Reliability and failure modes of narrow implant systems

Authors: Ronaldo Hirata, Estevam A. Bonfante, Rodolfo B. Anchieta, Lucas S. Machado, Gileade Freitas, Vinicius P. Fardin, Nick Tovar, Paulo G. Coelho

Published in: Clinical Oral Investigations | Issue 7/2016

Login to get access

Abstract

Objectives

Narrow implants are indicated in areas of limited bone width or when grafting is nonviable. However, the reduction of implant diameter may compromise their performance. This study evaluated the reliability of several narrow implant systems under fatigue, after restored with single-unit crowns.

Materials and methods

Narrow implant systems were divided (n = 18 each), as follows: Astra (ASC); BioHorizons (BSC); Straumann Roxolid (SNC), Intra-Lock (IMC), and Intra-Lock one-piece abutment (ILO). Maxillary central incisor crowns were cemented and subjected to step-stress accelerated life testing in water. Use level probability Weibull curves and reliability for a mission of 100,000 cycles at 130- and 180-N loads (90 % two-sided confidence intervals) were calculated. Scanning electron microscopy was used for fractography.

Results

Reliability for 100,000 cycles at 130 N was ∼99 % in group ASC, ∼99 % in BSC, ∼96 % in SNC, ∼99 % in IMC, and ∼100 % in ILO. At 180 N, reliability of ∼34 % resulted for the ASC group, ∼91 % for BSC, ∼53 % for SNC, ∼70 % for IMC, and ∼99 % for ILO. Abutment screw fracture was the main failure mode for all groups.

Conclusions

Reliability was not different between systems for 100,000 cycles at the 130-N load. A significant decrease was observed at the 180-N load for ASC, SNC, and IMC, whereas it was maintained for BSC and ILO.

Clinical relevance

The investigated narrow implants presented mechanical performance under fatigue that suggests their safe use as single crowns in the anterior region.
Literature
1.
go back to reference Papaspyridakos P, Chen CJ, Chuang SK, Weber HP, Gallucci GO (2012) A systematic review of biologic and technical complications with fixed implant rehabilitations for edentulous patients. Int J Oral Maxillofac Implants 27:102–110PubMed Papaspyridakos P, Chen CJ, Chuang SK, Weber HP, Gallucci GO (2012) A systematic review of biologic and technical complications with fixed implant rehabilitations for edentulous patients. Int J Oral Maxillofac Implants 27:102–110PubMed
5.
go back to reference Hirata R, Bonfante EA, Machado LS, Tovar N, Coelho PG (2014) Mechanical evaluation of four narrow-diameter implant systems. Int J Prosthodont 27:359–362CrossRefPubMed Hirata R, Bonfante EA, Machado LS, Tovar N, Coelho PG (2014) Mechanical evaluation of four narrow-diameter implant systems. Int J Prosthodont 27:359–362CrossRefPubMed
6.
go back to reference Cehreli MC, Akca K (2004) Narrow-diameter implants as terminal support for occlusal three-unit FPDs: a biomechanical analysis. Int J Periodontics Restorative Dent 24:513–519PubMed Cehreli MC, Akca K (2004) Narrow-diameter implants as terminal support for occlusal three-unit FPDs: a biomechanical analysis. Int J Periodontics Restorative Dent 24:513–519PubMed
8.
go back to reference Kitagawa T, Tanimoto Y, Odaki M, Nemoto K, Aida M (2005) Influence of implant/abutment joint designs on abutment screw loosening in a dental implant system. J Biomed Mater Res B Appl Biomater 75:457–463. doi:10.1002/jbm.b.30328 CrossRefPubMed Kitagawa T, Tanimoto Y, Odaki M, Nemoto K, Aida M (2005) Influence of implant/abutment joint designs on abutment screw loosening in a dental implant system. J Biomed Mater Res B Appl Biomater 75:457–463. doi:10.​1002/​jbm.​b.​30328 CrossRefPubMed
9.
go back to reference Freitas-Junior AC, Rocha EP, Bonfante EA, Almeida EO, Anchieta RB, Martini AP, Assuncao WG, Silva NR, Coelho PG (2012) Biomechanical evaluation of internal and external hexagon platform switched implant-abutment connections: an in vitro laboratory and three-dimensional finite element analysis. Dent Mater 28:e218–e228. doi:10.1016/j.dental.2012.05.004 CrossRefPubMed Freitas-Junior AC, Rocha EP, Bonfante EA, Almeida EO, Anchieta RB, Martini AP, Assuncao WG, Silva NR, Coelho PG (2012) Biomechanical evaluation of internal and external hexagon platform switched implant-abutment connections: an in vitro laboratory and three-dimensional finite element analysis. Dent Mater 28:e218–e228. doi:10.​1016/​j.​dental.​2012.​05.​004 CrossRefPubMed
10.
go back to reference Parrington R (2002) Fractography of metals and plastics. ASM International Parrington R (2002) Fractography of metals and plastics. ASM International
13.
go back to reference Bonfante EA, Almeida EO, Lorenzoni FC, Coelho PG (2015) Effects of implant diameter and prosthesis retention system on the reliability of single crowns. Int J Oral Maxillofac Implants 30:95–101CrossRefPubMed Bonfante EA, Almeida EO, Lorenzoni FC, Coelho PG (2015) Effects of implant diameter and prosthesis retention system on the reliability of single crowns. Int J Oral Maxillofac Implants 30:95–101CrossRefPubMed
17.
go back to reference Almeida EO, Freitas Jr AC, Bonfante EA, Marotta L, Silva NR, Coelho PG (2013) Mechanical testing of implant-supported anterior crowns with different implant/abutment connections. Int J Oral Maxillofac Implant 28:103–108CrossRef Almeida EO, Freitas Jr AC, Bonfante EA, Marotta L, Silva NR, Coelho PG (2013) Mechanical testing of implant-supported anterior crowns with different implant/abutment connections. Int J Oral Maxillofac Implant 28:103–108CrossRef
18.
go back to reference Freitas AC, Jr., Bonfante EA, Rocha EP, Silva NR, Marotta L, Coelho PG (2011) Effect of implant connection and restoration design (screwed vs. cemented) in reliability and failure modes of anterior crowns. Eur J Oral Sci 119:323–330. doi:10.1111/j.1600-0722.2011.00837.x Freitas AC, Jr., Bonfante EA, Rocha EP, Silva NR, Marotta L, Coelho PG (2011) Effect of implant connection and restoration design (screwed vs. cemented) in reliability and failure modes of anterior crowns. Eur J Oral Sci 119:323–330. doi:10.​1111/​j.​1600-0722.​2011.​00837.​x
19.
20.
go back to reference Abernethy R (2006) The new Weibull handbook. Dr. Robert B, Abernethy, North Palm Beach, Florida Abernethy R (2006) The new Weibull handbook. Dr. Robert B, Abernethy, North Palm Beach, Florida
22.
go back to reference Nelson W (2004) Accelerated testing: statistical models, test plans and data analysis. John Wiley & Sons, New York, NY Nelson W (2004) Accelerated testing: statistical models, test plans and data analysis. John Wiley & Sons, New York, NY
23.
go back to reference Manda MG, Psyllaki PP, Tsipas DN, Koidis PT (2009) Observations on an in-vivo failure of a titanium dental implant/abutment screw system: a case report. J Biomed Mater Res B Appl Biomater 89:264–273. doi:10.1002/jbm.b.31211 CrossRefPubMed Manda MG, Psyllaki PP, Tsipas DN, Koidis PT (2009) Observations on an in-vivo failure of a titanium dental implant/abutment screw system: a case report. J Biomed Mater Res B Appl Biomater 89:264–273. doi:10.​1002/​jbm.​b.​31211 CrossRefPubMed
25.
go back to reference Nelson W (1990) Book title. John Wiley & Sons, New York Nelson W (1990) Book title. John Wiley & Sons, New York
26.
go back to reference Barber HD, Seckinger RJ (1994) The role of the small-diameter dental implant: a preliminary report on the miniplant system. Compendium 15(1390):1392 Barber HD, Seckinger RJ (1994) The role of the small-diameter dental implant: a preliminary report on the miniplant system. Compendium 15(1390):1392
28.
go back to reference McCracken M (1999) Dental implant materials: commercially pure titanium and titanium alloys. J Prosthodont 8:40–43CrossRefPubMed McCracken M (1999) Dental implant materials: commercially pure titanium and titanium alloys. J Prosthodont 8:40–43CrossRefPubMed
29.
go back to reference Hirata R, Bonfante EA, Machado LS, Tovar N, Coelho PG (in press 2015) Mechanical evaluation of two titanium alloy grades used in implant dentistry. Int J Oral Maxillofac Implant. Hirata R, Bonfante EA, Machado LS, Tovar N, Coelho PG (in press 2015) Mechanical evaluation of two titanium alloy grades used in implant dentistry. Int J Oral Maxillofac Implant.
31.
go back to reference Roach M, Williamson RS, Zardiackas L (2005) Comparison of the corrosion fatigue characteristics of CP Ti grade 4, Ti-6Al-4V ELI, Ti-6Al-7Nb, and Ti-15Mo. In: International A (ed) Book title., Roach M, Williamson RS, Zardiackas L (2005) Comparison of the corrosion fatigue characteristics of CP Ti grade 4, Ti-6Al-4V ELI, Ti-6Al-7Nb, and Ti-15Mo. In: International A (ed) Book title.,
33.
go back to reference Almeida EO, Junior AC, Bonfante EA, Silva NR, Coelho PG (2012) Reliability evaluation of alumina-blasted/acid-etched versus laser-sintered dental implants. Lasers Med Sci. doi:10.1007/s10103-012-1170-8 Almeida EO, Junior AC, Bonfante EA, Silva NR, Coelho PG (2012) Reliability evaluation of alumina-blasted/acid-etched versus laser-sintered dental implants. Lasers Med Sci. doi:10.​1007/​s10103-012-1170-8
35.
go back to reference Freitas-Junior AC, Bonfante EA, Martins LM, Silva NR, Marotta L, Coelho PG (2011) Effect of implant diameter on reliability and failure modes of molar crowns. Int J Prosthodont 24:557–561PubMed Freitas-Junior AC, Bonfante EA, Martins LM, Silva NR, Marotta L, Coelho PG (2011) Effect of implant diameter on reliability and failure modes of molar crowns. Int J Prosthodont 24:557–561PubMed
Metadata
Title
Reliability and failure modes of narrow implant systems
Authors
Ronaldo Hirata
Estevam A. Bonfante
Rodolfo B. Anchieta
Lucas S. Machado
Gileade Freitas
Vinicius P. Fardin
Nick Tovar
Paulo G. Coelho
Publication date
01-09-2016
Publisher
Springer Berlin Heidelberg
Published in
Clinical Oral Investigations / Issue 7/2016
Print ISSN: 1432-6981
Electronic ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-015-1636-8

Other articles of this Issue 7/2016

Clinical Oral Investigations 7/2016 Go to the issue