Oxidized Zirconium Head on Crosslinked Polyethylene Liner in Total Hip Arthroplasty: A 7- to 12-year In Vivo Comparative Wear Study

Authors: George K. Karidakis, MD, Theofilos Karachalios, MD, DSc

Published in: Clinical Orthopaedics and Related Research® | Issue 12/2015

Login to get access

Abstract

Background

Osteolysis resulting from wear debris production from the bearing surfaces is a major factor limiting long-term survival of hip implants. Oxidized zirconium head on crosslinked polyethylene (XLPE) is a modern bearing coupling. However, midterm in vivo wear data of this coupling are not known.

Questions/purposes

The purpose of this study was to investigate in vivo whether the combination of an oxidized zirconium femoral head on XLPE produces less wear than a ceramic head on XLPE or a ceramic head on conventional polyethylene (CPE) couplings and whether any of these bearing combinations results in higher hip scores.

Methods

Between 2003 and 2007, we performed 356 total hip arthroplasties in 288 patients; of those, 199 (69.1%) patients (199 hips) were enrolled in what began as a randomized trial. Unfortunately, after the 57th patient, the randomization process was halted because of patients’ preference for the oxidized zirconium bearing instead of the ceramic after (as they were informed by the consent form), and after that, alternate allocation to the study groups was performed. Hips were allocated into four groups: in Group A, a 28-mm ceramic head on CPE was used; in Group B, a 28-mm ceramic head on XLPE; in Group C, a 28-mm Oxinium head on XLPE; and in Group D, a 32-mm Oxinium head on XLPE. The authors prospectively collected in vivo wear data (linear wear, linear wear rate, volumetric wear, and volumetric wear rate) using PolyWare software. Preoperative and postoperative clinical data, including Harris and Oxford hip scores, were also collected at regular intervals. Of those patients enrolled, 188 (95%) were available for final followup at a minimum of 7 years (mean, 9 years; range, 7–12 years).

Results

All bearing surfaces showed a varying high bedding-in effect (plastic deformation of the liner) up to the second postoperative year. At 5 years both oxidized zirconium on XLPE groups showed lower (p < 0.01) volumetric wear (mean ± SD mm3) and volumetric wear rates (mean ± SD mm3/year) (Group C: 310 ± 55–206 ± 55 mm3/year, Group D: 320 ± 58–205 ± 61 mm3/year) when compared with ceramic on CPE (Group A: 791 ± 124–306 ± 85 mm3/year) and ceramic on XLPE (Group B: 1420 ± 223–366 ± 88 mm3/year) groups. For those patients who had completed 10 years of followup (20 patients [44.5%] of Group A, 21 [45.7%] of Group B, 23 [47.9%] of Group C, and 22 [44.9%] of Group D), at 10 years, both oxidized zirconium on XLPE groups also showed lower (p < 0.01) volumetric wear (mean ± SD mm3) and volumetric wear rates (mean ± SD mm3/year) (Group C: 356 ± 64 to 215 ± 54 mm3/year, Group D: 354 ± 50 to 210 ± 64 mm3/year) when compared with ceramic on CPE (Group A: 895 ± 131 to 380 ± 80 mm3/year) and ceramic on XLPE (Group B: 1625 ± 253 to 480 ± 101 mm3/year) groups. When wear rates of both oxidized zirconium groups were compared, no differences were found at any time interval with the numbers available. Two hips (one from Group A and one from Group B) are scheduled for revision as a result of wear and osteolysis. There were no differences in hip scores among the groups with the numbers available.

Conclusions

In this study, in vivo wear parameters were lower when the combination of an oxidized zirconium head on XLPE liner was used at an average of 9 years (range, 7–12 years) followup. Further larger-scale clinical studies should confirm these findings and evaluate osteolysis and revision rates in association with the use of this bearing coupling.

Level of Evidence

Level II, therapeutic study.
Literature
1.
go back to reference Allain J, Roudot-Thorval F, Delecrin J, Anract P, Migaud H, Goutallier D. Revision total hip arthroplasty performed after fracture of a ceramic femoral head: a multicenter survivorship study. J Bone Joint Surg Am. 2003;85:825–830.PubMed Allain J, Roudot-Thorval F, Delecrin J, Anract P, Migaud H, Goutallier D. Revision total hip arthroplasty performed after fracture of a ceramic femoral head: a multicenter survivorship study. J Bone Joint Surg Am. 2003;85:825–830.PubMed
2.
go back to reference Almousa SA, Greidanus NV, Masri BA, Duncan CP, Garbuz DS. The natural history of inflammatory pseudotumors in asymptomatic patients after metal-on-metal hip arthroplasty. Clin Orthop Relat Res. 2013;471:3814–3821.PubMedCentralCrossRefPubMed Almousa SA, Greidanus NV, Masri BA, Duncan CP, Garbuz DS. The natural history of inflammatory pseudotumors in asymptomatic patients after metal-on-metal hip arthroplasty. Clin Orthop Relat Res. 2013;471:3814–3821.PubMedCentralCrossRefPubMed
4.
go back to reference Beksac B, Salas A, Della Valle AG, Salvati E. Wear is reduced in THA performed with highly cross-linked polyethylene. Clin Orthop Relat Res. 2009;467:1765–1772.PubMedCentralCrossRefPubMed Beksac B, Salas A, Della Valle AG, Salvati E. Wear is reduced in THA performed with highly cross-linked polyethylene. Clin Orthop Relat Res. 2009;467:1765–1772.PubMedCentralCrossRefPubMed
5.
go back to reference Benezra V, Mangin S, Treska M. Microstructural investigation of the oxide scale on Zr-2.5 Nb and its interface with the alloy substrate. Mater Res Soc Symp Proc. 1999;550:337–342.CrossRef Benezra V, Mangin S, Treska M. Microstructural investigation of the oxide scale on Zr-2.5 Nb and its interface with the alloy substrate. Mater Res Soc Symp Proc. 1999;550:337–342.CrossRef
6.
go back to reference Borlin N, Thien T, Karrholm J. The precision of radiostereometric measurements. Manual vs digital measurements. J Biomech. 2002;35:69–79.CrossRefPubMed Borlin N, Thien T, Karrholm J. The precision of radiostereometric measurements. Manual vs digital measurements. J Biomech. 2002;35:69–79.CrossRefPubMed
7.
go back to reference Burroughs BR, Hallstrom B, Golladay GJ, Hoeffel D, Harris WH. Range of motion and stability in total hip arthroplasty with 28, 32, 38, and 44 mm femoral head sizes: an in vitro study. J Arthroplasty. 2005;20:11–19.CrossRefPubMed Burroughs BR, Hallstrom B, Golladay GJ, Hoeffel D, Harris WH. Range of motion and stability in total hip arthroplasty with 28, 32, 38, and 44 mm femoral head sizes: an in vitro study. J Arthroplasty. 2005;20:11–19.CrossRefPubMed
8.
go back to reference Calvert GT, Devane PA, Fielden J, Adams K, Horne JG. A double-blind, prospective, randomized controlled trial comparing highly cross-linked and conventional polyethylene in primary total hip arthroplasty. J Arthroplasty. 2009;24:505–510.CrossRefPubMed Calvert GT, Devane PA, Fielden J, Adams K, Horne JG. A double-blind, prospective, randomized controlled trial comparing highly cross-linked and conventional polyethylene in primary total hip arthroplasty. J Arthroplasty. 2009;24:505–510.CrossRefPubMed
9.
go back to reference Devane PA, Bourne RB, Rorabeck CH, Hardie RM, Horne JG. Measurement of polyethylene wear in metal-backed acetabular cups. I. Three-dimensional technique. Clin Orthop Relat Res. 1995;319:303–316.PubMed Devane PA, Bourne RB, Rorabeck CH, Hardie RM, Horne JG. Measurement of polyethylene wear in metal-backed acetabular cups. I. Three-dimensional technique. Clin Orthop Relat Res. 1995;319:303–316.PubMed
10.
go back to reference Devane PA, Horne JG. Assessment of polyethylene wear in total hip replacement. Clin Orthop Relat Res. 1999;369:59–72.CrossRefPubMed Devane PA, Horne JG. Assessment of polyethylene wear in total hip replacement. Clin Orthop Relat Res. 1999;369:59–72.CrossRefPubMed
11.
go back to reference Ebramzadeh E, Sangiorgio SN, Lattuada F, Kang JS, Chiesa R, McKellop HA, Dorr LD. Accuracy of measurement of polyethylene wear with use of radiographs of total hip replacements. J Bone Joint Surg Am. 2003;85:2378–2384.PubMed Ebramzadeh E, Sangiorgio SN, Lattuada F, Kang JS, Chiesa R, McKellop HA, Dorr LD. Accuracy of measurement of polyethylene wear with use of radiographs of total hip replacements. J Bone Joint Surg Am. 2003;85:2378–2384.PubMed
12.
go back to reference Fukui K, Kaneuji A, Sugimori T, Ichiseki T, Kitamura K, Matsumoto T. Wear comparison between a highly cross-linked polyethylene and conventional polyethylene against a zirconia femoral head: minimum 5-year follow-up. J Arthroplasty. 2011;26:45–49.CrossRefPubMed Fukui K, Kaneuji A, Sugimori T, Ichiseki T, Kitamura K, Matsumoto T. Wear comparison between a highly cross-linked polyethylene and conventional polyethylene against a zirconia femoral head: minimum 5-year follow-up. J Arthroplasty. 2011;26:45–49.CrossRefPubMed
13.
go back to reference Garvin KL, Hartman CW, Mangla J, Murdoch N, Martell JM. Wear analysis in THA utilizing oxidized zirconium and crosslinked polyethylene. Clin Orthop Relat Res. 2009;467:141–145.PubMedCentralCrossRefPubMed Garvin KL, Hartman CW, Mangla J, Murdoch N, Martell JM. Wear analysis in THA utilizing oxidized zirconium and crosslinked polyethylene. Clin Orthop Relat Res. 2009;467:141–145.PubMedCentralCrossRefPubMed
14.
go back to reference Glyn-Jones S, Isaac S, Hauptfleisch J, McLardy-Smith P, Murray DW, Gill HS. Does highly cross-linked polyethylene wear less than conventional polyethylene in total hip arthroplasty? A double blind, randomized, and controlled trial using roentgen stereophotogrammetric analysis. J Arthroplasty. 2008;23:337–343.CrossRefPubMed Glyn-Jones S, Isaac S, Hauptfleisch J, McLardy-Smith P, Murray DW, Gill HS. Does highly cross-linked polyethylene wear less than conventional polyethylene in total hip arthroplasty? A double blind, randomized, and controlled trial using roentgen stereophotogrammetric analysis. J Arthroplasty. 2008;23:337–343.CrossRefPubMed
15.
go back to reference Good V, Ries M, Barrack RL, Widding K, Hunter G, Heuer D. Reduced wear with oxidized zirconium femoral heads. J Bone Joint Surg Am. 2003;85(Suppl 4):105–110.PubMed Good V, Ries M, Barrack RL, Widding K, Hunter G, Heuer D. Reduced wear with oxidized zirconium femoral heads. J Bone Joint Surg Am. 2003;85(Suppl 4):105–110.PubMed
16.
go back to reference Hasegawa M, Sudo A, Uchida A. Cobalt-chromium head wear following revision hip arthroplasty performed after ceramic fracture. A case report. Acta Orthop Scand. 2006;77:833–835.CrossRef Hasegawa M, Sudo A, Uchida A. Cobalt-chromium head wear following revision hip arthroplasty performed after ceramic fracture. A case report. Acta Orthop Scand. 2006;77:833–835.CrossRef
17.
go back to reference Heuer D, Good V, Widding K. Wear performance of damaged oxidized Zr-2.5 Nb modular femoral heads. Soc Biomat. 2003;366:7–10. Heuer D, Good V, Widding K. Wear performance of damaged oxidized Zr-2.5 Nb modular femoral heads. Soc Biomat. 2003;366:7–10.
18.
go back to reference Jasty MJ, Floyd WE, Schiller AL, Goldring SR, Harris WH. Localized osteolysis in stable, non-septic total hip replacement. J Bone Joint Surg Am. 1986;68:912–919.PubMed Jasty MJ, Floyd WE, Schiller AL, Goldring SR, Harris WH. Localized osteolysis in stable, non-septic total hip replacement. J Bone Joint Surg Am. 1986;68:912–919.PubMed
19.
go back to reference Kadar T, Furnes O, Aamodt A, Indrekvam K, Havelin LI, Haugan K, Espehaug B, Hallan G. The influence of acetabular inclination angle on the penetration of polyethylene and migration of the acetabular component: a prospective, radiostereometric study on cemented acetabular components. J Bone Joint Surg Br. 2012;94:302–307.CrossRefPubMed Kadar T, Furnes O, Aamodt A, Indrekvam K, Havelin LI, Haugan K, Espehaug B, Hallan G. The influence of acetabular inclination angle on the penetration of polyethylene and migration of the acetabular component: a prospective, radiostereometric study on cemented acetabular components. J Bone Joint Surg Br. 2012;94:302–307.CrossRefPubMed
20.
go back to reference Karachalios T, Hartofilakidis G, Zacharakis N, Tsekoura M. A 12- to 18- year radiographic follow-up study of Charnley low-friction arthroplasty. Clin Orthop Relat Res. 1993;296:140–147.PubMed Karachalios T, Hartofilakidis G, Zacharakis N, Tsekoura M. A 12- to 18- year radiographic follow-up study of Charnley low-friction arthroplasty. Clin Orthop Relat Res. 1993;296:140–147.PubMed
21.
go back to reference Kwon YM, Ostlere SJ, McLardy-Smith P, Athanasou NA, Gill HS, Murray DW. Asymptomatic pseudotumors after metal-on-metal hip resurfacing arthroplasty prevalence and metal ion study. J Arthroplasty. 2011;26:511–518.CrossRefPubMed Kwon YM, Ostlere SJ, McLardy-Smith P, Athanasou NA, Gill HS, Murray DW. Asymptomatic pseudotumors after metal-on-metal hip resurfacing arthroplasty prevalence and metal ion study. J Arthroplasty. 2011;26:511–518.CrossRefPubMed
22.
go back to reference Ladon D, Doherty A, Newson R, Turner J, Bhamra M, Case CP. Changes in metal levels and chromosome aberrations in the peripheral blood of patients after metal-on-metal hip arthroplasty. J Arthroplasty. 2004;19(Suppl 3):78–83.CrossRefPubMed Ladon D, Doherty A, Newson R, Turner J, Bhamra M, Case CP. Changes in metal levels and chromosome aberrations in the peripheral blood of patients after metal-on-metal hip arthroplasty. J Arthroplasty. 2004;19(Suppl 3):78–83.CrossRefPubMed
23.
go back to reference Lewis PM, Moore CA, Olsen M, Schemitsch EH, Waddell JP. Comparison of mid-term clinical outcomes after primary total hip arthroplasty with Oxinium vs cobalt chrome femoral heads. Orthopedics. 2008;31(Suppl 2):114–122. Lewis PM, Moore CA, Olsen M, Schemitsch EH, Waddell JP. Comparison of mid-term clinical outcomes after primary total hip arthroplasty with Oxinium vs cobalt chrome femoral heads. Orthopedics. 2008;31(Suppl 2):114–122.
24.
go back to reference Livermore J, Ilstrup D, Morrey B. Effect of femoral head size on wear of the polyethylene acetabular component. J Bone Joint Surg Am. 1990;72:518–528.PubMed Livermore J, Ilstrup D, Morrey B. Effect of femoral head size on wear of the polyethylene acetabular component. J Bone Joint Surg Am. 1990;72:518–528.PubMed
25.
go back to reference Mai K, Verioti C, Ezzet KA, Copp SN, Walker RH, Colwell CW Jr. Incidence of ‘squeaking’ after ceramic-on-ceramic total hip arthroplasty. Clin Orthop Relat Res. 2010;468:413–417.PubMedCentralCrossRefPubMed Mai K, Verioti C, Ezzet KA, Copp SN, Walker RH, Colwell CW Jr. Incidence of ‘squeaking’ after ceramic-on-ceramic total hip arthroplasty. Clin Orthop Relat Res. 2010;468:413–417.PubMedCentralCrossRefPubMed
26.
go back to reference Manning DW, Chiang PP, Martell JM, Galante JO, Harris WH. In vivo comparative wear study of traditional and highly crosslinked polyethylene in total hip arthroplasty. J Arthroplasty. 2005;20:880–886.CrossRefPubMed Manning DW, Chiang PP, Martell JM, Galante JO, Harris WH. In vivo comparative wear study of traditional and highly crosslinked polyethylene in total hip arthroplasty. J Arthroplasty. 2005;20:880–886.CrossRefPubMed
27.
go back to reference Martell JM, Berdia S. Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg Am. 1997;79:1635–1641.PubMed Martell JM, Berdia S. Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg Am. 1997;79:1635–1641.PubMed
28.
go back to reference McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty. A randomized controlled trial. J Bone Joint Surg Am. 2009;91:773–782.CrossRefPubMed McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty. A randomized controlled trial. J Bone Joint Surg Am. 2009;91:773–782.CrossRefPubMed
29.
go back to reference Morison ZA, Patil S, Khan HA, Bogoch ER, Schemitsch EH, Waddell JP. A randomized controlled trial comparing Oxinium and cobalt-chrome on standard and cross-linked polyethylene. J Arthroplasty. 2014;29(Suppl):164–168.CrossRefPubMed Morison ZA, Patil S, Khan HA, Bogoch ER, Schemitsch EH, Waddell JP. A randomized controlled trial comparing Oxinium and cobalt-chrome on standard and cross-linked polyethylene. J Arthroplasty. 2014;29(Suppl):164–168.CrossRefPubMed
30.
go back to reference Mu Z, Tian J, Wu T, Yang J, Pei F. A systematic review of radiological outcomes of highly cross-linked polyethylene versus conventional polyethylene in total hip arthroplasty. Int Orthop. 2008;33:599–604.CrossRef Mu Z, Tian J, Wu T, Yang J, Pei F. A systematic review of radiological outcomes of highly cross-linked polyethylene versus conventional polyethylene in total hip arthroplasty. Int Orthop. 2008;33:599–604.CrossRef
31.
go back to reference Muratoglu OK, Greenbaum ES, Bragdon CR, Jasty M, Freiberg AA, Harris WH. Surface analysis of early retrieved acetabular polyethylene liners: a comparison of conventional and highly crosslinked polyethylenes. J Arthroplasty. 2004;19:68–77.CrossRefPubMed Muratoglu OK, Greenbaum ES, Bragdon CR, Jasty M, Freiberg AA, Harris WH. Surface analysis of early retrieved acetabular polyethylene liners: a comparison of conventional and highly crosslinked polyethylenes. J Arthroplasty. 2004;19:68–77.CrossRefPubMed
32.
go back to reference Muratoglu OK, Merrill EW, Bragdon CR, O’Connor D, Hoeffel D, Burroughs B, Jasty M, Harris WH. Effect of radiation, heat, and aging on in vitro wear resistance of polyethylene. Clin Orthop Relat Res. 2003;417:253–262.PubMed Muratoglu OK, Merrill EW, Bragdon CR, O’Connor D, Hoeffel D, Burroughs B, Jasty M, Harris WH. Effect of radiation, heat, and aging on in vitro wear resistance of polyethylene. Clin Orthop Relat Res. 2003;417:253–262.PubMed
33.
go back to reference Mutimer J, Devane PA, Adams K, Horne JG. Highly crosslinked polyethylene reduces wear in total hip arthroplasty at 5 years. Clin Orthop Relat Res. 2010;468:3228–3233.PubMedCentralCrossRefPubMed Mutimer J, Devane PA, Adams K, Horne JG. Highly crosslinked polyethylene reduces wear in total hip arthroplasty at 5 years. Clin Orthop Relat Res. 2010;468:3228–3233.PubMedCentralCrossRefPubMed
34.
go back to reference Nakahara I, Nakamura N, Nishii T, Miki H, Sakai T, Sugano N. Minimum five-year follow-up wear measurement of Longevity highly cross-linked polyethylene cup against cobalt-chromium or zirconia heads. J Arthroplasty. 2010;8:1182–1187.CrossRef Nakahara I, Nakamura N, Nishii T, Miki H, Sakai T, Sugano N. Minimum five-year follow-up wear measurement of Longevity highly cross-linked polyethylene cup against cobalt-chromium or zirconia heads. J Arthroplasty. 2010;8:1182–1187.CrossRef
35.
go back to reference Rissanen P, Aro S, Slatis P, Sintonen H, Paavolainen P. Health and quality of life before and after hip or knee arthroplasty. J Arthroplasty. 1995;10:169–175.CrossRefPubMed Rissanen P, Aro S, Slatis P, Sintonen H, Paavolainen P. Health and quality of life before and after hip or knee arthroplasty. J Arthroplasty. 1995;10:169–175.CrossRefPubMed
36.
go back to reference Ritter MA, Albohm MJ, Keating EM, Faris PM, Meding JB. Comparative outcomes of total joint arthroplasty. J Arthroplasty. 1995;10:737–741.CrossRefPubMed Ritter MA, Albohm MJ, Keating EM, Faris PM, Meding JB. Comparative outcomes of total joint arthroplasty. J Arthroplasty. 1995;10:737–741.CrossRefPubMed
37.
go back to reference Santavirta S, Boehler M, Harris WH, Konttinen YT, Lappalainen R, Muratoglu O, Rieker C, Salzer M. Alternative materials to improve total hip replacement tribology. Acta Orthop Scand. 2003;74:380–388.CrossRefPubMed Santavirta S, Boehler M, Harris WH, Konttinen YT, Lappalainen R, Muratoglu O, Rieker C, Salzer M. Alternative materials to improve total hip replacement tribology. Acta Orthop Scand. 2003;74:380–388.CrossRefPubMed
38.
go back to reference Tai SM, Munir S, Walter WL, Pearce SJ, Walter WK, Zicat BA. Squeaking in large diameter ceramic-on-ceramic bearings in total hip arthroplasty. J Arthroplasty. 2015;30:282–285.CrossRefPubMed Tai SM, Munir S, Walter WL, Pearce SJ, Walter WK, Zicat BA. Squeaking in large diameter ceramic-on-ceramic bearings in total hip arthroplasty. J Arthroplasty. 2015;30:282–285.CrossRefPubMed
39.
go back to reference Thomas GE, Simpson DJ, Mehmood S, Taylor A, McLardy-Smith P, Gill HS, Murray DW, Glyn-Jones S. The seven-year wear of highly cross-linked polyethylene in total hip arthroplasty: a double-blind, randomized controlled trial using radiostereometric analysis. J Bone Joint Surg Am. 2011;93:716–722.CrossRefPubMed Thomas GE, Simpson DJ, Mehmood S, Taylor A, McLardy-Smith P, Gill HS, Murray DW, Glyn-Jones S. The seven-year wear of highly cross-linked polyethylene in total hip arthroplasty: a double-blind, randomized controlled trial using radiostereometric analysis. J Bone Joint Surg Am. 2011;93:716–722.CrossRefPubMed
40.
go back to reference Traina F, De Fine M, Bordini B, Toni A. Risk factors for ceramic liner fracture after total hip arthroplasty. Hip Int. 2012;22:607–614.CrossRefPubMed Traina F, De Fine M, Bordini B, Toni A. Risk factors for ceramic liner fracture after total hip arthroplasty. Hip Int. 2012;22:607–614.CrossRefPubMed
41.
go back to reference Urban JA, Garvin KL, Boese CK, Bryson K, Pedersen DR, Callaghan JJ, Miller RK. Ceramic-on-polyethylene bearing surfaces in total hip arthroplasty: seventeen to twenty-one-year results. J Bone Joint Surg Am. 2001;83:1688–1694.PubMed Urban JA, Garvin KL, Boese CK, Bryson K, Pedersen DR, Callaghan JJ, Miller RK. Ceramic-on-polyethylene bearing surfaces in total hip arthroplasty: seventeen to twenty-one-year results. J Bone Joint Surg Am. 2001;83:1688–1694.PubMed
42.
go back to reference Walter A. On the material and the tribology of alumina-alumina couplings for hip joint prostheses. Clin Orthop Relat Res. 1992;282:31–46.PubMed Walter A. On the material and the tribology of alumina-alumina couplings for hip joint prostheses. Clin Orthop Relat Res. 1992;282:31–46.PubMed
43.
go back to reference Whittaker JP, Charron KD, McCalden RW, Macdonald SJ, Bourne RB. Comparison of steady state femoral head penetration rates between two highly cross-linked polyethylenes in total hip arthroplasty. J Arthroplasty. 2010;25:680–686.CrossRefPubMed Whittaker JP, Charron KD, McCalden RW, Macdonald SJ, Bourne RB. Comparison of steady state femoral head penetration rates between two highly cross-linked polyethylenes in total hip arthroplasty. J Arthroplasty. 2010;25:680–686.CrossRefPubMed
44.
go back to reference Willert H, Bertram H, Buchhorn G. Osteolysis in alloarthroplasty of the hip: the role of bone cement fragmentation. Clin Orthop Relat Res. 1990;258:108–121.PubMed Willert H, Bertram H, Buchhorn G. Osteolysis in alloarthroplasty of the hip: the role of bone cement fragmentation. Clin Orthop Relat Res. 1990;258:108–121.PubMed
45.
go back to reference Willert H, Bertram H, Buchhorn G. Osteolysis in alloarthroplasty of the hip: the role of ultra high molecular weight polyethylene wear particles. Clin Orthop Relat Res. 1990;258:95–107.PubMed Willert H, Bertram H, Buchhorn G. Osteolysis in alloarthroplasty of the hip: the role of ultra high molecular weight polyethylene wear particles. Clin Orthop Relat Res. 1990;258:95–107.PubMed
46.
go back to reference Willert HG, Semlitsch M. Tissue reactions to plastic and metallic wear products of joint endoprostheses. Clin Orthop Relat Res. 1996;333:4–14.PubMed Willert HG, Semlitsch M. Tissue reactions to plastic and metallic wear products of joint endoprostheses. Clin Orthop Relat Res. 1996;333:4–14.PubMed
Metadata
Title
Oxidized Zirconium Head on Crosslinked Polyethylene Liner in Total Hip Arthroplasty: A 7- to 12-year In Vivo Comparative Wear Study
Authors
George K. Karidakis, MD
Theofilos Karachalios, MD, DSc
Publication date
01-12-2015
Publisher
Springer US
Published in
Clinical Orthopaedics and Related Research® / Issue 12/2015
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-015-4503-7

Other articles of this Issue 12/2015

Clinical Orthopaedics and Related Research® 12/2015 Go to the issue