A multi-institutional survey

Authors: M. Sc. Dipl.-Ing. (FH) Beatrice Steiniger, René Berger, Sabine Eilzer, Christine Kornhuber, Kathleen Lorenz, Torsten Peil, Carsten Reiffenstuhl, Johannes Schilz, Dirk Schröder, Michael Schwedas, Stephanie Pensold, Mathias Walke, Kirsten Weibert, Ulrich Wolf, Tilo Wiezorek

Published in: Strahlentherapie und Onkologie | Issue 1/2017

Login to get access

Abstract

Purpose

This project compares the different patient-related quality assurance systems for intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT) techniques currently used in the central Germany area with an independent measuring system.

Materials and methods

The participating institutions generated 21 treatment plans with different combinations of treatment planning systems (TPS) and linear accelerators (LINAC) for the QUASIMODO (Quality ASsurance of Intensity MODulated radiation Oncology) patient model. The plans were exposed to the ArcCHECK measuring system (Sun Nuclear Corporation, Melbourne, FL, USA). The dose distributions were analyzed using the corresponding software and a point dose measured at the isocenter with an ionization chamber.

Results

According to the generally used criteria of a 10 % threshold, 3 % difference, and 3 mm distance, the majority of plans investigated showed a gamma index exceeding 95 %. Only one plan did not fulfill the criteria and three of the plans did not comply with the commonly accepted tolerance level of ±3 % in point dose measurement.

Conclusion

Using only one of the two examined methods for patient-related quality assurance is not sufficiently significant in all cases.
Literature
1.
go back to reference Feygelman V, Zhang G, Stevens C, Nelms BE (2011) Evaluation of a new VMAT QA device, or the “X” and “O” array geometries. J Appl Clin Med Phys 12(2):3346PubMed Feygelman V, Zhang G, Stevens C, Nelms BE (2011) Evaluation of a new VMAT QA device, or the “X” and “O” array geometries. J Appl Clin Med Phys 12(2):3346PubMed
2.
3.
go back to reference Ezzell GA, Burmeister JW, Dogan N, LoSasso TJ, Mechalakos JG, Mihailidis D, Molineu A, Palta JR, Ramsey CR, Salter BJ, Shi J, Xia P, Yue NJ, Xiao Y (2009) IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys 36(11):5359–5373PubMedCrossRef Ezzell GA, Burmeister JW, Dogan N, LoSasso TJ, Mechalakos JG, Mihailidis D, Molineu A, Palta JR, Ramsey CR, Salter BJ, Shi J, Xia P, Yue NJ, Xiao Y (2009) IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys 36(11):5359–5373PubMedCrossRef
5.
go back to reference Low DA, Moran JM, Dempsey JF, Dong L, Oldham M (2011) Dosimetry tools and techniques for IMRT. Med Phys 38(3):1313–1338PubMedCrossRef Low DA, Moran JM, Dempsey JF, Dong L, Oldham M (2011) Dosimetry tools and techniques for IMRT. Med Phys 38(3):1313–1338PubMedCrossRef
6.
go back to reference DIN 6875-3:2008-03 Spezielle Bestrahlungseinrichtungen – Teil 3: Fluenzmodulierte Strahlentherapie – Kennmerkmale, Prüfmethoden und Regeln für den klinischen Einsatz. DIN 6875-3:2008-03 Spezielle Bestrahlungseinrichtungen – Teil 3: Fluenzmodulierte Strahlentherapie – Kennmerkmale, Prüfmethoden und Regeln für den klinischen Einsatz.
7.
go back to reference DIN 6875-4:2011-10 Spezielle Bestrahlungseinrichtungen – Teil 4: Fluenzmodulierte Strahlentherapie – Konstanzprüfungen. DIN 6875-4:2011-10 Spezielle Bestrahlungseinrichtungen – Teil 4: Fluenzmodulierte Strahlentherapie – Konstanzprüfungen.
8.
go back to reference DIN 6800-2:2008-03 Dosismessverfahren nach der Sondenmethode für Photonen- und Elektronenstrahlung – Teil 2: Dosimetrie hochenergetischer Photonen- und Elektronenstrahlung mit Ionisationskammern. DIN 6800-2:2008-03 Dosismessverfahren nach der Sondenmethode für Photonen- und Elektronenstrahlung – Teil 2: Dosimetrie hochenergetischer Photonen- und Elektronenstrahlung mit Ionisationskammern.
13.
go back to reference Bohsung J, Gillis S, Arrans R, Bakai A, De Wagter C, Knoos T, Mijnheer BJ, Paiusco M, Perrin BA, Welleweerd H, Williams P (2005) IMRT treatment planning: a comparative inter-system and inter-centre planning exercise of the ESTRO QUASIMODO group. Radiother Oncol 76(3):354–361. doi:10.1016/j.radonc.2005.08.003 PubMedCrossRef Bohsung J, Gillis S, Arrans R, Bakai A, De Wagter C, Knoos T, Mijnheer BJ, Paiusco M, Perrin BA, Welleweerd H, Williams P (2005) IMRT treatment planning: a comparative inter-system and inter-centre planning exercise of the ESTRO QUASIMODO group. Radiother Oncol 76(3):354–361. doi:10.​1016/​j.​radonc.​2005.​08.​003 PubMedCrossRef
15.
go back to reference Li G, Bai S, Chen N, Henderson L, Wu K, Xiao J, Zhang Y, Jiang Q, Jiang X (2013) Evaluation of the sensitivity of two 3D diode array dosimetry systems to setup error for quality assurance (QA) of volumetric-modulated arc therapy (VMAT). J Appl Clin Med Phys 14(5):13–24. doi:10.1120/jacmp.v14i5.3828 PubMed Li G, Bai S, Chen N, Henderson L, Wu K, Xiao J, Zhang Y, Jiang Q, Jiang X (2013) Evaluation of the sensitivity of two 3D diode array dosimetry systems to setup error for quality assurance (QA) of volumetric-modulated arc therapy (VMAT). J Appl Clin Med Phys 14(5):13–24. doi:10.​1120/​jacmp.​v14i5.​3828 PubMed
16.
go back to reference Chaswal V, Weldon M, Gupta N, Chakravarti A, Rong Y (2014) Commissioning and comprehensive evaluation of the ArcCHECK cylindrical diode array for VMAT pretreatment delivery QA. J Appl Clin Med Phys 15(4):4832. doi:10.1120/jacmp.v15i4.4832 PubMed Chaswal V, Weldon M, Gupta N, Chakravarti A, Rong Y (2014) Commissioning and comprehensive evaluation of the ArcCHECK cylindrical diode array for VMAT pretreatment delivery QA. J Appl Clin Med Phys 15(4):4832. doi:10.​1120/​jacmp.​v15i4.​4832 PubMed
20.
go back to reference Sun Nuclear Corporation (2012) ArcCHECK Reference Guide. Sun Nuclear Corporation (2012) ArcCHECK Reference Guide.
21.
go back to reference IAEA (2000) TRS No. 398 Absorbed dose determination in external beam radiotherapy. IAEA (2000) TRS No. 398 Absorbed dose determination in external beam radiotherapy.
24.
28.
go back to reference Melidis C, Bosch WR, Izewska J, Fidarova E, Zubizarreta E, Ishikura S, Followill D, Galvin J, Xiao Y, Ebert MA, Kron T, Clark CH, Miles EA, Aird EG, Weber DC, Ulin K, Verellen D, Hurkmans CW (2014) Radiation therapy quality assurance in clinical trials – Global Harmonisation Group. Radiother Oncol 111(3):327–329. doi:10.1016/j.radonc.2014.03.023 PubMedPubMedCentralCrossRef Melidis C, Bosch WR, Izewska J, Fidarova E, Zubizarreta E, Ishikura S, Followill D, Galvin J, Xiao Y, Ebert MA, Kron T, Clark CH, Miles EA, Aird EG, Weber DC, Ulin K, Verellen D, Hurkmans CW (2014) Radiation therapy quality assurance in clinical trials – Global Harmonisation Group. Radiother Oncol 111(3):327–329. doi:10.​1016/​j.​radonc.​2014.​03.​023 PubMedPubMedCentralCrossRef
29.
go back to reference Stasi M, Bresciani S, Miranti A, Maggio A, Sapino V, Gabriele P (2012) Pretreatment patient-specific IMRT quality assurance: a correlation study between gamma index and patient clinical dose volume histogram. Med Phys 39(12):7626–7634. doi:10.1118/1.4767763 PubMedCrossRef Stasi M, Bresciani S, Miranti A, Maggio A, Sapino V, Gabriele P (2012) Pretreatment patient-specific IMRT quality assurance: a correlation study between gamma index and patient clinical dose volume histogram. Med Phys 39(12):7626–7634. doi:10.​1118/​1.​4767763 PubMedCrossRef
30.
go back to reference Pasler M, Wirtz H, Lutterbach J (2011) Impact of gantry rotation time on plan quality and dosimetric verification-volumetric modulated arc therapy (VMAT) vs. intensity modulated radiotherapy (IMRT). Strahlenther Onkol 187(12):812–819. doi:10.1007/s00066-011-2263-1 PubMedCrossRef Pasler M, Wirtz H, Lutterbach J (2011) Impact of gantry rotation time on plan quality and dosimetric verification-volumetric modulated arc therapy (VMAT) vs. intensity modulated radiotherapy (IMRT). Strahlenther Onkol 187(12):812–819. doi:10.​1007/​s00066-011-2263-1 PubMedCrossRef
32.
Metadata
Title
Patient-related quality assurance with different combinations of treatment planning systems, techniques, and machines
A multi-institutional survey
Authors
M. Sc. Dipl.-Ing. (FH) Beatrice Steiniger
René Berger
Sabine Eilzer
Christine Kornhuber
Kathleen Lorenz
Torsten Peil
Carsten Reiffenstuhl
Johannes Schilz
Dirk Schröder
Michael Schwedas
Stephanie Pensold
Mathias Walke
Kirsten Weibert
Ulrich Wolf
Tilo Wiezorek
Publication date
01-01-2017
Publisher
Springer Berlin Heidelberg
Published in
Strahlentherapie und Onkologie / Issue 1/2017
Print ISSN: 0179-7158
Electronic ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-016-1064-y

Other articles of this Issue 1/2017

Strahlentherapie und Onkologie 1/2017 Go to the issue