Published in: Journal of Medical Ultrasonics 2/2018

01-04-2018 | Original Article

Optimal surface property for three-dimensional echo visualization

Authors: Shunsuke Asakura, Saki Nishikawa, Kazumasa Orihashi, Yoshio Inoue

Published in: Journal of Medical Ultrasonics | Issue 2/2018

Login to get access

Abstract

Purpose

Three-dimensional (3D) echo is a useful modality to guide off-pump intracardiac surgery, and this study aimed to determine the optimal surface property for minimizing echo dropout and grating lobes.

Methods

A total of 13 specimens were examined. The specimens were scanned while they were rotated from 0° to 90°, and en face views were recorded in the 3D zoom mode. Changes in brightness were compared with the surface profile, which was traced from microscopic images. The brightness and size of grating lobes were also examined.

Results

Decreases in brightness that were observed at a large rotation angle were classified into three groups (sharp, moderate, and gradual), and variation in the effective reflective plane size appeared to be a major determinant of this property. The same material showed different degrees of reflection depending on the surface profile. The size of grating lobes was affected by variation in the reflective plane size, but not by the intensity of reflection. The brightness of the grating lobes was similar among specimens.

Conclusions

These results suggest that echo dropout and grating lobes can be effectively eliminated by an appropriate surface profile; thus, a likely practical solution would be to manufacture a surface that provides an adequate reflective surface size at various incident angles.
Literature
1.
go back to reference Suchon E, Pieculewicz M, Tracz W, et al. Transcatheter closure as an alternative and equivalent method to the surgical treatment of atrial septal defect in adults: comparison of early and late results. Med Sci Monit. 2009;15:612–7. Suchon E, Pieculewicz M, Tracz W, et al. Transcatheter closure as an alternative and equivalent method to the surgical treatment of atrial septal defect in adults: comparison of early and late results. Med Sci Monit. 2009;15:612–7.
2.
go back to reference Kijima Y, Akagi T, Nakagawa K, et al. Catheter closure of patent foramen ovale in patients with cryptogenic cerebrovascular accidents: initial experiences in Japan. Cardiovasc Interv Ther. 2014;29:11–7.CrossRefPubMed Kijima Y, Akagi T, Nakagawa K, et al. Catheter closure of patent foramen ovale in patients with cryptogenic cerebrovascular accidents: initial experiences in Japan. Cardiovasc Interv Ther. 2014;29:11–7.CrossRefPubMed
3.
go back to reference Magee AG, Huggon IC, Seed PT, et al. Transcatheter coil occlusion of the arterial duct; results of the European Registry. Eur Heart J. 2001;22:1817–21.CrossRefPubMed Magee AG, Huggon IC, Seed PT, et al. Transcatheter coil occlusion of the arterial duct; results of the European Registry. Eur Heart J. 2001;22:1817–21.CrossRefPubMed
4.
go back to reference Sakata Y, Sayed Y, Salinger MH, et al. Percutaneous balloon aortic valvuloplasty: antegrade transseptal vs conventional retrograde transarterial approach. Catheter Cardiovasc Interv. 2005;64:314–21.CrossRefPubMed Sakata Y, Sayed Y, Salinger MH, et al. Percutaneous balloon aortic valvuloplasty: antegrade transseptal vs conventional retrograde transarterial approach. Catheter Cardiovasc Interv. 2005;64:314–21.CrossRefPubMed
5.
go back to reference Palacios IF, Sanchez PL, Harrell LC, et al. Which patients benefit from percutaneous mitral balloon valvuloplasty? Prevalvuloplasty and postvalvuloplasty variables that predict long-term outcome. Circulation. 2002;105:1465–71.CrossRefPubMed Palacios IF, Sanchez PL, Harrell LC, et al. Which patients benefit from percutaneous mitral balloon valvuloplasty? Prevalvuloplasty and postvalvuloplasty variables that predict long-term outcome. Circulation. 2002;105:1465–71.CrossRefPubMed
6.
go back to reference Nishimura RA, Otto CM, Bonow RO, et al. 2014 AHA/ACC guideline for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol. 2014;63:e57–185.CrossRefPubMed Nishimura RA, Otto CM, Bonow RO, et al. 2014 AHA/ACC guideline for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol. 2014;63:e57–185.CrossRefPubMed
7.
go back to reference Smith CR, Leon MB, Mack MJ, PARTNER Trial Investigators, et al. Transcatheter versus surgical aortic-valve replacement in high-risk patients. N Engl J Med. 2011;364:2187–98.CrossRefPubMed Smith CR, Leon MB, Mack MJ, PARTNER Trial Investigators, et al. Transcatheter versus surgical aortic-valve replacement in high-risk patients. N Engl J Med. 2011;364:2187–98.CrossRefPubMed
8.
go back to reference Sukernik MR, Mets B, Kachulis B, et al. The impact of newly diagnosed patent foramen ovale in patients undergoing off-pump coronary artery bypass grafting: case series of eleven patients. Anesth Analg. 2002;95:1142–6.CrossRefPubMed Sukernik MR, Mets B, Kachulis B, et al. The impact of newly diagnosed patent foramen ovale in patients undergoing off-pump coronary artery bypass grafting: case series of eleven patients. Anesth Analg. 2002;95:1142–6.CrossRefPubMed
9.
go back to reference Akhter M, Lajos TZ. Pitfalls of undetected patent foramen ovale in off-pump cases. Ann Thorac Surg. 1999;67:546–8.CrossRefPubMed Akhter M, Lajos TZ. Pitfalls of undetected patent foramen ovale in off-pump cases. Ann Thorac Surg. 1999;67:546–8.CrossRefPubMed
10.
go back to reference Cui YC, Hao XH, Huang FJ, et al. Coronary artery bypass grafting in adults with congenital heart disease. J Card Surg. 2010;25:629–32.CrossRefPubMed Cui YC, Hao XH, Huang FJ, et al. Coronary artery bypass grafting in adults with congenital heart disease. J Card Surg. 2010;25:629–32.CrossRefPubMed
11.
go back to reference Hao XH, Huang FJ, Wang SY, et al. Off-pump coronary artery bypass grafting with device closure of atrial septal defect. Ann Thorac Surg. 2010;90:1727–9.CrossRefPubMed Hao XH, Huang FJ, Wang SY, et al. Off-pump coronary artery bypass grafting with device closure of atrial septal defect. Ann Thorac Surg. 2010;90:1727–9.CrossRefPubMed
12.
go back to reference Ueno Y, Iguchi Y, Inoue T, et al. Paradoxical brain embolism may not be uncommon-prospective study in acute ischemic stroke. J Neurol. 2007;254:763–6.CrossRefPubMed Ueno Y, Iguchi Y, Inoue T, et al. Paradoxical brain embolism may not be uncommon-prospective study in acute ischemic stroke. J Neurol. 2007;254:763–6.CrossRefPubMed
13.
go back to reference Lange SA, Schoen SP, Braun MU, et al. Perforation of aortic root as secondary complication after implantation of patent foramen ovale occlusion device in a 31-year-old woman. J Interv Cardiol. 2006;19:166–9.CrossRefPubMed Lange SA, Schoen SP, Braun MU, et al. Perforation of aortic root as secondary complication after implantation of patent foramen ovale occlusion device in a 31-year-old woman. J Interv Cardiol. 2006;19:166–9.CrossRefPubMed
14.
go back to reference Crawford GB, Brindis RG, Krucoff MW, et al. Percutaneous atrial septal occluder devices and cardiac erosion: a review of the literature. Catheter Cardiovasc Interv. 2012;80:157–67.CrossRefPubMed Crawford GB, Brindis RG, Krucoff MW, et al. Percutaneous atrial septal occluder devices and cardiac erosion: a review of the literature. Catheter Cardiovasc Interv. 2012;80:157–67.CrossRefPubMed
15.
go back to reference Suematsu Y, Takamoto S, Kaneko Y, et al. Beating atrial septal defect closure monitored by epicardial real-time three-dimensional echocardiography without cardiopulmonary bypass. Circulation. 2003;11:785–90.CrossRef Suematsu Y, Takamoto S, Kaneko Y, et al. Beating atrial septal defect closure monitored by epicardial real-time three-dimensional echocardiography without cardiopulmonary bypass. Circulation. 2003;11:785–90.CrossRef
16.
go back to reference Huang J, Triedman JK, Vasilyev NV, et al. Imaging artifacts of medical instruments in ultrasound-guided interventions. J Ultrasound Med. 2007;26:1303–22.CrossRefPubMed Huang J, Triedman JK, Vasilyev NV, et al. Imaging artifacts of medical instruments in ultrasound-guided interventions. J Ultrasound Med. 2007;26:1303–22.CrossRefPubMed
17.
go back to reference Orihashi K, Sueda T, Okada K, et al. Real-time three dimensional echo-guided closure of atrial septal defect: an experimental model. Interact Cardiovasc Thorac Surg. 2005;4:391–5.CrossRefPubMed Orihashi K, Sueda T, Okada K, et al. Real-time three dimensional echo-guided closure of atrial septal defect: an experimental model. Interact Cardiovasc Thorac Surg. 2005;4:391–5.CrossRefPubMed
18.
go back to reference Orihashi K, Takahashi S, Ozawa M, et al. Three-dimensional echo-guided suture of atrial septal defect with Maniceps in an experimental model. Hiroshima J Med Sci. 2010;59:57–63.PubMed Orihashi K, Takahashi S, Ozawa M, et al. Three-dimensional echo-guided suture of atrial septal defect with Maniceps in an experimental model. Hiroshima J Med Sci. 2010;59:57–63.PubMed
19.
go back to reference Jandzinski DI, Carson N, Davis D, et al. Treated needles: do they facilitate sonographically guided biopsies? J Ultrasound Med. 2003;22:1233–7.CrossRefPubMed Jandzinski DI, Carson N, Davis D, et al. Treated needles: do they facilitate sonographically guided biopsies? J Ultrasound Med. 2003;22:1233–7.CrossRefPubMed
20.
go back to reference Lochab J, Singh VR. Acoustic behaviour of plastics for medical applications. Ind J Pure Appl Phys. 2004;42:595–9. Lochab J, Singh VR. Acoustic behaviour of plastics for medical applications. Ind J Pure Appl Phys. 2004;42:595–9.
21.
go back to reference Garu PK, Chaki TK. Acoustic and mechanical properties of neoprene rubber for encapsulation of underwater transducers. Int J Sci Eng Technol. 2012;1:231–7. Garu PK, Chaki TK. Acoustic and mechanical properties of neoprene rubber for encapsulation of underwater transducers. Int J Sci Eng Technol. 2012;1:231–7.
Metadata
Title
Optimal surface property for three-dimensional echo visualization
Authors
Shunsuke Asakura
Saki Nishikawa
Kazumasa Orihashi
Yoshio Inoue
Publication date
01-04-2018
Publisher
Springer Japan
Published in
Journal of Medical Ultrasonics / Issue 2/2018
Print ISSN: 1346-4523
Electronic ISSN: 1613-2254
DOI
https://doi.org/10.1007/s10396-017-0831-4

Other articles of this Issue 2/2018

Journal of Medical Ultrasonics 2/2018 Go to the issue