Quality control of bedside DR in neonatal chest radiography using a chest stabilization device and its clinical application

Authors: Xinqi Zhang, Yinfeng Zhang

Published in: BMC Pediatrics | Issue 1/2020

Login to get access

Abstract

Background

To develop a method for movement control during radiation exposure and to improve image quality of bedside thoracic DR in neonates.

Methods

Total 60 cases of neonates’ thoracic DR X-ray images, which were taken before and after neonates’ movement control, were compared and analyzed. X-ray exposure was set at 47 kV/1.4 mAs for all films that were taken without movement control, while various exposure conditions were used based on the neonate’s body weight when the neonate’s movement was controlled.

Results

The radiation dose of X-ray exposure was significantly lower after neonates’ movement control (7.32 ± 0.20 μGy) than that before the movement control (24.20 ± 0.82 μGy, P <  0.05), and it was decreased most dramatically in the neonates with lowest body weight (70%). After neonates’ movement control, image quality was significantly improved (44 cases out of 60, 73.3%) compared to that before movement control (only 5 out 60, 8.3%, P <  0.05). There was no significant difference in the score of image background noise before and after movement control (P <  0.05).

Conclusion

Movement control with simple device could not only significantly improve the image quality, but also remarkably reduce radiation exposure dose.
Literature
1.
go back to reference Guo Q. Application of bedside DR in the neonate's chest X-ray examination. J Intellegence Health (Chinese). 2018;5:139–40. Guo Q. Application of bedside DR in the neonate's chest X-ray examination. J Intellegence Health (Chinese). 2018;5:139–40.
2.
go back to reference Lu H, Xuan J, Zhang Y. Application of wireless flat board detector in the analysis of image qualities of neonates bedside chest X-ray. J Qiqihaer Med Coll (Chinese). 2017;38(03):267–9. Lu H, Xuan J, Zhang Y. Application of wireless flat board detector in the analysis of image qualities of neonates bedside chest X-ray. J Qiqihaer Med Coll (Chinese). 2017;38(03):267–9.
3.
go back to reference Xiang R, Zhang X. Application of graded CR exposure in the bedside chest X-ray in the neonates. Modern Med Hygiene. 2014;5:736–7. Xiang R, Zhang X. Application of graded CR exposure in the bedside chest X-ray in the neonates. Modern Med Hygiene. 2014;5:736–7.
4.
go back to reference Lin Q. Application of portable and digital X-ray device in the neonate bedside chest X-ray filming. Modern Med. 2014;21:151–2. Lin Q. Application of portable and digital X-ray device in the neonate bedside chest X-ray filming. Modern Med. 2014;21:151–2.
5.
go back to reference Zhang X, Zhang Y, Cui S. Application of low dose chest DR in clinic. J Hebei Med Univ. 2018;39:1316–20. Zhang X, Zhang Y, Cui S. Application of low dose chest DR in clinic. J Hebei Med Univ. 2018;39:1316–20.
6.
go back to reference Xu X, Hua H. Effect of automatic exposure and two-factor-control model on the chest DR X-ray image and radiation dose. Radiology Practice (Chinese). 2017;32:79–81. Xu X, Hua H. Effect of automatic exposure and two-factor-control model on the chest DR X-ray image and radiation dose. Radiology Practice (Chinese). 2017;32:79–81.
7.
go back to reference Wang G. Application of bedside DR chest X-ray in the neonate CCU. Domest Abroad Med. 2013;4:183–5. Wang G. Application of bedside DR chest X-ray in the neonate CCU. Domest Abroad Med. 2013;4:183–5.
8.
go back to reference Brenner D, Elliston C, Hall E, Berdon W. Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol. 2001;176:289–96.CrossRef Brenner D, Elliston C, Hall E, Berdon W. Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol. 2001;176:289–96.CrossRef
9.
go back to reference Berrington de Gonzalez A, Darby S. Risk of cancer from diagnostic X-rays: estimates for the UK and 14 other countries. Lancet. 2004;363:345–51.CrossRef Berrington de Gonzalez A, Darby S. Risk of cancer from diagnostic X-rays: estimates for the UK and 14 other countries. Lancet. 2004;363:345–51.CrossRef
10.
go back to reference Zhu J, Li H, Cai J. Application of Iterative Algorithms in the spiral CT image formation in neonate portal vein. Chin J Med Phys. 2015;32:728–32. Zhu J, Li H, Cai J. Application of Iterative Algorithms in the spiral CT image formation in neonate portal vein. Chin J Med Phys. 2015;32:728–32.
11.
go back to reference Zhang M, Li X, Li J. Evaluation on the factors associated with neonates bedside chest X-ray image quality. J Pract Medicine Image. 2012;13:359–62. Zhang M, Li X, Li J. Evaluation on the factors associated with neonates bedside chest X-ray image quality. J Pract Medicine Image. 2012;13:359–62.
Metadata
Title
Quality control of bedside DR in neonatal chest radiography using a chest stabilization device and its clinical application
Authors
Xinqi Zhang
Yinfeng Zhang
Publication date
01-12-2020
Publisher
BioMed Central
Published in
BMC Pediatrics / Issue 1/2020
Electronic ISSN: 1471-2431
DOI
https://doi.org/10.1186/s12887-020-02228-0

Other articles of this Issue 1/2020

BMC Pediatrics 1/2020 Go to the issue