Published in: Radiation Oncology 1/2012

Open Access 01-12-2012 | Research

Robotic-based carbon ion therapy and patient positioning in 6 degrees of freedom: setup accuracy of two standard immobilization devices used in carbon ion therapy and IMRT

Authors: Alexandra D Jensen, Marcus Winter, Sabine P Kuhn, Jürgen Debus, Olaf Nairz, Marc W Münter

Published in: Radiation Oncology | Issue 1/2012

Login to get access

Abstract

Purpose

To investigate repositioning accuracy in particle radiotherapy in 6 degrees of freedom (DOF) and intensity-modulated radiotherapy (IMRT, 3 DOF) for two immobilization devices (Scotchcast masks vs thermoplastic head masks) currently in use at our institution for fractionated radiation therapy in head and neck cancer patients.

Methods and materials

Position verifications in patients treated with carbon ion therapy and IMRT for head and neck malignancies were evaluated. Most patients received combined treatment regimen (IMRT plus carbon ion boost), immobilization was achieved with either Scotchcast or thermoplastic head masks. Position corrections in robotic-based carbon ion therapy allowing 6 DOF were compared to IMRT allowing corrections in 3 DOF for two standard immobilization devices. In total, 838 set-up controls of 38 patients were analyzed.

Results

Robotic-based position correction including correction of rotations was well tolerated and without discomfort. Standard deviations of translational components were between 0.5 and 0.8 mm for Scotchcast and 0.7 and 1.3 mm for thermoplastic masks in 6 DOF and 1.2 - 1.4 mm and 1.0 - 1.1 mm in 3 DOF respectively. Mean overall displacement vectors were between 2.1 mm (Scotchcast) and 2.9 mm (thermoplastic masks) in 6 DOF and 3.9 - 3.0 mm in 3 DOF respectively. Displacement vectors were lower when correction in 6 DOF was allowed as opposed to 3 DOF only, which was maintained at the traditional action level of > 3 mm for position correction in the pre-on-board imaging era.

Conclusion

Setup accuracy for both systems was within the expected range. Smaller shifts were required when 6 DOF were available for correction as opposed to 3 DOF. Where highest possible positioning accuracy is required, frequent image guidance is mandatory to achieve best possible plan delivery and maintenance of sharp gradients and optimal normal tissue sparing inherent in carbon ion therapy.
Appendix
Available only for authorised users
Literature
1.
go back to reference Meyer J, Wilbert J, Baier K, Guckenberger M, Richter A, Sauer O, Flentje M: Positioning accuracy of cone-beam computed tomography in combination with a HexaPOD robot treatment table. Int J Radiat Oncol Biol Phys 2007, 67: 1220-1228. 10.1016/j.ijrobp.2006.11.010CrossRefPubMed Meyer J, Wilbert J, Baier K, Guckenberger M, Richter A, Sauer O, Flentje M: Positioning accuracy of cone-beam computed tomography in combination with a HexaPOD robot treatment table. Int J Radiat Oncol Biol Phys 2007, 67: 1220-1228. 10.1016/j.ijrobp.2006.11.010CrossRefPubMed
2.
go back to reference Guckenberger M, Meyer J, Vordermark D, Baier K, Wilbert J, Flentje M: Magnitude and clinical relevance of translational and rotational patient setup errors: a cone-beam CT study. Int J Radiat Oncol Biol Phys 2006, 65: 934-942. 10.1016/j.ijrobp.2006.02.019CrossRefPubMed Guckenberger M, Meyer J, Vordermark D, Baier K, Wilbert J, Flentje M: Magnitude and clinical relevance of translational and rotational patient setup errors: a cone-beam CT study. Int J Radiat Oncol Biol Phys 2006, 65: 934-942. 10.1016/j.ijrobp.2006.02.019CrossRefPubMed
3.
go back to reference Allgower CE, Schreuder AN, Farr JB, Mascia AE: Experiences with an application of industrial robotics for accurate patient positioning in proton radiotherapy. Int J Med Robotics Comput Assist Surg 2007, 3: 72. 10.1002/rcs.128CrossRef Allgower CE, Schreuder AN, Farr JB, Mascia AE: Experiences with an application of industrial robotics for accurate patient positioning in proton radiotherapy. Int J Med Robotics Comput Assist Surg 2007, 3: 72. 10.1002/rcs.128CrossRef
4.
go back to reference Engelsman M, Rosenthal SJ, Michaud SL, Adams JA, Schneider RJ, Bradley SG, Flanz JB, Kooy HM: Intra- and interfractional patient motion for a variety of immobilization devices. Med Phys 2005, 32: 3468-3474. 10.1118/1.2089507CrossRefPubMed Engelsman M, Rosenthal SJ, Michaud SL, Adams JA, Schneider RJ, Bradley SG, Flanz JB, Kooy HM: Intra- and interfractional patient motion for a variety of immobilization devices. Med Phys 2005, 32: 3468-3474. 10.1118/1.2089507CrossRefPubMed
5.
go back to reference Kassaee A, Das IJ, Tochner Z, Rosenthal DI: Modification of Gill-Thomas-Cosman frame for extracranial head-and-neck stereotactic radiotherapy. Int J Radiat Oncol Biol Phys 2003, 57: 1192-1195. 10.1016/S0360-3016(03)00774-0CrossRefPubMed Kassaee A, Das IJ, Tochner Z, Rosenthal DI: Modification of Gill-Thomas-Cosman frame for extracranial head-and-neck stereotactic radiotherapy. Int J Radiat Oncol Biol Phys 2003, 57: 1192-1195. 10.1016/S0360-3016(03)00774-0CrossRefPubMed
6.
go back to reference Sweeney RA, Bale R, Auberger T, Vogele M, Foerster S, Nevinny-Stickel M, Lukas P: A Simple and non-invasive vacuum mouthpiece-based head fixation system for high-precision radiotherapy. Strahlenther Onkol 2001, 177: 43-47. 10.1007/PL00002357CrossRefPubMed Sweeney RA, Bale R, Auberger T, Vogele M, Foerster S, Nevinny-Stickel M, Lukas P: A Simple and non-invasive vacuum mouthpiece-based head fixation system for high-precision radiotherapy. Strahlenther Onkol 2001, 177: 43-47. 10.1007/PL00002357CrossRefPubMed
7.
go back to reference Han C, Radnay EH, Schultheiss TE: Evaluation of rotational and translational setup variations for brain tumor patients with mask immobilization system. Int J Radiat Oncol Biol Phys 2009, 75: S594. abstr. # 2921CrossRef Han C, Radnay EH, Schultheiss TE: Evaluation of rotational and translational setup variations for brain tumor patients with mask immobilization system. Int J Radiat Oncol Biol Phys 2009, 75: S594. abstr. # 2921CrossRef
8.
go back to reference Houweling AC, van der Meer S, van der Wal E, Terhaard CH, Raaijmakers CP: Improved immbilization using an individual head support in head-and-neck cancer patients. Radiother Oncol 2010, 96: 100. 10.1016/j.radonc.2010.04.014CrossRefPubMed Houweling AC, van der Meer S, van der Wal E, Terhaard CH, Raaijmakers CP: Improved immbilization using an individual head support in head-and-neck cancer patients. Radiother Oncol 2010, 96: 100. 10.1016/j.radonc.2010.04.014CrossRefPubMed
9.
go back to reference Gilbeau L, Octave-Prignot M, Loncol T, Renard L, Scalliet P, Grégoire V: Comparison of setup accuracy of three different thermoplastic masks fort he treatment of brain and head and neck tumors. Radiother Oncol 2001, 58: 155-162. 10.1016/S0167-8140(00)00280-2CrossRefPubMed Gilbeau L, Octave-Prignot M, Loncol T, Renard L, Scalliet P, Grégoire V: Comparison of setup accuracy of three different thermoplastic masks fort he treatment of brain and head and neck tumors. Radiother Oncol 2001, 58: 155-162. 10.1016/S0167-8140(00)00280-2CrossRefPubMed
10.
go back to reference Humphreys M, Guerrero-Urbano MT, Mubata C, Miles E, Harrington KJ, Bidmead M, Nutting CM: Assessment of a customised immobilisation system for head and neck IMRT using electronic portal imaging. Radiother Oncol 2005, 77: 39-44. 10.1016/j.radonc.2005.06.039CrossRefPubMed Humphreys M, Guerrero-Urbano MT, Mubata C, Miles E, Harrington KJ, Bidmead M, Nutting CM: Assessment of a customised immobilisation system for head and neck IMRT using electronic portal imaging. Radiother Oncol 2005, 77: 39-44. 10.1016/j.radonc.2005.06.039CrossRefPubMed
11.
go back to reference Fuss M, Salter BJ, Cheek D, Sadeghi A, Hevezi JM, Herman TS: Repositioning accuracy of a commercially available thermoplastic mask system. Radiother Oncol 2004, 71: 339-345. 10.1016/j.radonc.2004.03.003CrossRefPubMed Fuss M, Salter BJ, Cheek D, Sadeghi A, Hevezi JM, Herman TS: Repositioning accuracy of a commercially available thermoplastic mask system. Radiother Oncol 2004, 71: 339-345. 10.1016/j.radonc.2004.03.003CrossRefPubMed
12.
go back to reference Van Kranen S, van Beek S, Rasch Cvan Herk M, Sonke JJ: Setup uncertainties of anatomical sub-regions in head-and-neck cancer patients after offline CBCT guidance. Int J Radiat Oncol Biol Phys 2009, 73: 1566-1573. 10.1016/j.ijrobp.2008.11.035CrossRefPubMed Van Kranen S, van Beek S, Rasch Cvan Herk M, Sonke JJ: Setup uncertainties of anatomical sub-regions in head-and-neck cancer patients after offline CBCT guidance. Int J Radiat Oncol Biol Phys 2009, 73: 1566-1573. 10.1016/j.ijrobp.2008.11.035CrossRefPubMed
13.
go back to reference Velec M, Waldron JN, O'Sullivan B, Bayley A, Cummings B, Kim JJ, Ringash J, Breen SL, Lockwood GA, Dawson LA: Cone-beam CT assessment of interfraction and intrafraction setup error of two head-and-neck cancer thermoplastic masks. Int J Radiat Oncol Biol Phys 2010, 76: 949-955. 10.1016/j.ijrobp.2009.07.004CrossRefPubMed Velec M, Waldron JN, O'Sullivan B, Bayley A, Cummings B, Kim JJ, Ringash J, Breen SL, Lockwood GA, Dawson LA: Cone-beam CT assessment of interfraction and intrafraction setup error of two head-and-neck cancer thermoplastic masks. Int J Radiat Oncol Biol Phys 2010, 76: 949-955. 10.1016/j.ijrobp.2009.07.004CrossRefPubMed
14.
go back to reference Pehlivan B, Pichenot C, Castaing M, Auperin A, Lefkopoulos D, Arriagada R, Bourhis J: Interfractional set-up errors evaluation by daily electronic prtal imaging of IMRT in head and neck cancer patients. Acta Oncol 2009, 48: 440-445. 10.1080/02841860802400610CrossRefPubMed Pehlivan B, Pichenot C, Castaing M, Auperin A, Lefkopoulos D, Arriagada R, Bourhis J: Interfractional set-up errors evaluation by daily electronic prtal imaging of IMRT in head and neck cancer patients. Acta Oncol 2009, 48: 440-445. 10.1080/02841860802400610CrossRefPubMed
15.
go back to reference Li H, Zhu XR, Zhang L, Dong L, Tung S, Ahamad A, Chao KS, Morrison WH, Rosenthal DI, Schwartz DL, Mohan R, Garden AS: Comparison of 2D radiographic images and 3D cone beam computed tomography for positioning head-and-neck radiotherapy patients. Int J Radiat Oncol Biol Phys 2008, 71: 916-925. 10.1016/j.ijrobp.2008.01.008CrossRefPubMed Li H, Zhu XR, Zhang L, Dong L, Tung S, Ahamad A, Chao KS, Morrison WH, Rosenthal DI, Schwartz DL, Mohan R, Garden AS: Comparison of 2D radiographic images and 3D cone beam computed tomography for positioning head-and-neck radiotherapy patients. Int J Radiat Oncol Biol Phys 2008, 71: 916-925. 10.1016/j.ijrobp.2008.01.008CrossRefPubMed
16.
go back to reference Han C, Radnay EH, Schultheiss TE, Wong JYC: Evaluation of rotational and translational setup variations for brain tumor patients with mask immobilization system. Int J Radiat Oncol Biol Phys 2009, 75: S594. abstr. # 2921CrossRef Han C, Radnay EH, Schultheiss TE, Wong JYC: Evaluation of rotational and translational setup variations for brain tumor patients with mask immobilization system. Int J Radiat Oncol Biol Phys 2009, 75: S594. abstr. # 2921CrossRef
17.
go back to reference Karger CP, Jäkel O, Debus J, Kuhn S, Hartmann GH: Three-dimensional accuracy and interfractional reproducibility of patient fixation and positioning using a stereotactic head mask system. Int J Radiat Oncol Biol Phys 2001, 49: 1493-1504. 10.1016/S0360-3016(00)01562-5CrossRefPubMed Karger CP, Jäkel O, Debus J, Kuhn S, Hartmann GH: Three-dimensional accuracy and interfractional reproducibility of patient fixation and positioning using a stereotactic head mask system. Int J Radiat Oncol Biol Phys 2001, 49: 1493-1504. 10.1016/S0360-3016(00)01562-5CrossRefPubMed
18.
go back to reference Boda-Heggemann J, Walter C, Rahn A, Wertz H, Loeb I, Lohr F, Wenz F: Repositioning accuracy of two different mask systems - 3D revisited: comparison using true 3D/3D matching with cone-beam CT. Int J Radiat Oncol Biol Phys 2006, 66: 1568-1575. 10.1016/j.ijrobp.2006.08.054CrossRefPubMed Boda-Heggemann J, Walter C, Rahn A, Wertz H, Loeb I, Lohr F, Wenz F: Repositioning accuracy of two different mask systems - 3D revisited: comparison using true 3D/3D matching with cone-beam CT. Int J Radiat Oncol Biol Phys 2006, 66: 1568-1575. 10.1016/j.ijrobp.2006.08.054CrossRefPubMed
19.
go back to reference Rotondo RL, Sultanem K, Lavoie I, Skelly J, Raymond L: Comparison of repositioning accuracy of two commercially available immobilization systems for treatment of head-and-neck tumors using simulation computed tomography imaging. Int J Radiat Oncol Biol Phys 2008, 70: 1389-1396. 10.1016/j.ijrobp.2007.08.035CrossRefPubMed Rotondo RL, Sultanem K, Lavoie I, Skelly J, Raymond L: Comparison of repositioning accuracy of two commercially available immobilization systems for treatment of head-and-neck tumors using simulation computed tomography imaging. Int J Radiat Oncol Biol Phys 2008, 70: 1389-1396. 10.1016/j.ijrobp.2007.08.035CrossRefPubMed
20.
go back to reference Zeidan OA, Langen KM, Meeks SL, Manon RR, Wagner TH, Willoughby TR, Jenkins DW, Kupelian PA: Evaluation of image-guidance protocols in the treatment of head and neck cancers. Int J Radiat Oncol Biol Phys 2007, 67: 670-677. 10.1016/j.ijrobp.2006.09.040CrossRefPubMed Zeidan OA, Langen KM, Meeks SL, Manon RR, Wagner TH, Willoughby TR, Jenkins DW, Kupelian PA: Evaluation of image-guidance protocols in the treatment of head and neck cancers. Int J Radiat Oncol Biol Phys 2007, 67: 670-677. 10.1016/j.ijrobp.2006.09.040CrossRefPubMed
21.
go back to reference Haberer T, Becher W, Schardt D, Kraft G: Magnetic scanning system for heavy ion therapy. Nucl Instr Meth Phys Res 1993, 330: 296-305. 10.1016/0168-9002(93)91335-KCrossRef Haberer T, Becher W, Schardt D, Kraft G: Magnetic scanning system for heavy ion therapy. Nucl Instr Meth Phys Res 1993, 330: 296-305. 10.1016/0168-9002(93)91335-KCrossRef
22.
go back to reference Bentel GC, Marks LB, Hendren K, Brizel DM: Comparison of two head and neck immobilization devices. Int J Radiat Oncol Biol Phys 1997, 38: 867-873. 10.1016/S0360-3016(97)00075-8CrossRefPubMed Bentel GC, Marks LB, Hendren K, Brizel DM: Comparison of two head and neck immobilization devices. Int J Radiat Oncol Biol Phys 1997, 38: 867-873. 10.1016/S0360-3016(97)00075-8CrossRefPubMed
23.
go back to reference Van Herk M: Errors and margins in radiotherapy. Semin Radiat Oncol 2004, 14: 52-64. 10.1053/j.semradonc.2003.10.003CrossRefPubMed Van Herk M: Errors and margins in radiotherapy. Semin Radiat Oncol 2004, 14: 52-64. 10.1053/j.semradonc.2003.10.003CrossRefPubMed
24.
go back to reference Hurkmans CW, Remeijer P, Lebesque JV, Mijnheer BJ: Set-up verification using portal imaging: review of current clinical practice. Radiother Oncol 2001, 58: 105-120. 10.1016/S0167-8140(00)00260-7CrossRefPubMed Hurkmans CW, Remeijer P, Lebesque JV, Mijnheer BJ: Set-up verification using portal imaging: review of current clinical practice. Radiother Oncol 2001, 58: 105-120. 10.1016/S0167-8140(00)00260-7CrossRefPubMed
25.
go back to reference Spadea MF, Baroni G, Riboldi M, Luraschi R, Tagaste B, Garibaldi C, Catalano G, Orecchia R, Pedotti A: Benefits of six degrees of freedom for optically driven patient set-up corrections in SBRT. Technol Cancer res Treat 2008, 7: 187-195.CrossRefPubMed Spadea MF, Baroni G, Riboldi M, Luraschi R, Tagaste B, Garibaldi C, Catalano G, Orecchia R, Pedotti A: Benefits of six degrees of freedom for optically driven patient set-up corrections in SBRT. Technol Cancer res Treat 2008, 7: 187-195.CrossRefPubMed
26.
go back to reference Murphy MJ: Image-guided patient positioning: if one cannot correct for rotational offsets in external-beam radiotherapy setup, how should rotational offsets be managed? Med Phys 2007, 34: 1880-1883. 10.1118/1.2731485CrossRefPubMed Murphy MJ: Image-guided patient positioning: if one cannot correct for rotational offsets in external-beam radiotherapy setup, how should rotational offsets be managed? Med Phys 2007, 34: 1880-1883. 10.1118/1.2731485CrossRefPubMed
Metadata
Title
Robotic-based carbon ion therapy and patient positioning in 6 degrees of freedom: setup accuracy of two standard immobilization devices used in carbon ion therapy and IMRT
Authors
Alexandra D Jensen
Marcus Winter
Sabine P Kuhn
Jürgen Debus
Olaf Nairz
Marc W Münter
Publication date
01-12-2012
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2012
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/1748-717X-7-51

Other articles of this Issue 1/2012

Radiation Oncology 1/2012 Go to the issue