Skip to main content
Top
Published in: Journal of Cardiothoracic Surgery 1/2014

Open Access 01-12-2014 | Research article

Performance of steroid eluting bipolar epicardial leads in pediatric and congenital heart disease patients: 15 years of single center experience

Authors: Christian Paech, Martin Kostelka, Ingo Dähnert, Patrick Flosdorff, Frank Thomas Riede, Roman Antonin Gebauer

Published in: Journal of Cardiothoracic Surgery | Issue 1/2014

Login to get access

Abstract

Objective

Cardiac pacing is sometimes required for patients with congenital heart disease for various reasons. Because of complex anatomy, repetitive previous heart surgery and patient size, epicardial leads are of special importance in these patients. Using epicardial leads has been discussed controversly regarding pacing performance and lead survival. The aim of this study was to review the long-term data on pacing performance as well as lead survival of epicardial leads implanted in our center.

Methods

Retrospective chart review and review of the literature.

Results

82 consecutive pediatric patients or adult patients with congenital heart disease with 158 epicardial leads (Medtronic 4968, bipolar, steroid - eluting) were included. We found 1.2% pacemaker-related early postoperative complications. The incidence of lead dysfunction was 7.5% (12/158) for primary (i.e. directly related to the lead itself) lead dysfunction and 3.2% (5/158) of lead abandonment for reasons not directly related to the lead itself. Primary lead dysfunction occured after a median of 3.83 years. Lead survival at 2, 5 and 10 years was 98.7%, 93% and 92.4%. There were no infections reported. Stable median measurements for impedance (RA/RV/LV of 577/483/610 Ohm), sensing threshold (RA/RV/LV of 2.0/11.0/10.0 mV) and pacing threshold (RA/RV/LV of 0.75 V at 0.4 ms/1.0 V at 0.49 ms/1.0 V at 0.45 ms) indicated a good mid- to longterm performance. The only risk factor for primary lead dysfunction was young age at implantation.

Conclusion

The use of epicardial leads in pediatric and adult patients with congenital heart disease shows good longterm outcomes in terms of pacing performance and lead survival. The authors encourage using epicardial leads in patients with congenital heart disease based on the patient‘s individual characteristics.
Appendix
Available only for authorised users
Literature
1.
go back to reference Zartner PA, Wiebe W, Toussaint-Goetz N, Schneider MB: Lead removal in young patients in view of lifelong pacing. Europace. 2010, 12: 714-718. 10.1093/europace/euq059.CrossRefPubMed Zartner PA, Wiebe W, Toussaint-Goetz N, Schneider MB: Lead removal in young patients in view of lifelong pacing. Europace. 2010, 12: 714-718. 10.1093/europace/euq059.CrossRefPubMed
2.
go back to reference Janousek J, van Geldorp IE, Krupicková S, Rosenthal E, Nugent K, Tomaske M, Früh A, Elders J, Hiippala A, Kerst G, Gebauer RA, Kubus P, Frias P, Gabbarini F, Clur SA, Nagel B, Ganame J, Papagiannis J, Marek J, Tisma-Dupanovic S, Tsao S, Nürnberg JH, Wren C, Friedberg M, de Guillebon M, Volaufova J, Prinzen FW, Delhaas T: Permanent cardiac pacing in children: choosing the optimal pacing site: a multicenter study. Working Group for Cardiac Dysrhythmias and Electrophysiology of the Association for European Pediatric Cardiology. Circulation. 2013, 127 (5): 613-623. 10.1161/CIRCULATIONAHA.112.115428.CrossRefPubMed Janousek J, van Geldorp IE, Krupicková S, Rosenthal E, Nugent K, Tomaske M, Früh A, Elders J, Hiippala A, Kerst G, Gebauer RA, Kubus P, Frias P, Gabbarini F, Clur SA, Nagel B, Ganame J, Papagiannis J, Marek J, Tisma-Dupanovic S, Tsao S, Nürnberg JH, Wren C, Friedberg M, de Guillebon M, Volaufova J, Prinzen FW, Delhaas T: Permanent cardiac pacing in children: choosing the optimal pacing site: a multicenter study. Working Group for Cardiac Dysrhythmias and Electrophysiology of the Association for European Pediatric Cardiology. Circulation. 2013, 127 (5): 613-623. 10.1161/CIRCULATIONAHA.112.115428.CrossRefPubMed
3.
go back to reference Riede FT, Kostelka M, Dähnert I: Cardiac strangulation: a rare, but devastating complication of epicardial pacing causing progressive myocardial ischaemia. Eur Heart J. 2009, 30 (4): 435-CrossRefPubMed Riede FT, Kostelka M, Dähnert I: Cardiac strangulation: a rare, but devastating complication of epicardial pacing causing progressive myocardial ischaemia. Eur Heart J. 2009, 30 (4): 435-CrossRefPubMed
4.
go back to reference Tomaske M, Gerritse B, Kretzers L, Pretre R, Dodge-Kathami A, Rahn M, Bauersfeld U: A 12-year experience of bipolar steroid-eluting epicardial pacing leads in children. Ann Thorac Surg. 2008, 85 (5): 1704-1711. 10.1016/j.athoracsur.2008.02.016.CrossRefPubMed Tomaske M, Gerritse B, Kretzers L, Pretre R, Dodge-Kathami A, Rahn M, Bauersfeld U: A 12-year experience of bipolar steroid-eluting epicardial pacing leads in children. Ann Thorac Surg. 2008, 85 (5): 1704-1711. 10.1016/j.athoracsur.2008.02.016.CrossRefPubMed
5.
go back to reference Walker F, Siu SC, Woods S, Cameron DA, Webb GD, HARRIS L: Long-term outcomes of cardiac pacing in adults with congenital heart disease. J Am Coll Cardiol. 2004, 43 (10): 1894-1901. 10.1016/j.jacc.2003.12.044.CrossRefPubMed Walker F, Siu SC, Woods S, Cameron DA, Webb GD, HARRIS L: Long-term outcomes of cardiac pacing in adults with congenital heart disease. J Am Coll Cardiol. 2004, 43 (10): 1894-1901. 10.1016/j.jacc.2003.12.044.CrossRefPubMed
6.
go back to reference Lichtenstein BJ, Bichell DP, Connolly DM, Lamberti JJ, Shepard SM, Seslar SP: Surgical approaches to epicardial pacemaker placement: does pocket location affect lead survival?. Pediatr Cardiol. 2010, 31 (7): 1016-1024. 10.1007/s00246-010-9754-1.CrossRefPubMedPubMedCentral Lichtenstein BJ, Bichell DP, Connolly DM, Lamberti JJ, Shepard SM, Seslar SP: Surgical approaches to epicardial pacemaker placement: does pocket location affect lead survival?. Pediatr Cardiol. 2010, 31 (7): 1016-1024. 10.1007/s00246-010-9754-1.CrossRefPubMedPubMedCentral
7.
go back to reference Silvetti MS: Twenty years of paediatric cardiac pacing: 515 pacemakers and 480 leads implanted in 292 patients. Europace. 2006, 8 (7): 530-536. 10.1093/europace/eul062.CrossRefPubMed Silvetti MS: Twenty years of paediatric cardiac pacing: 515 pacemakers and 480 leads implanted in 292 patients. Europace. 2006, 8 (7): 530-536. 10.1093/europace/eul062.CrossRefPubMed
8.
go back to reference Post MC, Budts W, Bruaene A, Willems R, Meyns B, Rega F, Gewillig M: Failure of epicardial pacing leads in congenital heart disease: not uncommon and difficult to predict. Neth Heart J. 2011, 19 (7–8): 331-335.CrossRefPubMedPubMedCentral Post MC, Budts W, Bruaene A, Willems R, Meyns B, Rega F, Gewillig M: Failure of epicardial pacing leads in congenital heart disease: not uncommon and difficult to predict. Neth Heart J. 2011, 19 (7–8): 331-335.CrossRefPubMedPubMedCentral
9.
go back to reference Murayama H, Maeda M, Sakurai H, Usui A, Ueda Y: Predictors affecting durability of epicardial pacemaker leads in pediatric patients. J Thorac Cardiovasc Surg. 2008, 135 (2): 361-366. 10.1016/j.jtcvs.2007.09.002.CrossRefPubMed Murayama H, Maeda M, Sakurai H, Usui A, Ueda Y: Predictors affecting durability of epicardial pacemaker leads in pediatric patients. J Thorac Cardiovasc Surg. 2008, 135 (2): 361-366. 10.1016/j.jtcvs.2007.09.002.CrossRefPubMed
10.
go back to reference Odim J, Suckow B, Saedi B, Laks H, Shannon K: Equivalent performance of epicardial versus endocardial permanent pacing in children: a single institution and manufacturer experience. Ann Thorac Surg. 2008, 85 (4): 1412-1416. 10.1016/j.athoracsur.2007.12.075.CrossRefPubMed Odim J, Suckow B, Saedi B, Laks H, Shannon K: Equivalent performance of epicardial versus endocardial permanent pacing in children: a single institution and manufacturer experience. Ann Thorac Surg. 2008, 85 (4): 1412-1416. 10.1016/j.athoracsur.2007.12.075.CrossRefPubMed
11.
go back to reference Fortescue EB, Berul CI, Cecchin F, Walsh EP, Triedman JK, Alexander ME: Comparison of modern steroid-eluting epicardial and thin transvenous pacemaker leads in pediatric and congenital heart disease patients. J Interv Card Electrophysiol. 2005, 14 (1): 27-36. 10.1007/s10840-005-3797-x.CrossRefPubMed Fortescue EB, Berul CI, Cecchin F, Walsh EP, Triedman JK, Alexander ME: Comparison of modern steroid-eluting epicardial and thin transvenous pacemaker leads in pediatric and congenital heart disease patients. J Interv Card Electrophysiol. 2005, 14 (1): 27-36. 10.1007/s10840-005-3797-x.CrossRefPubMed
12.
go back to reference Batra AS, Balaji S: Pacing in adults with congenital heart disease. Expert Rev Cardiovasc Ther. 2006, 4 (5): 663-670. 10.1586/14779072.4.5.663.CrossRefPubMed Batra AS, Balaji S: Pacing in adults with congenital heart disease. Expert Rev Cardiovasc Ther. 2006, 4 (5): 663-670. 10.1586/14779072.4.5.663.CrossRefPubMed
13.
go back to reference Kwak JG, Kim SJ, Song JY, Choi EY, Lee SY, Shim WS, Lee CH, Lee C, Park CS: Permanent epicardial pacing in pediatric patients: 12-year experience at a single center. Ann Thorac Surg. 2012, 93 (2): 634-639. 10.1016/j.athoracsur.2011.09.072.CrossRefPubMed Kwak JG, Kim SJ, Song JY, Choi EY, Lee SY, Shim WS, Lee CH, Lee C, Park CS: Permanent epicardial pacing in pediatric patients: 12-year experience at a single center. Ann Thorac Surg. 2012, 93 (2): 634-639. 10.1016/j.athoracsur.2011.09.072.CrossRefPubMed
14.
go back to reference Kubus P, Materna O, Gebauer RA, Matejka T, Gebauer R, Tlaskal T, Janousek J: Permanent epicardial pacing in children: long-term results and factors modifying outcome. Europace. 2012, 14 (4): 509-514. 10.1093/europace/eur327.CrossRefPubMed Kubus P, Materna O, Gebauer RA, Matejka T, Gebauer R, Tlaskal T, Janousek J: Permanent epicardial pacing in children: long-term results and factors modifying outcome. Europace. 2012, 14 (4): 509-514. 10.1093/europace/eur327.CrossRefPubMed
15.
go back to reference Janousek J, van Geldorp IE, Krupickova S, Rosenthal E, Nugent K, Tomaske M, Früh A, Elders J, Hiippala A, Kerst G, Gebauer RA, Kubus P, Frias P, Gabbarini F, Clur SA, Nabel B, Ganame J, Papagiannis J, Marek J, Tisma-Dupanovic S, Tsao S, Nürnberg JH, Wren C, Friedberg M, de Guillebon M, Volaufova J, Prinzen FW, Delhaas T, Working Group for Cardiac Dysrhythmias and Electrophysiology of the Association for Pediatric Cardiology: Permanent cardiac pacing in children: choosing the optimal pacing site: a multicenter study. Circulation. 2013, 127 (5): 613-623. 10.1161/CIRCULATIONAHA.112.115428.CrossRefPubMed Janousek J, van Geldorp IE, Krupickova S, Rosenthal E, Nugent K, Tomaske M, Früh A, Elders J, Hiippala A, Kerst G, Gebauer RA, Kubus P, Frias P, Gabbarini F, Clur SA, Nabel B, Ganame J, Papagiannis J, Marek J, Tisma-Dupanovic S, Tsao S, Nürnberg JH, Wren C, Friedberg M, de Guillebon M, Volaufova J, Prinzen FW, Delhaas T, Working Group for Cardiac Dysrhythmias and Electrophysiology of the Association for Pediatric Cardiology: Permanent cardiac pacing in children: choosing the optimal pacing site: a multicenter study. Circulation. 2013, 127 (5): 613-623. 10.1161/CIRCULATIONAHA.112.115428.CrossRefPubMed
Metadata
Title
Performance of steroid eluting bipolar epicardial leads in pediatric and congenital heart disease patients: 15 years of single center experience
Authors
Christian Paech
Martin Kostelka
Ingo Dähnert
Patrick Flosdorff
Frank Thomas Riede
Roman Antonin Gebauer
Publication date
01-12-2014
Publisher
BioMed Central
Published in
Journal of Cardiothoracic Surgery / Issue 1/2014
Electronic ISSN: 1749-8090
DOI
https://doi.org/10.1186/1749-8090-9-84

Other articles of this Issue 1/2014

Journal of Cardiothoracic Surgery 1/2014 Go to the issue