Published in: Journal of Foot and Ankle Research 1/2009

Open Access 01-12-2009 | Methodology article

Reliability of measuring abductor hallucis muscle parameters using two different diagnostic ultrasound machines

Authors: Wayne A Hing, Keith Rome, Alyse FM Cameron

Published in: Journal of Foot and Ankle Research | Issue 1/2009

Login to get access

Abstract

Background

Diagnostic ultrasound provides a method of analysing soft tissue structures of the musculoskeletal system effectively and reliably. The aim of this study was to evaluate within and between session reliability of measuring muscle dorso-plantar thickness, medio-lateral length and cross-sectional area, of the abductor hallucis muscle using two different ultrasound machines, a higher end Philips HD11 Ultrasound machine and clinically orientated Chison 8300 Deluxe Digital Portable Ultrasound System.

Methods

The abductor hallucis muscle of both the left and right feet of thirty asymptomatic participants was imaged and then measured using both ultrasound machines. Interclass correlation coefficients (ICC) with 95% confidence intervals (CI) were used to calculate both within and between session intra-tester reliability. Standard error of the measurement (SEM) calculations were undertaken to assess difference between the actual measured score across trials and the smallest real difference (SRD) was calculated from the SEM to indicate the degree of change that would exceed the expected trial to trial variability.

Results

The ICCs, SEM and SRD for dorso-plantar thickness and medial-lateral length were shown to have excellent to high within and between-session reliability for both ultrasound machines. The between-session reliability indices for cross-sectional area were acceptable for both ultrasound machines.

Conclusion

The results of the current study suggest that regardless of the type ultrasound machine, intra-tester reliability for the measurement the abductor hallucis muscle parameters is very high.
Appendix
Available only for authorised users
Literature
1.
go back to reference Conaghan PG: Musculoskeletal ultrasonography: Improving our senses. Arthritis Care Res. 2005, 53: 639-642. 10.1002/art.21448.CrossRef Conaghan PG: Musculoskeletal ultrasonography: Improving our senses. Arthritis Care Res. 2005, 53: 639-642. 10.1002/art.21448.CrossRef
2.
go back to reference Norasteh A, Ebrahimi E, Salavati M, Rafiei J, Abbasnejad E: Reliability of B-mode ultrasonography for abdominal muscles in asymptomatic and patients with acute low back pain. J Bodyw Mov Ther. 2007, 11: 17-20. 10.1016/j.jbmt.2005.11.002.CrossRef Norasteh A, Ebrahimi E, Salavati M, Rafiei J, Abbasnejad E: Reliability of B-mode ultrasonography for abdominal muscles in asymptomatic and patients with acute low back pain. J Bodyw Mov Ther. 2007, 11: 17-20. 10.1016/j.jbmt.2005.11.002.CrossRef
3.
go back to reference Koppenhaver SL, Hebert JJ, Fritz JM, Parent EC, Teyhen DS, Magel JS: Reliability of rehabilitative ultrasound imaging of the transversus abdominis and lumbar multifidus muscles. Arch Phys Med Rehabil. 2009, 90: 87-94. 10.1016/j.apmr.2008.06.022.CrossRefPubMed Koppenhaver SL, Hebert JJ, Fritz JM, Parent EC, Teyhen DS, Magel JS: Reliability of rehabilitative ultrasound imaging of the transversus abdominis and lumbar multifidus muscles. Arch Phys Med Rehabil. 2009, 90: 87-94. 10.1016/j.apmr.2008.06.022.CrossRefPubMed
4.
go back to reference Cameron A, Rome K, Hing WA: Ultrasound evaluation of the abductor hallucis muscle: Reliability study. J Foot Ankle Res. 2008, 1: 12-10.1186/1757-1146-1-12.CrossRefPubMedPubMedCentral Cameron A, Rome K, Hing WA: Ultrasound evaluation of the abductor hallucis muscle: Reliability study. J Foot Ankle Res. 2008, 1: 12-10.1186/1757-1146-1-12.CrossRefPubMedPubMedCentral
5.
go back to reference Severinsen K, Andersen H: Evaluation of atrophy of foot muscles in diabetic neuropathy - a comparative study of nerve conduction studies and ultrasonography. J Clin Neurophysiol. 2007, 118: 2172-175. 10.1016/j.clinph.2007.06.019.CrossRef Severinsen K, Andersen H: Evaluation of atrophy of foot muscles in diabetic neuropathy - a comparative study of nerve conduction studies and ultrasonography. J Clin Neurophysiol. 2007, 118: 2172-175. 10.1016/j.clinph.2007.06.019.CrossRef
6.
go back to reference Arinci Incel N, Genc H, Erdem HR, Yorgancioglu ZR: Muscle imbalance in hallux valgus: an electromyographic study. Am J Phys Med Rehabil. 2003, 82: 345-9. 10.1097/00002060-200305000-00003.PubMed Arinci Incel N, Genc H, Erdem HR, Yorgancioglu ZR: Muscle imbalance in hallux valgus: an electromyographic study. Am J Phys Med Rehabil. 2003, 82: 345-9. 10.1097/00002060-200305000-00003.PubMed
7.
go back to reference Brenner E: Insertion of the abductor hallucis muscle in feet with and without hallux valgus. Anat Rec. 1999, 254: 429-34. 10.1002/(SICI)1097-0185(19990301)254:3<429::AID-AR14>3.0.CO;2-5.CrossRefPubMed Brenner E: Insertion of the abductor hallucis muscle in feet with and without hallux valgus. Anat Rec. 1999, 254: 429-34. 10.1002/(SICI)1097-0185(19990301)254:3<429::AID-AR14>3.0.CO;2-5.CrossRefPubMed
8.
go back to reference Greenman RL, Khaodhiar L, Lima C, Dinh T, Giurini JM, Veves A: Foot small muscle atrophy is present before the detection of clinical neuropathy. Diabetes Care. 2005, 28: 1425-30. 10.2337/diacare.28.6.1425.CrossRefPubMedPubMedCentral Greenman RL, Khaodhiar L, Lima C, Dinh T, Giurini JM, Veves A: Foot small muscle atrophy is present before the detection of clinical neuropathy. Diabetes Care. 2005, 28: 1425-30. 10.2337/diacare.28.6.1425.CrossRefPubMedPubMedCentral
9.
go back to reference Wong YS: Influence of the abductor hallucis muscle on the medial arch of the foot: a kinematic and anatomical cadaver study. Foot Ankle Int. 2007, 28: 617-20. 10.3113/FAI.2007.0617.CrossRefPubMed Wong YS: Influence of the abductor hallucis muscle on the medial arch of the foot: a kinematic and anatomical cadaver study. Foot Ankle Int. 2007, 28: 617-20. 10.3113/FAI.2007.0617.CrossRefPubMed
10.
go back to reference Wakefield RJ, D'Agostino M-A, Iagnocco A, Filippucci E, Backhaus M, Scheel AK, Joshua F, Naredo E, Schmidt WA, Grassi W, Moller I, Pineda C, Klauser A, Szkudlarek M, Terslev L, Balint P, Bruyn GAW, Swen WAA, Jousse-Joulin S, Kane D, Koski JM, O'Connor P, Milutinovic S, Conaghan P, Group OU: The OMERACT Ultrasound Group: status of current activities and research directions. J Rheumatol. 2007, 34: 848-51.PubMed Wakefield RJ, D'Agostino M-A, Iagnocco A, Filippucci E, Backhaus M, Scheel AK, Joshua F, Naredo E, Schmidt WA, Grassi W, Moller I, Pineda C, Klauser A, Szkudlarek M, Terslev L, Balint P, Bruyn GAW, Swen WAA, Jousse-Joulin S, Kane D, Koski JM, O'Connor P, Milutinovic S, Conaghan P, Group OU: The OMERACT Ultrasound Group: status of current activities and research directions. J Rheumatol. 2007, 34: 848-51.PubMed
11.
go back to reference Garrow AP, Papgeorgiou A, Silman AJ, Thomas E, Jayson MIV, Macfarlane GJ: The grading of the hallux valgus, the Manchester Scale. J Am Podiatr Med Assoc. 2001, 91: 74-78.CrossRefPubMed Garrow AP, Papgeorgiou A, Silman AJ, Thomas E, Jayson MIV, Macfarlane GJ: The grading of the hallux valgus, the Manchester Scale. J Am Podiatr Med Assoc. 2001, 91: 74-78.CrossRefPubMed
12.
go back to reference Vincent WJ: Statistics in kinesiology. 1999, Leeds: Human Kinetics Vincent WJ: Statistics in kinesiology. 1999, Leeds: Human Kinetics
13.
go back to reference Bruton A, Conway JH, Holgate ST: Reliability: what is it, and how is it measured?. Physiotherapy. 2000, 86: 94-95. 10.1016/S0031-9406(05)61211-4.CrossRef Bruton A, Conway JH, Holgate ST: Reliability: what is it, and how is it measured?. Physiotherapy. 2000, 86: 94-95. 10.1016/S0031-9406(05)61211-4.CrossRef
14.
go back to reference Rankin G, Stokes M: Reliability of assessment tools in rehabilitation: an illustration of appropriate statistical analyses. Clin Rehabil. 1998, 12: 197-199. 10.1191/026921598672178340.CrossRef Rankin G, Stokes M: Reliability of assessment tools in rehabilitation: an illustration of appropriate statistical analyses. Clin Rehabil. 1998, 12: 197-199. 10.1191/026921598672178340.CrossRef
15.
go back to reference Mantha S, Roizen M, Fleisher L, Thisted R, Foss J: Comparing methods of clinical measurement: reporting standards for Bland and Altman analysis. Anesth Analg. 2000, 90: 593-602. 10.1097/00000539-200003000-00018.CrossRefPubMed Mantha S, Roizen M, Fleisher L, Thisted R, Foss J: Comparing methods of clinical measurement: reporting standards for Bland and Altman analysis. Anesth Analg. 2000, 90: 593-602. 10.1097/00000539-200003000-00018.CrossRefPubMed
16.
go back to reference Bland J, Altman D: Applying the right statistics: analyses of measurement studies. Ultrasound Obstet Gynecol. 2003, 22: 85-93. 10.1002/uog.122.CrossRefPubMed Bland J, Altman D: Applying the right statistics: analyses of measurement studies. Ultrasound Obstet Gynecol. 2003, 22: 85-93. 10.1002/uog.122.CrossRefPubMed
17.
go back to reference Wallwork TL, Hides JA, Stanton WR: Intrarater and interrater reliability of assessment of lumbar multifidus muscle thickness using rehabilitative ultrasound imaging. J Orthop Sports Phys Ther. 2007, 37: 608-612.CrossRefPubMed Wallwork TL, Hides JA, Stanton WR: Intrarater and interrater reliability of assessment of lumbar multifidus muscle thickness using rehabilitative ultrasound imaging. J Orthop Sports Phys Ther. 2007, 37: 608-612.CrossRefPubMed
18.
go back to reference McPoil T, Vicenzino B, Cornwall M, Collins N, Warren M: Reliability and normative values for the foot mobility magnitude: a composite measure of vertical and medial-lateral mobility of the midfoot. J Foot Ankle Res. 2009, 2: 6-10.1186/1757-1146-2-6.CrossRefPubMedPubMedCentral McPoil T, Vicenzino B, Cornwall M, Collins N, Warren M: Reliability and normative values for the foot mobility magnitude: a composite measure of vertical and medial-lateral mobility of the midfoot. J Foot Ankle Res. 2009, 2: 6-10.1186/1757-1146-2-6.CrossRefPubMedPubMedCentral
19.
go back to reference Ota S, Ward SR, Chen YJ, Tsai YJ, Powers CM: Concurrent criterion-related validity and reliability of a clinical device used to assess lateral patellar displacement. J Orthop Sports Phys Ther. 2006, 36: 645-652.CrossRefPubMed Ota S, Ward SR, Chen YJ, Tsai YJ, Powers CM: Concurrent criterion-related validity and reliability of a clinical device used to assess lateral patellar displacement. J Orthop Sports Phys Ther. 2006, 36: 645-652.CrossRefPubMed
20.
go back to reference Reeves ND, Maganaris CN, Narici MV: Ultrasonographic assessment of human skeletal muscle size. Eur J Appl Physiol. 2004, 91-1: 116-118. 10.1007/s00421-003-0961-9.CrossRef Reeves ND, Maganaris CN, Narici MV: Ultrasonographic assessment of human skeletal muscle size. Eur J Appl Physiol. 2004, 91-1: 116-118. 10.1007/s00421-003-0961-9.CrossRef
21.
go back to reference Kane D, Grassi W, Sturrock R, Balint PV: A brief history of musculoskeletal ultrasound: 'From bats and ships to babies and hips'. Rheumatology. 2004, 43: 931-933. 10.1093/rheumatology/keh004.CrossRefPubMed Kane D, Grassi W, Sturrock R, Balint PV: A brief history of musculoskeletal ultrasound: 'From bats and ships to babies and hips'. Rheumatology. 2004, 43: 931-933. 10.1093/rheumatology/keh004.CrossRefPubMed
Metadata
Title
Reliability of measuring abductor hallucis muscle parameters using two different diagnostic ultrasound machines
Authors
Wayne A Hing
Keith Rome
Alyse FM Cameron
Publication date
01-12-2009
Publisher
BioMed Central
Published in
Journal of Foot and Ankle Research / Issue 1/2009
Electronic ISSN: 1757-1146
DOI
https://doi.org/10.1186/1757-1146-2-33

Other articles of this Issue 1/2009

Journal of Foot and Ankle Research 1/2009 Go to the issue