Abstract

Purpose

Traumatic disruption of the pelvis can lead to significant morbidity and mortality. ATLS® guidance advocates temporary stabilisation or ‘closure’ of the disrupted pelvis with a compression device or sheet. We undertook a best evidence equipment review to assess the ease and efficacy of the application of two leading commercially available devices, the T-POD® and the SAM Pelvic Sling™ II.

Methods

Fifty health care professionals and medical students participated in pelvic circumferential compression device (PCCD) education and assessment. Participants received a 10-min lecture on the epidemiology and aetiology of pelvic fractures and the principles of circumferential compression, followed by a practical demonstration. Three volunteers acted as trauma victims. Assessment included the time taken to secure the devices and whether this was achieved correctly. All participants completed a post-assessment survey.

Results

Both devices were applied correctly 100% of the time. The average time taken to secure the SAM Pelvic Sling™ II was 18 s and for the T-POD®, it was 31 s (p ≤ 0.0001). Forty-four participants (88%) agreed or strongly agreed that the SAM Pelvic Sling™ II was easy to use compared to 84% (n = 42) for the T-POD®. Thirty-nine participants (78%) reported that they preferred and, given the choice in the future, would select the T-POD® over the SAM Pelvic Sling™ II (n = 11, 22%).

Conclusions

The results of this study indicate that both PCCDs are easy and acceptable to use and, once learned, can be applied easily and rapidly. Participants applied both devices correctly 100% of the time, with successful application taking, on average, less than 60 s.
Literature
1.
go back to reference Croce MA, Magnotti LJ, Savage SA, Wood GW 2nd, Fabian TC. Emergent pelvic fixation in patients with exsanguinating pelvic fractures. J Am Coll Surg. 2007;205:935–9.PubMedCrossRef Croce MA, Magnotti LJ, Savage SA, Wood GW 2nd, Fabian TC. Emergent pelvic fixation in patients with exsanguinating pelvic fractures. J Am Coll Surg. 2007;205:935–9.PubMedCrossRef
2.
go back to reference White CE, Hsu JR, Holcomb JB. Haemodynamically unstable pelvic fractures. Injury. 2009;10:1023–30.CrossRef White CE, Hsu JR, Holcomb JB. Haemodynamically unstable pelvic fractures. Injury. 2009;10:1023–30.CrossRef
4.
go back to reference [No authors listed]. Advanced trauma life support course: student manual. 8th ed. Chicago: American College of Surgeons; 2008. [No authors listed]. Advanced trauma life support course: student manual. 8th ed. Chicago: American College of Surgeons; 2008.
5.
go back to reference Krieg JC, Mohr M, Ellis TJ, Simpson TS, Madey SM, Bottlang M. Emergent stabilization of pelvic ring injuries by controlled circumferential compression: a clinical trial. J Trauma. 2005;59:659–64.PubMedCrossRef Krieg JC, Mohr M, Ellis TJ, Simpson TS, Madey SM, Bottlang M. Emergent stabilization of pelvic ring injuries by controlled circumferential compression: a clinical trial. J Trauma. 2005;59:659–64.PubMedCrossRef
6.
go back to reference Spanjersberg WR, Knops SP, Schep NW, van Lieshout EM, Patka P, Schipper IB. Effectiveness and complications of pelvic circumferential compression devices in patients with unstable pelvic fractures: a systematic review of literature. Injury. 2009;40:1031–5.PubMedCrossRef Spanjersberg WR, Knops SP, Schep NW, van Lieshout EM, Patka P, Schipper IB. Effectiveness and complications of pelvic circumferential compression devices in patients with unstable pelvic fractures: a systematic review of literature. Injury. 2009;40:1031–5.PubMedCrossRef
7.
go back to reference Schaller TM, Sims S, Maxian T. Skin breakdown following circumferential pelvic antishock sheeting: a case report. J Orthop Trauma. 2005;19:661–5.PubMedCrossRef Schaller TM, Sims S, Maxian T. Skin breakdown following circumferential pelvic antishock sheeting: a case report. J Orthop Trauma. 2005;19:661–5.PubMedCrossRef
8.
9.
go back to reference DeAngelis NA, Wixted JJ, Drew J, Eskander MS, Eskander JP, French BG. Use of the trauma pelvic orthotic device (T-POD) for provisional stabilisation of anterior–posterior compression type pelvic fractures: a cadaveric study. Injury. 2008;39:903–6.PubMedCrossRef DeAngelis NA, Wixted JJ, Drew J, Eskander MS, Eskander JP, French BG. Use of the trauma pelvic orthotic device (T-POD) for provisional stabilisation of anterior–posterior compression type pelvic fractures: a cadaveric study. Injury. 2008;39:903–6.PubMedCrossRef
10.
go back to reference Bottlang M, Krieg JC, Mohr M, Simpson TS, Madey SM. Emergent management of pelvic ring fractures with use of circumferential compression. J Bone Joint Surg Am. 2002;84:43–7.PubMed Bottlang M, Krieg JC, Mohr M, Simpson TS, Madey SM. Emergent management of pelvic ring fractures with use of circumferential compression. J Bone Joint Surg Am. 2002;84:43–7.PubMed
Metadata
Title
Pelvic circumferential compression devices (PCCDs): a best evidence equipment review
Authors
D. J. Bryson
R. Davidson
R. Mackenzie
Publication date
01-08-2012
Publisher
Springer-Verlag
Published in
European Journal of Trauma and Emergency Surgery / Issue 4/2012
Print ISSN: 1863-9933
Electronic ISSN: 1863-9941
DOI
https://doi.org/10.1007/s00068-012-0180-3

Other articles of this Issue 4/2012

European Journal of Trauma and Emergency Surgery 4/2012 Go to the issue