Quality of basic life support when using different commercially available public access defibrillators

Authors: Michael P. Müller, Cynthia Poenicke, Maxi Kurth, Torsten Richter, Thea Koch, Carolin Eisold, Adrian Pfältzer, Axel R. Heller

Published in: Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine | Issue 1/2015

Login to get access

Abstract

Background

Basic life support (BLS) guidelines focus on chest compressions with a minimal no-flow fraction (NFF), early defibrillation, and a short perishock pause. By using an automated external defibrillator (AED) lay persons are guided through the process of attaching electrodes and initiating defibrillation. It is unclear, however, to what extent the voice instructions given by the AED might influence the quality of initial resuscitation.

Methods

Using a patient simulator, 8 different commercially available AEDs were evaluated within two different BLS scenarios (ventricular fibrillation vs. asystole). A BLS certified instructor acted according to the current European Resuscitation Council 2010 Guidelines and followed all of the AED voice prompts. In a second set of scenarios, the rescuer anticipated the appropriate actions and started already before the AED stopped speaking. A BLS scenario without AED served as the control. All scenarios were run three times.

Results

The time until the first chest compression was 25 ± 2 seconds without the AED and ranged from 50 ± 3 to 148 ± 13 seconds with the AED depending on the model used. The NFF was .26 ± .01 without the AED and between .37 ± .01 and .72 ± .01 when an AED was used. The perishock pause ranged from 12 ± 0 to 46 ± 0 seconds. The optimized sequence of actions reduced the NFF, which ranged now from .32 ± .01 to .41 ± .01, and the perishock pause ranging from 1 ± 1 to 19 ± 1 seconds.

Conclusions

Voice prompts given by commercially available AED merely meet the requirements of current evidence in basic life support. Furthermore, there is a significant difference between devices with regard to time until the first chest compression, perishock pause, no-flow fraction and other objective measures of the quality of BLS. However, the BLS quality may be improved with optimized handling of the AED. Thus, rescuers should be trained on the respective AED devices, and manufacturers should expend more effort in improving user guidance to shorten the NFF and perishock pause.
Appendix
Available only for authorised users
Literature
1.
go back to reference Aufderheide TP, Nichol G, Rea TD, Brown SP, Leroux BG, Pepe PE, et al. A trial of an impedance threshold device in out-of-hospital cardiac arrest. N Engl J Med. 2011;365:798–806.PubMedCentralPubMedCrossRef Aufderheide TP, Nichol G, Rea TD, Brown SP, Leroux BG, Pepe PE, et al. A trial of an impedance threshold device in out-of-hospital cardiac arrest. N Engl J Med. 2011;365:798–806.PubMedCentralPubMedCrossRef
2.
go back to reference Berg RA, Hemphill R, Abella BS, Aufderheide TP, Cave DM, Hazinski MF, et al. Part 5: adult basic life support: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2010;122:S685–705.PubMedCrossRef Berg RA, Hemphill R, Abella BS, Aufderheide TP, Cave DM, Hazinski MF, et al. Part 5: adult basic life support: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2010;122:S685–705.PubMedCrossRef
3.
go back to reference Nolan JP, Soar J, Zideman DA, Biarent D, Bossaert LL, Deakin C, et al. European Resuscitation Council Guidelines for Resuscitation 2010 Section 1. Executive summary. Resuscitation. 2010;81:1219–76.PubMedCrossRef Nolan JP, Soar J, Zideman DA, Biarent D, Bossaert LL, Deakin C, et al. European Resuscitation Council Guidelines for Resuscitation 2010 Section 1. Executive summary. Resuscitation. 2010;81:1219–76.PubMedCrossRef
4.
go back to reference Valenzuela TD, Roe DJ, Cretin S, Spaite DW, Larsen MP. Estimating effectiveness of cardiac arrest interventions: a logistic regression survival model. Circulation. 1997;96:3308–13.PubMedCrossRef Valenzuela TD, Roe DJ, Cretin S, Spaite DW, Larsen MP. Estimating effectiveness of cardiac arrest interventions: a logistic regression survival model. Circulation. 1997;96:3308–13.PubMedCrossRef
5.
go back to reference Valenzuela TD, Roe DJ, Nichol G, Clark LL, Spaite DW, Hardman RG. Outcomes of rapid defibrillation by security officers after cardiac arrest in casinos. N Engl J Med. 2000;343:1206–9.PubMedCrossRef Valenzuela TD, Roe DJ, Nichol G, Clark LL, Spaite DW, Hardman RG. Outcomes of rapid defibrillation by security officers after cardiac arrest in casinos. N Engl J Med. 2000;343:1206–9.PubMedCrossRef
6.
go back to reference Fleischhackl R, Losert H, Haugk M, Eisenburger P, Sterz F, Laggner AN, et al. Differing operational outcomes with six commercially available automated external defibrillators. Resuscitation. 2004;62:167–74.PubMedCrossRef Fleischhackl R, Losert H, Haugk M, Eisenburger P, Sterz F, Laggner AN, et al. Differing operational outcomes with six commercially available automated external defibrillators. Resuscitation. 2004;62:167–74.PubMedCrossRef
7.
go back to reference Vaillancourt C, Everson-Stewart S, Christenson J, Andrusiek D, Powell J, Nichol G, et al. The impact of increased chest compression fraction on return of spontaneous circulation for out-of-hospital cardiac arrest patients not in ventricular fibrillation. Resuscitation. 2011;82:1501–7.PubMedCentralPubMedCrossRef Vaillancourt C, Everson-Stewart S, Christenson J, Andrusiek D, Powell J, Nichol G, et al. The impact of increased chest compression fraction on return of spontaneous circulation for out-of-hospital cardiac arrest patients not in ventricular fibrillation. Resuscitation. 2011;82:1501–7.PubMedCentralPubMedCrossRef
8.
go back to reference Christenson J, Andrusiek D, Everson-Stewart S, Kudenchuk P, Hostler D, Powell J, et al. Chest compression fraction determines survival in patients with out-of-hospital ventricular fibrillation. Circulation. 2009;120:1241–7.PubMedCentralPubMedCrossRef Christenson J, Andrusiek D, Everson-Stewart S, Kudenchuk P, Hostler D, Powell J, et al. Chest compression fraction determines survival in patients with out-of-hospital ventricular fibrillation. Circulation. 2009;120:1241–7.PubMedCentralPubMedCrossRef
9.
go back to reference Rea TD, Stickney RE, Doherty A, Lank P. Performance of chest compressions by laypersons during the Public Access Defibrillation Trial. Resuscitation. 2010;81:293–6.PubMedCrossRef Rea TD, Stickney RE, Doherty A, Lank P. Performance of chest compressions by laypersons during the Public Access Defibrillation Trial. Resuscitation. 2010;81:293–6.PubMedCrossRef
10.
go back to reference Valenzuela TD, Kern KB, Clark LL, Berg RA, Berg MD, Berg DD, et al. Interruptions of chest compressions during emergency medical systems resuscitation. Circulation. 2005;112:1259–65.PubMedCrossRef Valenzuela TD, Kern KB, Clark LL, Berg RA, Berg MD, Berg DD, et al. Interruptions of chest compressions during emergency medical systems resuscitation. Circulation. 2005;112:1259–65.PubMedCrossRef
11.
go back to reference Wik L, Kramer-Johansen J, Myklebust H, Sørebø H, Svensson L, Fellows B, et al. Quality of cardiopulmonary resuscitation during out-of-hospital cardiac arrest. JAMA. 2005;293:299–304.PubMedCrossRef Wik L, Kramer-Johansen J, Myklebust H, Sørebø H, Svensson L, Fellows B, et al. Quality of cardiopulmonary resuscitation during out-of-hospital cardiac arrest. JAMA. 2005;293:299–304.PubMedCrossRef
12.
go back to reference Abella BS, Alvarado JP, Myklebust H, Edelson DP, Barry A, O’Hearn N, et al. Quality of cardiopulmonary resuscitation during in-hospital cardiac arrest. JAMA. 2005;293:305–10.PubMedCrossRef Abella BS, Alvarado JP, Myklebust H, Edelson DP, Barry A, O’Hearn N, et al. Quality of cardiopulmonary resuscitation during in-hospital cardiac arrest. JAMA. 2005;293:305–10.PubMedCrossRef
13.
go back to reference Müller MP, Richter T, Papkalla N, Poenicke C, Herkner C, Osmers A, et al. Effects of a mandatory basic life support training programme on the no-flow fraction during in-hospital cardiac resuscitation: An observational study. Resuscitation. 2014;85:874–8.PubMedCrossRef Müller MP, Richter T, Papkalla N, Poenicke C, Herkner C, Osmers A, et al. Effects of a mandatory basic life support training programme on the no-flow fraction during in-hospital cardiac resuscitation: An observational study. Resuscitation. 2014;85:874–8.PubMedCrossRef
14.
go back to reference Neukamm J, Gräsner J-T, Schewe J-C, Breil M, Bahr J, Heister U, et al. The impact of response time reliability on CPR incidence and resuscitation success: a benchmark study from the German Resuscitation Registry. Crit Care. 2011;15:R282.PubMedCentralPubMedCrossRef Neukamm J, Gräsner J-T, Schewe J-C, Breil M, Bahr J, Heister U, et al. The impact of response time reliability on CPR incidence and resuscitation success: a benchmark study from the German Resuscitation Registry. Crit Care. 2011;15:R282.PubMedCentralPubMedCrossRef
15.
go back to reference Meaney PA, Nadkarni VM, Kern KB, Indik JH, Halperin HR, Berg RA. Rhythms and outcomes of adult in-hospital cardiac arrest. Crit Care Med. 2010;38:101–8.PubMedCrossRef Meaney PA, Nadkarni VM, Kern KB, Indik JH, Halperin HR, Berg RA. Rhythms and outcomes of adult in-hospital cardiac arrest. Crit Care Med. 2010;38:101–8.PubMedCrossRef
16.
go back to reference Cheskes S, Schmicker RH, Christenson J, Salcido DD, Rea T, Powell J, et al. Perishock pause: an independent predictor of survival from out-of-hospital shockable cardiac arrest. Circulation. 2011;124:58–66.PubMedCentralPubMedCrossRef Cheskes S, Schmicker RH, Christenson J, Salcido DD, Rea T, Powell J, et al. Perishock pause: an independent predictor of survival from out-of-hospital shockable cardiac arrest. Circulation. 2011;124:58–66.PubMedCentralPubMedCrossRef
17.
go back to reference Berg MD, Clark LL, Valenzuela TD, Kern KB, Berg RA. Post-shock chest compression delays with automated external defibrillator use. Resuscitation. 2005;64:287–91.PubMedCrossRef Berg MD, Clark LL, Valenzuela TD, Kern KB, Berg RA. Post-shock chest compression delays with automated external defibrillator use. Resuscitation. 2005;64:287–91.PubMedCrossRef
18.
go back to reference Snyder D, Morgan C. Wide variation in cardiopulmonary resuscitation interruption intervals among commercially available automated external defibrillators may affect survival despite high defibrillation efficacy. Crit Care Med. 2004;32(Supplement):S421–4.PubMedCrossRef Snyder D, Morgan C. Wide variation in cardiopulmonary resuscitation interruption intervals among commercially available automated external defibrillators may affect survival despite high defibrillation efficacy. Crit Care Med. 2004;32(Supplement):S421–4.PubMedCrossRef
19.
go back to reference Gilmore CM, Rea TD, Becker LJ, Eisenberg MS. Three-Phase Model of Cardiac Arrest: Time-Dependent Benefit of Bystander Cardiopulmonary Resuscitation. Am J Cardiol. 2006;98:497–9.PubMedCrossRef Gilmore CM, Rea TD, Becker LJ, Eisenberg MS. Three-Phase Model of Cardiac Arrest: Time-Dependent Benefit of Bystander Cardiopulmonary Resuscitation. Am J Cardiol. 2006;98:497–9.PubMedCrossRef
Metadata
Title
Quality of basic life support when using different commercially available public access defibrillators
Authors
Michael P. Müller
Cynthia Poenicke
Maxi Kurth
Torsten Richter
Thea Koch
Carolin Eisold
Adrian Pfältzer
Axel R. Heller
Publication date
01-12-2015
Publisher
BioMed Central
DOI
https://doi.org/10.1186/s13049-015-0123-1

Other articles of this Issue 1/2015

Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine 1/2015 Go to the issue