Published in: European Radiology 6/2019

01-06-2019 | Cardiac

Reducing both radiation and contrast doses in coronary CT angiography in lean patients on a 16-cm wide-detector CT using 70 kVp and ASiR-V algorithm, in comparison with the conventional 100-kVp protocol

Authors: Yuhuan Chen, Zhentang Liu, Michelle Li, Yong Yu, Yongjun Jia, Guangming Ma, Zhijun Hu, DongHong Wei, Dou Li, Taiping He

Published in: European Radiology | Issue 6/2019

Login to get access

Abstract

Objective

To investigate the value of prospectively ECG-triggered coronary CT angiography (CCTA) for lean patients with body mass index (BMI) ≤ 23 kg/m2 using 70 kVp and high-level volume-based adaptive statistical iterative reconstruction (ASiR-V) algorithm on a 16-cm wide-detector CT system for reducing both radiation and contrast doses in comparison with the conventional 100-kVp protocol.

Materials and methods

Thirty patients (group A) were prospectively enrolled to undergo 70-kVp CCTA on a 16-cm wide-detector CT scanner with noise index (NI) of 36 HU and at weight-dependent contrast dose rate of 16 mg I/kg/s for 9-s injection. Images were reconstructed with 80% ASiR-V. Radiation dose, contrast dose, and image quality were statistically compared with 30 patients (group B) in database with matching BMI who underwent conventional 100-kVp CCTA with NI of 25 HU, and at 25 mg I/kg/s rate for 10-s injection and reconstructed with 60% ASiR-V.

Results

There was no significant difference in patient demographics between the two groups (all p > 0.05). The two groups also had similar mean CT values and contrast-noise ratio (CNR) and subjective image quality (all p > 0.05). However, group A with 70 kVp reduced the effective dose by 75.3% compared with group B (0.43 ± 0.20 mSv vs. 1.74 ± 1.01 mSv, p < 0.001), and required 42.4% less contrast dose than group B (22.46 ± 2.94 ml vs. 38.99 ± 5.10 ml, p < 0.001).

Conclusions

Prospectively ECG-triggered CCTA using 70 kVp and high-level ASiR-V on a 16-cm wide-detector CT system provides diagnostic images with substantial reduction in both radiation and contrast doses for patients with BMI ≤ 23 kg/m2 compared to the conventional 100-kVp protocol.

Key Points

• 70-kVp CCTA produces excellent images at sub-millisievert radiation.
• 70-kVp CCTA reduces both radiation and contrast doses over conventional protocol.
• Achieving low-dose CCTA with combined low kVp and high-level ASIR-V.
Literature
1.
go back to reference Di Cesare E, Gennarell A, Di Sibio A et al (2015) Image quality and radiation dose of single heartbeat 640-slice coronary CT angiography: a comparison between patients with chronic atrial fibrillation and subjects in normal sinus rhythm by propensity analysis. Eur J Radiol 84(4):631–636 Di Cesare E, Gennarell A, Di Sibio A et al (2015) Image quality and radiation dose of single heartbeat 640-slice coronary CT angiography: a comparison between patients with chronic atrial fibrillation and subjects in normal sinus rhythm by propensity analysis. Eur J Radiol 84(4):631–636
2.
go back to reference Pontone G, Bertella E, Mushtaq S et al (2016) Coronary artery disease: diagnostic accuracy of CT coronary angiography -- a comparison of high and standard spatial resolution scanning. Radiology 271(3):688–694CrossRef Pontone G, Bertella E, Mushtaq S et al (2016) Coronary artery disease: diagnostic accuracy of CT coronary angiography -- a comparison of high and standard spatial resolution scanning. Radiology 271(3):688–694CrossRef
3.
go back to reference Wang SY, Gao W, Zhong YM et al (2017) Prospective ECG-triggering cardiac CT for infants with complex congenital heart disease using low-dose contrast medium, low tube voltage, and adaptive statistical iterative reconstruction. Clin Radiol 72(6):502–507CrossRefPubMed Wang SY, Gao W, Zhong YM et al (2017) Prospective ECG-triggering cardiac CT for infants with complex congenital heart disease using low-dose contrast medium, low tube voltage, and adaptive statistical iterative reconstruction. Clin Radiol 72(6):502–507CrossRefPubMed
4.
go back to reference Lee HS, Suh YJ, Han K et al (2018) Effectiveness of automatic tube potential selection with tube current modulation in coronary CT angiography for obese patients: comparison with a body mass index-based protocol using the propensity score matching method. PLoS One. https://doi.org/10.1371/journal.pone.0190584. eCollection 2018 Lee HS, Suh YJ, Han K et al (2018) Effectiveness of automatic tube potential selection with tube current modulation in coronary CT angiography for obese patients: comparison with a body mass index-based protocol using the propensity score matching method. PLoS One. https://​doi.​org/​10.​1371/​journal.​pone.​0190584. eCollection 2018
5.
go back to reference Gopal A, Mao SS, Karlsberg D et al (2009) Radiation reduction with prospective ECG-triggering acquisition using 64-multidetector computed tomographic angiography. Int J Cardiovasc Imaging 25(4):405–416CrossRefPubMed Gopal A, Mao SS, Karlsberg D et al (2009) Radiation reduction with prospective ECG-triggering acquisition using 64-multidetector computed tomographic angiography. Int J Cardiovasc Imaging 25(4):405–416CrossRefPubMed
6.
go back to reference Messerli M, Dewes P, Scholtz JE et al (2017) Evaluation of an adaptive detector collimation for prospectively ECG-triggered coronary CT angiography with third-generation dual-source CT. Eur Radiol 28(5):2143–2150CrossRefPubMed Messerli M, Dewes P, Scholtz JE et al (2017) Evaluation of an adaptive detector collimation for prospectively ECG-triggered coronary CT angiography with third-generation dual-source CT. Eur Radiol 28(5):2143–2150CrossRefPubMed
7.
go back to reference Fuchs TA, Fiechter M, Gebhard C et al (2013) CT coronary angiography: impact of adapted statistical iterative reconstruction (ASiR) on coronary stenosis and plaque composition analysis. Int J Cardiovasc Imaging 29(3):719–724CrossRefPubMed Fuchs TA, Fiechter M, Gebhard C et al (2013) CT coronary angiography: impact of adapted statistical iterative reconstruction (ASiR) on coronary stenosis and plaque composition analysis. Int J Cardiovasc Imaging 29(3):719–724CrossRefPubMed
8.
go back to reference Li S, Tang X, Peng L et al (2015) A head-to-head comparison of homocysteine and cystatin c as pre-procedure predictors for contrast-induced nephropathy in patients undergoing coronary computed tomography angiography. Clin Chim Acta 444:86–91 Li S, Tang X, Peng L et al (2015) A head-to-head comparison of homocysteine and cystatin c as pre-procedure predictors for contrast-induced nephropathy in patients undergoing coronary computed tomography angiography. Clin Chim Acta 444:86–91
9.
go back to reference Formosa A, Santos DM, Marcuzzi D, Common AA, Prabhudesai V (2016) Low contrast dose catheter-directed CT angiography (CCTA). Cardiovasc Intervent Radiol 39(4):606–610CrossRefPubMed Formosa A, Santos DM, Marcuzzi D, Common AA, Prabhudesai V (2016) Low contrast dose catheter-directed CT angiography (CCTA). Cardiovasc Intervent Radiol 39(4):606–610CrossRefPubMed
10.
go back to reference Tan SK, Yeong CH, Raja Aman RRA et al (2018) Low tube voltage prospectively ECG-triggered coronary CT angiography: a systematic review of image quality and radiation dose. Br J Radiol 91(1088):20170874 Tan SK, Yeong CH, Raja Aman RRA et al (2018) Low tube voltage prospectively ECG-triggered coronary CT angiography: a systematic review of image quality and radiation dose. Br J Radiol 91(1088):20170874
14.
go back to reference Matveeva A, Schmitt RR, Edtinger K et al (2018) Coronary CT angiography in patients with atrial fibrillation: standard-dose and low-dose imaging with a high-resolution whole-heart CT scanner. Eur Radiol 28(8):3432–3440CrossRefPubMed Matveeva A, Schmitt RR, Edtinger K et al (2018) Coronary CT angiography in patients with atrial fibrillation: standard-dose and low-dose imaging with a high-resolution whole-heart CT scanner. Eur Radiol 28(8):3432–3440CrossRefPubMed
15.
go back to reference Hong C, Becker CR, Huber A et al (2001) ECG-gated reconstructed multi-detector row CT coronary angiography: effect of varying trigger delay on image quality. Radiology 220(3):712–717CrossRefPubMed Hong C, Becker CR, Huber A et al (2001) ECG-gated reconstructed multi-detector row CT coronary angiography: effect of varying trigger delay on image quality. Radiology 220(3):712–717CrossRefPubMed
16.
go back to reference Andreini D, Pontone G, Mushtaq S et al (2018) Image quality and radiation dose of coronary CT angiography performed with whole-heart coverage CT scanner with intra-cycle motion correction algorithm in patients with atrial fibrillation. Eur Radiol 28(4):1383–1392CrossRefPubMed Andreini D, Pontone G, Mushtaq S et al (2018) Image quality and radiation dose of coronary CT angiography performed with whole-heart coverage CT scanner with intra-cycle motion correction algorithm in patients with atrial fibrillation. Eur Radiol 28(4):1383–1392CrossRefPubMed
17.
go back to reference Akai H, Kiryu S, Shibata E et al (2016) Reducing CT radiation exposure with organ effective modulation: a retrospective clinical study. Eur J Radiol 85(9):1569–1573CrossRefPubMed Akai H, Kiryu S, Shibata E et al (2016) Reducing CT radiation exposure with organ effective modulation: a retrospective clinical study. Eur J Radiol 85(9):1569–1573CrossRefPubMed
18.
go back to reference Gordic S, Desbiolles L, Sedlmair M et al (2016) Optimizing radiation dose by using advanced modelled iterative reconstruction in high-pitch coronary CT angiography. Eur J Radiol 26(2):459–468 Gordic S, Desbiolles L, Sedlmair M et al (2016) Optimizing radiation dose by using advanced modelled iterative reconstruction in high-pitch coronary CT angiography. Eur J Radiol 26(2):459–468
19.
go back to reference Goitein O, Matetzky S, Eshet Y et al (2011) Coronary CT angiography for acute chest pain triage: techniques for radiation exposure reduction; 128 vs. 64 multidetector CT. Acta Radiol 52(8):840–845CrossRefPubMed Goitein O, Matetzky S, Eshet Y et al (2011) Coronary CT angiography for acute chest pain triage: techniques for radiation exposure reduction; 128 vs. 64 multidetector CT. Acta Radiol 52(8):840–845CrossRefPubMed
20.
go back to reference Pontone G, Andreini D, Bartorelli AL et al (2012) Radiation dose and diagnostic accuracy of multidetector computed tomography for the detection of significant coronary artery stenoses: a meta-analysis. Int J Cardiol 160(3):155–164CrossRefPubMed Pontone G, Andreini D, Bartorelli AL et al (2012) Radiation dose and diagnostic accuracy of multidetector computed tomography for the detection of significant coronary artery stenoses: a meta-analysis. Int J Cardiol 160(3):155–164CrossRefPubMed
21.
go back to reference Albrecht MH, Nance JW, Schoepf UJ et al (2017) Diagnostic accuracy of low and high tube voltage coronary CT angiography using an x-ray tube potential-tailored contrast medium injection protocol. Eur Radiol 28(5):1–9 Albrecht MH, Nance JW, Schoepf UJ et al (2017) Diagnostic accuracy of low and high tube voltage coronary CT angiography using an x-ray tube potential-tailored contrast medium injection protocol. Eur Radiol 28(5):1–9
22.
go back to reference Zhang LJ, Zhao YE, Schoepf UJ et al (2015) Seventy-peak kilovoltage high-pitch thoracic aortic CT angiography without ECG gating: evaluation of image quality and radiation dose. Acad Radiol 22(7):890–897CrossRefPubMed Zhang LJ, Zhao YE, Schoepf UJ et al (2015) Seventy-peak kilovoltage high-pitch thoracic aortic CT angiography without ECG gating: evaluation of image quality and radiation dose. Acad Radiol 22(7):890–897CrossRefPubMed
23.
go back to reference Cho ES, Chung TS, Ahn SJ, Chong K, Baek JH, Suh SH (2015) Cerebral computed tomography angiography using a 70 kVp protocol: improved vascular enhancement with a reduced volume of contrast medium and radiation dose. Eur Radiol 25(5):1421–1430 Cho ES, Chung TS, Ahn SJ, Chong K, Baek JH, Suh SH (2015) Cerebral computed tomography angiography using a 70 kVp protocol: improved vascular enhancement with a reduced volume of contrast medium and radiation dose. Eur Radiol 25(5):1421–1430
24.
go back to reference Qi L, Zhao Y, Zhou CS et al (2015) Image quality and radiation dose of lower extremity CT angiography at 70 kVp on an integrated circuit detector dual-source computed tomography. Acta Radiol 56(6):659–665CrossRefPubMed Qi L, Zhao Y, Zhou CS et al (2015) Image quality and radiation dose of lower extremity CT angiography at 70 kVp on an integrated circuit detector dual-source computed tomography. Acta Radiol 56(6):659–665CrossRefPubMed
25.
go back to reference Zhang LJ, Qi L, Wang J et al (2014) Feasibility of prospectively ECG-triggered high-pitch coronary CT angiography with 30 ml iodinated contrast agent at 70 kVp: initial experience. Eur Radiol 24(7):1537–1546CrossRefPubMed Zhang LJ, Qi L, Wang J et al (2014) Feasibility of prospectively ECG-triggered high-pitch coronary CT angiography with 30 ml iodinated contrast agent at 70 kVp: initial experience. Eur Radiol 24(7):1537–1546CrossRefPubMed
27.
go back to reference Feng R, Tong J, Liu X, Zhao Y, Zhang L (2017) High-pitch coronary CT angiography at 70 kVp adopting a protocol of low injection speed and low volume of contrast medium. Korean J Radiol 18(5):763–772CrossRefPubMedPubMedCentral Feng R, Tong J, Liu X, Zhao Y, Zhang L (2017) High-pitch coronary CT angiography at 70 kVp adopting a protocol of low injection speed and low volume of contrast medium. Korean J Radiol 18(5):763–772CrossRefPubMedPubMedCentral
28.
go back to reference Wang W, Zhao YE, Qi L et al (2017) Prospectively ECG-triggered high-pitch coronary CT angiography at 70 kVp with 30ml contrast agent: an intraindividual comparison with sequential scanning at 120 kvp with 60ml contrast agent. Eur J Radiol 90:97–105CrossRefPubMed Wang W, Zhao YE, Qi L et al (2017) Prospectively ECG-triggered high-pitch coronary CT angiography at 70 kVp with 30ml contrast agent: an intraindividual comparison with sequential scanning at 120 kvp with 60ml contrast agent. Eur J Radiol 90:97–105CrossRefPubMed
29.
go back to reference Benz DC, Gräni C, Mikulicic F et al (2016) Adaptive statistical iterative reconstruction-V: impact on image quality in ultralow-dose coronary computed tomography angiography. J Comput Assist Tomogr 40(6):958–963CrossRefPubMed Benz DC, Gräni C, Mikulicic F et al (2016) Adaptive statistical iterative reconstruction-V: impact on image quality in ultralow-dose coronary computed tomography angiography. J Comput Assist Tomogr 40(6):958–963CrossRefPubMed
30.
go back to reference Kok M, Mihl C, Hendriks BM et al (2016) Optimizing contrast media application in coronary CT angiography at lower tube voltage: evaluation in a circulation phantom and sixty patients. Eur J Radiol 85(6):1068–1074CrossRefPubMed Kok M, Mihl C, Hendriks BM et al (2016) Optimizing contrast media application in coronary CT angiography at lower tube voltage: evaluation in a circulation phantom and sixty patients. Eur J Radiol 85(6):1068–1074CrossRefPubMed
31.
go back to reference Geschwind J, Dake M (1988) Abrams’ angiography: interventional radiology. Lippincott Williams & Wilkins (LWW), Philadelphia Geschwind J, Dake M (1988) Abrams’ angiography: interventional radiology. Lippincott Williams & Wilkins (LWW), Philadelphia
Metadata
Title
Reducing both radiation and contrast doses in coronary CT angiography in lean patients on a 16-cm wide-detector CT using 70 kVp and ASiR-V algorithm, in comparison with the conventional 100-kVp protocol
Authors
Yuhuan Chen
Zhentang Liu
Michelle Li
Yong Yu
Yongjun Jia
Guangming Ma
Zhijun Hu
DongHong Wei
Dou Li
Taiping He
Publication date
01-06-2019
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 6/2019
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5837-9

Other articles of this Issue 6/2019

European Radiology 6/2019 Go to the issue