Abstract

To preserve residual hearing in cochlear implant candidates, the atraumatic insertion of the cochlea electrode has become a focus of cochlea implant research. In a previous study, intracochlear pressure changes during the opening of the round window membrane were investigated. In the current study, intracochlear pressure changes during opening of the round window membrane under dry and transfluid conditions were investigated. Round window openings were performed in an artificial cochlear model. Intracochlear pressure changes were measured using a micro-optical pressure sensor, which was placed in the apex. Openings of the round window membrane were performed under dry and wet conditions using a cannula and a diode laser. Statistically significant differences in the intracochlear pressure changes were seen between the different methods used for opening of the round window membrane. Lower pressure changes were seen by opening the round window membrane with the diode laser than with the cannula. A significant difference was seen between the dry and wet conditions. The atraumatic approach to the cochlea is assumed to be essential for the preservation of residual hearing. Opening of the round window under wet conditions produce a significant advantage on intracochlear pressure changes in comparison to dry conditions by limiting negative outward pressure.
Literature
1.
3.
go back to reference Lehnhardt E (1993) Intracochlear placement of cochlear implant electrodes in soft surgery technique. HNO 41(7):356–359PubMed Lehnhardt E (1993) Intracochlear placement of cochlear implant electrodes in soft surgery technique. HNO 41(7):356–359PubMed
4.
5.
go back to reference Fishman AJ, Moreno LE, Rivera A, Richter CP (2010) CO(2) laser fiber soft cochleostomy: development of a technique using human temporal bones and a guinea pig model. Lasers Surg Med 42(3):245–256. doi:10.1002/lsm.20902 CrossRefPubMed Fishman AJ, Moreno LE, Rivera A, Richter CP (2010) CO(2) laser fiber soft cochleostomy: development of a technique using human temporal bones and a guinea pig model. Lasers Surg Med 42(3):245–256. doi:10.​1002/​lsm.​20902 CrossRefPubMed
7.
go back to reference Todt I, Ernst A, Mittmann P (2016) Effects of round window opening size and moisturized electrodes on the intracochlear pressure related to the insertion of a cochlear implant electrode. Audiol Neurotol Extra 16(6):1–8. doi:10.1159/000442515 CrossRef Todt I, Ernst A, Mittmann P (2016) Effects of round window opening size and moisturized electrodes on the intracochlear pressure related to the insertion of a cochlear implant electrode. Audiol Neurotol Extra 16(6):1–8. doi:10.​1159/​000442515 CrossRef
9.
go back to reference Olson ES (1998) Observing middle and inner ear mechanics with novel intracochlear pressure sensors. J Acoust Soc Am 103(6):3445–3463CrossRefPubMed Olson ES (1998) Observing middle and inner ear mechanics with novel intracochlear pressure sensors. J Acoust Soc Am 103(6):3445–3463CrossRefPubMed
10.
go back to reference Adunka O, Unkelbach MH, Mack M, Hambek M, Gstoettner W, Kiefer J (2004) Cochlear implantation via the round window membrane minimizes trauma to cochlear structures: a histologically controlled insertion study. Acta Otolaryngol 124(7):807–812. doi:10.1080/00016480410018179 CrossRefPubMed Adunka O, Unkelbach MH, Mack M, Hambek M, Gstoettner W, Kiefer J (2004) Cochlear implantation via the round window membrane minimizes trauma to cochlear structures: a histologically controlled insertion study. Acta Otolaryngol 124(7):807–812. doi:10.​1080/​0001648041001817​9 CrossRefPubMed
12.
13.
go back to reference Briggs RJ, Tykocinski M, Xu J, Risi F, Svehla M, Cowan R, Stover T, Erfurt P, Lenarz T (2006) Comparison of round window and cochleostomy approaches with a prototype hearing preservation electrode. Audiol Neurootol 11(Suppl 1):42–48. doi:10.1159/000095613 CrossRefPubMed Briggs RJ, Tykocinski M, Xu J, Risi F, Svehla M, Cowan R, Stover T, Erfurt P, Lenarz T (2006) Comparison of round window and cochleostomy approaches with a prototype hearing preservation electrode. Audiol Neurootol 11(Suppl 1):42–48. doi:10.​1159/​000095613 CrossRefPubMed
14.
go back to reference Yoon YJ, Puria S, Steele CR (2006) Intracochlear pressure and organ of corti impedance from a linear active three-dimensional model. ORL J Otorhinolaryngol Relat Spec 68(6):365–372. doi:10.1159/000095279 CrossRefPubMed Yoon YJ, Puria S, Steele CR (2006) Intracochlear pressure and organ of corti impedance from a linear active three-dimensional model. ORL J Otorhinolaryngol Relat Spec 68(6):365–372. doi:10.​1159/​000095279 CrossRefPubMed
15.
go back to reference de la Rochefoucauld O, Decraemer WF, Khanna SM, Olson ES (2008) Simultaneous measurements of ossicular velocity and intracochlear pressure leading to the cochlear input impedance in gerbil. J Assoc Res Otolaryngol: JARO 9(2):161–177. doi:10.1007/s10162-008-0115-1 CrossRefPubMed de la Rochefoucauld O, Decraemer WF, Khanna SM, Olson ES (2008) Simultaneous measurements of ossicular velocity and intracochlear pressure leading to the cochlear input impedance in gerbil. J Assoc Res Otolaryngol: JARO 9(2):161–177. doi:10.​1007/​s10162-008-0115-1 CrossRefPubMed
17.
go back to reference Nedzelnitsky V (1980) Sound pressures in the basal turn of the cat cochlea. J Acoust Soc Am 68(6):1676–1689CrossRefPubMed Nedzelnitsky V (1980) Sound pressures in the basal turn of the cat cochlea. J Acoust Soc Am 68(6):1676–1689CrossRefPubMed
20.
go back to reference Graham JM, Ashcroft P (1999) Direct measurement of cerebrospinal fluid pressure through the cochlea in a congenitally deaf child with Mondini dysplasia undergoing cochlear implantation. Am J Otol 20(2):205–208PubMed Graham JM, Ashcroft P (1999) Direct measurement of cerebrospinal fluid pressure through the cochlea in a congenitally deaf child with Mondini dysplasia undergoing cochlear implantation. Am J Otol 20(2):205–208PubMed
21.
go back to reference Park JJ, Shen A, Keil S, Kraemer N, Westhofen M (2014) Radiological findings of the cochlear aqueduct in patients with Meniere’s disease using high-resolution CT and high-resolution MRI. Eur Arch Otorhinolaryngol 271(12):3325–3331. doi:10.1007/s00405-014-3199-8 CrossRefPubMed Park JJ, Shen A, Keil S, Kraemer N, Westhofen M (2014) Radiological findings of the cochlear aqueduct in patients with Meniere’s disease using high-resolution CT and high-resolution MRI. Eur Arch Otorhinolaryngol 271(12):3325–3331. doi:10.​1007/​s00405-014-3199-8 CrossRefPubMed
Metadata
Title
Optimisation of the round window opening in cochlear implant surgery in wet and dry conditions: impact on intracochlear pressure changes
Authors
Philipp Mittmann
A. Ernst
M. Mittmann
I. Todt
Publication date
01-11-2016
Publisher
Springer Berlin Heidelberg
Published in
European Archives of Oto-Rhino-Laryngology / Issue 11/2016
Print ISSN: 0937-4477
Electronic ISSN: 1434-4726
DOI
https://doi.org/10.1007/s00405-016-3993-6

Other articles of this Issue 11/2016

European Archives of Oto-Rhino-Laryngology 11/2016 Go to the issue