Background

Surgical intervention for spontaneous pneumothorax typically includes mechanical pleurodesis that frequently utilizes a Bovie scratch pad given its universal presence, low cost and ease of use. The pad is folded on itself after dividing it in half, allowing easier passage through the smaller incisions.
However, unintentional foreign body retention may occur during the course of an operation leading to reoperations or even worse complications. This case is reported to raise awareness that dividing the scratch pad may allow the embedded radio-opaque marker to fall out and become retained as a foreign body.

Case presentation

The patient is a 41 year-old female who presented with shortness of breath secondary to spontaneous pneumothorax. Chest CT scan showed apical blebs. The patient underwent video assisted thorascopic surgery (VATS) with bleb resection and mechanical pleurodesis using a divided and folded bovie scratch pad. Postoperative chest x-ray showed a retained foreign body. Reoperation confirmed this to be the radio-opaque marker of the scratch pad and was removed. The patient did well and was discharged the following day.

Conclusion

Dividing the bovie scratch pad may damage and “weaken” the product allowing the radio- opaque marker to fall out during its use for pleurodesis and should be discouraged. Recommendation is made of using the scratch pad as a whole and not dividing it.

Graphical abstract

Retained radio-opaque marker of bovie scratch pad during VATS mechanical pleurodesis.
Literature
6.
go back to reference Society of Thoracic Surgery Database v.2.3 Training Manual. Vol 2.3.; 2017. Society of Thoracic Surgery Database v.2.3 Training Manual. Vol 2.3.; 2017.
Metadata
Title
Retention of Bovie scratch pad radio-opaque marker during VATS Pleurodesis: case report
Authors
Deena Akras
Daniel Raymond
Rami Akhrass
Sudish Murthy
Publication date
01-12-2021
Publisher
BioMed Central
Published in
Journal of Cardiothoracic Surgery / Issue 1/2021
Electronic ISSN: 1749-8090
DOI
https://doi.org/10.1186/s13019-021-01497-9

Other articles of this Issue 1/2021

Journal of Cardiothoracic Surgery 1/2021 Go to the issue