Abstract

Background

To analyse the changes in surface and nickel ion release characteristics of fractured root canal shaping instruments in a simulated body fluid environment.

Methods

A total of 54 new instruments were studied. The instrument groups consisted of five different NiTi alloys and a stainless-steel alloy. To standardize instrument fracture, a torsional type of failure was created on each instrument. The fractured specimens of each instrument group were randomly divided into three static immersion subgroups of 1 h, 7-day, and 30-day (n = 3). Simulated body fluid (SBF) was prepared to mimic human blood plasma by Kokubo&Takadama protocol for ex situ static immersions at 37ºC. The surfaces were examined via scanning electron microscopy coupled with energy-dispersive X-ray spectroscopy. To determine the quantitative ion release, the retrieved SBFs were analyzed using inductively coupled plasma mass spectrometry. Two-way ANOVA and Tukey post hoc tests sought the statistical significance of the nickel ion values(p < 0.05).

Results

In 1 h of immersion, the newly formed structures, exhibiting mostly oxygen signals, were widespread and evident on NiTi surfaces. In contrast, fewer structures were detected on the SS surface in that subgroup. In 7 days of immersion, a tendency for a decrease in the density of the new structures was revealed in NiTi groups. The oxygen signals on NiTi group surfaces significantly increased, contrary to their decrease in SS. Signals of sodium, chlorine, and calcium were detected, indicating salt precipitates in groups. In 30 days of immersion, salt precipitates continued to form. The Ni-ion release values in all instrument groups presented significant differences in comparison to the SBF control in all immersion periods(p < 0.001). No significant differences were observed in immersion time periods or instrument groups(p > 0.05).

Conclusions

Within the limitations of the presented study, it was concluded that the fractured SS and NiTi root canal instruments release Ni ions in contact with body fluid. However, the Ni ion release values determined during the observation periods are lower than the critical toxic or allergic thresholds defined for the human body. This was due to the ionic dissolution cycle reaching a stable state from 1-hour to 30-day exposure to the body fluid of fractured instruments.
Literature
1.
go back to reference Siqueira JF Jr, Pérez AR, Marceliano-Alves MF, Provenzano JC, Silva SG, Pires FR, Vieira GCS, Rôças IN, Alves FRF. What happens to unprepared root canal walls: a correlative analysis using micro-computed tomography and histology/scanning electron microscopy. Int Endod J. 2018;51:501–8.PubMed Siqueira JF Jr, Pérez AR, Marceliano-Alves MF, Provenzano JC, Silva SG, Pires FR, Vieira GCS, Rôças IN, Alves FRF. What happens to unprepared root canal walls: a correlative analysis using micro-computed tomography and histology/scanning electron microscopy. Int Endod J. 2018;51:501–8.PubMed
2.
go back to reference Liu H, Shen Y, Wang Z, Haapasalo M. The ability of different irrigation methods to remove mixtures of calcium hydroxide and barium sulphate from isthmuses in 3D printed transparent root canal models. Odontology. 2022;110:27–34.PubMed Liu H, Shen Y, Wang Z, Haapasalo M. The ability of different irrigation methods to remove mixtures of calcium hydroxide and barium sulphate from isthmuses in 3D printed transparent root canal models. Odontology. 2022;110:27–34.PubMed
3.
go back to reference Li Y, Wang Z, Bao P, Meng T, Liu M, Li H, Shen Y, Liu D, Jia Z, Liu H. Cleaning and disinfecting oval-shaped root canals: ex vivo evaluation of three rotary instrumentation systems with passive ultrasonic irrigation. Medicina. 2023;59:962.PubMedPubMedCentral Li Y, Wang Z, Bao P, Meng T, Liu M, Li H, Shen Y, Liu D, Jia Z, Liu H. Cleaning and disinfecting oval-shaped root canals: ex vivo evaluation of three rotary instrumentation systems with passive ultrasonic irrigation. Medicina. 2023;59:962.PubMedPubMedCentral
4.
go back to reference Lui K, Liu H, Wang H, Yang X, Huang D, Zhou X, Gao Y, Shen Y. An application framework of 3D assessment image registration accuracy and untouched surface area in canal instrumentation laboratory research with micro-computed tomography. Clin Oral Investig. 2023;27:715–25.PubMed Lui K, Liu H, Wang H, Yang X, Huang D, Zhou X, Gao Y, Shen Y. An application framework of 3D assessment image registration accuracy and untouched surface area in canal instrumentation laboratory research with micro-computed tomography. Clin Oral Investig. 2023;27:715–25.PubMed
5.
go back to reference Ertuğrul İF, Orhan EO. Cyclic fatigue and energy-dispersive X-ray spectroscopy examination of the novel ROTATE instrument. Microsc Res Tech. 2019;82:2042–8.PubMed Ertuğrul İF, Orhan EO. Cyclic fatigue and energy-dispersive X-ray spectroscopy examination of the novel ROTATE instrument. Microsc Res Tech. 2019;82:2042–8.PubMed
6.
go back to reference Zupanc J, Vahdat-Pajouh N, Schäfer E. New thermomechanically treated NiTi alloys - a review. Int Endod J. 2018;51:1088–103.PubMed Zupanc J, Vahdat-Pajouh N, Schäfer E. New thermomechanically treated NiTi alloys - a review. Int Endod J. 2018;51:1088–103.PubMed
7.
go back to reference Elnaghy AM, Elsaka SE, Mandorah AO. In vitro comparison of cyclic fatigue resistance of TruNatomy in single and double curvature canals compared with different nickel-titanium rotary instruments. BMC Oral Health. 2020;20:38.PubMedPubMedCentral Elnaghy AM, Elsaka SE, Mandorah AO. In vitro comparison of cyclic fatigue resistance of TruNatomy in single and double curvature canals compared with different nickel-titanium rotary instruments. BMC Oral Health. 2020;20:38.PubMedPubMedCentral
8.
go back to reference Gouédard C, Pino L, Arbab-Chirani R, Arbab-Chirani S, Chevalier V. Comparison of the cyclic fatigue resistance of one curve, F6 skytaper, Protaper Next, and Hyflex CM endodontic files. Restor Dent Endod. 2022;47:e16.PubMedPubMedCentral Gouédard C, Pino L, Arbab-Chirani R, Arbab-Chirani S, Chevalier V. Comparison of the cyclic fatigue resistance of one curve, F6 skytaper, Protaper Next, and Hyflex CM endodontic files. Restor Dent Endod. 2022;47:e16.PubMedPubMedCentral
9.
go back to reference Wang H, Yang X, Zou L, Huang D, Zhou X, Xu J, Gao Y. Shaping outcome of ProTaper NEXT for root canal preparation in mandibular incisors: a micro-CT study. BMC Oral Health. 2022;22:302.PubMedPubMedCentral Wang H, Yang X, Zou L, Huang D, Zhou X, Xu J, Gao Y. Shaping outcome of ProTaper NEXT for root canal preparation in mandibular incisors: a micro-CT study. BMC Oral Health. 2022;22:302.PubMedPubMedCentral
10.
go back to reference Orhan EO, Ertuğrul İF, Tosun S. Monotonic torsional resistance and fatigue resistance of novel SCOPE RS instruments. Microsc Res Tech. 2021;84:62–70.PubMed Orhan EO, Ertuğrul İF, Tosun S. Monotonic torsional resistance and fatigue resistance of novel SCOPE RS instruments. Microsc Res Tech. 2021;84:62–70.PubMed
11.
go back to reference Ruddle CJ. Nonsurgical Endodontic Retreatment. In: Cohen S, Burns RC, editors. Pathways of the Pulp, 8th edition, St. Louis, MO, USA: Mosby;2002. p. 875–930. Ruddle CJ. Nonsurgical Endodontic Retreatment. In: Cohen S, Burns RC, editors. Pathways of the Pulp, 8th edition, St. Louis, MO, USA: Mosby;2002. p. 875–930.
12.
go back to reference Walton RE, Ardjmand K. Histological evaluation of the presence of bacteria in induced periapical lesions in monkeys. J Endod. 1992;18:216–27.PubMed Walton RE, Ardjmand K. Histological evaluation of the presence of bacteria in induced periapical lesions in monkeys. J Endod. 1992;18:216–27.PubMed
13.
go back to reference Siqueira JF Jr, Lopes HP. Bacteria on the apical root surfaces of untreated teeth with periradicular lesions: a scanning electron microscopy study. Int Endod J. 2001;34:216–20.PubMed Siqueira JF Jr, Lopes HP. Bacteria on the apical root surfaces of untreated teeth with periradicular lesions: a scanning electron microscopy study. Int Endod J. 2001;34:216–20.PubMed
14.
go back to reference Gorni FG, Gagliani MM. The outcome of endodontic retreatment: a 2-yr follow-up. J Endod. 2004;30:1–4.PubMed Gorni FG, Gagliani MM. The outcome of endodontic retreatment: a 2-yr follow-up. J Endod. 2004;30:1–4.PubMed
15.
go back to reference Madarati AA, Hunter MJ, Dummer PM. Management of intracanal separated instruments. J Endod. 2013;39:569–81.PubMed Madarati AA, Hunter MJ, Dummer PM. Management of intracanal separated instruments. J Endod. 2013;39:569–81.PubMed
16.
go back to reference McGuigan MB, Louca C, Duncan HF. Clinical decision-making after endodontic instrument fracture. Br Dent J. 2012;214:395–400. McGuigan MB, Louca C, Duncan HF. Clinical decision-making after endodontic instrument fracture. Br Dent J. 2012;214:395–400.
17.
go back to reference Solomonov M, Webber M, Keinan D. Fractured endodontic instrument: a clinical dilemma. Retrieve, bypass or entomb? N Y State Dent J. 2014;80:50–2.PubMed Solomonov M, Webber M, Keinan D. Fractured endodontic instrument: a clinical dilemma. Retrieve, bypass or entomb? N Y State Dent J. 2014;80:50–2.PubMed
20.
go back to reference Bekmurzayeva A, Duncanson WJ, Azevedo HS, Kanayeva D. Surface modification stainless steel for biomedical applications: revisiting a century-old material. Mater Sci Eng C. 2018;93:1073–89. Bekmurzayeva A, Duncanson WJ, Azevedo HS, Kanayeva D. Surface modification stainless steel for biomedical applications: revisiting a century-old material. Mater Sci Eng C. 2018;93:1073–89.
21.
go back to reference Hanawa T. Metal ion release from metal implants. Mater Sci Eng C. 2004;24:745–52. Hanawa T. Metal ion release from metal implants. Mater Sci Eng C. 2004;24:745–52.
22.
go back to reference Toker SM, Canadinc D, Maier HJ, Birer O. Evaluation of passive oxide layer formation-biocompatibility relationship in NiTi shape memory alloys: geometry and body location dependency. Mater Sci Eng C. 2014;36:118–29. Toker SM, Canadinc D, Maier HJ, Birer O. Evaluation of passive oxide layer formation-biocompatibility relationship in NiTi shape memory alloys: geometry and body location dependency. Mater Sci Eng C. 2014;36:118–29.
23.
go back to reference Toker SM, Canadinc D. Evaluation of the biocompatibility of NiTi dental wires: a comparison of laboratory experiments and clinical conditions. Mater Sci Eng C. 2014;40:142–7. Toker SM, Canadinc D. Evaluation of the biocompatibility of NiTi dental wires: a comparison of laboratory experiments and clinical conditions. Mater Sci Eng C. 2014;40:142–7.
24.
go back to reference Uzer B, Gumus B, Toker SM, Sahbazoglu D, Saher D, Yildirim C. A critical approach to the biocompatibility testing of Niti orthodontic archwires. Int J Metall Met Phys. 2016;1:1–7. Uzer B, Gumus B, Toker SM, Sahbazoglu D, Saher D, Yildirim C. A critical approach to the biocompatibility testing of Niti orthodontic archwires. Int J Metall Met Phys. 2016;1:1–7.
25.
go back to reference Bhatia S, Nagendrababu V, Peters OA, Fawzy A, Daood U. Evaluation of usage-induced degradation of different endodontic file systems. Sci Rep. 2021;11:9027.PubMedPubMedCentral Bhatia S, Nagendrababu V, Peters OA, Fawzy A, Daood U. Evaluation of usage-induced degradation of different endodontic file systems. Sci Rep. 2021;11:9027.PubMedPubMedCentral
26.
go back to reference Kovac V, Poljsak B, Bergant M, Scancar J, Mezeg U, Primozic J. Differences in metal ions released from orthodontic appliances in an in vitro and in vivo setting. Coat. 2022;12:190. Kovac V, Poljsak B, Bergant M, Scancar J, Mezeg U, Primozic J. Differences in metal ions released from orthodontic appliances in an in vitro and in vivo setting. Coat. 2022;12:190.
27.
go back to reference Nagendrababu V, Murray PE, Ordinola-Zapata R, Peters OA, Rôças IN, Siqueira JF Jr, Priya E, Jayaraman JJ, Pulikkotil S, Camilleri J, Boutsioukis C, Rossi-Fedele G, Dummer P. PRILE 2021 guidelines for reporting laboratory studies in endodontology: a consensus-based development. Int Endod J. 2021;54:1482–90.PubMed Nagendrababu V, Murray PE, Ordinola-Zapata R, Peters OA, Rôças IN, Siqueira JF Jr, Priya E, Jayaraman JJ, Pulikkotil S, Camilleri J, Boutsioukis C, Rossi-Fedele G, Dummer P. PRILE 2021 guidelines for reporting laboratory studies in endodontology: a consensus-based development. Int Endod J. 2021;54:1482–90.PubMed
28.
go back to reference International Organization for Standardization. ISO 3630 part 1:2008. Dentistry—Root-canal Instruments—Part 1: General requirements and test methods. Accessed 21 March 2023. International Organization for Standardization. ISO 3630 part 1:2008. Dentistry—Root-canal Instruments—Part 1: General requirements and test methods. Accessed 21 March 2023.
29.
go back to reference Kokubo T, Takadama H. How useful is SBF in predicting in vivo bone bioactivity? Biomater. 2006;27:2907–15. Kokubo T, Takadama H. How useful is SBF in predicting in vivo bone bioactivity? Biomater. 2006;27:2907–15.
30.
go back to reference Thompson SA. An overview of nickel-titanium alloys used in dentistry. Int Endod J. 2000;33:297–310.PubMed Thompson SA. An overview of nickel-titanium alloys used in dentistry. Int Endod J. 2000;33:297–310.PubMed
31.
go back to reference Marsicovetere ES, Clement DJ, del Rio CE. Morphometric video analysis of the engine-driven nickel-titanium Lightspeed instrument system. J Endod. 1996;22:231–5.PubMed Marsicovetere ES, Clement DJ, del Rio CE. Morphometric video analysis of the engine-driven nickel-titanium Lightspeed instrument system. J Endod. 1996;22:231–5.PubMed
32.
go back to reference Eggert C, Peters O, Barbakow F. Wear of nickel-titanium lightspeed instruments evaluated by scanning electron microscopy. J Endod. 1999;25:494–7.PubMed Eggert C, Peters O, Barbakow F. Wear of nickel-titanium lightspeed instruments evaluated by scanning electron microscopy. J Endod. 1999;25:494–7.PubMed
33.
go back to reference Kuhn G, Tavernier B, Jordan L. Influence of structure on nickel-titanium endodontic instruments failure. J Endod. 2001;27:516–20.PubMed Kuhn G, Tavernier B, Jordan L. Influence of structure on nickel-titanium endodontic instruments failure. J Endod. 2001;27:516–20.PubMed
34.
go back to reference Tripi TR, Bonaccorso A, Tripi V, Condorelli GG, Rapisarda E. Defects in GT rotary instruments after use: an SEM study. J Endod. 2001;27:782–5.PubMed Tripi TR, Bonaccorso A, Tripi V, Condorelli GG, Rapisarda E. Defects in GT rotary instruments after use: an SEM study. J Endod. 2001;27:782–5.PubMed
35.
go back to reference Martins RC, Bahia MG, Buono VT. Surface analysis of ProFile instruments by scanning electron microscopy and X-ray energy-dispersive spectroscopy: a preliminary study. Int Endod J. 2002;35:848–53.PubMed Martins RC, Bahia MG, Buono VT. Surface analysis of ProFile instruments by scanning electron microscopy and X-ray energy-dispersive spectroscopy: a preliminary study. Int Endod J. 2002;35:848–53.PubMed
36.
go back to reference Chianello G, Specian VL, Hardt LC, Raldi DP, Lage-Marques JL, Habitante SM. Surface finishing of unused rotary endodontic instruments: a SEM study. Braz Dent J. 2008;19:109–13.PubMed Chianello G, Specian VL, Hardt LC, Raldi DP, Lage-Marques JL, Habitante SM. Surface finishing of unused rotary endodontic instruments: a SEM study. Braz Dent J. 2008;19:109–13.PubMed
37.
go back to reference Toker SM, Gerstein G, Maier HJ, Canadinc D. Effects of microstructural mechanisms on the localized oxidation behavior of NiTi shape memory alloys in simulated body fluid. J Mater Sci. 2018;53:948–58. Toker SM, Gerstein G, Maier HJ, Canadinc D. Effects of microstructural mechanisms on the localized oxidation behavior of NiTi shape memory alloys in simulated body fluid. J Mater Sci. 2018;53:948–58.
39.
go back to reference Wataha JC, O’Dell NL, Singh BB, Ghazi M, Whitford GM, Lockwood PE. Relating nickel-induced tissue inflammation to nickel release in vivo. J Biomed Mater Res. 2001;58:537–44.PubMed Wataha JC, O’Dell NL, Singh BB, Ghazi M, Whitford GM, Lockwood PE. Relating nickel-induced tissue inflammation to nickel release in vivo. J Biomed Mater Res. 2001;58:537–44.PubMed
40.
go back to reference Badhe RV, Akinfosile O, Bijukumar D, Barba M, Mathew MT. Systemic toxicity eliciting metal ion levels from metallic implants and orthopedic devices - a mini review. Toxicol Lett. 2021;350:213–24.PubMed Badhe RV, Akinfosile O, Bijukumar D, Barba M, Mathew MT. Systemic toxicity eliciting metal ion levels from metallic implants and orthopedic devices - a mini review. Toxicol Lett. 2021;350:213–24.PubMed
41.
go back to reference Huang HH, Chiu YH, Lee TH, Wu SC, Yang HW, Su KH, Hsu CC. Ion release from NiTi orthodontic wires in artificial saliva with various acidities. Biomater. 2003;24:3585–92. Huang HH, Chiu YH, Lee TH, Wu SC, Yang HW, Su KH, Hsu CC. Ion release from NiTi orthodontic wires in artificial saliva with various acidities. Biomater. 2003;24:3585–92.
42.
go back to reference Trowbridge HO, Emling RC. Inflammation: a review of the process. 5th ed. Quintessence Publishing Co. Inc.; 2019. Trowbridge HO, Emling RC. Inflammation: a review of the process. 5th ed. Quintessence Publishing Co. Inc.; 2019.
43.
go back to reference Adorno CG, Yoshioka T, Suda H. Crack initiation on the apical root surface caused by three different nickel-titanium rotary files at different working lengths. J Endod. 2011;37:522–5.PubMed Adorno CG, Yoshioka T, Suda H. Crack initiation on the apical root surface caused by three different nickel-titanium rotary files at different working lengths. J Endod. 2011;37:522–5.PubMed
44.
go back to reference Ha JH, Park SS. Influence of glide path on the screw-in effect and torque of nickel-titanium rotary files in simulated resin root canals. Restor Dent Endod. 2012;37:215–9.PubMedPubMedCentral Ha JH, Park SS. Influence of glide path on the screw-in effect and torque of nickel-titanium rotary files in simulated resin root canals. Restor Dent Endod. 2012;37:215–9.PubMedPubMedCentral
45.
go back to reference Ha JH, Lee CJ, Kwak SW, El Abed R, Ha D, Kim HC. Geometric optimization for development of glide path preparation nickel-titanium rotary instrument. J Endod. 2015;41:916–9.PubMed Ha JH, Lee CJ, Kwak SW, El Abed R, Ha D, Kim HC. Geometric optimization for development of glide path preparation nickel-titanium rotary instrument. J Endod. 2015;41:916–9.PubMed
46.
go back to reference Spagnuolo G, Ametrano G, D’Antò V, Rengo C, Simeone M, Riccitiello F, Amato M. Effect of autoclaving on the surfaces of TiN -coated and conventional nickel-titanium rotary instruments. Int Endod J. 2012;45:1148–55.PubMed Spagnuolo G, Ametrano G, D’Antò V, Rengo C, Simeone M, Riccitiello F, Amato M. Effect of autoclaving on the surfaces of TiN -coated and conventional nickel-titanium rotary instruments. Int Endod J. 2012;45:1148–55.PubMed
47.
go back to reference Kwak SW, Shen Y, Liu H, Kim H-C, Haapasalo M. Torque Generation of the Endodontic Instruments: a narrative review. Materials. 2022;15(2):664.PubMedPubMedCentral Kwak SW, Shen Y, Liu H, Kim H-C, Haapasalo M. Torque Generation of the Endodontic Instruments: a narrative review. Materials. 2022;15(2):664.PubMedPubMedCentral
Metadata
Title
Nickel ion release and surface analyses on instrument fragments fractured beyond the apex: a laboratory investigation
Authors
Sıdıka Mine Toker
Ekim Onur Orhan
Arzu Beklen
Publication date
01-12-2023
Publisher
BioMed Central
Published in
BMC Oral Health / Issue 1/2023
Electronic ISSN: 1472-6831
DOI
https://doi.org/10.1186/s12903-023-03434-9

Other articles of this Issue 1/2023

BMC Oral Health 1/2023 Go to the issue