Relationship between the image quality and noise-equivalent count in time-of-flight positron emission tomography

Authors: Akira Maebatake, Go Akamatsu, Kenta Miwa, Yuji Tsutsui, Kazuhiko Himuro, Shingo Baba, Masayuki Sasaki

Published in: Annals of Nuclear Medicine | Issue 1/2016

Login to get access

Abstract

Objective

The aim of this study was to investigate the relationship between the NEC and TOF-PET image quality.

Methods

The National Electrical Manufactures Association and International Electrical Commission (NEMA IEC) body phantom with a 10-mm diameter sphere was filled with an 18F-FDG solution with a 4:1 radioactivity ratio. The PET data were acquired in the three-dimensional list mode for 20 min. We created frame data ranging from 1 to 5 min acquisition time, which were then reconstructed using the baseline ordered-subsets expectation maximization (OSEM), the OSEM + point spread function (PSF) algorithm, OSEM + time-of-flight (TOF) algorithm and OSEM + PSF + TOF algorithm. The PET images were analyzed according to the noise-equivalent count (NEC), the coefficients of variance of the background (CVBG), the maximum count (CVmax) and the contrast (CVCONT). The results were compared with the recommended value according to the guidelines for the oncology FDG-PET/CT protocol.

Results

The NEC was higher than the recommended value at 3 min or longer acquisition time. The CVBG lower than 15 % were obtained at 3 min acquisition time without TOF and at 2 min acquisition time with TOF. The CVBG of 10 % or lower were obtained at 5 min or longer acquisition time without TOF and at 4 min or longer acquisition time with TOF. Both the CVmax and CVCONT lower than 10 % were obtained at 3 min or longer acquisition time without TOF and at 1 min acquisition or longer with TOF. No particular relationships were observed between the frame number and degree of the variation in the image quality. The CVCONT significantly correlated with the NEC for the data reconstructed without TOF information, while there were no significant correlations between these useful metrics for the data reconstructed with TOF.

Conclusion

This study demonstrated that the NEC is not a useful metric for the evaluation of the image quality on TOF-PET images.
Literature
1.
go back to reference Rohren E, Turkington T, Coleman R. Clinical applications of PET in oncology. Radiology. 2004;231:305–32.PubMedCrossRef Rohren E, Turkington T, Coleman R. Clinical applications of PET in oncology. Radiology. 2004;231:305–32.PubMedCrossRef
2.
go back to reference Fischer B, Lassen U, Mortensen J, Mortensen J, Larsen S, Loft A, et al. Preoperative staging of lung cancer with combined PET-CT. N Engl J Med. 2009;361:32–9.PubMedCrossRef Fischer B, Lassen U, Mortensen J, Mortensen J, Larsen S, Loft A, et al. Preoperative staging of lung cancer with combined PET-CT. N Engl J Med. 2009;361:32–9.PubMedCrossRef
3.
go back to reference Hicks RJ, Kalff V, MacManus MP, Ware RE, McKenzie AF, Matthews JP, et al. The utility of (18)F-FDG PET for suspected recurrent non-small cell lung cancer after potentially curative therapy: impact on management and prognostic stratification. J Nucl Med. 2001;42(11):1605–13.PubMed Hicks RJ, Kalff V, MacManus MP, Ware RE, McKenzie AF, Matthews JP, et al. The utility of (18)F-FDG PET for suspected recurrent non-small cell lung cancer after potentially curative therapy: impact on management and prognostic stratification. J Nucl Med. 2001;42(11):1605–13.PubMed
4.
go back to reference Beyer T, Townsend DW, Brun T, Kinahan PE, Charron M, Roddy R, et al. A combined PET/CT scanner for clinical oncology. J Nucl Med. 2000;41(8):1369–79.PubMed Beyer T, Townsend DW, Brun T, Kinahan PE, Charron M, Roddy R, et al. A combined PET/CT scanner for clinical oncology. J Nucl Med. 2000;41(8):1369–79.PubMed
5.
go back to reference Masuda Y, Kondo C, Matsuo Y, Uetani M, Kusakabe K. Comparison of imaging protocols for 18F-FDG PET/CT in overweight patients: optimizing scan duration versus administered dose. J Nucl Med. 2009;29:844–8.CrossRef Masuda Y, Kondo C, Matsuo Y, Uetani M, Kusakabe K. Comparison of imaging protocols for 18F-FDG PET/CT in overweight patients: optimizing scan duration versus administered dose. J Nucl Med. 2009;29:844–8.CrossRef
6.
go back to reference Strother SC, Casey ME, Hoffman EJ. Measuring PET scanner sensitivity: relating count rates to image signal-to-noise ratios using noise equivalent counts. IEEE Trans Nucl Sci. 1990;37:783–8.CrossRef Strother SC, Casey ME, Hoffman EJ. Measuring PET scanner sensitivity: relating count rates to image signal-to-noise ratios using noise equivalent counts. IEEE Trans Nucl Sci. 1990;37:783–8.CrossRef
7.
go back to reference Watson CC. Count rate dependence of local signal-to-noise ratio in positron emission tomography. IEEE Trans Nucl Sci. 2004;51:2670–80.CrossRef Watson CC. Count rate dependence of local signal-to-noise ratio in positron emission tomography. IEEE Trans Nucl Sci. 2004;51:2670–80.CrossRef
8.
go back to reference Dahlbom M, Schiepers C, Czernin J. Comparison of noise equivalent count rates and image noise. IEEE Trans Nucl Sci. 2005;52:1386–90.CrossRef Dahlbom M, Schiepers C, Czernin J. Comparison of noise equivalent count rates and image noise. IEEE Trans Nucl Sci. 2005;52:1386–90.CrossRef
9.
go back to reference Fukukita H, Senda M, Terauchi T, Suzuki K, Daisaki H, Matsumoto K, et al. Japanese guideline for the oncology FDG-PET/CT data acquisition protocol: synopsis of version 1.0. Ann Nucl Med. 2010;24:325–34.PubMedCrossRef Fukukita H, Senda M, Terauchi T, Suzuki K, Daisaki H, Matsumoto K, et al. Japanese guideline for the oncology FDG-PET/CT data acquisition protocol: synopsis of version 1.0. Ann Nucl Med. 2010;24:325–34.PubMedCrossRef
10.
go back to reference Tong S, Alessio AM, Kinahan PE. Noise and signal properties in PSF-based fully 3D PET image reconstruction: an experimental evaluation. Phys Med Biol. 2010;55:1453–73.PubMedPubMedCentralCrossRef Tong S, Alessio AM, Kinahan PE. Noise and signal properties in PSF-based fully 3D PET image reconstruction: an experimental evaluation. Phys Med Biol. 2010;55:1453–73.PubMedPubMedCentralCrossRef
11.
go back to reference Surti S, Karp JS, Popescu LM, Daube-Witherspoon ME, Werner M. Investigation of time-of-flight benefit for fully 3-D PET. IEEE Trans Med Imaging. 2006;25:529–38.PubMedCrossRef Surti S, Karp JS, Popescu LM, Daube-Witherspoon ME, Werner M. Investigation of time-of-flight benefit for fully 3-D PET. IEEE Trans Med Imaging. 2006;25:529–38.PubMedCrossRef
12.
go back to reference Malti-Climent JM, Prieto E, Dominguez-Prado I, Garcia-Velloso MJ, Rodriguez-Fraile M, Arbizu J, et al. Contribution of time of flight and point spread function modeling to the performance characteristics of the PET/CT Biograph mCT scanner. Rev Esp Med Nucl Imagen Mol. 2013;32:13–21. Malti-Climent JM, Prieto E, Dominguez-Prado I, Garcia-Velloso MJ, Rodriguez-Fraile M, Arbizu J, et al. Contribution of time of flight and point spread function modeling to the performance characteristics of the PET/CT Biograph mCT scanner. Rev Esp Med Nucl Imagen Mol. 2013;32:13–21.
13.
go back to reference Akamatsu G, Ishikawa K, Mitsumoto K, Taniguchi T, Ohya N, Baba S, et al. Improvement in PET/CT Image quality with a combination of point-spread function and time-of-flight in relation to reconstruction parameters. J Nucl Med. 2012;53:1716–22.PubMedCrossRef Akamatsu G, Ishikawa K, Mitsumoto K, Taniguchi T, Ohya N, Baba S, et al. Improvement in PET/CT Image quality with a combination of point-spread function and time-of-flight in relation to reconstruction parameters. J Nucl Med. 2012;53:1716–22.PubMedCrossRef
14.
go back to reference Taniguchi T, Akamatsu G, Kasahara Y, Mitsumoto K, Baba S, Tsutsui Y, et al. Improvement in PET/CT image quality in overweight patients with PSF and TOF. Ann Nucl Med. 2015;29:71–7.PubMedPubMedCentralCrossRef Taniguchi T, Akamatsu G, Kasahara Y, Mitsumoto K, Baba S, Tsutsui Y, et al. Improvement in PET/CT image quality in overweight patients with PSF and TOF. Ann Nucl Med. 2015;29:71–7.PubMedPubMedCentralCrossRef
15.
go back to reference Watson CC, Casey ME, Bendriem B, Carney JP, Townsend DW, Eberl S, et al. Optimizing injected dose in clinical PET by accurately modeling the counting-rate response functions specific to individual patient scans. J Nucl Med. 2005;46:1825–34.PubMed Watson CC, Casey ME, Bendriem B, Carney JP, Townsend DW, Eberl S, et al. Optimizing injected dose in clinical PET by accurately modeling the counting-rate response functions specific to individual patient scans. J Nucl Med. 2005;46:1825–34.PubMed
16.
go back to reference Brown C, Dempsey MF, Gillen G, Elliott AT. Investigation of 18F-FDG 3D mode PET image quality versus acquisition time. Nucl Med Commun. 2010;31:254–9.PubMedCrossRef Brown C, Dempsey MF, Gillen G, Elliott AT. Investigation of 18F-FDG 3D mode PET image quality versus acquisition time. Nucl Med Commun. 2010;31:254–9.PubMedCrossRef
17.
go back to reference Lois C, Jakoby BW, Long MJ, Hubner KF, Barker DW, Casey ME, et al. An assessment of the impact of incorporating time-of-flight information into clinical PET/CT imaging. J Nucl Med. 2010;51:237–45.PubMedPubMedCentralCrossRef Lois C, Jakoby BW, Long MJ, Hubner KF, Barker DW, Casey ME, et al. An assessment of the impact of incorporating time-of-flight information into clinical PET/CT imaging. J Nucl Med. 2010;51:237–45.PubMedPubMedCentralCrossRef
18.
go back to reference Conti M, Hong I, Michel C. Reconstruction of scattered and unscattered PET coincidences using TOF and energy information. Phys Med Biol. 2012;57:N307–17.PubMedCrossRef Conti M, Hong I, Michel C. Reconstruction of scattered and unscattered PET coincidences using TOF and energy information. Phys Med Biol. 2012;57:N307–17.PubMedCrossRef
19.
go back to reference Vandenberghe S, Elmbt LV, Guerchaft M, Clementel E, Verhaeghe J, Bol A, et al. Optimization of time-of-flight reconstruction on Philips GEMINI TF. Eur J Nucl Med Mol Imaging. 2009;36:1994–2001.PubMedCrossRef Vandenberghe S, Elmbt LV, Guerchaft M, Clementel E, Verhaeghe J, Bol A, et al. Optimization of time-of-flight reconstruction on Philips GEMINI TF. Eur J Nucl Med Mol Imaging. 2009;36:1994–2001.PubMedCrossRef
20.
go back to reference Thielemans K, Asma E, Ahn Sm, RM Manjeshwar, T Deller, Ross SG, et al. Impact of PSF modeling on the convergence rate and edge behavior of EM images in PET. In: IEEE Nuclear Science Symposium Conference Record (NSS/MIC), Knoxville, TN; 2010. p. 3267–72. Thielemans K, Asma E, Ahn Sm, RM Manjeshwar, T Deller, Ross SG, et al. Impact of PSF modeling on the convergence rate and edge behavior of EM images in PET. In: IEEE Nuclear Science Symposium Conference Record (NSS/MIC), Knoxville, TN; 2010. p. 3267–72.
21.
go back to reference Boellaard R, Krak NC, Hoekstra OS, Lammertsma AA. Effects of noise, image resolution, and ROI definition on the accuracy of standard uptake values: a simulation study. J Nucl Med. 2004;45:1519–27.PubMed Boellaard R, Krak NC, Hoekstra OS, Lammertsma AA. Effects of noise, image resolution, and ROI definition on the accuracy of standard uptake values: a simulation study. J Nucl Med. 2004;45:1519–27.PubMed
22.
go back to reference Lodge MA, Chaudhry MA, Wahl RL. Noise considerations for PET quantification using maximum and peak standardized uptake value. J Nucl Med. 2012;53(7):1041–7.PubMedPubMedCentralCrossRef Lodge MA, Chaudhry MA, Wahl RL. Noise considerations for PET quantification using maximum and peak standardized uptake value. J Nucl Med. 2012;53(7):1041–7.PubMedPubMedCentralCrossRef
23.
go back to reference Schwartz J, Humm JL, Gonen M, Kalaigian H, Schoder H, Larson SM, et al. Repeatability of SUV measurements in serial PET. Med Phys. 2011;38:2629–38.PubMedCrossRef Schwartz J, Humm JL, Gonen M, Kalaigian H, Schoder H, Larson SM, et al. Repeatability of SUV measurements in serial PET. Med Phys. 2011;38:2629–38.PubMedCrossRef
24.
go back to reference Karp JS, Surti S, Daube-Witherspoon ME, Muehllehner G. Benefit of time-of-flight in PET: experimental and clinical results. J Nucl Med. 2008;49:462–70.PubMedPubMedCentralCrossRef Karp JS, Surti S, Daube-Witherspoon ME, Muehllehner G. Benefit of time-of-flight in PET: experimental and clinical results. J Nucl Med. 2008;49:462–70.PubMedPubMedCentralCrossRef
25.
go back to reference Molina-Duran F, Dinter D, Schoenahl F, Shoenberg SO, Glatting G. Dependence of image quality on acquisition time for the PET/CT Biograph mCT. Z Med Phys. 2014;24:73–9.PubMedCrossRef Molina-Duran F, Dinter D, Schoenahl F, Shoenberg SO, Glatting G. Dependence of image quality on acquisition time for the PET/CT Biograph mCT. Z Med Phys. 2014;24:73–9.PubMedCrossRef
26.
go back to reference Mullani NA, Markham J, Ter-Pogossian MM. Feasibility of time-of-flight reconstruction in positron emission tomography. J Nucl Med. 1980;21:1095–7.PubMed Mullani NA, Markham J, Ter-Pogossian MM. Feasibility of time-of-flight reconstruction in positron emission tomography. J Nucl Med. 1980;21:1095–7.PubMed
27.
go back to reference Budinger TF. Time-of-flight positron emission tomography: status relative to conventional PET. J Nucl Med. 1983;24:73–8.PubMed Budinger TF. Time-of-flight positron emission tomography: status relative to conventional PET. J Nucl Med. 1983;24:73–8.PubMed
28.
go back to reference Conti M. Effect of randoms on signal-to-noise ratio in TOF PET. IEEE Trans Nucl Sci. 2006;53:1188–93.CrossRef Conti M. Effect of randoms on signal-to-noise ratio in TOF PET. IEEE Trans Nucl Sci. 2006;53:1188–93.CrossRef
Metadata
Title
Relationship between the image quality and noise-equivalent count in time-of-flight positron emission tomography
Authors
Akira Maebatake
Go Akamatsu
Kenta Miwa
Yuji Tsutsui
Kazuhiko Himuro
Shingo Baba
Masayuki Sasaki
Publication date
01-01-2016
Publisher
Springer Japan
Published in
Annals of Nuclear Medicine / Issue 1/2016
Print ISSN: 0914-7187
Electronic ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-015-1032-0

Other articles of this Issue 1/2016

Annals of Nuclear Medicine 1/2016 Go to the issue