Published in: Journal of Orthopaedic Surgery and Research 1/2021

Open Access 01-12-2021 | Research article

Polymethyl methacrylate cure time in simulated in vivo total knee arthroplasty versus in vitro conditions

Authors: Daniel A. Funk, Quang-Viet Nguyen, Michael Swank

Published in: Journal of Orthopaedic Surgery and Research | Issue 1/2021

Login to get access

Abstract

Background

The present means of confirming the cure of intra-operative polymethyl methacrylate (PMMA) cement are to wait for the remainder cement to harden. To our knowledge, there is no available technique to determine the precise moment of cure for in-vivo cement beneath the tibial tray. This study uses a novel means to determine cement curing time in two environments. One environment represents the operating theater, and the other environment attempts to model cement conditions under the tibial tray during surgery.

Materials and methods

We determined the temperature-versus-time plot of cement curing using the following two temperature sensors: one in a simulated implanted tibial tray and another in the remainder cement. We performed 55 tests using dental methyl methacrylate cement mixed in the same ratio as the orthopedic cement. To simulate in vivo conditions, a simulated stainless-steel tibial tray was implanted on a cancellous bone substitute (Sawbones, Vashon Island, WA, USA) using standard cement technique and subsequently placed in a 90°F (32.2 °C) circulating water bath. We positioned a temperature sensor in the cement mantel and positioned a second sensor in a portion of the remaining cement. The temperature from both sensors was measured simultaneously, beginning at 5 min after mixing and continuing for 20 min. The first derivative of the temperature provided the precise curing time for each condition. We analyzed the results of 55 repeated experiments with an independent samples t-test.

Results

With the described technique, we were able to accurately determine the moment of cure of the cement beneath the simulated tray. There was a mean difference between cure time of 5 min and 26 s (p value < 0.001) between the two conditions.

Conclusions

We validated that our technique was successful in determining the precise time to cure in two different environments.

Level of evidence

This was not a clinical trial and did not involve patients as such the level of evidence was Grade A: Consistent 1 and 2.
Literature
1.
go back to reference ASTM International. Designation: F451-16, standard specification for acrylic bone cement. West Conshohocken, PA: ASTM International; 2019. ASTM International. Designation: F451-16, standard specification for acrylic bone cement. West Conshohocken, PA: ASTM International; 2019.
2.
go back to reference DePuySynthes. Guidance for cementing total knee replacements. Warsaw, IN: DePuy Orthopedics; 2018. DePuySynthes. Guidance for cementing total knee replacements. Warsaw, IN: DePuy Orthopedics; 2018.
3.
go back to reference Billi F, Kavanaugh A, Schmalzried H, Schmalzried T. Factors influencing the initial strength of the tibial tray–PMMA cement bond [poster number 1854]. ORS annual meeting; 2014. Billi F, Kavanaugh A, Schmalzried H, Schmalzried T. Factors influencing the initial strength of the tibial tray–PMMA cement bond [poster number 1854]. ORS annual meeting; 2014.
4.
go back to reference Maag C, Peckenpaugh E, Metcalfe A, Langhorn J, Heldreth M. Influence of intra-operative lipid/marrow infiltration and intra-operative motion upon cemented tibial implant fixation [poster no. 1788]. ORS annual meeting; 2019. Maag C, Peckenpaugh E, Metcalfe A, Langhorn J, Heldreth M. Influence of intra-operative lipid/marrow infiltration and intra-operative motion upon cemented tibial implant fixation [poster no. 1788]. ORS annual meeting; 2019.
5.
go back to reference Cohen J. Statistical power analysis for the behavioral sciences. 2nd ed. Hillside: Lawrence Erlbaum Associates; 1988. Cohen J. Statistical power analysis for the behavioral sciences. 2nd ed. Hillside: Lawrence Erlbaum Associates; 1988.
6.
go back to reference Field A. Discovering statistics using IBM SPSS statistics. London: Sage; 2013. Field A. Discovering statistics using IBM SPSS statistics. London: Sage; 2013.
7.
go back to reference Meyer PR Jr, Lautenschlager EP, Moore BK. On the setting properties of acrylic bone cement. J Bone Joint Surg Am. 1973;55:149–56.CrossRef Meyer PR Jr, Lautenschlager EP, Moore BK. On the setting properties of acrylic bone cement. J Bone Joint Surg Am. 1973;55:149–56.CrossRef
8.
go back to reference Dunne NJ, Orr JF. Curing characteristics of acrylic bone cement. J Mater Sci Mater Med. 2002;13:17–22.CrossRef Dunne NJ, Orr JF. Curing characteristics of acrylic bone cement. J Mater Sci Mater Med. 2002;13:17–22.CrossRef
9.
go back to reference Li C, Mason J, Yakimicki D. Thermal characterization of PMMA-based bone cement curing. J Mater Sci Mater Med. 2004;15:85–9.CrossRef Li C, Mason J, Yakimicki D. Thermal characterization of PMMA-based bone cement curing. J Mater Sci Mater Med. 2004;15:85–9.CrossRef
10.
go back to reference Kühn K-D. PMMA cements. Berlin: Springer; 2014. p. 115–24. Kühn K-D. PMMA cements. Berlin: Springer; 2014. p. 115–24.
11.
go back to reference DiPisa JA, Sih GS, Berman AT. The temperature problem at the bone-acrylic cement interface of the total hip replacement. Clin Orthop Relat Res. 1976;121:95–8. DiPisa JA, Sih GS, Berman AT. The temperature problem at the bone-acrylic cement interface of the total hip replacement. Clin Orthop Relat Res. 1976;121:95–8.
12.
go back to reference Linder L. Reaction of bone to the acute chemical trauma of bone cement. J Bone Joint Surg Am. 1977;59:82–7.CrossRef Linder L. Reaction of bone to the acute chemical trauma of bone cement. J Bone Joint Surg Am. 1977;59:82–7.CrossRef
13.
go back to reference Deramond H, Wright NT, Belkoff SM. Temperature elevation caused by bone cement polymerization during vertebroplasty. Bone. 1999;25:17S-21S.CrossRef Deramond H, Wright NT, Belkoff SM. Temperature elevation caused by bone cement polymerization during vertebroplasty. Bone. 1999;25:17S-21S.CrossRef
14.
go back to reference Harving S, Søballe K, Bünger C. A method for bone–cement interface thermometry. An in-vitro comparison between low temperature curing cement Palavit® and Surgical Simplex® P. Acta Orthop Scand. 1991;62:546–8.CrossRef Harving S, Søballe K, Bünger C. A method for bone–cement interface thermometry. An in-vitro comparison between low temperature curing cement Palavit® and Surgical Simplex® P. Acta Orthop Scand. 1991;62:546–8.CrossRef
15.
go back to reference Vertullo CJ, Zbrojkiewicz D, Vizesi F, Walsh WR. Thermal analysis of the tibial cement interface with modern cementing technique. Open Orthop J. 2016;10:19–25.CrossRef Vertullo CJ, Zbrojkiewicz D, Vizesi F, Walsh WR. Thermal analysis of the tibial cement interface with modern cementing technique. Open Orthop J. 2016;10:19–25.CrossRef
16.
go back to reference Li C, Schmid S, Mason J. Effects of pre-cooling and pre-heating procedures on cement polymerization and thermal osteonecrosis in cemented hip replacements. Med Eng Phys. 2003;25:559–64.CrossRef Li C, Schmid S, Mason J. Effects of pre-cooling and pre-heating procedures on cement polymerization and thermal osteonecrosis in cemented hip replacements. Med Eng Phys. 2003;25:559–64.CrossRef
17.
go back to reference Koh BT, Tan JH, Ramruttun AK, Wang W. Effect of storage temperature and equilibration time on polymethyl methacrylate (PMMA) bone cement polymerization in joint replacement surgery. J Orthop Surg Res. 2015;10:178.CrossRef Koh BT, Tan JH, Ramruttun AK, Wang W. Effect of storage temperature and equilibration time on polymethyl methacrylate (PMMA) bone cement polymerization in joint replacement surgery. J Orthop Surg Res. 2015;10:178.CrossRef
18.
go back to reference Seeger JB, Jaeger S, Bitsch RG, Mohr G, Rohner E, Clarius M. The effect of bone lavage on femoral cement penetration and interface temperature during Oxford unicompartmental knee arthroplasty with cement. J Bone Joint Surg Am. 2013;95:48–53.CrossRef Seeger JB, Jaeger S, Bitsch RG, Mohr G, Rohner E, Clarius M. The effect of bone lavage on femoral cement penetration and interface temperature during Oxford unicompartmental knee arthroplasty with cement. J Bone Joint Surg Am. 2013;95:48–53.CrossRef
19.
go back to reference Lombardi Jr AV, Berend KR, Adams JB. Why knee replacements fail in 2013: patient, surgeon, or implant? Bone Joint J. 2014;96-B:101–104. Lombardi Jr AV, Berend KR, Adams JB. Why knee replacements fail in 2013: patient, surgeon, or implant? Bone Joint J. 2014;96-B:101–104.
21.
go back to reference Arsoy D, Pagnano MW, Lewallen DG, Hanssen AD, Sierra RJ. Aseptic tibial debonding as a cause of early failure in a modern total knee arthroplasty design. Clin Orthop Relat Res. 2013;471:94–101.CrossRef Arsoy D, Pagnano MW, Lewallen DG, Hanssen AD, Sierra RJ. Aseptic tibial debonding as a cause of early failure in a modern total knee arthroplasty design. Clin Orthop Relat Res. 2013;471:94–101.CrossRef
22.
go back to reference Mason J, Behnam Y, Fehring T, Clary C. Simultaneous femoral and tibial cementation negatively effects tibial fixation in total knee arthroplasty [scientific exhibit 15]. AAOS annual meeting, 2018. Mason J, Behnam Y, Fehring T, Clary C. Simultaneous femoral and tibial cementation negatively effects tibial fixation in total knee arthroplasty [scientific exhibit 15]. AAOS annual meeting, 2018.
Metadata
Title
Polymethyl methacrylate cure time in simulated in vivo total knee arthroplasty versus in vitro conditions
Authors
Daniel A. Funk
Quang-Viet Nguyen
Michael Swank
Publication date
01-12-2021
Publisher
BioMed Central
Published in
Journal of Orthopaedic Surgery and Research / Issue 1/2021
Electronic ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-021-02790-y

Other articles of this Issue 1/2021

Journal of Orthopaedic Surgery and Research 1/2021 Go to the issue