Published in: BMC Musculoskeletal Disorders 1/2013

Open Access 01-12-2013 | Research article

Reliability of the standard goniometry and diagrammatic recording of finger joint angles: a comparative study with healthy subjects and non-professional raters

Author: Valdas Macionis

Published in: BMC Musculoskeletal Disorders | Issue 1/2013

Login to get access

Abstract

Background

Diagrammatic recording of finger joint angles by using two criss-crossed paper strips can be a quick substitute to the standard goniometry. As a preliminary step toward clinical validation of the diagrammatic technique, the current study employed healthy subjects and non-professional raters to explore whether reliability estimates of the diagrammatic goniometry are comparable with those of the standard procedure.

Methods

The study included two procedurally different parts, which were replicated by assigning 24 medical students to act interchangeably as 12 subjects and 12 raters. A larger component of the study was designed to compare goniometers side-by-side in measurement of finger joint angles varying from subject to subject. In the rest of the study, the instruments were compared by parallel evaluations of joint angles similar for all subjects in a situation of simulated change of joint range of motion over time. The subjects used special guides to position the joints of their left ring finger at varying angles of flexion and extension. The obtained diagrams of joint angles were converted to numerical values by computerized measurements. The statistical approaches included calculation of appropriate intraclass correlation coefficients, standard errors of measurements, proportions of measurement differences of 5 or less degrees, and significant differences between paired observations.

Results

Reliability estimates were similar for both goniometers. Intra-rater and inter-rater intraclass correlation coefficients ranged from 0.69 to 0.93. The corresponding standard errors of measurements ranged from 2.4 to 4.9 degrees. Repeated measurements of a considerable number of raters fell within clinically non-meaningful 5 degrees of each other in proportions comparable with a criterion value of 0.95. Data collected with both instruments could be similarly interpreted in a simulated situation of change of joint range of motion over time.

Conclusions

The paper goniometer and the standard goniometer can be used interchangeably by non-professional raters for evaluation of normal finger joints. The obtained results warrant further research to assess clinical performance of the paper strip technique.
Appendix
Available only for authorised users
Literature
1.
go back to reference Robins RH: Hand assessment charts. J Hand Surg Br. 1986, 11: 287-298. 10.1016/0266-7681(86)90287-1.CrossRefPubMed Robins RH: Hand assessment charts. J Hand Surg Br. 1986, 11: 287-298. 10.1016/0266-7681(86)90287-1.CrossRefPubMed
2.
go back to reference Ellis B, Bruton A, Goddard JR: Joint angle measurement: a comparative study of the reliability of goniometry and wire tracing for the hand. Clin Rehabil. 1997, 11: 314-320. 10.1177/026921559701100408.CrossRefPubMed Ellis B, Bruton A, Goddard JR: Joint angle measurement: a comparative study of the reliability of goniometry and wire tracing for the hand. Clin Rehabil. 1997, 11: 314-320. 10.1177/026921559701100408.CrossRefPubMed
3.
go back to reference Macionis V: A technique for graphical recording of range of motion using an improvised paper goniometer. J Hand Ther. 2011, 24: 374-377. 10.1016/j.jht.2011.05.003.CrossRefPubMed Macionis V: A technique for graphical recording of range of motion using an improvised paper goniometer. J Hand Ther. 2011, 24: 374-377. 10.1016/j.jht.2011.05.003.CrossRefPubMed
4.
go back to reference De Vet HC, Terwee CB, Knol DL, Bouter LM: When to use agreement versus reliability measures. J Clin Epidemiol. 2006, 59: 1033-1039. 10.1016/j.jclinepi.2005.10.015.CrossRefPubMed De Vet HC, Terwee CB, Knol DL, Bouter LM: When to use agreement versus reliability measures. J Clin Epidemiol. 2006, 59: 1033-1039. 10.1016/j.jclinepi.2005.10.015.CrossRefPubMed
5.
go back to reference Kottner J, Audigé L, Brorson S, Donner A, Gajewski BJ, Hróbjartsson A, Roberts C, Shoukri M, Streiner DL: Guidelines for Reporting Reliability and Agreement Studies (GRRAS) were proposed. J Clin Epidemiol. 2011, 64: 96-106. 10.1016/j.jclinepi.2010.03.002.CrossRefPubMed Kottner J, Audigé L, Brorson S, Donner A, Gajewski BJ, Hróbjartsson A, Roberts C, Shoukri M, Streiner DL: Guidelines for Reporting Reliability and Agreement Studies (GRRAS) were proposed. J Clin Epidemiol. 2011, 64: 96-106. 10.1016/j.jclinepi.2010.03.002.CrossRefPubMed
6.
go back to reference Hamilton GF, Lachenbruch PA: Reliability of goniometers in assessing finger joint angle. Phys Ther. 1969, 49: 465-469.PubMed Hamilton GF, Lachenbruch PA: Reliability of goniometers in assessing finger joint angle. Phys Ther. 1969, 49: 465-469.PubMed
7.
go back to reference Giudice ML: Effects of continuous passive motion and elevation on hand oedema. Am J Occup Ther. 1990, 44: 914-921. 10.5014/ajot.44.10.914.CrossRefPubMed Giudice ML: Effects of continuous passive motion and elevation on hand oedema. Am J Occup Ther. 1990, 44: 914-921. 10.5014/ajot.44.10.914.CrossRefPubMed
8.
go back to reference Breger-Lee D, Voelker ET, Giurintano D, Novick A, Browder L: Reliability of torque range of motion: a preliminary study. J Hand Ther. 1993, 6: 29-34. 10.1016/S0894-1130(12)80178-5.CrossRefPubMed Breger-Lee D, Voelker ET, Giurintano D, Novick A, Browder L: Reliability of torque range of motion: a preliminary study. J Hand Ther. 1993, 6: 29-34. 10.1016/S0894-1130(12)80178-5.CrossRefPubMed
9.
go back to reference Dijkstra PU, de Bont LG, van der Weele LT, Boering G: Joint mobility measurements: reliability of a standardized method. Cranio. 1994, 12: 52-57.CrossRefPubMed Dijkstra PU, de Bont LG, van der Weele LT, Boering G: Joint mobility measurements: reliability of a standardized method. Cranio. 1994, 12: 52-57.CrossRefPubMed
10.
go back to reference Weiss PL, August S, Peters G, Sampalis J: Using the Exos Handmaster to measure digital range of motion: reliability and validity. Med Eng Phys. 1994, 16: 323-328. 10.1016/1350-4533(94)90059-0.CrossRefPubMed Weiss PL, August S, Peters G, Sampalis J: Using the Exos Handmaster to measure digital range of motion: reliability and validity. Med Eng Phys. 1994, 16: 323-328. 10.1016/1350-4533(94)90059-0.CrossRefPubMed
11.
go back to reference Flowers KR, LaStayo P: Effect of total end range time on improving passive range of motion. J Hand Ther. 1994, 7: 150-157. 10.1016/S0894-1130(12)80056-1.CrossRefPubMed Flowers KR, LaStayo P: Effect of total end range time on improving passive range of motion. J Hand Ther. 1994, 7: 150-157. 10.1016/S0894-1130(12)80056-1.CrossRefPubMed
12.
go back to reference Chiu HY: A method of two-dimensional measurement for evaluating finger motion impairment. A description of the method and comparison with angular measurement. J Hand Surg Br. 1995, 20: 691-695. 10.1016/S0266-7681(05)80138-X.CrossRefPubMed Chiu HY: A method of two-dimensional measurement for evaluating finger motion impairment. A description of the method and comparison with angular measurement. J Hand Surg Br. 1995, 20: 691-695. 10.1016/S0266-7681(05)80138-X.CrossRefPubMed
13.
go back to reference Goldsmith N, Juzl E: Inter-rater reliability of two trained raters using a goniometer for the measurement of finger joints. Br J Hand Ther. 1998, 3: 11-12. cited by Burr et al. [26] Goldsmith N, Juzl E: Inter-rater reliability of two trained raters using a goniometer for the measurement of finger joints. Br J Hand Ther. 1998, 3: 11-12. cited by Burr et al. [26]
14.
go back to reference Chiu HY, Su FC, Wang ST, Hsu HY: The motion analysis system and goniometry of the finger joints. J Hand Surg Br. 1998, 23: 788-791. 10.1016/S0266-7681(98)80098-3.CrossRefPubMed Chiu HY, Su FC, Wang ST, Hsu HY: The motion analysis system and goniometry of the finger joints. J Hand Surg Br. 1998, 23: 788-791. 10.1016/S0266-7681(98)80098-3.CrossRefPubMed
15.
go back to reference Bruton A, Ellis B, Goddard J: Comparison of visual estimation and goniometry for assessment of metacarpophalangeal joint angle. Physiotherapy. 1999, 85: 2011-2018. Bruton A, Ellis B, Goddard J: Comparison of visual estimation and goniometry for assessment of metacarpophalangeal joint angle. Physiotherapy. 1999, 85: 2011-2018.
16.
go back to reference Williams NW, Penrose JM, Caddy CM, Barnes E, Hose DR, Harley P: A goniometric glove for clinical hand assessment. Construction, calibration and validation. J Hand Surg Br. 2000, 25: 200-207.CrossRefPubMed Williams NW, Penrose JM, Caddy CM, Barnes E, Hose DR, Harley P: A goniometric glove for clinical hand assessment. Construction, calibration and validation. J Hand Surg Br. 2000, 25: 200-207.CrossRefPubMed
17.
go back to reference Brown A, Cramer LD, Eckhaus D, Schmidt J, Ware L, MacKenzie E: Validity and reliability of the Dexter hand evaluation and therapy system in hand-injured patients. J Hand Ther. 2000, 13: 37-45. 10.1016/S0894-1130(00)80051-4.CrossRefPubMed Brown A, Cramer LD, Eckhaus D, Schmidt J, Ware L, MacKenzie E: Validity and reliability of the Dexter hand evaluation and therapy system in hand-injured patients. J Hand Ther. 2000, 13: 37-45. 10.1016/S0894-1130(00)80051-4.CrossRefPubMed
18.
go back to reference Catalano LW, Browne RH, Carter PR, Frobish AC, Ezaki M, Littler JW: The Littler line method and the area under a Gaussian curve: a new method of assessing digital range of motion. Hand Surg Am. 2001, 26: 23-30. 10.1053/jhsu.2001.21512.CrossRef Catalano LW, Browne RH, Carter PR, Frobish AC, Ezaki M, Littler JW: The Littler line method and the area under a Gaussian curve: a new method of assessing digital range of motion. Hand Surg Am. 2001, 26: 23-30. 10.1053/jhsu.2001.21512.CrossRef
19.
go back to reference Groth GN, VanDeven KM, Phillips EC, Ehretsman RL: Goniometry of the proximal and distal interphalangeal joints, Part II: placement preferences, interrater reliability, and concurrent validity. J Hand Ther. 2001, 14: 23-29. 10.1016/S0894-1130(01)80021-1.CrossRefPubMed Groth GN, VanDeven KM, Phillips EC, Ehretsman RL: Goniometry of the proximal and distal interphalangeal joints, Part II: placement preferences, interrater reliability, and concurrent validity. J Hand Ther. 2001, 14: 23-29. 10.1016/S0894-1130(01)80021-1.CrossRefPubMed
20.
go back to reference Lefevre-Colau MM, Poiraudeau S, Fermanian J, Mayoux-Benhamou MA, Bargy F, Revel M: Reliability of two goniometers in assessing rheumatoid finger mobility: Relationship between mobility and disability. Eura Medicophys. 2001, 37: 3-10. Lefevre-Colau MM, Poiraudeau S, Fermanian J, Mayoux-Benhamou MA, Bargy F, Revel M: Reliability of two goniometers in assessing rheumatoid finger mobility: Relationship between mobility and disability. Eura Medicophys. 2001, 37: 3-10.
21.
go back to reference Macdermid JC, Fox E, Richards RS, Roth JH: Validity of pulp-to-palm distance as a measure of finger flexion. J Hand Surg Br. 2001, 26: 432-435. 10.1054/jhsb.2001.0612.CrossRefPubMed Macdermid JC, Fox E, Richards RS, Roth JH: Validity of pulp-to-palm distance as a measure of finger flexion. J Hand Surg Br. 2001, 26: 432-435. 10.1054/jhsb.2001.0612.CrossRefPubMed
22.
go back to reference Georgeu GA, Mayfield S, Logan AM: Lateral digital photography with computer-aided goniometry versus standard goniometry for recording finger joint angles. J Hand Surg Br. 2002, 27: 184-186. 10.1054/jhsb.2001.0692.CrossRefPubMed Georgeu GA, Mayfield S, Logan AM: Lateral digital photography with computer-aided goniometry versus standard goniometry for recording finger joint angles. J Hand Surg Br. 2002, 27: 184-186. 10.1054/jhsb.2001.0692.CrossRefPubMed
23.
go back to reference Ellis B, Bruton A: A study to compare the reliability of composite finger flexion with goniometry for measurement of range of motion in the hand. Clin Rehabil. 2002, 16: 562-570. 10.1191/0269215502cr513oa.CrossRefPubMed Ellis B, Bruton A: A study to compare the reliability of composite finger flexion with goniometry for measurement of range of motion in the hand. Clin Rehabil. 2002, 16: 562-570. 10.1191/0269215502cr513oa.CrossRefPubMed
24.
go back to reference Rose V, Nduka CC, Pereira JA, Pickford MA, Belcher HJ: Visual estimation of finger angles: do we need goniometers?. J Hand Surg Br. 2002, 27: 382-384.CrossRefPubMed Rose V, Nduka CC, Pereira JA, Pickford MA, Belcher HJ: Visual estimation of finger angles: do we need goniometers?. J Hand Surg Br. 2002, 27: 382-384.CrossRefPubMed
25.
go back to reference Glasgow C, Wilton J, Tooth L: Optimal daily total end range time for contracture: resolution in hand splinting. J Hand Ther. 2003, 16: 207-218. 10.1016/S0894-1130(03)00036-X.CrossRefPubMed Glasgow C, Wilton J, Tooth L: Optimal daily total end range time for contracture: resolution in hand splinting. J Hand Ther. 2003, 16: 207-218. 10.1016/S0894-1130(03)00036-X.CrossRefPubMed
26.
go back to reference Burr N, Pratt AL, Stott D: Inter-rater and intra-rater reliability when measuring interphalangeal joints: comparison between three hand-held goniometers. Physiotherapy. 2003, 89: 641-652. 10.1016/S0031-9406(05)60097-1.CrossRef Burr N, Pratt AL, Stott D: Inter-rater and intra-rater reliability when measuring interphalangeal joints: comparison between three hand-held goniometers. Physiotherapy. 2003, 89: 641-652. 10.1016/S0031-9406(05)60097-1.CrossRef
27.
go back to reference Glasgow C, James M, O‘Sullivan J, Tooth LR: Measurement of joint stiffness in the hand: a preliminary investigation of the reliability and validity of torque angle curves. Br J Hand Ther. 2004, 9: 11-12. Glasgow C, James M, O‘Sullivan J, Tooth LR: Measurement of joint stiffness in the hand: a preliminary investigation of the reliability and validity of torque angle curves. Br J Hand Ther. 2004, 9: 11-12.
28.
go back to reference Pratt AL, Burr N, Stott D: An investigation into the degree of precision achieved by a team of hand therapists and surgeons using hand goniometry with a standardised protocol. Hand Ther. 2004, 9: 116-121. Pratt AL, Burr N, Stott D: An investigation into the degree of precision achieved by a team of hand therapists and surgeons using hand goniometry with a standardised protocol. Hand Ther. 2004, 9: 116-121.
29.
go back to reference Stam HJ, Ardon MS, den Ouden AC, Schreuders TA, Roebroeck ME: The compangle: a new goniometer for joint angle measurements of the hand. a technical note. Eura Medicophys. 2006, 42: 37-40.PubMed Stam HJ, Ardon MS, den Ouden AC, Schreuders TA, Roebroeck ME: The compangle: a new goniometer for joint angle measurements of the hand. a technical note. Eura Medicophys. 2006, 42: 37-40.PubMed
30.
go back to reference Cook JR, Baker NA, Cham R, Hale E, Redfern MS: Measurements of wrist and finger postures: a comparison of goniometric and motion capture techniques. J Appl Biomech. 2007, 23: 70-78.PubMed Cook JR, Baker NA, Cham R, Hale E, Redfern MS: Measurements of wrist and finger postures: a comparison of goniometric and motion capture techniques. J Appl Biomech. 2007, 23: 70-78.PubMed
31.
go back to reference Kato M, Echigo A, Ohta H, Ishiai S, Aoki M, Tsubota S, Uchiyama E: The accuracy of goniometric measurements of proximal interphalangeal joints in fresh cadavers: comparison between methods of measurement, types of goniometers, and fingers. J Hand Ther. 2007, 20: 12-18. 10.1197/j.jht.2006.11.015.CrossRefPubMed Kato M, Echigo A, Ohta H, Ishiai S, Aoki M, Tsubota S, Uchiyama E: The accuracy of goniometric measurements of proximal interphalangeal joints in fresh cadavers: comparison between methods of measurement, types of goniometers, and fingers. J Hand Ther. 2007, 20: 12-18. 10.1197/j.jht.2006.11.015.CrossRefPubMed
32.
go back to reference Smith RP, Dias JJ, Ullah A, Bhowal B: Visual and computer software-aided estimates of Dupuytren's contractures: correlation with clinical goniometric measurements. Ann R Coll Surg Engl. 2009, 91: 296-300.CrossRefPubMedPubMedCentral Smith RP, Dias JJ, Ullah A, Bhowal B: Visual and computer software-aided estimates of Dupuytren's contractures: correlation with clinical goniometric measurements. Ann R Coll Surg Engl. 2009, 91: 296-300.CrossRefPubMedPubMedCentral
33.
go back to reference Lewis E, Fors L, Tharion WJ: Interrater and intrarater reliability of finger goniometric measurements. Am J Occup Ther. 2010, 64: 555-561. 10.5014/ajot.2010.09028.CrossRefPubMed Lewis E, Fors L, Tharion WJ: Interrater and intrarater reliability of finger goniometric measurements. Am J Occup Ther. 2010, 64: 555-561. 10.5014/ajot.2010.09028.CrossRefPubMed
34.
go back to reference Torok KS, Baker NA, Lucas M, Domsic RT, Boudreau R, Medsger TA: Reliability and validity of the delta finger-to-palm (FTP), a new measure of finger range of motion in systemic sclerosis. Clin Exp Rheumatol. 2010, 28: S28-36.PubMedPubMedCentral Torok KS, Baker NA, Lucas M, Domsic RT, Boudreau R, Medsger TA: Reliability and validity of the delta finger-to-palm (FTP), a new measure of finger range of motion in systemic sclerosis. Clin Exp Rheumatol. 2010, 28: S28-36.PubMedPubMedCentral
35.
go back to reference Li K, Chen IM, Yeo SH, Lim CK: Development of finger-motion capturing device based on optical linear encoder. J Rehabil Res Dev. 2011, 48: 69-82. 10.1682/JRRD.2010.02.0013.CrossRefPubMed Li K, Chen IM, Yeo SH, Lim CK: Development of finger-motion capturing device based on optical linear encoder. J Rehabil Res Dev. 2011, 48: 69-82. 10.1682/JRRD.2010.02.0013.CrossRefPubMed
36.
go back to reference Oess NP, Wanek J, Curt A: Design and evaluation of a low-cost instrumented glove for hand function assessment. J Neuroeng Rehabil. 2012, 9: 2-10.1186/1743-0003-9-2.CrossRefPubMedPubMedCentral Oess NP, Wanek J, Curt A: Design and evaluation of a low-cost instrumented glove for hand function assessment. J Neuroeng Rehabil. 2012, 9: 2-10.1186/1743-0003-9-2.CrossRefPubMedPubMedCentral
37.
go back to reference Engstrand C, Krevers B, Kvist J: Interrater reliability in finger joint goniometer measurement in Dupuytren’s disease. Am J Occup Ther. 2012, 66: 98-103. 10.5014/ajot.2012.001925.CrossRefPubMed Engstrand C, Krevers B, Kvist J: Interrater reliability in finger joint goniometer measurement in Dupuytren’s disease. Am J Occup Ther. 2012, 66: 98-103. 10.5014/ajot.2012.001925.CrossRefPubMed
38.
go back to reference Portney LG, Watkins MP: Foundations of Clinical Research: Applications to Practice. 2008, NJ, Prentice Hall: Upper Saddle River, 3rd Portney LG, Watkins MP: Foundations of Clinical Research: Applications to Practice. 2008, NJ, Prentice Hall: Upper Saddle River, 3rd
39.
go back to reference Bruton A, Conway JH, Holgate ST: Reliability: What is it and how is it measured?. Physiotherapy. 2000, 86: 94-99. 10.1016/S0031-9406(05)61211-4.CrossRef Bruton A, Conway JH, Holgate ST: Reliability: What is it and how is it measured?. Physiotherapy. 2000, 86: 94-99. 10.1016/S0031-9406(05)61211-4.CrossRef
40.
go back to reference Weir JP: Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res. 2005, 19: 231-240.PubMed Weir JP: Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res. 2005, 19: 231-240.PubMed
41.
go back to reference Barnhart HX, Haber MJ, Lin LI: An overview on assessing agreement with continuous measurements. J Biopharm Stat. 2007, 17: 529-569. 10.1080/10543400701376480.CrossRefPubMed Barnhart HX, Haber MJ, Lin LI: An overview on assessing agreement with continuous measurements. J Biopharm Stat. 2007, 17: 529-569. 10.1080/10543400701376480.CrossRefPubMed
42.
go back to reference Shrout PE, Fleiss JL: Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979, 86: 420-428.CrossRefPubMed Shrout PE, Fleiss JL: Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979, 86: 420-428.CrossRefPubMed
43.
go back to reference Eliasziw M, Young SL, Woodbury MG, Fryday-Field K: Statistical methodology for the concurrent assessment of interrater and intrarater reliability: using goniometric measurements as an example. Phys Ther. 1994, 74: 777-788.PubMed Eliasziw M, Young SL, Woodbury MG, Fryday-Field K: Statistical methodology for the concurrent assessment of interrater and intrarater reliability: using goniometric measurements as an example. Phys Ther. 1994, 74: 777-788.PubMed
44.
go back to reference Hayen A, Dennis RJ, Finch CF: Determining the intra- and inter-observer reliability of screening tools used in sports injury research. J Sci Med Sport. 2007, 10: 201-210. 10.1016/j.jsams.2006.09.002.CrossRefPubMed Hayen A, Dennis RJ, Finch CF: Determining the intra- and inter-observer reliability of screening tools used in sports injury research. J Sci Med Sport. 2007, 10: 201-210. 10.1016/j.jsams.2006.09.002.CrossRefPubMed
46.
go back to reference Hopkins WG: Measures of reliability in sports medicine and science. Sports Med. 2000, 30: 1-15. 10.2165/00007256-200030010-00001.CrossRefPubMed Hopkins WG: Measures of reliability in sports medicine and science. Sports Med. 2000, 30: 1-15. 10.2165/00007256-200030010-00001.CrossRefPubMed
47.
48.
go back to reference Walter SD, Eliasziw M, Donner A: Sample size and optimal designs for reliability studies. Stat Med. 1998, 17: 101-110. 10.1002/(SICI)1097-0258(19980115)17:1<101::AID-SIM727>3.0.CO;2-E.CrossRefPubMed Walter SD, Eliasziw M, Donner A: Sample size and optimal designs for reliability studies. Stat Med. 1998, 17: 101-110. 10.1002/(SICI)1097-0258(19980115)17:1<101::AID-SIM727>3.0.CO;2-E.CrossRefPubMed
49.
go back to reference Crawshaw J, Chambers J: A concise course in advanced level statistics with worked examples. 2001, Cheltenham, UK: Nelson Thornes, 469-470. 4th Crawshaw J, Chambers J: A concise course in advanced level statistics with worked examples. 2001, Cheltenham, UK: Nelson Thornes, 469-470. 4th
50.
go back to reference Maxwell SE, Delaney HD: Designing Experiments and Analyzing Data: A Model Comparison Perspective. 2004, Mahwah, NJ: Lawrence Erlbaum Associates, 2nd Maxwell SE, Delaney HD: Designing Experiments and Analyzing Data: A Model Comparison Perspective. 2004, Mahwah, NJ: Lawrence Erlbaum Associates, 2nd
Metadata
Title
Reliability of the standard goniometry and diagrammatic recording of finger joint angles: a comparative study with healthy subjects and non-professional raters
Author
Valdas Macionis
Publication date
01-12-2013
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2013
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/1471-2474-14-17

Other articles of this Issue 1/2013

BMC Musculoskeletal Disorders 1/2013 Go to the issue