Overstuffing in resurfacing hemiarthroplasty is a potential risk for failure

Authors: Pieter C. Geervliet, Jore H. Willems, Inger N. Sierevelt, Cornelis P. J. Visser, Arthur van Noort

Published in: Journal of Orthopaedic Surgery and Research | Issue 1/2019

Login to get access

Abstract

Purpose

Literature describes the concern of an overstuffed shoulder joint after a resurfacing humeral head implant (RHHI). The purpose of this study was to evaluate inter-observer variability of (1) the critical shoulder angle (CSA), (2) the length of the gleno-humeral offset (LGHO), and (3) the anatomic center of rotation (COR) in a patient population operated with a Global Conservative Anatomic Prosthesis (CAP) RHHI. The measurements were compared between the revision and non-revision groups to find predictive indicators for failure.

Methods

Pre- and postoperative radiographs were retrieved from 48 patients who underwent RHHI from 2007 to 2009 using a Global CAP hemiarthroplasty for end-stage osteoarthritis. This cohort consisted of 36 females (12 men) with a mean age of 77 years (SD 7.5). Two musculoskeletal radiologist and two specialized shoulder orthopedic surgeons measured the CSA, LGHO, and COR of all patients.

Results

The inter-observer reliability showed excellent reliability for the CSA, LGHO, and the COR, varying between 0.91 and 0.98. The mean COR of the non-revision group was 4.9 mm (SD 2.5) compared to mean COR of the revision group, 8 mm (SD 2.2) (p < 0.01). The COR is the predictor of failure (OR 1.90 (95%Cl 1.19–3.02)) with a cut of point of 5.8 mm. The mean CSA was 29.8° (SD 3.9) There was no significant difference between the revision and non-revision groups (p = 0.34). The mean LGHO was 2.6 mm (SD 3.3) post-surgery. The mean LGHO of the revision group was 3.9 (SD 1.7) (p = 0.04) post-surgery. Despite the difference in mean LGHO, this is not a predictor for failure.

Conclusion

The CSA, LGHO, and COR can be used on radiographs and have a high inter-observer agreement. In contrast with the CSA and LGHO, we found a correlation between clinical failure and revision surgery in case of a deviation of the COR greater than 5 mm.

Trial registration

Institutional review board, number: ACLU 2016.0054, Ethical Committee number: CBP M1330348. Registered 7 November 2006.
Literature
1.
go back to reference Levy O, Funk L, Sforza G, Copeland SA. Copeland surface replacement arthroplasty of the shoulder in rheumatoid arthritis. J Bone Joint Surg - Series A. 2004;86:512–8.CrossRef Levy O, Funk L, Sforza G, Copeland SA. Copeland surface replacement arthroplasty of the shoulder in rheumatoid arthritis. J Bone Joint Surg - Series A. 2004;86:512–8.CrossRef
2.
go back to reference Levy O, Copeland SA. Cementless surface replacement arthroplasty (Copeland CSRA) for osteoarthritis of the shoulder. J Shoulder Elb Surg. 2004;13(3):266.CrossRef Levy O, Copeland SA. Cementless surface replacement arthroplasty (Copeland CSRA) for osteoarthritis of the shoulder. J Shoulder Elb Surg. 2004;13(3):266.CrossRef
3.
go back to reference Thomas SR, Wilson AJ, Chambler A, Harding I, Thomas M. Outcome of Copeland surface replacement shoulder arthroplasty. J Shoulder Elb Surg. 2005;14(5):485.CrossRef Thomas SR, Wilson AJ, Chambler A, Harding I, Thomas M. Outcome of Copeland surface replacement shoulder arthroplasty. J Shoulder Elb Surg. 2005;14(5):485.CrossRef
4.
go back to reference Mullett H, Levy O, Raj D, Even T, Abraham R, Copeland SA. Copeland surface replacement of the shoulder: results of an hydroxyapatite-coated cementless impant in patients over 80 years of age. J Bone and Joint Surg - British Volume. 2007;89(11):1466.CrossRef Mullett H, Levy O, Raj D, Even T, Abraham R, Copeland SA. Copeland surface replacement of the shoulder: results of an hydroxyapatite-coated cementless impant in patients over 80 years of age. J Bone and Joint Surg - British Volume. 2007;89(11):1466.CrossRef
5.
go back to reference Pritchett JW. Long-term results and patient satisfaction after shoulder resurfacing. J Shoulder Elb Surg. 2011;20(5):771.CrossRef Pritchett JW. Long-term results and patient satisfaction after shoulder resurfacing. J Shoulder Elb Surg. 2011;20(5):771.CrossRef
6.
go back to reference Al-Hadithy N, Domos P, Sewell MD, Naleem A, Papanna MC, Pandit R. Cementless surface replacement arthroplasty of the shoulder for osteoarthritis: results of fifty mark III Copeland prosthesis from an independent center with four-year mean follow-up. J Shoulder Elb Surg. 2012;21:1776–81.CrossRef Al-Hadithy N, Domos P, Sewell MD, Naleem A, Papanna MC, Pandit R. Cementless surface replacement arthroplasty of the shoulder for osteoarthritis: results of fifty mark III Copeland prosthesis from an independent center with four-year mean follow-up. J Shoulder Elb Surg. 2012;21:1776–81.CrossRef
7.
go back to reference Rasmussen JV, Polk A, Sorensen AK, Olsen BS, Brorson S. Outcome, revision rate and indication for revision following resurfacing hemiarthroplasty for osteoarthritis of the shoulder: 837 operations reported to the Danish shoulder arthroplasty registry. Bone Joint J. 2014;96:519–25.PubMedCrossRef Rasmussen JV, Polk A, Sorensen AK, Olsen BS, Brorson S. Outcome, revision rate and indication for revision following resurfacing hemiarthroplasty for osteoarthritis of the shoulder: 837 operations reported to the Danish shoulder arthroplasty registry. Bone Joint J. 2014;96:519–25.PubMedCrossRef
8.
go back to reference Alolabi B, Youderian AR, Napolitano L, Szerlip BW, Evans PJ, Nowinski RJ, et al. Radiographic assessment of prosthetic humeral head size after anatomic shoulder arthroplasty. J Shoulder Elb Surg. 2014;23(11):1740.CrossRef Alolabi B, Youderian AR, Napolitano L, Szerlip BW, Evans PJ, Nowinski RJ, et al. Radiographic assessment of prosthetic humeral head size after anatomic shoulder arthroplasty. J Shoulder Elb Surg. 2014;23(11):1740.CrossRef
9.
go back to reference Copeland S. The continuing development of shoulder replacement: “Reaching the surface”. J Bone Joint Surg - Series A. 2006;88(4):900. Copeland S. The continuing development of shoulder replacement: “Reaching the surface”. J Bone Joint Surg - Series A. 2006;88(4):900.
10.
go back to reference Mechlenburg I, Amstrup A, Klebe T, Jacobsen SS, Teichert G, Stilling M. The Copeland resurfacing humeral head implant does not restore humeral head anatomy. A retrospective study. Arch Orthop Trauma Surg. 2013;133:615–9.PubMedCrossRef Mechlenburg I, Amstrup A, Klebe T, Jacobsen SS, Teichert G, Stilling M. The Copeland resurfacing humeral head implant does not restore humeral head anatomy. A retrospective study. Arch Orthop Trauma Surg. 2013;133:615–9.PubMedCrossRef
12.
go back to reference Alizadehkhaiyat O, Kyriakos A, Singer MS, Frostick SP. Outcome of Copeland shoulder resurfacing arthroplasty with a 4-year mean follow-up. J Shoulder Elb Surg. 2013;22:1352–8.CrossRef Alizadehkhaiyat O, Kyriakos A, Singer MS, Frostick SP. Outcome of Copeland shoulder resurfacing arthroplasty with a 4-year mean follow-up. J Shoulder Elb Surg. 2013;22:1352–8.CrossRef
13.
go back to reference Geervliet PC, van den Bekerom MPJ, Spruyt P, Curvers M, van Noort A, Visser CPJ. Outcome and revision rate of uncemented glenohumeral resurfacing (C.A.P.) after 5–8 years. Archives of Orthopaedic and Trauma Surgery. 2017. Springer Berlin Heidelberg; [Cited 2017 Apr 22];1–8. Available from: http://link.springer.com/10.1007/s00402-017-2688-9 Geervliet PC, van den Bekerom MPJ, Spruyt P, Curvers M, van Noort A, Visser CPJ. Outcome and revision rate of uncemented glenohumeral resurfacing (C.A.P.) after 5–8 years. Archives of Orthopaedic and Trauma Surgery. 2017. Springer Berlin Heidelberg; [Cited 2017 Apr 22];1–8. Available from: http://​link.​springer.​com/​10.​1007/​s00402-017-2688-9
14.
go back to reference Geervliet P, Van Den Bekerom M, Spruyt P, Curvers M, Visser C, van Noort A. Short-term results of the global C.A.P. uncemented resurfacing shoulder prosthesis. Orthopedics. 2014;37:42–7.PubMedCrossRef Geervliet P, Van Den Bekerom M, Spruyt P, Curvers M, Visser C, van Noort A. Short-term results of the global C.A.P. uncemented resurfacing shoulder prosthesis. Orthopedics. 2014;37:42–7.PubMedCrossRef
15.
go back to reference Engelsma Y, Buma P, Geervliet PC, Noort AV. Septic failure is not a septic loosening: a case report of a failed shoulder prosthesis. I J Shoulder Surg. 2012;6:1.CrossRef Engelsma Y, Buma P, Geervliet PC, Noort AV. Septic failure is not a septic loosening: a case report of a failed shoulder prosthesis. I J Shoulder Surg. 2012;6:1.CrossRef
17.
go back to reference Walch G, Badet R, Boulahia A, Khoury A. Morphologic study of the glenoid in primary glenohumeral osteoarthritis. J Arthroplast. 1999;14:756–60.CrossRef Walch G, Badet R, Boulahia A, Khoury A. Morphologic study of the glenoid in primary glenohumeral osteoarthritis. J Arthroplast. 1999;14:756–60.CrossRef
19.
go back to reference Moor BK, Bouaicha S, Rothenfluh DA, Sukthankar A, Gerber C. Is there an association between the individual anatomy of the scapula and the development of rotator cuff tears or osteoarthritis of the glenohumeral joint? A radiological study of the critical shoulder angle. Bone Joint J. 2013;95(7):935.PubMedCrossRef Moor BK, Bouaicha S, Rothenfluh DA, Sukthankar A, Gerber C. Is there an association between the individual anatomy of the scapula and the development of rotator cuff tears or osteoarthritis of the glenohumeral joint? A radiological study of the critical shoulder angle. Bone Joint J. 2013;95(7):935.PubMedCrossRef
22.
go back to reference Iannotti JP, Gabriel JP, Schneck SL, Evans BG, Misra S. The normal glenohumeral relationships. An anatomical study of one hundred and forty shoulders. J Bone Joint Surg - Series A. 1992;74(4):491.CrossRef Iannotti JP, Gabriel JP, Schneck SL, Evans BG, Misra S. The normal glenohumeral relationships. An anatomical study of one hundred and forty shoulders. J Bone Joint Surg - Series A. 1992;74(4):491.CrossRef
23.
go back to reference de Vet HCW, Terwee CB, Knol DL, Bouter LM. When to use agreement versus reliability measures. J Clin Epidemiol. 2006;59(10):1033.PubMedCrossRef de Vet HCW, Terwee CB, Knol DL, Bouter LM. When to use agreement versus reliability measures. J Clin Epidemiol. 2006;59(10):1033.PubMedCrossRef
24.
go back to reference Streiner DL, Norman GR. Health measurement scales: a practical guide to their development and use. Health measurement scales: a practical guide to their development and use. USA: Oxford University Press; 2008.CrossRef Streiner DL, Norman GR. Health measurement scales: a practical guide to their development and use. Health measurement scales: a practical guide to their development and use. USA: Oxford University Press; 2008.CrossRef
25.
go back to reference Weir JP. Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res. 2005;19(1):231.PubMed Weir JP. Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res. 2005;19(1):231.PubMed
26.
go back to reference de Vet HC, Terwee CB, Ostelo RW, Beckerman H, Knol DL, Bouter LM. Minimal changes in health status questionnaires: distinction between minimally detectable change and minimally important change. Health Qual Life Outcomes. 2006;4(1):54.PubMedPubMedCentralCrossRef de Vet HC, Terwee CB, Ostelo RW, Beckerman H, Knol DL, Bouter LM. Minimal changes in health status questionnaires: distinction between minimally detectable change and minimally important change. Health Qual Life Outcomes. 2006;4(1):54.PubMedPubMedCentralCrossRef
27.
go back to reference Favre P, Moor B, Snedeker JG, Gerber C. Influence of component positioning on impingement in conventional total shoulder arthroplasty. Clin Biomech. 2008;23(2):175.CrossRef Favre P, Moor B, Snedeker JG, Gerber C. Influence of component positioning on impingement in conventional total shoulder arthroplasty. Clin Biomech. 2008;23(2):175.CrossRef
28.
go back to reference Harryman DT, Sidles JA, Harris SL, Lippitt SB, Matsen FA. The effect of articular conformity and the size of the humeral head component on laxity and motion after glenohumeral arthroplasty: a study in cadavera. J Bone Joint Surg - Series A. 1995;77(4):555.CrossRef Harryman DT, Sidles JA, Harris SL, Lippitt SB, Matsen FA. The effect of articular conformity and the size of the humeral head component on laxity and motion after glenohumeral arthroplasty: a study in cadavera. J Bone Joint Surg - Series A. 1995;77(4):555.CrossRef
29.
go back to reference Nyffeler RW, Sheikh R, Jacob HAC, Gerber C. Influence of humeral prosthesis height on biomechanics of glenohumeral abduction: an in vitro study. J Bone Joint Surg - Series A. 2004;77(4):555. Nyffeler RW, Sheikh R, Jacob HAC, Gerber C. Influence of humeral prosthesis height on biomechanics of glenohumeral abduction: an in vitro study. J Bone Joint Surg - Series A. 2004;77(4):555.
30.
go back to reference Terrier A, Ramondetti S, Merlini F, Pioletti DD, Farron A. Biomechanical consequences of humeral component malpositioning after anatomical total shoulder arthroplasty. J Shoulder Elb Surg. 2010;19(8):1184.CrossRef Terrier A, Ramondetti S, Merlini F, Pioletti DD, Farron A. Biomechanical consequences of humeral component malpositioning after anatomical total shoulder arthroplasty. J Shoulder Elb Surg. 2010;19(8):1184.CrossRef
31.
go back to reference Thomas SR, Sforza G, Levy O. Copeland SA. Geometrical analysis of Copeland surface replacement shoulder arthroplasty in relation to normal anatomy. J Shoulder Elb Surg. 2005;14(2):186.CrossRef Thomas SR, Sforza G, Levy O. Copeland SA. Geometrical analysis of Copeland surface replacement shoulder arthroplasty in relation to normal anatomy. J Shoulder Elb Surg. 2005;14(2):186.CrossRef
32.
go back to reference Williams GR, Wong KL, Pepe MD, Tan V, Silverberg D, Ramsey ML, et al. The effect of articular malposition after total shoulder arthroplasty on glenohumeral translations, range of motion, and subacromial impingement. J Shoulder Elb Surg. 2001;10(5):399.CrossRef Williams GR, Wong KL, Pepe MD, Tan V, Silverberg D, Ramsey ML, et al. The effect of articular malposition after total shoulder arthroplasty on glenohumeral translations, range of motion, and subacromial impingement. J Shoulder Elb Surg. 2001;10(5):399.CrossRef
35.
go back to reference Moor BK, Kuster R, Osterhoff G, Baumgartner D, CML W, Zumstein MA, et al. Inclination-dependent changes of the critical shoulder angle significantly influence superior glenohumeral joint stability. Clin Biomech. 2016;32:268.CrossRef Moor BK, Kuster R, Osterhoff G, Baumgartner D, CML W, Zumstein MA, et al. Inclination-dependent changes of the critical shoulder angle significantly influence superior glenohumeral joint stability. Clin Biomech. 2016;32:268.CrossRef
36.
go back to reference Gerber C, Snedeker JG, Baumgartner D, Viehöfer AF. Supraspinatus tendon load during abduction is dependent on the size of the critical shoulder angle: a biomechanical analysis. J Orthop Res. 2014;32(7):952.PubMedCrossRef Gerber C, Snedeker JG, Baumgartner D, Viehöfer AF. Supraspinatus tendon load during abduction is dependent on the size of the critical shoulder angle: a biomechanical analysis. J Orthop Res. 2014;32(7):952.PubMedCrossRef
37.
go back to reference Mantell MT, Nelson R, Lowe JT, Endrizzi DP, Jawa A. Critical shoulder angle is associated with full-thickness rotator cuff tears in patients with glenohumeral osteoarthritis. J Shoulder Elb Surg. 2017;26(12):e376.CrossRef Mantell MT, Nelson R, Lowe JT, Endrizzi DP, Jawa A. Critical shoulder angle is associated with full-thickness rotator cuff tears in patients with glenohumeral osteoarthritis. J Shoulder Elb Surg. 2017;26(12):e376.CrossRef
38.
go back to reference Watling JP, Sanchez JE, Heilbroner SP, Levine WN, Bigliani LU, Jobin CM. Glenoid component loosening associated with increased critical shoulder angle at midterm follow-up. J Shoulder Elb Surg. 2018;27(3):449.CrossRef Watling JP, Sanchez JE, Heilbroner SP, Levine WN, Bigliani LU, Jobin CM. Glenoid component loosening associated with increased critical shoulder angle at midterm follow-up. J Shoulder Elb Surg. 2018;27(3):449.CrossRef
39.
go back to reference Lehtinen JT, Belt EA, Kauppi MJ, Kaarela K, Kuusela PP, Kautiainen HJ, et al. Bone destruction, upward migration, and medialisation of rheumatoid shoulder: a 15 year follow up study. Ann Rheum Dis. 2001;60(4):322.PubMedPubMedCentralCrossRef Lehtinen JT, Belt EA, Kauppi MJ, Kaarela K, Kuusela PP, Kautiainen HJ, et al. Bone destruction, upward migration, and medialisation of rheumatoid shoulder: a 15 year follow up study. Ann Rheum Dis. 2001;60(4):322.PubMedPubMedCentralCrossRef
40.
go back to reference Pearl ML, Kurutz S, Robertson DD, Yamaguchi K. Geometric analysis of selected press fit prosthetic systems for proximal humeral replacement. J Orthop Res. 2002;20(2):192.PubMedCrossRef Pearl ML, Kurutz S, Robertson DD, Yamaguchi K. Geometric analysis of selected press fit prosthetic systems for proximal humeral replacement. J Orthop Res. 2002;20(2):192.PubMedCrossRef
41.
go back to reference Pearl ML, Kurutz S. Geometric analysis of commonly used prosthetic systems for proximal humeral replacement. J Bone Joint Surg (Am Vol). 1999;81(5):660.CrossRef Pearl ML, Kurutz S. Geometric analysis of commonly used prosthetic systems for proximal humeral replacement. J Bone Joint Surg (Am Vol). 1999;81(5):660.CrossRef
42.
go back to reference Levy O, Tsvieli O, Merchant J, Young L, Trimarchi A, Dattani R, et al. Surface replacement arthroplasty for glenohumeral arthropathy in patients aged younger than fifty years: results after a minimum ten-year follow-up. Journal of shoulder and elbow surgery. Elsevier Ltd. 2015;24:1049–60. Levy O, Tsvieli O, Merchant J, Young L, Trimarchi A, Dattani R, et al. Surface replacement arthroplasty for glenohumeral arthropathy in patients aged younger than fifty years: results after a minimum ten-year follow-up. Journal of shoulder and elbow surgery. Elsevier Ltd. 2015;24:1049–60.
43.
go back to reference Levy O, Copeland S. Cementless surface replacement arthroplasty of the shoulder: 5- to 10-year results with the Copeland mark-2 prosthesis. J Bone Joint Surg Br Vol. 2001;83(2):213.CrossRef Levy O, Copeland S. Cementless surface replacement arthroplasty of the shoulder: 5- to 10-year results with the Copeland mark-2 prosthesis. J Bone Joint Surg Br Vol. 2001;83(2):213.CrossRef
44.
go back to reference Spiegl UJ, Horan MP, Smith SW, Ho CP, Millett PJ. The critical shoulder angle is associated with rotator cuff tears and shoulder osteoarthritis and is better assessed with radiographs over MRI. Knee Surg Sports Traumatol Arthrosc. 2016;24(7):2244.PubMedCrossRef Spiegl UJ, Horan MP, Smith SW, Ho CP, Millett PJ. The critical shoulder angle is associated with rotator cuff tears and shoulder osteoarthritis and is better assessed with radiographs over MRI. Knee Surg Sports Traumatol Arthrosc. 2016;24(7):2244.PubMedCrossRef
45.
go back to reference Bouaicha S, Ehrmann C, Slankamenac K, Regan WD, Moor BK. Comparison of the critical shoulder angle in radiographs and computed tomography. Skelet Radiol. 2014;43(8):1053.CrossRef Bouaicha S, Ehrmann C, Slankamenac K, Regan WD, Moor BK. Comparison of the critical shoulder angle in radiographs and computed tomography. Skelet Radiol. 2014;43(8):1053.CrossRef
46.
go back to reference Youderian AR, Ricchetti ET, Drews M, Iannotti JP. Determination of humeral head size in anatomic shoulder replacement for glenohumeral osteoarthritis. J Shoulder Elb Surg. 2014;23(7):955.CrossRef Youderian AR, Ricchetti ET, Drews M, Iannotti JP. Determination of humeral head size in anatomic shoulder replacement for glenohumeral osteoarthritis. J Shoulder Elb Surg. 2014;23(7):955.CrossRef
47.
go back to reference von Engelhardt LV, Manzke M, Breil-Wirth A, Filler TJ, Jerosch J. Restoration of the joint geometry and outcome after stemless TESS shoulder arthroplasty. World J Orthop. 2017;8(10):790.CrossRef von Engelhardt LV, Manzke M, Breil-Wirth A, Filler TJ, Jerosch J. Restoration of the joint geometry and outcome after stemless TESS shoulder arthroplasty. World J Orthop. 2017;8(10):790.CrossRef
48.
go back to reference Chalmers PN, Granger EK, Orvets ND, Patterson BM, Chamberlain AM, Keener JD, et al. Does prosthetic humeral articular surface positioning associate with outcome after total shoulder arthroplasty? J Shoulder Elb Surg. 2018;27(5):863.CrossRef Chalmers PN, Granger EK, Orvets ND, Patterson BM, Chamberlain AM, Keener JD, et al. Does prosthetic humeral articular surface positioning associate with outcome after total shoulder arthroplasty? J Shoulder Elb Surg. 2018;27(5):863.CrossRef
49.
go back to reference Pearl ML, Kurutz S, Postachini R. Geometric variables in anatomic replacement of the proximal humerus: how much prosthetic geometry is necessary? J Shoulder Elb Surg. 2009;18(3):366.CrossRef Pearl ML, Kurutz S, Postachini R. Geometric variables in anatomic replacement of the proximal humerus: how much prosthetic geometry is necessary? J Shoulder Elb Surg. 2009;18(3):366.CrossRef
50.
go back to reference Jeong J, Bryan J, Iannotti JP. Effect of a variable prosthetic neck-shaft angle and the surgical technique on replication of normal humeral anatomy. J BoneJoint Surg - Series A. 2009;91(8):1932.CrossRef Jeong J, Bryan J, Iannotti JP. Effect of a variable prosthetic neck-shaft angle and the surgical technique on replication of normal humeral anatomy. J BoneJoint Surg - Series A. 2009;91(8):1932.CrossRef
Metadata
Title
Overstuffing in resurfacing hemiarthroplasty is a potential risk for failure
Authors
Pieter C. Geervliet
Jore H. Willems
Inger N. Sierevelt
Cornelis P. J. Visser
Arthur van Noort
Publication date
01-12-2019
Publisher
BioMed Central
Published in
Journal of Orthopaedic Surgery and Research / Issue 1/2019
Electronic ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-019-1522-1

Other articles of this Issue 1/2019

Journal of Orthopaedic Surgery and Research 1/2019 Go to the issue