New implant designs for fracture fixation in osteoporotic bone

Authors: J. Goldhahn, J. Seebeck, R. Frei, B. Frenz, I. Antoniadis, E. Schneider

Published in: Osteoporosis International | Special Issue 2/2005

Login to get access

Abstract

Screws are one of the limiting factors for fixation of implants, particularly in poor bone quality. A class of new implants with an implant–bone–interface optimized regarding load transition by increasing the peripheral area might improve the anchorage of implants in osteoporotic bone. However, the shape of these implants requires new technologies for insertion. The goal of the work presented here was to analyze the relevant parameters regarding implant geometry and to demonstrate the effect of new procedures for their insertion. The investigation was divided into three parts: 1) implant design optimisation, 2) efficiency of cortical bone ablation, and 3) implant insertion technology. Finite element analysis (FEA) was performed to investigate the influence of the number of lobes, the radius of the outer curvature and additional milling to remove any sharp changes of section around the lobe. Opening of the cortical bone with an Er:YAG laser was studied using calf cortex from 2 to 7 mm thickness. The effect of a) pulse energy and pulse duration, b) cortical thickness, c) wet or dry boundary conditions on volume and geometry of ablated bone, time required to penetrate the cortical bone and local bone tissue damage was quantified. Pneumatic and ultrasound based insertion were compared in the third experiment. The cortical bone was prepared in the following ways: a) no opening, b) predrilling of three holes (1 mm diameter each) and c) exact pre-cutting of the whole contour. Increasing the radius of the outer curvature from 2 to 5 mm reduces the peak stresses during loading in all planes in the implant as well as in the adjacent cortical bone by about 30–40%. An increase in the number of lobes from two to three decreases the mean peak stress by about 46% (α<0.001) and the range between the minimal and maximal peak stresses for different loading directions by about 83%. Penetration of cortical bone with an Er:YAG laser was possible up to a cortical thickness of 6 mm with fewer than 100 pulses. The ablation rate per pulse increased more with increasing duration than with increasing energy. Signs of bone damage such as melting were only visible when high pulse energies and durations were used. Insertion of the prototype was possible with all devices, but only when the whole contour was cut out of the cortical bone. However, the use of the ultrasound vibrator led to heating up of the tissue fluid and subsequently to water evaporation and tissue damage. Insertion of the prototype was possible with both pneumatic vibrators, but only when the whole contour was cut out of the cortical bone. New implant designs may lead to reduced stress peaks in the surrounding bone and might be inserted with the help of new insertion technologies, namely laser cutting of cortical bone and pneumatic vibration. Further studies are required to optimize these technologies prior to clinical use.
Literature
1.
go back to reference Cordey J, Borgeaud M, Perren SM (2000) Force transfer between the plate and the bone: relative importance of the bending stiffness of the screws friction between plate and bone. Injury 31:C21–28CrossRefPubMed Cordey J, Borgeaud M, Perren SM (2000) Force transfer between the plate and the bone: relative importance of the bending stiffness of the screws friction between plate and bone. Injury 31:C21–28CrossRefPubMed
2.
go back to reference Chapman JR, Harrington RM, Lee KM, Anderson PA, Tencer AF, Kowalski D (1996) Factors affecting the pullout strength of cancellous bone screws. J Biomech Eng 118:391–398PubMed Chapman JR, Harrington RM, Lee KM, Anderson PA, Tencer AF, Kowalski D (1996) Factors affecting the pullout strength of cancellous bone screws. J Biomech Eng 118:391–398PubMed
3.
go back to reference DeCoster TA, Heetderks DB, Downey DJ, Ferries JS, Jones W (1990) Optimizing bone screw pullout force. J Orthop Trauma 4:169–174PubMed DeCoster TA, Heetderks DB, Downey DJ, Ferries JS, Jones W (1990) Optimizing bone screw pullout force. J Orthop Trauma 4:169–174PubMed
4.
go back to reference Thompson JD, Benjamin JB, Szivek JA (1997) Pullout strengths of cannulated and noncannulated cancellous bone screws. Clin Orthop 341:241–249CrossRefPubMed Thompson JD, Benjamin JB, Szivek JA (1997) Pullout strengths of cannulated and noncannulated cancellous bone screws. Clin Orthop 341:241–249CrossRefPubMed
5.
go back to reference Westmoreland GL, McLaurin TM, Hutton WC (2002) Screw pullout strength: a biomechanical comparison of large-fragment and small-fragment fixation in the tibial plateau. J Orthop Trauma 16:178–181CrossRefPubMed Westmoreland GL, McLaurin TM, Hutton WC (2002) Screw pullout strength: a biomechanical comparison of large-fragment and small-fragment fixation in the tibial plateau. J Orthop Trauma 16:178–181CrossRefPubMed
6.
go back to reference Murphy TP, Hill CM, Kapatkin AS, Radin A, Shofer FS, Smith GK (2001) Pullout properties of 3.5-mm AO/ASIF self-tapping and cortex screws in a uniform synthetic material and in canine bone. Vet Surg 30:253–260CrossRefPubMed Murphy TP, Hill CM, Kapatkin AS, Radin A, Shofer FS, Smith GK (2001) Pullout properties of 3.5-mm AO/ASIF self-tapping and cortex screws in a uniform synthetic material and in canine bone. Vet Surg 30:253–260CrossRefPubMed
7.
go back to reference Saka B (2000) Mechanical and biomechanical measurements of five currently available osteosynthesis systems of self-tapping screws. Br J Oral Maxillofac Surg 38:70–75CrossRefPubMed Saka B (2000) Mechanical and biomechanical measurements of five currently available osteosynthesis systems of self-tapping screws. Br J Oral Maxillofac Surg 38:70–75CrossRefPubMed
8.
go back to reference Colgan SA, Hecker AT, Kirker-Head CA, Hayes WC (1998) A comparison of the Synthes 4.5-mm cannulated screw and the Synthes 4.5-mm standard cortex screw systems in equine bone. Vet Surg 27:540–546PubMed Colgan SA, Hecker AT, Kirker-Head CA, Hayes WC (1998) A comparison of the Synthes 4.5-mm cannulated screw and the Synthes 4.5-mm standard cortex screw systems in equine bone. Vet Surg 27:540–546PubMed
9.
go back to reference Gausepohl T, Mohring R, Pennig D, Koebke J (2001) Fine thread versus coarse thread. A comparison of the maximum holding power. Injury 32:SD1–7 Gausepohl T, Mohring R, Pennig D, Koebke J (2001) Fine thread versus coarse thread. A comparison of the maximum holding power. Injury 32:SD1–7
10.
go back to reference Asnis SE, Ernberg JJ, Bostrom MP, Wright TM, Harrington RM, Tencer A, Peterson M (1996) Cancellous bone screw thread design and holding power. J Orthop Trauma 10:462–469CrossRefPubMed Asnis SE, Ernberg JJ, Bostrom MP, Wright TM, Harrington RM, Tencer A, Peterson M (1996) Cancellous bone screw thread design and holding power. J Orthop Trauma 10:462–469CrossRefPubMed
11.
go back to reference Scott WA, Allum RL, Wright KW (1985) Implant-induced trabecular damage in cadaveric femoral necks. Acta Orthop Scand 56:145–146PubMed Scott WA, Allum RL, Wright KW (1985) Implant-induced trabecular damage in cadaveric femoral necks. Acta Orthop Scand 56:145–146PubMed
12.
go back to reference Schutz M, Sudkamp NP (2003) Revolution in plate osteosynthesis: new internal fixator systems. J Orthop Sci 8:252–258CrossRefPubMed Schutz M, Sudkamp NP (2003) Revolution in plate osteosynthesis: new internal fixator systems. J Orthop Sci 8:252–258CrossRefPubMed
13.
go back to reference Frigg R. Locking compression plate (LCP) (2001) An osteosynthesis plate based on the dynamic compression plate and the point contact fixator (PC-Fix). Injury 32 (Suppl 2):63–66CrossRefPubMed Frigg R. Locking compression plate (LCP) (2001) An osteosynthesis plate based on the dynamic compression plate and the point contact fixator (PC-Fix). Injury 32 (Suppl 2):63–66CrossRefPubMed
14.
go back to reference Perren SM (2002) Evolution of the internal fixation of long bone fractures. The scientific basis of biological internal fixation: choosing a new balance between stability and biology. J Bone Joint Surg [Br] 84:1093–1110 Perren SM (2002) Evolution of the internal fixation of long bone fractures. The scientific basis of biological internal fixation: choosing a new balance between stability and biology. J Bone Joint Surg [Br] 84:1093–1110
15.
go back to reference Marius Nygaard Smith-Petersen (1953) Lancet 265:145–146 Marius Nygaard Smith-Petersen (1953) Lancet 265:145–146
16.
go back to reference Goldhahn J, Frei R, Seebeck J (2001) Device for fixing implants on or in a bone. PCT/CH/00/00229(WO 01/80754 A1)1-11-2001 Goldhahn J, Frei R, Seebeck J (2001) Device for fixing implants on or in a bone. PCT/CH/00/00229(WO 01/80754 A1)1-11-2001
17.
go back to reference Raftopoulos D, Katsamanis E, Saul F, Liu W, Saddemi S (1993) An intermediate loading rate technique for the determination of mechanical properties of human femoral cortical bone. J Biomed Eng 15:60–66PubMed Raftopoulos D, Katsamanis E, Saul F, Liu W, Saddemi S (1993) An intermediate loading rate technique for the determination of mechanical properties of human femoral cortical bone. J Biomed Eng 15:60–66PubMed
18.
go back to reference Fried NM, Fried D (2001) Comparison of Er:YAG and 9.6-microm TE CO(2) lasers for ablation of skull tissue. Lasers Surg Med 28:335–343CrossRefPubMed Fried NM, Fried D (2001) Comparison of Er:YAG and 9.6-microm TE CO(2) lasers for ablation of skull tissue. Lasers Surg Med 28:335–343CrossRefPubMed
19.
go back to reference Gonzalez C, van de Merwe WP, Smith M, Reinisch L (1990) Comparison of the erbium-yttrium aluminum garnet and carbon dioxide lasers for in vitro bone and cartilage ablation. Laryngoscope 100:14–17PubMed Gonzalez C, van de Merwe WP, Smith M, Reinisch L (1990) Comparison of the erbium-yttrium aluminum garnet and carbon dioxide lasers for in vitro bone and cartilage ablation. Laryngoscope 100:14–17PubMed
20.
go back to reference Muller R (2003) Bone microarchitecture assessment: current and future trends. Osteoporos Int 14 (Suppl 5):89–99CrossRefPubMed Muller R (2003) Bone microarchitecture assessment: current and future trends. Osteoporos Int 14 (Suppl 5):89–99CrossRefPubMed
21.
go back to reference Smith-Petersen MC, Vangorder GW (1931) Intracapsular fractures of the neck of the femur. Arch Surg 23:715–759 Smith-Petersen MC, Vangorder GW (1931) Intracapsular fractures of the neck of the femur. Arch Surg 23:715–759
22.
go back to reference Sehr JR, Szabo RM (1988) Semitubular blade plate for fixation in the proximal humerus. J Orthop Trauma 2:327–332PubMed Sehr JR, Szabo RM (1988) Semitubular blade plate for fixation in the proximal humerus. J Orthop Trauma 2:327–332PubMed
23.
go back to reference Ring D, McKee MD, Perey BH, Jupiter JB (2001) The use of a blade plate and autogenous cancellous bone graft in the treatment of ununited fractures of the proximal humerus. J Shoulder Elbow Surg 10:501–507CrossRefPubMed Ring D, McKee MD, Perey BH, Jupiter JB (2001) The use of a blade plate and autogenous cancellous bone graft in the treatment of ununited fractures of the proximal humerus. J Shoulder Elbow Surg 10:501–507CrossRefPubMed
24.
go back to reference Palmer SH, Handley R, Willett K (2000) The use of interlocked “customised” blade plates in the treatment of metaphyseal fractures in patients with poor bone stock. Injury 31:187–191CrossRefPubMed Palmer SH, Handley R, Willett K (2000) The use of interlocked “customised” blade plates in the treatment of metaphyseal fractures in patients with poor bone stock. Injury 31:187–191CrossRefPubMed
25.
go back to reference Milch H (1954) Blade plate for extension osteotomy of the femur. J Bone Joint Surg [Am] 36-A:155–158 Milch H (1954) Blade plate for extension osteotomy of the femur. J Bone Joint Surg [Am] 36-A:155–158
26.
go back to reference Carpenter CA, Jupiter JB (1996) Blade plate reconstruction of metaphyseal nonunion of the tibia. Clin Orthop 332:23–28CrossRefPubMed Carpenter CA, Jupiter JB (1996) Blade plate reconstruction of metaphyseal nonunion of the tibia. Clin Orthop 332:23–28CrossRefPubMed
27.
go back to reference Heidemann W, Gerlach KL, Grobel KH, Kollner HG (1998) Influence of different pilot hole sizes on torque measurements and pullout analysis of osteosynthesis screws. J Craniomaxillofac Surg 26:50–55PubMed Heidemann W, Gerlach KL, Grobel KH, Kollner HG (1998) Influence of different pilot hole sizes on torque measurements and pullout analysis of osteosynthesis screws. J Craniomaxillofac Surg 26:50–55PubMed
28.
go back to reference Gantous A, Phillips JH (1995) The effects of varying pilot hole size on the holding power of miniscrews and microscrews. Plast Reconstr Surg 95:1165–1169PubMed Gantous A, Phillips JH (1995) The effects of varying pilot hole size on the holding power of miniscrews and microscrews. Plast Reconstr Surg 95:1165–1169PubMed
29.
go back to reference Lovick DS, Ryken TC, Traynelis VC, Dexter F (1997) Assessment of primary and salvage lateral mass screw insertion torque in a cadaveric model. J Spinal Disord 10:431–435PubMed Lovick DS, Ryken TC, Traynelis VC, Dexter F (1997) Assessment of primary and salvage lateral mass screw insertion torque in a cadaveric model. J Spinal Disord 10:431–435PubMed
30.
go back to reference Walsh LJ (2003) The current status of laser applications in dentistry. Aust Dent J 48:146–155PubMed Walsh LJ (2003) The current status of laser applications in dentistry. Aust Dent J 48:146–155PubMed
31.
go back to reference Azizkhan RG (2003) Laser surgery: new applications for pediatric skin and airway lesions. Curr Opin Pediatr 15:243–247CrossRefPubMed Azizkhan RG (2003) Laser surgery: new applications for pediatric skin and airway lesions. Curr Opin Pediatr 15:243–247CrossRefPubMed
32.
go back to reference Small IA, Osborn TP, Fuller T, Hussain M, Kobernick S (1979) Observations of carbon dioxide laser and bone bur in the osteotomy of the rabbit tibia. J Oral Surg 37:159–166PubMed Small IA, Osborn TP, Fuller T, Hussain M, Kobernick S (1979) Observations of carbon dioxide laser and bone bur in the osteotomy of the rabbit tibia. J Oral Surg 37:159–166PubMed
33.
go back to reference Tauber C, Farine I, Horoszowski H, Gassner S (1979) Fracture healing in rabbits after osteotomy using the CO2 laser. Acta Orthop Scand 50:385–390PubMed Tauber C, Farine I, Horoszowski H, Gassner S (1979) Fracture healing in rabbits after osteotomy using the CO2 laser. Acta Orthop Scand 50:385–390PubMed
34.
go back to reference Peavy GM, Reinisch L, Payne JT, Venugopalan V (1999) Comparison of cortical bone ablations by using infrared laser wavelengths 2.9 to 9.2 microm. Lasers Surg Med 25:421–434CrossRefPubMed Peavy GM, Reinisch L, Payne JT, Venugopalan V (1999) Comparison of cortical bone ablations by using infrared laser wavelengths 2.9 to 9.2 microm. Lasers Surg Med 25:421–434CrossRefPubMed
35.
go back to reference Hibst R (1992) Mechanical effects of erbium:YAG laser bone ablation. Lasers Surg Med 12:125–130PubMed Hibst R (1992) Mechanical effects of erbium:YAG laser bone ablation. Lasers Surg Med 12:125–130PubMed
36.
go back to reference Abouzgia MB, James DF (1997) Temperature rise during drilling through bone. Int J Oral Maxillofac Implants 12:342–353PubMed Abouzgia MB, James DF (1997) Temperature rise during drilling through bone. Int J Oral Maxillofac Implants 12:342–353PubMed
37.
go back to reference Cordioli G, Majzoub Z (1997) Heat generation during implant site preparation: an in vitro study. Int J Oral Maxillofac Implants 12:186–193PubMed Cordioli G, Majzoub Z (1997) Heat generation during implant site preparation: an in vitro study. Int J Oral Maxillofac Implants 12:186–193PubMed
38.
go back to reference Eriksson AR, Albrektsson T, Albrektsson B (1984) Heat caused by drilling cortical bone. Temperature measured in vivo in patients and animals. Acta Orthop Scand 55:629–631PubMed Eriksson AR, Albrektsson T, Albrektsson B (1984) Heat caused by drilling cortical bone. Temperature measured in vivo in patients and animals. Acta Orthop Scand 55:629–631PubMed
39.
go back to reference Krause WR, Bradbury DW, Kelly JE, Lunceford EM (1982) Temperature elevations in orthopaedic cutting operations. J Biomech 15:267–275CrossRefPubMed Krause WR, Bradbury DW, Kelly JE, Lunceford EM (1982) Temperature elevations in orthopaedic cutting operations. J Biomech 15:267–275CrossRefPubMed
40.
go back to reference Toksvig-Larsen S, Ryd L, Lindstrand A (1992) Temperature influence in different orthopaedic saw blades. J Arthroplasty 7:21–24CrossRefPubMed Toksvig-Larsen S, Ryd L, Lindstrand A (1992) Temperature influence in different orthopaedic saw blades. J Arthroplasty 7:21–24CrossRefPubMed
41.
go back to reference Watanabe F, Tawada Y, Komatsu S, Hata Y (1992) Heat distribution in bone during preparation of implant sites: heat analysis by real-time thermography. Int J Oral Maxillofac Implants 7:212–219PubMed Watanabe F, Tawada Y, Komatsu S, Hata Y (1992) Heat distribution in bone during preparation of implant sites: heat analysis by real-time thermography. Int J Oral Maxillofac Implants 7:212–219PubMed
42.
go back to reference Frentzen M, Gotz W, Ivanenko M, Afilal S, Werner M, Hering P (2003) Osteotomy with 80-micros CO2 laser pulses—histological results. Lasers Med Sci 18:119–124CrossRefPubMed Frentzen M, Gotz W, Ivanenko M, Afilal S, Werner M, Hering P (2003) Osteotomy with 80-micros CO2 laser pulses—histological results. Lasers Med Sci 18:119–124CrossRefPubMed
43.
go back to reference Hafez MI, Coombs RR, Zhou S, McCarthy ID (2002) Ablation of bone, cartilage, and facet joint capsule using Ho:YAG laser. J Clin Laser Med Surg 20:251–255CrossRefPubMed Hafez MI, Coombs RR, Zhou S, McCarthy ID (2002) Ablation of bone, cartilage, and facet joint capsule using Ho:YAG laser. J Clin Laser Med Surg 20:251–255CrossRefPubMed
44.
go back to reference Ivanenko MM, Fahimi-Weber S, Mitra T, Wierich W, Hering P (2002) Bone tissue ablation with sub-microS pulses of a Q-switch CO(2) laser: histological examination of thermal side effects. Lasers Med Sci 17:258–264CrossRefPubMed Ivanenko MM, Fahimi-Weber S, Mitra T, Wierich W, Hering P (2002) Bone tissue ablation with sub-microS pulses of a Q-switch CO(2) laser: histological examination of thermal side effects. Lasers Med Sci 17:258–264CrossRefPubMed
45.
go back to reference Rayan GM, Stanfield DT, Cahill S, Kosanke SD, Kopta JA (1992) Effects of rapid pulsed CO2 laser beam on cortical bone in vivo. Lasers Surg Med 12:615–620PubMed Rayan GM, Stanfield DT, Cahill S, Kosanke SD, Kopta JA (1992) Effects of rapid pulsed CO2 laser beam on cortical bone in vivo. Lasers Surg Med 12:615–620PubMed
46.
go back to reference Hesjedal T, Behme G (2002) The origin of ultrasound-induced friction reduction in microscopic mechanical contacts. IEEE Trans Ultrason Ferroelectr Freq Control 49:356–364CrossRefPubMed Hesjedal T, Behme G (2002) The origin of ultrasound-induced friction reduction in microscopic mechanical contacts. IEEE Trans Ultrason Ferroelectr Freq Control 49:356–364CrossRefPubMed
47.
go back to reference Feller AM, Domres B (1984) [Thermal skin and bone necrosis following injury with an ultrasound bonding machine]. Aktuelle Traumatol 14:37–39PubMed Feller AM, Domres B (1984) [Thermal skin and bone necrosis following injury with an ultrasound bonding machine]. Aktuelle Traumatol 14:37–39PubMed
Metadata
Title
New implant designs for fracture fixation in osteoporotic bone
Authors
J. Goldhahn
J. Seebeck
R. Frei
B. Frenz
I. Antoniadis
E. Schneider
Publication date
01-03-2005
Publisher
Springer-Verlag
Published in
Osteoporosis International / Issue Special Issue 2/2005
Print ISSN: 0937-941X
Electronic ISSN: 1433-2965
DOI
https://doi.org/10.1007/s00198-004-1765-4

Other articles of this Special Issue 2/2005

Osteoporosis International 2/2005 Go to the issue