01-10-2001 | Leading Article

Post-Marketing Surveillance and Vigilance for Medical Devices

The European Approach

Author: Dr Hazel Randall

Published in: Drug Safety | Issue 12/2001

Login to get access

Abstract

The extent to which the medical device manufacturers are responsible for actively monitoring the performance of their products after they have successfully passed the rigorous pre-market approval process has always been a matter of diverse opinion. Within Europe, the law is unhelpfully vague on this point. While there are some comparatively clear obligations for reporting incidents to the authorities (known as the ‘vigilance system’), little detail is given on how diligently the manufacturer should try to find out about such incidents.
In the early stages of the European Community Directives covering medical devices, there was much emphasis upon formulating guidance to help interpret the vigilance reporting requirements. It is, however, only recently that attention has turned to attempting to clarify what is expected from post-marketing surveillance (PMS) in its broader sense.
This article discuses both the vigilance and PMS processes and outlines the currently available European, and particularly UK, guidance documents which are aimed at promoting a more level playing field across industry where these activities are concerned. In particular, it explains the principle differences between vigilance and post-marketing surveillance: the former being the reporting of adverse incidents by manufacturers to the regulatory authorities and their subsequent sharing of key incident data between each other; the latter being the process by which information on overall devise performance is captured, analysed and acted upon. Nevertheless, it is still a struggle to gain widespread appreciation that these two activities are not in fact one and the same.
Literature
1.
go back to reference Council Directive: 90/285/EEC concerning Active Implantable Medical Devices, OJ L189; 1990 Jun 10 Council Directive: 90/285/EEC concerning Active Implantable Medical Devices, OJ L189; 1990 Jun 10
2.
go back to reference Council Directive: 93/42/EEC concerning Medical Devices, OJ L169; Volume 36; 1993 Jul 12 Council Directive: 93/42/EEC concerning Medical Devices, OJ L169; Volume 36; 1993 Jul 12
3.
go back to reference Council Directive: 98/79/EC concerning In Vitro Diagnostic Medical Devices OJ L331/1; 1998 Oct 27 Council Directive: 98/79/EC concerning In Vitro Diagnostic Medical Devices OJ L331/1; 1998 Oct 27
4.
go back to reference The European Commission guidelines on a medical devices vigilance system, (MEDDEV 2.12/1). 3 rev. 1998 Mar The European Commission guidelines on a medical devices vigilance system, (MEDDEV 2.12/1). 3 rev. 1998 Mar
5.
go back to reference Guidance on the medical devices vigilance system for CE marked joint replacement implants. London: Medical Devices Agency; 1998 Sep. MDA Publication Guidance on the medical devices vigilance system for CE marked joint replacement implants. London: Medical Devices Agency; 1998 Sep. MDA Publication
6.
go back to reference Guidance on the medical devices vigilance system for CE marked artificial heart valves. Version 1. London: Medical Devices Agency; 1998 Oct. MDA Publication Guidance on the medical devices vigilance system for CE marked artificial heart valves. Version 1. London: Medical Devices Agency; 1998 Oct. MDA Publication
7.
go back to reference Guidance on the medical devices vigilance system for CE marked breast implants. Version 1. London:Medical Devices Agency; 1999 Aug. MDA Publication Guidance on the medical devices vigilance system for CE marked breast implants. Version 1. London:Medical Devices Agency; 1999 Aug. MDA Publication
8.
go back to reference Guidance on the medical devices vigilance system for CE marked coronary stents. Version 1. London:Medical Devices Agency; 2001 Jan. MDA Publication Guidance on the medical devices vigilance system for CE marked coronary stents. Version 1. London:Medical Devices Agency; 2001 Jan. MDA Publication
9.
go back to reference Post-marketing surveillance (PMS) post market/production. NB-MED/2.12/Rec 1 Rev 11: (Accepted 2000 Feb 29) Post-marketing surveillance (PMS) post market/production. NB-MED/2.12/Rec 1 Rev 11: (Accepted 2000 Feb 29)
10.
go back to reference BS EN 46001: 1994 application of EN 29001 (BS 5750: part 1) to the manufacture of medical devices. BS EN 46001: 1994 application of EN 29001 (BS 5750: part 1) to the manufacture of medical devices.
11.
go back to reference BS EN 9004-1: 1994 quality management and quality system elements part 1 guidelines. BS EN 9004-1: 1994 quality management and quality system elements part 1 guidelines.
12.
go back to reference BS EN 50103: 1996 guidance on the application of EN 29001 and EN46001 and of EN 29002 and EN 46002 for the active (including active implantable) medical device industry. BS EN 50103: 1996 guidance on the application of EN 29001 and EN46001 and of EN 29002 and EN 46002 for the active (including active implantable) medical device industry.
13.
go back to reference Post-market surveillance of CE marked joint replacement implants including guidance to manufacturers on post-market clinical studies. London:Medical Devices Agency; 2000 Sep. MDA Publication Post-market surveillance of CE marked joint replacement implants including guidance to manufacturers on post-market clinical studies. London:Medical Devices Agency; 2000 Sep. MDA Publication
Metadata
Title
Post-Marketing Surveillance and Vigilance for Medical Devices
The European Approach
Author
Dr Hazel Randall
Publication date
01-10-2001
Publisher
Springer International Publishing
Published in
Drug Safety / Issue 12/2001
Print ISSN: 0114-5916
Electronic ISSN: 1179-1942
DOI
https://doi.org/10.2165/00002018-200124120-00001

Other articles of this Issue 12/2001

Drug Safety 12/2001 Go to the issue