Needle aspiration for treating iatrogenic pneumothorax after cardiac electronic device implantation: a pilot study

Authors: Dominika Domokos, Andras Szabo, Gyongyver Banhegyi, Balazs Polgar, Zsolt Bari, Peter Bogyi, Istvan Marczell, Leticia Papp, Robert Gabor Kiss, Gabor Zoltan Duray, Bela Merkely, Istvan Hizoh

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 2/2020

Login to get access

Abstract

Purpose

Pneumothorax (PTX) following cardiac implantable electronic device procedures is traditionally treated with chest tube drainage (CTD). We hypothesized that, in a subset of patients, the less invasive needle aspiration (NA) may also be effective. We compared the strategy of primary NA with that of primary CTD in a single-center observational study.

Methods

Of the 970 procedures with subclavian venous access between January 2016 and June 2018, 23 patients had PTX requiring intervention. Beginning with March 2017, the traditional primary CTD (9 cases) has been replaced by the “NA first” strategy (14 patients). Outcome measures were procedural success rate and duration of hospitalization evaluated both as time to event (log-rank test) and as a discrete variable (Wilcoxon-Mann-Whitney test).

Results

Needle aspiration was successful in 8/14 (57.1%) of the cases (95% CI 28.9–82.3%), whereas PTX resolved in all patients after CTD was 9/9 (100%, 95% CI 66.4–100.0%, p = 0.0481). Regarding length of hospital stay, intention to treat time to event analysis showed no difference between the two approaches (p = 0.73). Also, the median difference was not statistically significant (− 2.0 days, p = 0.17). In contrast, per protocol evaluation revealed reduced risk of prolonged hospitalization for NA patients (p = 0.0025) with a median difference of − 4.0 days (p = 0.0012). Failure of NA did not result in a meaningful delay in discharge timing as median difference was 1.5 days (p = 0.28).

Conclusions

Our data suggest that in a number of patients iatrogenic PTX may be successfully treated with NA resulting in shorter hospitalization without the risk of meaningful discharge delay in unsuccessful cases.
Literature
15.
go back to reference Franco E, Rodriguez Muñoz D, Matía R, Hernandez-Madrid A, Carbonell San Román A, Sánchez I, et al. Wireless ultrasound-guided axillary vein cannulation for the implantation of cardiovascular implantable electric devices. J Cardiovasc Electrophysiol. 2016;27:482–7. https://doi.org/10.1111/jce.12917.CrossRefPubMed Franco E, Rodriguez Muñoz D, Matía R, Hernandez-Madrid A, Carbonell San Román A, Sánchez I, et al. Wireless ultrasound-guided axillary vein cannulation for the implantation of cardiovascular implantable electric devices. J Cardiovasc Electrophysiol. 2016;27:482–7. https://​doi.​org/​10.​1111/​jce.​12917.CrossRefPubMed
16.
go back to reference Andrivet P, Djedaini K, Teboul JL, Brochard L, Dreyfuss D. Spontaneous pneumothorax. Comparison of thoracic drainage vs immediate or delayed needle aspiration. Chest. 1995;108:335–9.CrossRef Andrivet P, Djedaini K, Teboul JL, Brochard L, Dreyfuss D. Spontaneous pneumothorax. Comparison of thoracic drainage vs immediate or delayed needle aspiration. Chest. 1995;108:335–9.CrossRef
Metadata
Title
Needle aspiration for treating iatrogenic pneumothorax after cardiac electronic device implantation: a pilot study
Authors
Dominika Domokos
Andras Szabo
Gyongyver Banhegyi
Balazs Polgar
Zsolt Bari
Peter Bogyi
Istvan Marczell
Leticia Papp
Robert Gabor Kiss
Gabor Zoltan Duray
Bela Merkely
Istvan Hizoh
Publication date
01-03-2020
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 2/2020
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-019-00596-x

Other articles of this Issue 2/2020

Journal of Interventional Cardiac Electrophysiology 2/2020 Go to the issue