Published in: European Journal of Nuclear Medicine and Molecular Imaging 3/2020

01-03-2020 | Pyelonephritis | Short Communication

Planar scan vs. SPECT/low-dose CT for estimating split renal function by 99mTc-DMSA scintigraphy in children

Authors: Michala Holm Reichkendler, Ronan M. G. Berg, Robin de Nijs, Hanne Nørgaard, Ida Maria Schmidt, Lise Borgwardt

Published in: European Journal of Nuclear Medicine and Molecular Imaging | Issue 3/2020

Login to get access

Abstract

In the present study, we compared estimates of split renal function (SRF) in paediatric patients of various diagnostic subgroups by 99mTc-dimercaptosuccinic acid (DMSA) scintigraphy using either geometric mean (GM) based on planar scans or a volume of interest (VOI)-based analysis on single photon emission tomography combined with low-dose CT (SPECT/ldCT). Two experienced physicians blinded to patient diagnosis retrospectively analysed all paediatric 99mTc-DMSA scintigraphies that were conducted in our department between 2011 and 2016 and which included both a planar scan and SPECT/ldCT. All scintigraphies were performed on either a Phillips Precedence 16 slice CT or a Siemens Symbia 16 slice CT. SRF was estimated from planar scintigraphy using the geometric mean (GM), while the VOI-based analysis (VBA) was used for kidney segmentation on SPECT/ldCT.

Results

A total of 68 scintigraphies were included. A Bland-Altman plot-based analysis showed a bias for SRF of 2.1% with limits of agreement from − 7.5 to + 11.7% for the whole data set but showed larger differences between the two methods outside the normal range of 45–55%. In the GM-based SRF analyses, 29 cases were found to be outside the normal range, and in seven of these, VBA showed normal SRF. In the remaining 39 cases, VBA showed an abnormal SRF in only one case.

Conclusion

Approximately a quarter of planar DMSA scintigraphies that show an abnormal SRF in paediatric patients may be normal when assessed by SPECT/ldCT, which likely reflects underestimation of the kidney with the poorest function when assessed by GM due to the lack of attenuation correction. Planar scans that show an abnormal SRF in paediatric patients should thus preferably be supplemented by SPECT/ldCT.
Literature
2.
go back to reference Mendichovszky I, Solar BT, Smeulders N, Easty M, Biassoni L. Nuclear medicine in pediatric nephro-urology: an overview. Semin Nucl Med. 2017;47:204–28.CrossRef Mendichovszky I, Solar BT, Smeulders N, Easty M, Biassoni L. Nuclear medicine in pediatric nephro-urology: an overview. Semin Nucl Med. 2017;47:204–28.CrossRef
3.
go back to reference Lim R. Vesicoureteral reflux and urinary tract infection: evolving practices and current controversies in pediatric imaging. AJR Am J Roentgenol. 2009;192:1197–208.CrossRef Lim R. Vesicoureteral reflux and urinary tract infection: evolving practices and current controversies in pediatric imaging. AJR Am J Roentgenol. 2009;192:1197–208.CrossRef
4.
go back to reference Brenner M, Bonta D, Eslamy H, Ziessman HA. Comparison of 99mTc-DMSA dual-head SPECT versus high-resolution parallel-hole planar imaging for the detection of renal cortical defects. AJR Am J Roentgenol. 2009;193:333–7.CrossRef Brenner M, Bonta D, Eslamy H, Ziessman HA. Comparison of 99mTc-DMSA dual-head SPECT versus high-resolution parallel-hole planar imaging for the detection of renal cortical defects. AJR Am J Roentgenol. 2009;193:333–7.CrossRef
5.
go back to reference Chroustová D, Trnka J, Šírová V, Urbanová I, Langer J, Kubinyi J. Comparison of planar DMSA scan with an evaluation based on SPECT imaging in the split renal function assessment. Nucl Med Rev Cent East Eur. 2016;19:12–7.CrossRef Chroustová D, Trnka J, Šírová V, Urbanová I, Langer J, Kubinyi J. Comparison of planar DMSA scan with an evaluation based on SPECT imaging in the split renal function assessment. Nucl Med Rev Cent East Eur. 2016;19:12–7.CrossRef
7.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRef Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRef
8.
go back to reference Rutland MD. A comprehensive analysis of renal DTPA studies. I. Theory and normal values. Nucl Med Commun. 1985;6(1):11–20.CrossRef Rutland MD. A comprehensive analysis of renal DTPA studies. I. Theory and normal values. Nucl Med Commun. 1985;6(1):11–20.CrossRef
9.
go back to reference Cao X, Xu X, Grant FD, Treves ST. Estimation of split renal function with 99mTc-DMSA SPECT: comparison between 3D volumetric assessment and 2D coronal projection imaging. AJR Am J Roentgenol. 2016;207:1324–8.CrossRef Cao X, Xu X, Grant FD, Treves ST. Estimation of split renal function with 99mTc-DMSA SPECT: comparison between 3D volumetric assessment and 2D coronal projection imaging. AJR Am J Roentgenol. 2016;207:1324–8.CrossRef
Metadata
Title
Planar scan vs. SPECT/low-dose CT for estimating split renal function by 99mTc-DMSA scintigraphy in children
Authors
Michala Holm Reichkendler
Ronan M. G. Berg
Robin de Nijs
Hanne Nørgaard
Ida Maria Schmidt
Lise Borgwardt
Publication date
01-03-2020
Publisher
Springer Berlin Heidelberg
Published in
European Journal of Nuclear Medicine and Molecular Imaging / Issue 3/2020
Print ISSN: 1619-7070
Electronic ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-019-04575-2

Other articles of this Issue 3/2020

European Journal of Nuclear Medicine and Molecular Imaging 3/2020 Go to the issue