Open Access 01-01-2018 | Laboratory Investigation

Needle Tip Visibility in 3D Ultrasound Images

Authors: Muhammad Arif, Adriaan Moelker, Theo van Walsum

Published in: CardioVascular and Interventional Radiology | Issue 1/2018

Login to get access

Abstract

Aim

Needle visibility is crucial for effective and safe ultrasound-guided interventional procedures. Several studies have investigated needle visibility in 2D ultrasound imaging, but less information is available for 3D ultrasound imaging, a modality that has great potential for image guidance interventions. We performed a prospective study, to quantitatively compare the echogenicity of various commercially available needles in 3D ultrasound images used in clinical practice under freehand needle introduction.

Materials and Methods

A set of seven needles, containing biopsy needles, a TIPS needle, an ablation needle and a puncture needle, were included in the study. A liver-mimicking phantom and cow liver were punctured by each needle. 3D sweeps and real-time 3D data were acquired at three different angles (20°, 55° and 90°). Needle visibility was quantified by calculating contrast-to-noise ratio.

Results

In the liver-mimicking phantom, all needles showed better visibility than in the cow liver. At large angles, contrast-to-noise ratio and needle visibility were almost similar in both cases, but at lower angles differences in visibility were observed with different types of needles.

Conclusion

The contrast-to-noise ratio increased with the increase in angle of insonation. The difference in visibility of different needles is more pronounced at 20° angle. The echogenic properties of inhomogeneous cow liver tissues make the needles visibility worse as compared to a homogenous phantom. The needle visibility becomes worse in 3D real-time data as compared to 3D ultrasound sweeps.
Literature
1.
go back to reference Chapman GA, Johnson D, Bodenham AR. Visualisation of needle position using ultrasonography. Anaesthesia. 2006;61(2):148–58.CrossRefPubMed Chapman GA, Johnson D, Bodenham AR. Visualisation of needle position using ultrasonography. Anaesthesia. 2006;61(2):148–58.CrossRefPubMed
2.
go back to reference Culp WC, McCowan TC, Goertzen TC, Habbe TG, Hummel MM, LeVeen RF, et al. Relative ultrasonographic echogenicity of standard, dimpled, and polymeric-coated needles. J Vasc Interv Radiol JVIR. 2000;11(3):351–8.CrossRefPubMed Culp WC, McCowan TC, Goertzen TC, Habbe TG, Hummel MM, LeVeen RF, et al. Relative ultrasonographic echogenicity of standard, dimpled, and polymeric-coated needles. J Vasc Interv Radiol JVIR. 2000;11(3):351–8.CrossRefPubMed
3.
go back to reference Nichols K, Wright LB, Spencer T, Culp WC. Changes in ultrasonographic echogenicity and visibility of needles with changes in angles of insonation. J Vasc Interv Radiol. 2003;14(12):1553–7.CrossRefPubMed Nichols K, Wright LB, Spencer T, Culp WC. Changes in ultrasonographic echogenicity and visibility of needles with changes in angles of insonation. J Vasc Interv Radiol. 2003;14(12):1553–7.CrossRefPubMed
4.
go back to reference Kawai N, Minamiguchi H, Sato M, Nakai M, Sanda H, Tanaka T, et al. Evaluation of vascular puncture needles with specific modifications for enhanced ultrasound visibility: in vitro study. World J Radiol. 2012;4(6):273–7.CrossRefPubMedPubMedCentral Kawai N, Minamiguchi H, Sato M, Nakai M, Sanda H, Tanaka T, et al. Evaluation of vascular puncture needles with specific modifications for enhanced ultrasound visibility: in vitro study. World J Radiol. 2012;4(6):273–7.CrossRefPubMedPubMedCentral
5.
go back to reference Maecken T, Zenz M, Grau T. Ultrasound characteristics of needles for regional anesthesia. Reg Anesth Pain Med. 2007;32(5):440–7.CrossRefPubMed Maecken T, Zenz M, Grau T. Ultrasound characteristics of needles for regional anesthesia. Reg Anesth Pain Med. 2007;32(5):440–7.CrossRefPubMed
6.
go back to reference Banerjee J, Klink C, Niessen WJ, Moelker A, van Walsum T. 4D Ultrasound tracking of liver and its verification for tips guidance. IEEE Trans Med Imaging. 2016;35(1):52–62.CrossRefPubMed Banerjee J, Klink C, Niessen WJ, Moelker A, van Walsum T. 4D Ultrasound tracking of liver and its verification for tips guidance. IEEE Trans Med Imaging. 2016;35(1):52–62.CrossRefPubMed
7.
go back to reference Arif M, Idzenga T, van Mastrigt R, de Korte CL. Estimation of urinary flow velocity in models of obstructed and unobstructed urethras by decorrelation of ultrasound radiofrequency signals. Ultrasound Med Biol. 2014;40(5):938–46.CrossRefPubMed Arif M, Idzenga T, van Mastrigt R, de Korte CL. Estimation of urinary flow velocity in models of obstructed and unobstructed urethras by decorrelation of ultrasound radiofrequency signals. Ultrasound Med Biol. 2014;40(5):938–46.CrossRefPubMed
8.
go back to reference Pluymen LH. Liver tissue mimicking materials for image-guided needle interventions [Thesis]: TU Delft; 2016. Pluymen LH. Liver tissue mimicking materials for image-guided needle interventions [Thesis]: TU Delft; 2016.
9.
go back to reference Hopkins RE, Bradley M. In-vitro visualization of biopsy needles with ultrasound: a comparative study of standard and echogenic needles using an ultrasound phantom. Clin Radiol. 2001;56(6):499–502.CrossRefPubMed Hopkins RE, Bradley M. In-vitro visualization of biopsy needles with ultrasound: a comparative study of standard and echogenic needles using an ultrasound phantom. Clin Radiol. 2001;56(6):499–502.CrossRefPubMed
10.
go back to reference Cohnen M, Saleh A, Lüthen R, Bode J, Mödder U. Improvement of sonographic needle visibility in cirrhotic livers during transjugular intrahepatic portosystemic stent-shunt procedures with use of real-time compound imaging. J Vasc Interv Radiol. 2003;14(1):103–6.CrossRefPubMed Cohnen M, Saleh A, Lüthen R, Bode J, Mödder U. Improvement of sonographic needle visibility in cirrhotic livers during transjugular intrahepatic portosystemic stent-shunt procedures with use of real-time compound imaging. J Vasc Interv Radiol. 2003;14(1):103–6.CrossRefPubMed
11.
go back to reference Jones CD, McGahan JP, Clark KJ. Color Doppler ultrasonographic detection of a vibrating needle system. J Ultrasound Med. 1997;16(4):269–74.CrossRefPubMed Jones CD, McGahan JP, Clark KJ. Color Doppler ultrasonographic detection of a vibrating needle system. J Ultrasound Med. 1997;16(4):269–74.CrossRefPubMed
12.
go back to reference Bradley MJ. An in vitro study to understand successful free-hand ultrasound guided intervention. Clin Radiol. 2001;56(6):495–8.CrossRefPubMed Bradley MJ. An in vitro study to understand successful free-hand ultrasound guided intervention. Clin Radiol. 2001;56(6):495–8.CrossRefPubMed
Metadata
Title
Needle Tip Visibility in 3D Ultrasound Images
Authors
Muhammad Arif
Adriaan Moelker
Theo van Walsum
Publication date
01-01-2018
Publisher
Springer US
Published in
CardioVascular and Interventional Radiology / Issue 1/2018
Print ISSN: 0174-1551
Electronic ISSN: 1432-086X
DOI
https://doi.org/10.1007/s00270-017-1798-7

Other articles of this Issue 1/2018

CardioVascular and Interventional Radiology 1/2018 Go to the issue