Potential of dosage reduction in cone-beam-computed tomography (CBCT) for radiological diagnostics of the paranasal sinuses

Authors: C. Güldner, A. Ningo, J. Voigt, I. Diogo, J. Heinrichs, R. Weber, T. Wilhelm, M. Fiebich

Published in: European Archives of Oto-Rhino-Laryngology | Issue 4/2013

Login to get access

Abstract

More than 10 years ago, cone-beam-computed tomography (CBCT) was introduced in ENT radiology. Until now, the focus of research was to evaluate clinical limits of this technique. The aim of this work is the evaluation of specific dosages and the identification of potential optimization in the performance of CBCT of the paranasal sinuses. Based on different tube parameters (tube current, tube voltage, and rotation angles), images of the nose and the paranasal sinuses were taken on a phantom head with the Accu-I-tomo F17 (Morita, Kyoto, Japan). The dosages applied to the lens and parotid gland were measured with OSL dosimetry. The imaging quality was evaluated by independent observers. All datasets were reviewed according to a checklist of surgically important anatomic structures. Even for lowest radiation exposure (4 mA, 76 kV, 180°, computed tomography dosage index (CTDI) = 1.8 mGy), the imaging quality was sufficient. Of course a significant reduction of the imaging quality could be seen, so a reliable mean was set for 4 mA, 84 kV, and 180° rotation angle (CTDI = 2.4 mGy). In this combination, a reduction of 92 % in lens-dose and of 77 % of dosage at the parotid gland was observed in comparison to the maximal possible adjustments (8 mA, 90 kV, 360°, CTDI = 10.9 mGy). There is potential for optimization in CBCT. Changing the rotation angle (180° instead of 360°) leads to a dose reduction of 50 %. Furthermore from clinical point of view in case of chronic rhinosinusitis a relevant reduction of dosage is possible. Therefore, it is necessary to intensify the interdisciplinary discussion about the disease specifics required quality of imaging.
Literature
1.
go back to reference Stuck BA, C. Bachert P. Federspil et al. (2011) Rhinosinusitis guidelines-unabridged version: S2 guidelines from the German Society of Otorhinolaryngology, Head and Neck Surgery. HNO Stuck BA, C. Bachert P. Federspil et al. (2011) Rhinosinusitis guidelines-unabridged version: S2 guidelines from the German Society of Otorhinolaryngology, Head and Neck Surgery. HNO
2.
go back to reference Bremke M, R. Leppek and J.A. Werner (2010) Digital volume tomography in ENT medicine. HNO Bremke M, R. Leppek and J.A. Werner (2010) Digital volume tomography in ENT medicine. HNO
3.
go back to reference Güldner C, Diogo I, Windfuhr J et al (2011) Analysis of the fossa olfactoria using cone beam tomography (CBT). Acta Otolaryngol 131(1):72–78PubMedCrossRef Güldner C, Diogo I, Windfuhr J et al (2011) Analysis of the fossa olfactoria using cone beam tomography (CBT). Acta Otolaryngol 131(1):72–78PubMedCrossRef
4.
go back to reference Kurzweg T, Dalchow CV, Bremke M et al (2011) The value of digital volume tomography in assessing the position of cochlear implant arrays in temporal bone specimens. Ear Hear 31(3):413–419CrossRef Kurzweg T, Dalchow CV, Bremke M et al (2011) The value of digital volume tomography in assessing the position of cochlear implant arrays in temporal bone specimens. Ear Hear 31(3):413–419CrossRef
5.
go back to reference Savvateeva DM, Güldner C, Murthum T et al (2010) Digital volume tomography (DVT) measurements of the olfactory cleft and olfactory fossa. Acta Otolaryngol 130(3):398–404PubMedCrossRef Savvateeva DM, Güldner C, Murthum T et al (2010) Digital volume tomography (DVT) measurements of the olfactory cleft and olfactory fossa. Acta Otolaryngol 130(3):398–404PubMedCrossRef
6.
go back to reference Offergeld C, Kromeier J, Aschendorff A et al (2007) Rotational tomography of the normal and reconstructed middle ear in temporal bones: an experimental study. Eur Arch Otorhinolaryngol 264(4):345–351PubMedCrossRef Offergeld C, Kromeier J, Aschendorff A et al (2007) Rotational tomography of the normal and reconstructed middle ear in temporal bones: an experimental study. Eur Arch Otorhinolaryngol 264(4):345–351PubMedCrossRef
7.
go back to reference Kyriakou Y, Kolditz D, Langner O, Krause J, Kalender W (2011) Digital volume tomography (DVT) and multislice spiral CT (MSCT): an objective examination of dose and image quality. Rofo 183:144–153PubMedCrossRef Kyriakou Y, Kolditz D, Langner O, Krause J, Kalender W (2011) Digital volume tomography (DVT) and multislice spiral CT (MSCT): an objective examination of dose and image quality. Rofo 183:144–153PubMedCrossRef
8.
go back to reference Yukihara EG, McKeever SW (2008) Optically stimulated luminescence (OSL) dosimetry in medicine. Phys Med Biol 53(20):R351–R379PubMedCrossRef Yukihara EG, McKeever SW (2008) Optically stimulated luminescence (OSL) dosimetry in medicine. Phys Med Biol 53(20):R351–R379PubMedCrossRef
9.
go back to reference Al-Senan RM, Hatab MR (2011) Characteristics of an OSLD in the diagnostic energy range. Med Phys 38(7):4396–4405PubMedCrossRef Al-Senan RM, Hatab MR (2011) Characteristics of an OSLD in the diagnostic energy range. Med Phys 38(7):4396–4405PubMedCrossRef
10.
go back to reference Damet J, Sans-Merce M, Mieville F et al (2010) Comparison of organ doses and image quality between CT and flat panel XperCT scans in wrist and inner ear examinations. Radiat Prot Dosimetry 139(1–3):164–168PubMedCrossRef Damet J, Sans-Merce M, Mieville F et al (2010) Comparison of organ doses and image quality between CT and flat panel XperCT scans in wrist and inner ear examinations. Radiat Prot Dosimetry 139(1–3):164–168PubMedCrossRef
11.
go back to reference Zuckerkandl E (1882) Normale und pathologische Anatomie der Nasenhöhle und ihrer pneumatischen Anhänge. In: W. Braumüller (ed) vol. 1. Vienna Zuckerkandl E (1882) Normale und pathologische Anatomie der Nasenhöhle und ihrer pneumatischen Anhänge. In: W. Braumüller (ed) vol. 1. Vienna
12.
go back to reference Keros P (1962) On the practical value of differences in the level of the lamina cribrosa of the ethmoid. Z Laryngol Rhinol Otol 41:809–813PubMed Keros P (1962) On the practical value of differences in the level of the lamina cribrosa of the ethmoid. Z Laryngol Rhinol Otol 41:809–813PubMed
13.
go back to reference Kainz J, Stammberger H (1988) The roof of the anterior ethmoid: a locus minoris resistentiae in the skull base. Laryngol Rhinol Otol (Stuttg) 67(4):142–149CrossRef Kainz J, Stammberger H (1988) The roof of the anterior ethmoid: a locus minoris resistentiae in the skull base. Laryngol Rhinol Otol (Stuttg) 67(4):142–149CrossRef
14.
go back to reference Ohnishi T, Yanagisawa E (1994) Endoscopic anatomy of the anterior ethmoidal artery. Ear Nose Throat J 73(9):634–636PubMed Ohnishi T, Yanagisawa E (1994) Endoscopic anatomy of the anterior ethmoidal artery. Ear Nose Throat J 73(9):634–636PubMed
15.
go back to reference Simmen D, Raghavan U, Briner HR et al (2006) The surgeon’s view of the anterior ethmoid artery. Clin Otolaryngol 31(3):187–191PubMedCrossRef Simmen D, Raghavan U, Briner HR et al (2006) The surgeon’s view of the anterior ethmoid artery. Clin Otolaryngol 31(3):187–191PubMedCrossRef
16.
go back to reference Bremke M, Wiegand S, Sesterhenn AM et al (2007) Digital volume tomography in the diagnosis of nasal bone fractures. Rhinology 47(2):126–131 Bremke M, Wiegand S, Sesterhenn AM et al (2007) Digital volume tomography in the diagnosis of nasal bone fractures. Rhinology 47(2):126–131
17.
go back to reference Bremke M, Sesterhenn AM, Murthum T et al (2009) Digital volume tomography (DVT) as a diagnostic modality of the anterior skull base. Acta Otolaryngol 129(10):1106–1114PubMedCrossRef Bremke M, Sesterhenn AM, Murthum T et al (2009) Digital volume tomography (DVT) as a diagnostic modality of the anterior skull base. Acta Otolaryngol 129(10):1106–1114PubMedCrossRef
18.
go back to reference Balbach L, Trinkel V, Guldner C et al (2011) Radiological examinations of the anatomy of the inferior turbinate using digital volume tomography (DVT). Rhinology 49(2):248–252PubMed Balbach L, Trinkel V, Guldner C et al (2011) Radiological examinations of the anatomy of the inferior turbinate using digital volume tomography (DVT). Rhinology 49(2):248–252PubMed
19.
go back to reference ICRP (2007) The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP 37(2–4):1–332 ICRP (2007) The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP 37(2–4):1–332
20.
go back to reference Prins R, Dauer LT, Colosi DC et al (2011) Significant reduction in dental cone beam computed tomography (CBCT) eye dose through the use of leaded glasses. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 112(4):502–507PubMedCrossRef Prins R, Dauer LT, Colosi DC et al (2011) Significant reduction in dental cone beam computed tomography (CBCT) eye dose through the use of leaded glasses. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 112(4):502–507PubMedCrossRef
Metadata
Title
Potential of dosage reduction in cone-beam-computed tomography (CBCT) for radiological diagnostics of the paranasal sinuses
Authors
C. Güldner
A. Ningo
J. Voigt
I. Diogo
J. Heinrichs
R. Weber
T. Wilhelm
M. Fiebich
Publication date
01-03-2013
Publisher
Springer-Verlag
Published in
European Archives of Oto-Rhino-Laryngology / Issue 4/2013
Print ISSN: 0937-4477
Electronic ISSN: 1434-4726
DOI
https://doi.org/10.1007/s00405-012-2177-2

Other articles of this Issue 4/2013

European Archives of Oto-Rhino-Laryngology 4/2013 Go to the issue