Patient-specific quality assurance for proton depth dose distribution using a multi-layer ionization chamber in a single-ring wobbling method

Authors: Takahiro Kato, Kazuhiro Arai, Tatsuhiko Sagara, Ryohei Kato, Yuhei Yamazaki, Sho Oyama

Published in: Radiological Physics and Technology | Issue 3/2019

Login to get access

Abstract

The use of a multi-layer ionization chamber, Zebra, in patient-specific quality assurance (QA) for proton depth dose distributions in a single-ring wobbling method is investigated. The depth dose distributions measured using Zebra are compared with those calculated using the treatment planning system (TPS), XiO-M, and measured using an ionization chamber with a motorized water phantom system. Because the TPS only provides point doses, the average doses are calculated using in-house software. The detector size-corrected depth dose distributions are obtained by determining the average of the dose distributions from the TPS over a cylindrical region similar to the size of the Zebra detectors. The calculated depth dose distributions from the cases with a simple compensator shape are in good agreement with those obtained from the TPS without performing volume averaging; however, a 15% difference was shown when compared with those from the cases with a complex compensator shape. Then, the measurements are compared with the detector size-corrected depth dose distributions, showing an improved agreement within 3% for the highly steep dose gradient regions. Although there are some field size limitations, the Zebra system is a useful device for the fast measurement of patient-specific QA for depth dose distributions in wobbled proton beams. However, careful consideration is required for complex dose distribution fields, because the measurements obtained using Zebra cannot be directly compared to the depth dose distributions from the TPS owing to the finite detector size of the Zebra chamber.
Literature
1.
go back to reference IAEA. Absorbed dose Determination in External Beam Radiotherapy-An international Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water Technical Report Series No 398. Vienne: International Atomic Energy Agency; 2000. IAEA. Absorbed dose Determination in External Beam Radiotherapy-An international Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water Technical Report Series No 398. Vienne: International Atomic Energy Agency; 2000.
2.
go back to reference Nichiporov D, Solberg K, Hsi W, His W, Wolanski M, Mascia A, Farr J, Schreuder A. Multichannel detectors for profile measurements in clinical proton fields. Med Phys. 2007;34:2683–90.CrossRefPubMed Nichiporov D, Solberg K, Hsi W, His W, Wolanski M, Mascia A, Farr J, Schreuder A. Multichannel detectors for profile measurements in clinical proton fields. Med Phys. 2007;34:2683–90.CrossRefPubMed
3.
go back to reference Yajima K, Kanai T, Kusano Y, Shimojyu T. Development of a multi-layer ionization chamber for heavy-ion radiotherapy. Phys Med Biol. 2009;54:N107–14.CrossRefPubMed Yajima K, Kanai T, Kusano Y, Shimojyu T. Development of a multi-layer ionization chamber for heavy-ion radiotherapy. Phys Med Biol. 2009;54:N107–14.CrossRefPubMed
4.
go back to reference Dhanesar S, Sahoo N, Kerr M, Taylor B, Summers P, Zhu XR, Poenisch F, Gillin M. Quality assurance of proton beams using a multilayer ionization chamber system. Med Phys. 2013;40:092102.CrossRefPubMed Dhanesar S, Sahoo N, Kerr M, Taylor B, Summers P, Zhu XR, Poenisch F, Gillin M. Quality assurance of proton beams using a multilayer ionization chamber system. Med Phys. 2013;40:092102.CrossRefPubMed
5.
go back to reference Rinaldi I, Brons S, Jäkel O, Voss B, Parodi K. A method to increase the nominal range resolution of a stack of parallel-plate ionization chambers. Phys Med Biol. 2014;59:5501–15.CrossRefPubMed Rinaldi I, Brons S, Jäkel O, Voss B, Parodi K. A method to increase the nominal range resolution of a stack of parallel-plate ionization chambers. Phys Med Biol. 2014;59:5501–15.CrossRefPubMed
6.
go back to reference Takayanagi T, Nihongi H, Nishiuchi H, Tadokoro M, Ito Y, Nakashima C, Fujitaka S, Umezawa M, Matsuda K, Sakae T, Terunuma T. Dual ring multilayer ionization chamber and theory-based correction technique for scanning proton therapy. Med Phys. 2016;43:4150–62.CrossRefPubMed Takayanagi T, Nihongi H, Nishiuchi H, Tadokoro M, Ito Y, Nakashima C, Fujitaka S, Umezawa M, Matsuda K, Sakae T, Terunuma T. Dual ring multilayer ionization chamber and theory-based correction technique for scanning proton therapy. Med Phys. 2016;43:4150–62.CrossRefPubMed
7.
go back to reference Mirandola A, Magro G, Lavagno M, Mairani A, Molinelli S, Russo S, Mastella E, Vai A, Maestri D, Rossa V, Ciocca M. Characterization of a multilayer ionization chamber prototype for fast verification of relative depth ionization curves and spread-out-Bragg-peaks in light ion beam therapy. Med Phys. 2018;45:2266–77.CrossRefPubMed Mirandola A, Magro G, Lavagno M, Mairani A, Molinelli S, Russo S, Mastella E, Vai A, Maestri D, Rossa V, Ciocca M. Characterization of a multilayer ionization chamber prototype for fast verification of relative depth ionization curves and spread-out-Bragg-peaks in light ion beam therapy. Med Phys. 2018;45:2266–77.CrossRefPubMed
8.
9.
go back to reference Koehler AM, Schneider RJ, Sisterson JM. Flattening of proton dose distributions for large-field radiotherapy. Med Phys. 1977;4:297–301.CrossRefPubMed Koehler AM, Schneider RJ, Sisterson JM. Flattening of proton dose distributions for large-field radiotherapy. Med Phys. 1977;4:297–301.CrossRefPubMed
10.
go back to reference Himukai T, Takada Y, Hotta K, Hara Y, Komori M, Kanai T, Kohno R. Analytical design method of optimum ridge filters for wobbled and collimated proton beams. Jpn J Med Phys. 2008;28:57–69. Himukai T, Takada Y, Hotta K, Hara Y, Komori M, Kanai T, Kohno R. Analytical design method of optimum ridge filters for wobbled and collimated proton beams. Jpn J Med Phys. 2008;28:57–69.
11.
go back to reference Kase Y, Yamashita H, Numano M, Sakama M, Mizota M, Maeda Y, Tameshige Y, Murayama S. A model-based analysis of a simplified beam-specific dose output in proton therapy with a single-ring wobbling system. Phys Med Biol. 2015;60:359–74.CrossRefPubMed Kase Y, Yamashita H, Numano M, Sakama M, Mizota M, Maeda Y, Tameshige Y, Murayama S. A model-based analysis of a simplified beam-specific dose output in proton therapy with a single-ring wobbling system. Phys Med Biol. 2015;60:359–74.CrossRefPubMed
12.
go back to reference Tansho R, Takada Y, Kohno R, Hotta K, Hara Y, Mizutani S, Akimoto T. Experimental verification of dose calculation using the simplified Monte Carlo method with an improved initial beam model for a beam-wobbling system. Phys Med Biol. 2013;58:6047–64.CrossRefPubMed Tansho R, Takada Y, Kohno R, Hotta K, Hara Y, Mizutani S, Akimoto T. Experimental verification of dose calculation using the simplified Monte Carlo method with an improved initial beam model for a beam-wobbling system. Phys Med Biol. 2013;58:6047–64.CrossRefPubMed
13.
go back to reference Yonai S, Kanematsu N, Komori M, Kanai T, Takei Y, Takahashi O, Isobe Y, Tashiro M, Koikegami H, Tomita H. Evaluation of beam wobbling methods for heavy-ion radiotherapy. Med Phys. 2008;35:927–38.CrossRefPubMed Yonai S, Kanematsu N, Komori M, Kanai T, Takei Y, Takahashi O, Isobe Y, Tashiro M, Koikegami H, Tomita H. Evaluation of beam wobbling methods for heavy-ion radiotherapy. Med Phys. 2008;35:927–38.CrossRefPubMed
14.
go back to reference Lomax AJ, Böhringer T, Bolsi A, Coray D, Emert F, Goitein G, Jermann M, Lin S, Pedroni E, Rutz H, Stadelmann O, Timmermann B, Verwey J, Weber DC. Treatment planning and verification of proton therapy using spot scanning: initial experiences. Med Phys. 2004;31:3150–7.CrossRefPubMed Lomax AJ, Böhringer T, Bolsi A, Coray D, Emert F, Goitein G, Jermann M, Lin S, Pedroni E, Rutz H, Stadelmann O, Timmermann B, Verwey J, Weber DC. Treatment planning and verification of proton therapy using spot scanning: initial experiences. Med Phys. 2004;31:3150–7.CrossRefPubMed
Metadata
Title
Patient-specific quality assurance for proton depth dose distribution using a multi-layer ionization chamber in a single-ring wobbling method
Authors
Takahiro Kato
Kazuhiro Arai
Tatsuhiko Sagara
Ryohei Kato
Yuhei Yamazaki
Sho Oyama
Publication date
01-09-2019
Publisher
Springer Singapore
Published in
Radiological Physics and Technology / Issue 3/2019
Print ISSN: 1865-0333
Electronic ISSN: 1865-0341
DOI
https://doi.org/10.1007/s12194-019-00524-8

Other articles of this Issue 3/2019

Radiological Physics and Technology 3/2019 Go to the issue