A novel bone cement injector augments Chinese osteoporotic lumbar pedicle screw channel: a biomechanical investigation

Authors: Suochao Fu, Yu Zhang, Fuzhi Ai, Jianhua Wang, Zenghui Wu, Xiangyang Ma, Zixiang Wu, Zheng Wang, Wei Lei, Hong Xia

Published in: BMC Musculoskeletal Disorders | Issue 1/2022

Login to get access

Abstract

Background

The study aimed to (1) create a series of pedicle injectors with different number of holes on the sheath especially for the Chinese elderly patients and (2) further investigate the effects of the injectors on the augmentation of pedicle screw among osteoporotic lumbar pedicle channel.

Methods

This study used the biomechanical test module of polyurethane (Pacific Research Laboratory Corp, USA) to simulate the mechanical properties of human osteoporotic cancellous bone. The bone cement injectors were invented based on anatomical parameters of lumbar pedicle in Chinese elderly patients. Mechanical test experiments were performed on the bone cement injectors according to the three groups, namely, a local augmentation group, a full-length augmentation group, and a control group. The local augmentation group included three subgroups including 4-hole group, 6-hole group, and 8-hole group. All holes were laterally placed. The full-length augmentation group was a straight-hole injector. The control group was defined that pedicle screws were inserted without any cement augmentation. Six screws were inserted in each group and the maximum insertion torque was recorded. After 24 h of injecting acrylic bone cement, routine X-ray and CT examinations were performed to evaluate the distribution of bone cement. The axial pull-out force of screws was tested with the help of the material testing system 858 (MTS-858) mechanical tester.

Results

The bone cement injectors were consisted of the sheaths and the steel rods and the sheaths had different number of lateral holes. The control group had the lowest maximum insertion torque as compared with the 4-hole, 6-hole, 8-hole, and straight pore groups (P < 0.01), but the differences between the 4-hole, 6-hole, 8-hole, and straight pore groups were no statistical significance. The control group had the lowest maximum axial pull-out force as compared with the other four groups (P < 0.01). Subgroup analysis showed the 8-hole group (161.35 ± 27.17 N) had the lower maximum axial pull-out force as compared with the 4-hole group (217.29 ± 49.68 N), 6-hole group (228.39 ± 57.83 N), and straight pore group (237.55 ± 35.96 N) (P < 0.01). Bone cement was mainly distributed in 1/3 of the distal end of the screw among the 4-hole group, in the middle 1/3 and distal end of the screw among the 6-hole group, in the proximal 1/3 of the screw among the 8-hole group, and along the long axis of the whole screw body in the straight pore group. It might indicate that the 8-hole and straight-hole groups were more vulnerable to spinal canal cement leakage. After pullout, bone cement was also closely connected with the screw without any looseness or fragmentation.

Conclusions

The bone cement injectors with different number of holes can be used to augment the pedicle screw channel. The pedicle screw augmented by the 4-hole or 6-hole sheath may have similar effects to the straight pore injector. However, the 8-hole injector may result in relatively lower pull-out strength and the straight pore injector has the risks of cement leakage as well as cement solidarization near the screw head.
Literature
1.
go back to reference Hernlund E, Svedbom A, Ivergard M, Compston J, Cooper C, Stenmark J, McCloskey EV, Jonsson B, Kanis JA. Osteoporosis in the European Union: medical management, epidemiology and economic burden. A report prepared in collaboration with the International Osteoporosis Foundation (IOF) and the European Federation of Pharmaceutical Industry Associations (EFPIA). Arch Osteoporos. 2013;8:136.CrossRef Hernlund E, Svedbom A, Ivergard M, Compston J, Cooper C, Stenmark J, McCloskey EV, Jonsson B, Kanis JA. Osteoporosis in the European Union: medical management, epidemiology and economic burden. A report prepared in collaboration with the International Osteoporosis Foundation (IOF) and the European Federation of Pharmaceutical Industry Associations (EFPIA). Arch Osteoporos. 2013;8:136.CrossRef
2.
go back to reference Edwards MH, Dennison EM, Aihie Sayer A, Fielding R, Cooper C. Osteoporosis and sarcopenia in older age. Bone. 2015;80:126–30.CrossRef Edwards MH, Dennison EM, Aihie Sayer A, Fielding R, Cooper C. Osteoporosis and sarcopenia in older age. Bone. 2015;80:126–30.CrossRef
3.
go back to reference Teoh SH, Chui CK. Bone material properties and fracture analysis: needle insertion for spinal surgery. J Mech Behav Biomed Mater. 2008;1(2):115–39.CrossRef Teoh SH, Chui CK. Bone material properties and fracture analysis: needle insertion for spinal surgery. J Mech Behav Biomed Mater. 2008;1(2):115–39.CrossRef
4.
go back to reference Looker AC, Orwoll ES, Johnston CC Jr, Lindsay RL, Wahner HW, Dunn WL, Calvo MS, Harris TB, Heyse SP. Prevalence of low femoral bone density in older U.S. adults from NHANES III. J Bone Miner Res. 1997;12(11):1761–8.CrossRef Looker AC, Orwoll ES, Johnston CC Jr, Lindsay RL, Wahner HW, Dunn WL, Calvo MS, Harris TB, Heyse SP. Prevalence of low femoral bone density in older U.S. adults from NHANES III. J Bone Miner Res. 1997;12(11):1761–8.CrossRef
5.
go back to reference Kanis JA, Johnell O, Oden A, Jonsson B, De Laet C, Dawson A. Risk of hip fracture according to the World Health Organization criteria for osteopenia and osteoporosis. Bone. 2000;27(5):585–90.CrossRef Kanis JA, Johnell O, Oden A, Jonsson B, De Laet C, Dawson A. Risk of hip fracture according to the World Health Organization criteria for osteopenia and osteoporosis. Bone. 2000;27(5):585–90.CrossRef
6.
go back to reference Lin X, Xiong D, Peng YQ, Sheng ZF, Wu XY, Wu XP, Wu F, Yuan LQ, Liao EY. Epidemiology and management of osteoporosis in the People’s Republic of China: current perspectives. Clin Interv Aging. 2015;10:1017–33.PubMedPubMedCentral Lin X, Xiong D, Peng YQ, Sheng ZF, Wu XY, Wu XP, Wu F, Yuan LQ, Liao EY. Epidemiology and management of osteoporosis in the People’s Republic of China: current perspectives. Clin Interv Aging. 2015;10:1017–33.PubMedPubMedCentral
7.
go back to reference Zhang J, Dennison E, Prieto-Alhambra D. Osteoporosis epidemiology using international cohorts. Curr Opin Rheumatol. 2020;32(4):387–93.CrossRef Zhang J, Dennison E, Prieto-Alhambra D. Osteoporosis epidemiology using international cohorts. Curr Opin Rheumatol. 2020;32(4):387–93.CrossRef
8.
go back to reference van Staa TP, Dennison EM, Leufkens HG, Cooper C. Epidemiology of fractures in England and Wales. Bone. 2001;29(6):517–22.CrossRef van Staa TP, Dennison EM, Leufkens HG, Cooper C. Epidemiology of fractures in England and Wales. Bone. 2001;29(6):517–22.CrossRef
9.
go back to reference Ballane G, Cauley JA, Luckey MM, El-Hajj Fuleihan G. Worldwide prevalence and incidence of osteoporotic vertebral fractures. Osteoporos Int. 2017;28(5):1531–42.CrossRef Ballane G, Cauley JA, Luckey MM, El-Hajj Fuleihan G. Worldwide prevalence and incidence of osteoporotic vertebral fractures. Osteoporos Int. 2017;28(5):1531–42.CrossRef
10.
go back to reference Si L, Winzenberg TM, Jiang Q, Chen M, Palmer AJ. Projection of osteoporosis-related fractures and costs in China: 2010–2050. Osteoporos Int. 2015;26(7):1929–37.CrossRef Si L, Winzenberg TM, Jiang Q, Chen M, Palmer AJ. Projection of osteoporosis-related fractures and costs in China: 2010–2050. Osteoporos Int. 2015;26(7):1929–37.CrossRef
11.
go back to reference Kutsal FY, Ergin Ergani GO. Vertebral compression fractures: still an unpredictable aspect of osteoporosis. Turk J Med Sci. 2021;51(2):393–9.CrossRef Kutsal FY, Ergin Ergani GO. Vertebral compression fractures: still an unpredictable aspect of osteoporosis. Turk J Med Sci. 2021;51(2):393–9.CrossRef
12.
go back to reference Clynes MA, Harvey NC, Curtis EM, Fuggle NR, Dennison EM, Cooper C. The epidemiology of osteoporosis. Br Med Bull. 2020;133(1):105–17.PubMed Clynes MA, Harvey NC, Curtis EM, Fuggle NR, Dennison EM, Cooper C. The epidemiology of osteoporosis. Br Med Bull. 2020;133(1):105–17.PubMed
13.
go back to reference Wood KB, Li W, Lebl DR, Ploumis A. Management of thoracolumbar spine fractures. Spine J. 2014;14(1):145–64.CrossRef Wood KB, Li W, Lebl DR, Ploumis A. Management of thoracolumbar spine fractures. Spine J. 2014;14(1):145–64.CrossRef
14.
go back to reference Halvorson TL, Kelley LA, Thomas KA, Whitecloud TS 3rd, Cook SD. Effects of bone mineral density on pedicle screw fixation. Spine (Phila Pa 1976). 1994;19(21):2415–20.CrossRef Halvorson TL, Kelley LA, Thomas KA, Whitecloud TS 3rd, Cook SD. Effects of bone mineral density on pedicle screw fixation. Spine (Phila Pa 1976). 1994;19(21):2415–20.CrossRef
15.
go back to reference Ponnusamy KE, Iyer S, Gupta G, Khanna AJ. Instrumentation of the osteoporotic spine: biomechanical and clinical considerations. Spine J. 2011;11(1):54–63.CrossRef Ponnusamy KE, Iyer S, Gupta G, Khanna AJ. Instrumentation of the osteoporotic spine: biomechanical and clinical considerations. Spine J. 2011;11(1):54–63.CrossRef
16.
go back to reference Zong ZW, Qin H, Chen SX, Yang JZ, Yang L, Zhang L, Du WQ, Zhong X, Zhou RJ, Tan D, et al. Chinese expert consensus on the treatment of modern combat-related spinal injuries. Mil Med Res. 2019;6(1):6.PubMedPubMedCentral Zong ZW, Qin H, Chen SX, Yang JZ, Yang L, Zhang L, Du WQ, Zhong X, Zhou RJ, Tan D, et al. Chinese expert consensus on the treatment of modern combat-related spinal injuries. Mil Med Res. 2019;6(1):6.PubMedPubMedCentral
17.
go back to reference Jain P, Rana M, Biswas JK, Khan MR. Biomechanics of spinal implants-a review. Biomed Phys Eng Express. 2020;6(4):042002.CrossRef Jain P, Rana M, Biswas JK, Khan MR. Biomechanics of spinal implants-a review. Biomed Phys Eng Express. 2020;6(4):042002.CrossRef
18.
go back to reference Hoppe S, Keel MJ. Pedicle screw augmentation in osteoporotic spine: indications, limitations and technical aspects. Eur J Trauma Emerg Surg. 2017;43(1):3–8.CrossRef Hoppe S, Keel MJ. Pedicle screw augmentation in osteoporotic spine: indications, limitations and technical aspects. Eur J Trauma Emerg Surg. 2017;43(1):3–8.CrossRef
19.
go back to reference El Saman A, Meier S, Sander A, Kelm A, Marzi I, Laurer H. Reduced loosening rate and loss of correction following posterior stabilization with or without PMMA augmentation of pedicle screws in vertebral fractures in the elderly. Eur J Trauma Emerg Surg. 2013;39(5):455–60.CrossRef El Saman A, Meier S, Sander A, Kelm A, Marzi I, Laurer H. Reduced loosening rate and loss of correction following posterior stabilization with or without PMMA augmentation of pedicle screws in vertebral fractures in the elderly. Eur J Trauma Emerg Surg. 2013;39(5):455–60.CrossRef
20.
go back to reference Polly DW Jr, Orchowski JR, Ellenbogen RG. Revision pedicle screws. Bigger, longer shims–what is best? Spine (Phila Pa 1976). 1998;23(12):1374–9.CrossRef Polly DW Jr, Orchowski JR, Ellenbogen RG. Revision pedicle screws. Bigger, longer shims–what is best? Spine (Phila Pa 1976). 1998;23(12):1374–9.CrossRef
21.
go back to reference Brasiliense LB, Lazaro BC, Reyes PM, Newcomb AG, Turner JL, Crandall DG, Crawford NR. Characteristics of immediate and fatigue strength of a dual-threaded pedicle screw in cadaveric spines. Spine J. 2013;13(8):947–56.CrossRef Brasiliense LB, Lazaro BC, Reyes PM, Newcomb AG, Turner JL, Crandall DG, Crawford NR. Characteristics of immediate and fatigue strength of a dual-threaded pedicle screw in cadaveric spines. Spine J. 2013;13(8):947–56.CrossRef
22.
go back to reference Wang W, Liu C, Li J, Li H, Wu J, Liu H, Li C, Zhou Y. Comparison of the fenestrated pedicle screw and conventional pedicle screw in minimally percutaneous fixation for the treatment of spondylolisthesis with osteoporotic spine. Clin Neurol Neurosurg. 2019;183:105377.CrossRef Wang W, Liu C, Li J, Li H, Wu J, Liu H, Li C, Zhou Y. Comparison of the fenestrated pedicle screw and conventional pedicle screw in minimally percutaneous fixation for the treatment of spondylolisthesis with osteoporotic spine. Clin Neurol Neurosurg. 2019;183:105377.CrossRef
23.
go back to reference Jain P, Khan MR. Selection of suitable pedicle screw for degenerated cortical and cancellous bone of human lumbar spine: A finite element study. Int J Artif Organs. 2021;44(5):361–6.CrossRef Jain P, Khan MR. Selection of suitable pedicle screw for degenerated cortical and cancellous bone of human lumbar spine: A finite element study. Int J Artif Organs. 2021;44(5):361–6.CrossRef
24.
go back to reference Jain P, Khan MR. Biomechanical Study of Fused Lumbar Spine Considering Bone Degeneracy Using FEA. Arab J Sci Eng. 2018;43(3):1325–34. Jain P, Khan MR. Biomechanical Study of Fused Lumbar Spine Considering Bone Degeneracy Using FEA. Arab J Sci Eng. 2018;43(3):1325–34.
25.
go back to reference Rohmiller MT, Schwalm D, Glattes RC, Elalayli TG, Spengler DM. Evaluation of calcium sulfate paste for augmentation of lumbar pedicle screw pullout strength. Spine J. 2002;2(4):255–60.CrossRef Rohmiller MT, Schwalm D, Glattes RC, Elalayli TG, Spengler DM. Evaluation of calcium sulfate paste for augmentation of lumbar pedicle screw pullout strength. Spine J. 2002;2(4):255–60.CrossRef
26.
go back to reference Chen LH, Tai CL, Lee DM, Lai PL, Lee YC, Niu CC, Chen WJ. Pullout strength of pedicle screws with cement augmentation in severe osteoporosis: a comparative study between cannulated screws with cement injection and solid screws with cement pre-filling. BMC Musculoskelet Disord. 2011;12:33.CrossRef Chen LH, Tai CL, Lee DM, Lai PL, Lee YC, Niu CC, Chen WJ. Pullout strength of pedicle screws with cement augmentation in severe osteoporosis: a comparative study between cannulated screws with cement injection and solid screws with cement pre-filling. BMC Musculoskelet Disord. 2011;12:33.CrossRef
27.
go back to reference Hashemi A, Bednar D, Ziada S. Pullout strength of pedicle screws augmented with particulate calcium phosphate: an experimental study. Spine J. 2009;9(5):404–10.CrossRef Hashemi A, Bednar D, Ziada S. Pullout strength of pedicle screws augmented with particulate calcium phosphate: an experimental study. Spine J. 2009;9(5):404–10.CrossRef
28.
go back to reference Chen LH, Tai CL, Lai PL, Lee DM, Tsai TT, Fu TS, Niu CC, Chen WJ. Pullout strength for cannulated pedicle screws with bone cement augmentation in severely osteoporotic bone: influences of radial hole and pilot hole tapping. Clin Biomech (Bristol, Avon). 2009;24(8):613–8.CrossRef Chen LH, Tai CL, Lai PL, Lee DM, Tsai TT, Fu TS, Niu CC, Chen WJ. Pullout strength for cannulated pedicle screws with bone cement augmentation in severely osteoporotic bone: influences of radial hole and pilot hole tapping. Clin Biomech (Bristol, Avon). 2009;24(8):613–8.CrossRef
29.
go back to reference Hsu CC, Chao CK, Wang JL, Hou SM, Tsai YT, Lin J. Increase of pullout strength of spinal pedicle screws with conical core: biomechanical tests and finite element analyses. J Orthop Res. 2005;23(4):788–94.CrossRef Hsu CC, Chao CK, Wang JL, Hou SM, Tsai YT, Lin J. Increase of pullout strength of spinal pedicle screws with conical core: biomechanical tests and finite element analyses. J Orthop Res. 2005;23(4):788–94.CrossRef
Metadata
Title
A novel bone cement injector augments Chinese osteoporotic lumbar pedicle screw channel: a biomechanical investigation
Authors
Suochao Fu
Yu Zhang
Fuzhi Ai
Jianhua Wang
Zenghui Wu
Xiangyang Ma
Zixiang Wu
Zheng Wang
Wei Lei
Hong Xia
Publication date
01-12-2022
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2022
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-022-05181-4

Other articles of this Issue 1/2022

BMC Musculoskeletal Disorders 1/2022 Go to the issue