000, 45: 1229-1241. 10.1088/0031-9155/45/5/311CrossRefPubMed Jakel O, Debus J: Selection of beam angles for radiotherapy of skull base tumours using charged particles. Phys Med Biol 2000, 45: 1229-1241. 10.1088/0031-9155/45/5/311CrossRefPubMed
26.
go back to reference Mellenberg DE: Dose behind various immobilization and beam-modifying devices. Int J Radiat Oncol Biol Phys 1995, 32: 1193-1197.CrossRefPubMed Mellenberg DE: Dose behind various immobilization and beam-modifying devices. Int J Radiat Oncol Biol Phys 1995, 32: 1193-1197.CrossRefPubMed
27.
go back to reference Carl J, Vestergaard A: Skin damage probabilities using fixation materials in high-energy photon beams. Radiother Oncol 2000, 55: 191-198. 10.1016/S0167-8140(00)00177-8CrossRefPubMed Carl J, Vestergaard A: Skin damage probabilities using fixation materials in high-energy photon beams. Radiother Oncol 2000, 55: 191-198. 10.1016/S0167-8140(00)00177-8CrossRefPubMed
28.
go back to reference Martin A, Bale RJ, Vogele M, Gunkel AR, Thumfart WF, Freysinger W: Vogele-Bale-Hohner mouthpiece: registration device for frameless stereotactic surgery. Radiology 1998, 208: 261-265.CrossRefPubMed Martin A, Bale RJ, Vogele M, Gunkel AR, Thumfart WF, Freysinger W: Vogele-Bale-Hohner mouthpiece: registration device for frameless stereotactic surgery. Radiology 1998, 208: 261-265.CrossRefPubMed
29.
go back to reference Meeks SL, Bova FJ, Friedman WA, Buatti JM, Moore RD, Mendenhall WM: IRLED-based patient localization for linac radiosurgery. Int J Radiat Oncol Biol Phys 1998, 41: 433-439.CrossRefPubMed Meeks SL, Bova FJ, Friedman WA, Buatti JM, Moore RD, Mendenhall WM: IRLED-based patient localization for linac radiosurgery. Int J Radiat Oncol Biol Phys 1998, 41: 433-439.CrossRefPubMed
30.
go back to reference Furuse M, Aoki T, Takagi T, Takahashi JA, Ishikawa M: Frameless stereotactic radiosurgery with a bite-plate: our experience with brain metastases. Minim Invasive Neurosurg 2008, 51: 333-335. 10.1055/s-0028-1085445CrossRefPubMed Furuse M, Aoki T, Takagi T, Takahashi JA, Ishikawa M: Frameless stereotactic radiosurgery with a bite-plate: our experience with brain metastases. Minim Invasive Neurosurg 2008, 51: 333-335. 10.1055/s-0028-1085445CrossRefPubMed
31.
go back to reference Wiersma RD, Wen Z, Sadinski M, Farrey K, Yenice KM: Development of a frameless stereotactic radiosurgery system based on real-time 6D position monitoring and adaptive head motion compensation. Phys Med Biol 55: 389-401. 10.1088/0031-9155/55/2/004 Wiersma RD, Wen Z, Sadinski M, Farrey K, Yenice KM: Development of a frameless stereotactic radiosurgery system based on real-time 6D position monitoring and adaptive head motion compensation. Phys Med Biol 55: 389-401. 10.1088/0031-9155/55/2/004
32.
go back to reference Sweeney R, Bale R, Vogele M, Nevinny-Stickel M, Bluhm A, Auer T, Hessenberger G, Lukas P: Repositioning accuracy: comparison of a noninvasive head holder with thermoplastic mask for fractionated radiotherapy and a case report. Int J Radiat Oncol Biol Phys 1998, 41: 475-483.CrossRefPubMed Sweeney R, Bale R, Vogele M, Nevinny-Stickel M, Bluhm A, Auer T, Hessenberger G, Lukas P: Repositioning accuracy: comparison of a noninvasive head holder with thermoplastic mask for fractionated radiotherapy and a case report. Int J Radiat Oncol Biol Phys 1998, 41: 475-483.CrossRefPubMed
33.
go back to reference Fahrig A, Ganslandt O, Lambrecht U, Grabenbauer G, Kleinert G, Sauer R, Hamm K: Hypofractionated stereotactic radiotherapy for brain metastases--results from three different dose concepts. Strahlenther Onkol 2007, 183: 625-630. 10.1007/s00066-007-1714-1CrossRefPubMed Fahrig A, Ganslandt O, Lambrecht U, Grabenbauer G, Kleinert G, Sauer R, Hamm K: Hypofractionated stereotactic radiotherapy for brain metastases--results from three different dose concepts. Strahlenther Onkol 2007, 183: 625-630. 10.1007/s00066-007-1714-1CrossRefPubMed
34.
go back to reference Gilhuijs KG, Drukker K, Touw A, Ven PJ, van Herk M: Interactive three dimensional inspection of patient setup in radiation therapy using digital portal images and computed tomography data. Int J Radiat Oncol Biol Phys 1996, 34: 873-885.CrossRefPubMed Gilhuijs KG, Drukker K, Touw A, Ven PJ, van Herk M: Interactive three dimensional inspection of patient setup in radiation therapy using digital portal images and computed tomography data. Int J Radiat Oncol Biol Phys 1996, 34: 873-885.CrossRefPubMed
35.
go back to reference Amer A, Marchant T, Sykes J, Czajka J, Moore C: Imaging doses from the Elekta Synergy X-ray cone beam CT system. Br J Radiol 2007, 80: 476-482. 10.1259/bjr/80446730CrossRefPubMed Amer A, Marchant T, Sykes J, Czajka J, Moore C: Imaging doses from the Elekta Synergy X-ray cone beam CT system. Br J Radiol 2007, 80: 476-482. 10.1259/bjr/80446730CrossRefPubMed
36.
go back to reference Willner J, Flentje M, Bratengeier K: CT simulation in stereotactic brain radiotherapy--analysis of isocenter reproducibility with mask fixation. Radiother Oncol 1997, 45: 83-88. 10.1016/S0167-8140(97)00135-7CrossRefPubMed Willner J, Flentje M, Bratengeier K: CT simulation in stereotactic brain radiotherapy--analysis of isocenter reproducibility with mask fixation. Radiother Oncol 1997, 45: 83-88. 10.1016/S0167-8140(97)00135-7CrossRefPubMed
37.
go back to reference Masi L, Casamassima F, Polli C, Menichelli C, Bonucci I, Cavedon C: Cone beam CT image guidance for intracranial stereotactic treatments: comparison with a frame guided set-up. Int J Radiat Oncol Biol Phys 2008, 71: 926-933.CrossRefPubMed Masi L, Casamassima F, Polli C, Menichelli C, Bonucci I, Cavedon C: Cone beam CT image guidance for intracranial stereotactic treatments: comparison with a frame guided set-up. Int J Radiat Oncol Biol Phys 2008, 71: 926-933.CrossRefPubMed
38.
go back to reference Fuss M, Salter BJ, Cheek D, Sadeghi A, Hevezi JM, Herman TS: Repositioning accuracy of a commercially available thermoplastic mask system. Radiother Oncol 2004, 71: 339-345. 10.1016/j.radonc.2004.03.003CrossRefPubMed Fuss M, Salter BJ, Cheek D, Sadeghi A, Hevezi JM, Herman TS: Repositioning accuracy of a commercially available thermoplastic mask system. Radiother Oncol 2004, 71: 339-345. 10.1016/j.radonc.2004.03.003CrossRefPubMed
39.
go back to reference Salter BJ, Fuss M, Vollmer DG, Sadeghi A, Bogaev CA, Cheek DA, Herman TS, Hevezi JM: The TALON removable head frame system for stereotactic radiosurgery/radiotherapy: measurement of the repositioning accuracy. Int J Radiat Oncol Biol Phys 2001, 51: 555-562.CrossRefPubMed Salter BJ, Fuss M, Vollmer DG, Sadeghi A, Bogaev CA, Cheek DA, Herman TS, Hevezi JM: The TALON removable head frame system for stereotactic radiosurgery/radiotherapy: measurement of the repositioning accuracy. Int J Radiat Oncol Biol Phys 2001, 51: 555-562.CrossRefPubMed
Metadata
Title
Semi-robotic 6 degree of freedom positioning for intracranial high precision radiotherapy; first phantom and clinical results
Authors
Jürgen Wilbert
Matthias Guckenberger
Bülent Polat
Otto Sauer
Michael Vogele
Michael Flentje
Reinhart A Sweeney
Publication date
01-12-2010
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2010
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/1748-717X-5-42

Other articles of this Issue 1/2010

Radiation Oncology 1/2010 Go to the issue