01-08-2021 | ORIGINAL ARTICLE

Noise heterogeneity in attenuation-corrected cardiac SPECT images increases perfusion value uncertainty near the base of the heart

Authors: Sarah G. Cuddy-Walsh, PhD, R. Glenn Wells, PhD

Published in: Journal of Nuclear Cardiology | Issue 4/2021

Login to get access

Abstract

Background

Dedicated cardiac SPECT cameras which employ multi-pinhole detectors have variable photon sensitivity within the camera’s field-of-view such that a lower number of photon counts is typically detected from the base of the heart than from the apex. Consequently, the noise in a reconstructed image is expected to be higher at the base than at the apex of the heart.

Methods

Patient emission images were resampled to create statistical replicates which were reconstructed with and without attenuation correction. Noise images were computed using one standard deviation of the replicated images. These were evaluated for 93 patients with normal study results, each imaged with both a dual-headed parallel-hole camera and a multi-pinhole camera. Statistics for a normal database (NDB) of images from the 93 patients were also calculated.

Results

Image noise (1.7-fold) and NDB uncertainty (1.3-fold) increase significantly from the apex-to-the base of the heart in attenuation-corrected multi-pinhole SPECT images. The differences for non-attenuation-corrected images or those acquired with a parallel-hole camera were not significant.

Conclusions

For best interpretation of attenuation-corrected images acquired with multi-pinhole cameras, knowledge of NDB uncertainty gradients should be taken into consideration.
Appendix
Available only for authorised users
Literature
1.
go back to reference Dorbala S, Ananthasubramaniam K, Armstrong IS, et al. Single Photon Emission Computed Tomography (SPECT) myocardial perfusion imaging guidelines: Instrumentation, acquisition, processing, and interpretation. J Nucl Cardiol. 2018;25:1784.CrossRef Dorbala S, Ananthasubramaniam K, Armstrong IS, et al. Single Photon Emission Computed Tomography (SPECT) myocardial perfusion imaging guidelines: Instrumentation, acquisition, processing, and interpretation. J Nucl Cardiol. 2018;25:1784.CrossRef
2.
go back to reference Wells RG, Marvin B, Poirier M, Renaud JM, DeKemp RA, Ruddy TD. Optimization of SPECT measurement of myocardial blood flow with corrections for attenuation, motion, and blood-binding compared to PET. J Nucl Med. 2017;58:2013–9.CrossRef Wells RG, Marvin B, Poirier M, Renaud JM, DeKemp RA, Ruddy TD. Optimization of SPECT measurement of myocardial blood flow with corrections for attenuation, motion, and blood-binding compared to PET. J Nucl Med. 2017;58:2013–9.CrossRef
3.
go back to reference Kennedy JA, Israel O, Frenkel A. 3D iteratively reconstructed spatial resolution map and sensitivity characterization of a dedicated cardiac SPECT camera. J Nucl Cardiol. 2014;21:443–52.CrossRef Kennedy JA, Israel O, Frenkel A. 3D iteratively reconstructed spatial resolution map and sensitivity characterization of a dedicated cardiac SPECT camera. J Nucl Cardiol. 2014;21:443–52.CrossRef
4.
go back to reference Cuddy-Walsh SG, Wells RG. Patient-specific estimation of spatially variant image noise for a pinhole cardiac SPECT camera. Med Phys. 2018;45:2033–47.CrossRef Cuddy-Walsh SG, Wells RG. Patient-specific estimation of spatially variant image noise for a pinhole cardiac SPECT camera. Med Phys. 2018;45:2033–47.CrossRef
5.
go back to reference White JA, Lawson RS. A Poisson resampling method for simulating reduced counts in nuclear medicine images. Phys Med Biol. 2015;60:N167–76.CrossRef White JA, Lawson RS. A Poisson resampling method for simulating reduced counts in nuclear medicine images. Phys Med Biol. 2015;60:N167–76.CrossRef
6.
go back to reference Buvat I. A non-parametric bootstrap approach for analysing the statistical properties of SPECT and PET images. Phys Med Biol. 2002;47:1761–75.CrossRef Buvat I. A non-parametric bootstrap approach for analysing the statistical properties of SPECT and PET images. Phys Med Biol. 2002;47:1761–75.CrossRef
7.
go back to reference Qi J. A unified noise analysis for iterative image estimation. Phys Med Biol. 2003;48:3505–19.CrossRef Qi J. A unified noise analysis for iterative image estimation. Phys Med Biol. 2003;48:3505–19.CrossRef
8.
go back to reference Barrett HH, Wilson DW, Tsui BM. Noise properties of the EM algorithm: I. Theory. Phys Med Biol. 1994;39:833–46.CrossRef Barrett HH, Wilson DW, Tsui BM. Noise properties of the EM algorithm: I. Theory. Phys Med Biol. 1994;39:833–46.CrossRef
9.
go back to reference Rahmim A, Tang J. Noise propagation in resolution modeled PET imaging and its impact on detectability. Phys Med Biol. 2013;58:6945–68.CrossRef Rahmim A, Tang J. Noise propagation in resolution modeled PET imaging and its impact on detectability. Phys Med Biol. 2013;58:6945–68.CrossRef
11.
go back to reference Pourmoghaddas A, Vanderwerf K, Ruddy TD, Glenn Wells R, Wells RG. Scatter correction improves concordance in SPECT MPI with a dedicated cardiac SPECT solid-state camera. J Nucl Cardiol. 2015;22:334–43.CrossRef Pourmoghaddas A, Vanderwerf K, Ruddy TD, Glenn Wells R, Wells RG. Scatter correction improves concordance in SPECT MPI with a dedicated cardiac SPECT solid-state camera. J Nucl Cardiol. 2015;22:334–43.CrossRef
12.
go back to reference Abdi H. The Bonferonni and Šidák corrections for multiple comparisons. In: Salkind N, editor. Encyclopedia of measurement and statistics. Thousand Oaks: Sage; 2007. Abdi H. The Bonferonni and Šidák corrections for multiple comparisons. In: Salkind N, editor. Encyclopedia of measurement and statistics. Thousand Oaks: Sage; 2007.
13.
go back to reference Grossman GB, Garcia EV, Bateman TM, et al. Quantitative Tc-99m sestamibi attenuation-corrected SPECT: Development and multicenter trial validation of myocardial perfusion stress gender-independent normal database in an obese population. J Nucl Cardiol. 2004;11:263–72.CrossRef Grossman GB, Garcia EV, Bateman TM, et al. Quantitative Tc-99m sestamibi attenuation-corrected SPECT: Development and multicenter trial validation of myocardial perfusion stress gender-independent normal database in an obese population. J Nucl Cardiol. 2004;11:263–72.CrossRef
14.
go back to reference Dvorak RA, Brown RKJ, Corbett JR. Interpretation of SPECT/CT myocardial perfusion images: Common artifacts and quality control techniques. Radiographics. 2011;31:2041–57.CrossRef Dvorak RA, Brown RKJ, Corbett JR. Interpretation of SPECT/CT myocardial perfusion images: Common artifacts and quality control techniques. Radiographics. 2011;31:2041–57.CrossRef
15.
go back to reference Timmins R, Ruddy TD, Wells RG. Patient position alters attenuation effects in multipinhole cardiac SPECT. Med Phys. 2015;42:1233.CrossRef Timmins R, Ruddy TD, Wells RG. Patient position alters attenuation effects in multipinhole cardiac SPECT. Med Phys. 2015;42:1233.CrossRef
16.
go back to reference Hindorf C, Oddstig J, Hedeer F, Hansson MJ, Jögi J, Engblom H. Importance of correct patient positioning in myocardial perfusion SPECT when using a CZT camera. J Nucl Cardiol. 2014;21:695–702.CrossRef Hindorf C, Oddstig J, Hedeer F, Hansson MJ, Jögi J, Engblom H. Importance of correct patient positioning in myocardial perfusion SPECT when using a CZT camera. J Nucl Cardiol. 2014;21:695–702.CrossRef
17.
go back to reference Gimelli A, Bottai M, Quaranta A, Giorgetti A, Genovesi D, Marzullo P. Gender differences in the evaluation of coronary artery disease with a cadmium-zinc telluride camera. Eur J Nucl Med Mol Imaging. 2013;40:1542–8.CrossRef Gimelli A, Bottai M, Quaranta A, Giorgetti A, Genovesi D, Marzullo P. Gender differences in the evaluation of coronary artery disease with a cadmium-zinc telluride camera. Eur J Nucl Med Mol Imaging. 2013;40:1542–8.CrossRef
Metadata
Title
Noise heterogeneity in attenuation-corrected cardiac SPECT images increases perfusion value uncertainty near the base of the heart
Authors
Sarah G. Cuddy-Walsh, PhD
R. Glenn Wells, PhD
Publication date
01-08-2021
Publisher
Springer International Publishing
Published in
Journal of Nuclear Cardiology / Issue 4/2021
Print ISSN: 1071-3581
Electronic ISSN: 1532-6551
DOI
https://doi.org/10.1007/s12350-019-01821-w

Other articles of this Issue 4/2021

Journal of Nuclear Cardiology 4/2021 Go to the issue