Reproducibility of the mfERG between instruments

Authors: Wendy W. Harrison, Marcus A. Bearse Jr., Jason S. Ng, Shirin Barez, Marilyn E. Schneck, Anthony J. Adams

Published in: Documenta Ophthalmologica | Issue 1/2009

Login to get access

Abstract

Purpose First, to examine both the reproducibility of the multifocal electroretinogram (mfERG) recorded on different versions of the same instrument, and the repeatability of the mfERG recorded on a single instrument using two different amplifiers. Second, to demonstrate a means by which multicenter and longitudinal studies that use more than one recording instrument can compare and combine data effectively. Methods Three different amplifiers and two mfERG setups, one using VERIS™ 4.3 software (mfERG1) and another using VERIS™ Pro 5.2 software (mfERG2), were evaluated. A total of 73 subjects with normal vision were tested in three groups. Group 1 (n = 42) was recorded using two amplifiers in parallel on mfERG1. Group 2 (n = 52) was recorded on mfERG2 using a single amplifier. Group 3 was a subgroup of 21 subjects from groups 1 and 2 that were tested sequentially on both instruments. A fourth group of 26 subjects with diabetes were also recorded using the two parallel amplifiers on mfERG1. P1 implicit times and N1-P1 amplitudes of the 103 local first order mfERGs were measured, and the differences between the instruments and amplifiers were evaluated as raw scores and Z-scores based on normative data. Measurements of individual responses and measurements averaged over the 103 responses were analyzed. Results Simultaneous recordings made on mfERG1 with the two different amplifiers showed differences in implicit times but similar amplitudes. There was a mean implicit time difference of 2.5 ms between the amplifiers but conversion to Z-scores improved their agreement. Recordings made on different days with the two instruments produced similar but more variable results, with amplitudes differing between them more than implicit times. For local response implicit times, the 95% confidence interval of the difference between instruments was approximately ±1 Z-score (±0.9 ms) in either direction. For local response amplitude, it was approximately ±1.6 Z-scores (±0.3 μV). Conclusions Different amplifiers can yield quite different mfERG P1 implicit times, even with identical band-pass settings. However, the reproducibility of mfERG Z-scores across recording instrumentation is relatively high. Comparison of data across systems and laboratories, necessary for multicenter or longitudinal investigations, is facilitated if raw data are converted into Z-scores based on normative data.
Literature
2.
go back to reference Janaky M, Palffy A, Deak A, Szilagyi M, Benedek G (2007) Multifocal ERG reveals several patterns of cone degeneration in retinitis pigmentosa with concentric narrowing of the visual field. Invest Ophthalmol Vis Sci 48:383–389. doi:10.1167/iovs.06-0661 PubMedCrossRef Janaky M, Palffy A, Deak A, Szilagyi M, Benedek G (2007) Multifocal ERG reveals several patterns of cone degeneration in retinitis pigmentosa with concentric narrowing of the visual field. Invest Ophthalmol Vis Sci 48:383–389. doi:10.​1167/​iovs.​06-0661 PubMedCrossRef
3.
go back to reference Han Y, Bearse MA Jr, Schneck ME, Barez S, Jacobsen CH, Adams AJ (2004) Multifocal electroretinogram delays predict sites of subsequent diabetic retinopathy. Invest Ophthalmol Vis Sci 45:948–954. doi:10.1167/iovs.03-1101 PubMedCrossRef Han Y, Bearse MA Jr, Schneck ME, Barez S, Jacobsen CH, Adams AJ (2004) Multifocal electroretinogram delays predict sites of subsequent diabetic retinopathy. Invest Ophthalmol Vis Sci 45:948–954. doi:10.​1167/​iovs.​03-1101 PubMedCrossRef
4.
go back to reference Han Y, Schneck ME, Bearse MA Jr, Barez S, Jacobsen CH, Jewell NP et al (2004) Formulation and evaluation of a predictive model to identify the sites of future diabetic retinopathy. Invest Ophthalmol Vis Sci 45:4106–4112. doi:10.1167/iovs.04-0405 PubMedCrossRef Han Y, Schneck ME, Bearse MA Jr, Barez S, Jacobsen CH, Jewell NP et al (2004) Formulation and evaluation of a predictive model to identify the sites of future diabetic retinopathy. Invest Ophthalmol Vis Sci 45:4106–4112. doi:10.​1167/​iovs.​04-0405 PubMedCrossRef
13.
go back to reference Gundogan FC, Sobaci G, Bayraktar MZ (2008) Intra-sessional and inter-sessional variability of multifocal electroretinogram. Doc Ophthalmol Gundogan FC, Sobaci G, Bayraktar MZ (2008) Intra-sessional and inter-sessional variability of multifocal electroretinogram. Doc Ophthalmol
17.
go back to reference Hood DC, Li J (1997) A technique for measuring individual multifocal ERG records. Trends Opt Photon 11:280–293 Hood DC, Li J (1997) A technique for measuring individual multifocal ERG records. Trends Opt Photon 11:280–293
19.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed
25.
go back to reference Chan HL, Siu AW (2003) Effect of optical defocus on multifocal ERG responses. Clin Exp Optom 86:317–322PubMedCrossRef Chan HL, Siu AW (2003) Effect of optical defocus on multifocal ERG responses. Clin Exp Optom 86:317–322PubMedCrossRef
27.
go back to reference Han Y, Bearse MA Jr, Schneck ME, Barez S, Jacobsen C, Adams AJ (2004) Towards optimal filtering of “standard” multifocal electroretinogram (mfERG) recordings: findings in normal and diabetic subjects. Br J Ophthalmol 88:543–550. doi:10.1136/bjo.2003.026625 PubMedCrossRef Han Y, Bearse MA Jr, Schneck ME, Barez S, Jacobsen C, Adams AJ (2004) Towards optimal filtering of “standard” multifocal electroretinogram (mfERG) recordings: findings in normal and diabetic subjects. Br J Ophthalmol 88:543–550. doi:10.​1136/​bjo.​2003.​026625 PubMedCrossRef
29.
go back to reference Fortune B, Schneck ME, Adams AJ (1999) Multifocal electroretinogram delays reveal local retinal dysfunction in early diabetic retinopathy. Invest Ophthalmol Vis Sci 40:2638–2651PubMed Fortune B, Schneck ME, Adams AJ (1999) Multifocal electroretinogram delays reveal local retinal dysfunction in early diabetic retinopathy. Invest Ophthalmol Vis Sci 40:2638–2651PubMed
Metadata
Title
Reproducibility of the mfERG between instruments
Authors
Wendy W. Harrison
Marcus A. Bearse Jr.
Jason S. Ng
Shirin Barez
Marilyn E. Schneck
Anthony J. Adams
Publication date
01-08-2009
Publisher
Springer-Verlag
Published in
Documenta Ophthalmologica / Issue 1/2009
Print ISSN: 0012-4486
Electronic ISSN: 1573-2622
DOI
https://doi.org/10.1007/s10633-009-9171-z

Other articles of this Issue 1/2009

Documenta Ophthalmologica 1/2009 Go to the issue