Published in: International Journal of Computer Assisted Radiology and Surgery 6/2014

01-11-2014 | Original Article

Needle deflection estimation: prostate brachytherapy phantom experiments

Authors: Hossein Sadjadi, Keyvan Hashtrudi-Zaad, Gabor Fichtinger

Published in: International Journal of Computer Assisted Radiology and Surgery | Issue 6/2014

Login to get access

Abstract

Purpose

The performance of a fusion-based needle deflection estimation method was experimentally evaluated using prostate brachytherapy phantoms. The accuracy of the needle deflection estimation was determined. The robustness of the approach with variations in needle insertion speed and soft tissue biomechanical properties was investigated.

Methods

A needle deflection estimation method was developed to determine the amount of needle bending during insertion into deformable tissue by combining a kinematic deflection model with measurements taken from two electromagnetic trackers placed at the tip and the base of the needle. Experimental verification of this method for use in prostate brachytherapy needle insertion procedures was performed. A total of 21 beveled tip, 18 ga, 200 mm needles were manually inserted at various speeds through a template and toward different targets distributed within 3 soft tissue mimicking polyvinyl chloride prostate phantoms of varying stiffness. The tracked positions of both the needle tip and base were recorded, and Kalman filters were applied to fuse the sensory information. The estimation results were validated using ground truth obtained from fluoroscopy images.

Results

The manual insertion speed ranged from 8 to 34 mm/s, needle deflection ranged from 5 to 8 mm at an insertion depth of 76 mm, and the elastic modulus of the soft tissue ranged from 50 to 150 kPa. The accuracy and robustness of the estimation method were verified within these ranges. When compared to purely model-based estimation, we observed a reduction in needle tip position estimation error by \(52\pm 17\) % (mean \(\pm \) SD) and the cumulative deflection error by \(57\pm 19\) %.

Conclusions

Fusion of electromagnetic sensors demonstrated significant improvement in estimating needle deflection compared to model-based methods. The method has potential clinical applicability in the guidance of needle placement medical interventions, particularly prostate brachytherapy.
Literature
1.
go back to reference Park W, Liu Y, Zhou Y, Moses M, Chirikjian GS (2008) Kinematic state estimation and motion planning for stochastic nonholonomic systems using the exponential map. Robotica 26(4):419–434PubMedCrossRefPubMedCentral Park W, Liu Y, Zhou Y, Moses M, Chirikjian GS (2008) Kinematic state estimation and motion planning for stochastic nonholonomic systems using the exponential map. Robotica 26(4):419–434PubMedCrossRefPubMedCentral
2.
go back to reference Abolhassani N, Patel R, Moallem M (2007) Needle insertion into soft tissue: a survey. Med Eng Phys 29(4):413–431PubMedCrossRef Abolhassani N, Patel R, Moallem M (2007) Needle insertion into soft tissue: a survey. Med Eng Phys 29(4):413–431PubMedCrossRef
3.
go back to reference Sadjadi H, Hashtrudi-Zaad K, Fichtinger G (2012) Needle deflection estimation using fusion of electromagnetic trackers. In: International conference of the IEEE engineering in medicine and biology society (EMBC). San Diego, pp 952–955 Sadjadi H, Hashtrudi-Zaad K, Fichtinger G (2012) Needle deflection estimation using fusion of electromagnetic trackers. In: International conference of the IEEE engineering in medicine and biology society (EMBC). San Diego, pp 952–955
4.
go back to reference Sadjadi H, Hashtrudi-Zaad K, Fichtinger G (Oct 2013) Fusion of electromagnetic trackers to improve needle deflection estimation: simulation study. IEEE Trans Biomed Eng (TBME) 60(10):2706–2715 Sadjadi H, Hashtrudi-Zaad K, Fichtinger G (Oct 2013) Fusion of electromagnetic trackers to improve needle deflection estimation: simulation study. IEEE Trans Biomed Eng (TBME) 60(10):2706–2715
5.
go back to reference Sadjadi H, Hashtrud-Zaad K, Xu H, Fichtinger G (2013) Experimental evaluation of needle deflection estimation for brachytherapy. In: Computer assisted radiology and surgery (CARS 2013). Heidelberg (abstract in International Journal of Computer Assisted Surgery and Radiology, 8(1):391–392) Sadjadi H, Hashtrud-Zaad K, Xu H, Fichtinger G (2013) Experimental evaluation of needle deflection estimation for brachytherapy. In: Computer assisted radiology and surgery (CARS 2013). Heidelberg (abstract in International Journal of Computer Assisted Surgery and Radiology, 8(1):391–392)
6.
go back to reference Hinnen KA, Battermann JJ, van Roermund JGH, Moerland MA, Jurgenliemk-Schulz IM, Frank SJ, van Vulpen M (2010) Long-term biochemical and survival outcome of 921 patients treated with I-125 permanent prostate brachytherapy. Int J Radiat Oncol Biol Phys 76(5):1433–1438PubMedCrossRef Hinnen KA, Battermann JJ, van Roermund JGH, Moerland MA, Jurgenliemk-Schulz IM, Frank SJ, van Vulpen M (2010) Long-term biochemical and survival outcome of 921 patients treated with I-125 permanent prostate brachytherapy. Int J Radiat Oncol Biol Phys 76(5):1433–1438PubMedCrossRef
7.
go back to reference Wallner K, Merrick G, Dattoli MJ, Blasko JC (2007) Prostate brachytherapy made complicated. SmartMedicine Press, p 600 Wallner K, Merrick G, Dattoli MJ, Blasko JC (2007) Prostate brachytherapy made complicated. SmartMedicine Press, p 600
8.
go back to reference Van Veen YRJ, Jahya A, Misra S (2012) Macroscopic and microscopic observations of needle insertion into gels. Eng Med 226(6):441–449CrossRef Van Veen YRJ, Jahya A, Misra S (2012) Macroscopic and microscopic observations of needle insertion into gels. Eng Med 226(6):441–449CrossRef
9.
go back to reference Podder T, Clark D, Sherman J, Fuller D, Messing E, Rubens D, Strang J, Zhang Y, O’Dell W, Ng W, Yu Y (2005) Effects of tip geometry of surgical needles: an assessment of force and deflection. In: European medical and biological engineering conference (EMBEC05). Prague, pp 1727–1983 Podder T, Clark D, Sherman J, Fuller D, Messing E, Rubens D, Strang J, Zhang Y, O’Dell W, Ng W, Yu Y (2005) Effects of tip geometry of surgical needles: an assessment of force and deflection. In: European medical and biological engineering conference (EMBEC05). Prague, pp 1727–1983
10.
go back to reference Podder T, Sherman J, Rubens D, Messing E, Strang J, Ng WS, Yu Y (2008) Methods for prostate stabilization during transperineal LDR brachytherapy. Phys Med Biol 53(6):1563PubMedCrossRef Podder T, Sherman J, Rubens D, Messing E, Strang J, Ng WS, Yu Y (2008) Methods for prostate stabilization during transperineal LDR brachytherapy. Phys Med Biol 53(6):1563PubMedCrossRef
11.
go back to reference Nath S, Chen Z, Yue N, Trumpore S, Peschel R (2000) Dosimetric effects of needle divergence in prostate seed implant using 125I and 103Pd radioactive seeds. Med Phys 27(5):1058PubMedCrossRef Nath S, Chen Z, Yue N, Trumpore S, Peschel R (2000) Dosimetric effects of needle divergence in prostate seed implant using 125I and 103Pd radioactive seeds. Med Phys 27(5):1058PubMedCrossRef
12.
go back to reference Lagerburg V, Moerland M, Konings M, Vosse R, JaBJ Lagendijk (2006) Development of a tapping device: a new needle insertion method for prostate brachytherapy. Phys Med Biol 51:891PubMedCrossRef Lagerburg V, Moerland M, Konings M, Vosse R, JaBJ Lagendijk (2006) Development of a tapping device: a new needle insertion method for prostate brachytherapy. Phys Med Biol 51:891PubMedCrossRef
13.
go back to reference Abolhassani N, Patel R, Ayazi F (2007) Effects of different insertion methods on reducing needle deflection. In: 29th Annual international conference of the IEEE engineering in medicine and biology society, EMBS 2007, pp 491–494 Abolhassani N, Patel R, Ayazi F (2007) Effects of different insertion methods on reducing needle deflection. In: 29th Annual international conference of the IEEE engineering in medicine and biology society, EMBS 2007, pp 491–494
14.
go back to reference McGill CS, Schwartz JA, Moore JZ, McLaughlin PW, Shih AJ (2011) Precision grid and hand motion for accurate needle insertion in brachytherapy. Med Phys 38:4749PubMedCrossRef McGill CS, Schwartz JA, Moore JZ, McLaughlin PW, Shih AJ (2011) Precision grid and hand motion for accurate needle insertion in brachytherapy. Med Phys 38:4749PubMedCrossRef
15.
go back to reference Majewicz A, Marra S, van Vledder M, Lin M, Choti M, Song D, Okamura A (2012) Behavior of tip-steerable needles in ex vivo and in vivo tissue. IEEE Trans Biomed Eng 59(10):2705–2715PubMedCrossRefPubMedCentral Majewicz A, Marra S, van Vledder M, Lin M, Choti M, Song D, Okamura A (2012) Behavior of tip-steerable needles in ex vivo and in vivo tissue. IEEE Trans Biomed Eng 59(10):2705–2715PubMedCrossRefPubMedCentral
16.
go back to reference Goksel O, Dehghan E, Salcudean S (2009) Modeling and simulation of flexible needles. Med Eng Phys 31(9):1069–1078 Goksel O, Dehghan E, Salcudean S (2009) Modeling and simulation of flexible needles. Med Eng Phys 31(9):1069–1078
17.
go back to reference Glozman D, Shoham M (2007) Image-guided robotic flexible needle steering. IEEE Trans Robot 23(3):459–467CrossRef Glozman D, Shoham M (2007) Image-guided robotic flexible needle steering. IEEE Trans Robot 23(3):459–467CrossRef
18.
go back to reference Asadian A, Kermani M, Patel R (2011) An analytical model for deflection of flexible needles during needle insertion. In: IEEE/RSJ international conference on intelligent robots and systems (IROS), pp 2551–2556 Asadian A, Kermani M, Patel R (2011) An analytical model for deflection of flexible needles during needle insertion. In: IEEE/RSJ international conference on intelligent robots and systems (IROS), pp 2551–2556
19.
go back to reference Okamura AM, Simone C, O’Leary MD (2004) Force modeling for needle insertion into soft tissue. IEEE Trans Biomed Eng 51(10):1707–1716PubMedCrossRef Okamura AM, Simone C, O’Leary MD (2004) Force modeling for needle insertion into soft tissue. IEEE Trans Biomed Eng 51(10):1707–1716PubMedCrossRef
20.
go back to reference DiMaio SP, Salcudean S (2003) Needle insertion modeling and simulation. IEEE Trans Robot Autom 19(5):864–875CrossRef DiMaio SP, Salcudean S (2003) Needle insertion modeling and simulation. IEEE Trans Robot Autom 19(5):864–875CrossRef
21.
go back to reference Podder T, Clark D, Sherman J, Fuller D, Messing E, Rubens D, Strang J, Brasacchio R, Liao L, Ng WS et al (2006) In vivo motion and force measurement of surgical needle intervention during prostate brachytherapy. Med Phys 33:2915PubMedCrossRef Podder T, Clark D, Sherman J, Fuller D, Messing E, Rubens D, Strang J, Brasacchio R, Liao L, Ng WS et al (2006) In vivo motion and force measurement of surgical needle intervention during prostate brachytherapy. Med Phys 33:2915PubMedCrossRef
22.
go back to reference Yan K, Podder T, Xiao D, Yu Y, Liu TI, Cheng C, Ng W (2006) An improved needle steering model with online parameter estimator. Int J Comput Assist Radiol Surg 1(4):205–212CrossRef Yan K, Podder T, Xiao D, Yu Y, Liu TI, Cheng C, Ng W (2006) An improved needle steering model with online parameter estimator. Int J Comput Assist Radiol Surg 1(4):205–212CrossRef
23.
go back to reference Jafari AH, Dhaouadi R, Jhemi A (2013) Nonlinear friction estimation in elastic drive systems using a dynamic neural network-based observer. J Adv Comput Intell Intell Inform 17(4):637–646 Jafari AH, Dhaouadi R, Jhemi A (2013) Nonlinear friction estimation in elastic drive systems using a dynamic neural network-based observer. J Adv Comput Intell Intell Inform 17(4):637–646
24.
go back to reference DiMaio SP, Salcudean SE (2005) Needle steering and motion planning in soft tissues. IEEE Trans Biomed Eng 52(6):965–974PubMedCrossRef DiMaio SP, Salcudean SE (2005) Needle steering and motion planning in soft tissues. IEEE Trans Biomed Eng 52(6):965–974PubMedCrossRef
25.
go back to reference Webster RJ, Kim JS, Cowan NJ, Chirikjian GS, Okamura AM (2006) Nonholonomic modeling of needle steering. Int J Robot Res 25(5):509CrossRef Webster RJ, Kim JS, Cowan NJ, Chirikjian GS, Okamura AM (2006) Nonholonomic modeling of needle steering. Int J Robot Res 25(5):509CrossRef
26.
go back to reference Minhas DS, Engh JA, Fenske MM, Riviere CN (2007) Modeling of needle steering via duty-cycled spinning. In: International conference of the IEEE engineering in medicine and biology society (EMBC), pp 2756–2759 Minhas DS, Engh JA, Fenske MM, Riviere CN (2007) Modeling of needle steering via duty-cycled spinning. In: International conference of the IEEE engineering in medicine and biology society (EMBC), pp 2756–2759
27.
go back to reference Bernardes MC, Adorno BV, Poignet P, Borges GA (2013) Robot-assisted automatic insertion of steerable needles with closed-loop imaging feedback and intraoperative trajectory replanning. Mechatronics 23(6):630–645CrossRef Bernardes MC, Adorno BV, Poignet P, Borges GA (2013) Robot-assisted automatic insertion of steerable needles with closed-loop imaging feedback and intraoperative trajectory replanning. Mechatronics 23(6):630–645CrossRef
28.
go back to reference Park W, Reed KB, Okamura AM, Chirikjian GS (2010) Estimation of model parameters for steerable needles. In: IEEE International conference on robotics and automation (ICRA), pp 3703–3708 Park W, Reed KB, Okamura AM, Chirikjian GS (2010) Estimation of model parameters for steerable needles. In: IEEE International conference on robotics and automation (ICRA), pp 3703–3708
29.
go back to reference Maier-Hein L, Franz A, Birkfellner W, Hummel J, Gergel I, Wegner I, Meinzer HP (2012) Standardized assessment of new electromagnetic field generators in an interventional radiology setting. Med Phys 39:3424PubMedCrossRef Maier-Hein L, Franz A, Birkfellner W, Hummel J, Gergel I, Wegner I, Meinzer HP (2012) Standardized assessment of new electromagnetic field generators in an interventional radiology setting. Med Phys 39:3424PubMedCrossRef
30.
go back to reference Aboofazeli M, Abolmaesumi P, Mousavi P, Fichtinger G (2009) A new scheme for curved needle segmentation in three-dimensional ultrasound images. In: IEEE international symposium on biomedical imaging: from nano to macro, (ISBI), pp 1067–1070 Aboofazeli M, Abolmaesumi P, Mousavi P, Fichtinger G (2009) A new scheme for curved needle segmentation in three-dimensional ultrasound images. In: IEEE international symposium on biomedical imaging: from nano to macro, (ISBI), pp 1067–1070
31.
go back to reference Ren H, Rank D, Merdes M, Stallkamp J, Kazanzides P (2012) Multisensor data fusion in an integrated tracking system for endoscopic surgery. IEEE Trans Inf Technol Biomed 16(1):106–111PubMedCrossRef Ren H, Rank D, Merdes M, Stallkamp J, Kazanzides P (2012) Multisensor data fusion in an integrated tracking system for endoscopic surgery. IEEE Trans Inf Technol Biomed 16(1):106–111PubMedCrossRef
32.
go back to reference Dehghan E, Wen X, Zahiri-Azar R, Marchal M, Salcudean SE (2008) Needle-tissue interaction modeling using ultrasound-based motion estimation: phantom study. Computer Aided Surg 13(5):265–280 Dehghan E, Wen X, Zahiri-Azar R, Marchal M, Salcudean SE (2008) Needle-tissue interaction modeling using ultrasound-based motion estimation: phantom study. Computer Aided Surg 13(5):265–280
33.
go back to reference Datla NV, Konh B, Koo JJ, Choi DJ, Yu Y, Dicker AP, Podder TK, Darvish K, Hutapea P (2014) Polyacrylamide phantom for self-actuating needle-tissue interaction studies. Med Eng Phys 36(1):140–145PubMedCrossRef Datla NV, Konh B, Koo JJ, Choi DJ, Yu Y, Dicker AP, Podder TK, Darvish K, Hutapea P (2014) Polyacrylamide phantom for self-actuating needle-tissue interaction studies. Med Eng Phys 36(1):140–145PubMedCrossRef
34.
go back to reference Hungr N, Long JA, Beix V, Troccaz J (2012) A realistic deformable prostate phantom for multimodal imaging and needle-insertion procedures. Med Phys 39(4):2031–2041PubMedCrossRef Hungr N, Long JA, Beix V, Troccaz J (2012) A realistic deformable prostate phantom for multimodal imaging and needle-insertion procedures. Med Phys 39(4):2031–2041PubMedCrossRef
35.
go back to reference Betrouni N, Nevoux P, Leroux B, Colin P, Puech P, Mordon S (2013) An anatomically realistic and adaptable prostate phantom for laser thermotherapy treatment planning. Med Phys 40:022701PubMedCrossRef Betrouni N, Nevoux P, Leroux B, Colin P, Puech P, Mordon S (2013) An anatomically realistic and adaptable prostate phantom for laser thermotherapy treatment planning. Med Phys 40:022701PubMedCrossRef
36.
go back to reference Bo LE, Leira HO, Tangen GA, Hofstad EF, Amundsen T, Lango T (2012) Accuracy of electromagnetic tracking with a prototype field generator in an interventional OR setting. Med Phys 39(1):399PubMedCrossRef Bo LE, Leira HO, Tangen GA, Hofstad EF, Amundsen T, Lango T (2012) Accuracy of electromagnetic tracking with a prototype field generator in an interventional OR setting. Med Phys 39(1):399PubMedCrossRef
37.
go back to reference Buyyounouski MK, Davis BJ, Prestidge BR, Shanahan TG, Stock RG, Grimm PD, Demanes DJ, Zaider M, Horwitz EM (2012) A survey of current clinical practice in permanent and temporary prostate brachytherapy: 2010 update. Brachytherapy 11(4):299PubMedCrossRef Buyyounouski MK, Davis BJ, Prestidge BR, Shanahan TG, Stock RG, Grimm PD, Demanes DJ, Zaider M, Horwitz EM (2012) A survey of current clinical practice in permanent and temporary prostate brachytherapy: 2010 update. Brachytherapy 11(4):299PubMedCrossRef
38.
go back to reference Seeberger R, Kane G, Hoffmann J, Eggers G (2012) Accuracy assessment for navigated maxillo-facial surgery using an electromagnetic tracking device. J Craniomaxillofac Surg 40(2):156–161 PubMedCrossRef Seeberger R, Kane G, Hoffmann J, Eggers G (2012) Accuracy assessment for navigated maxillo-facial surgery using an electromagnetic tracking device. J Craniomaxillofac Surg 40(2):156–161 PubMedCrossRef
39.
go back to reference Lugez E, Pichora DR, Akl SG, Ellis RE (2013) Accuracy of electromagnetic tracking in an operating-room setting. In: Computer assisted radiology and surgery (CARS 2013). Heidelberg (abstract in International Journal of Computer Assisted Surgery and Radiology, vol. 8(1):147–148) Lugez E, Pichora DR, Akl SG, Ellis RE (2013) Accuracy of electromagnetic tracking in an operating-room setting. In: Computer assisted radiology and surgery (CARS 2013). Heidelberg (abstract in International Journal of Computer Assisted Surgery and Radiology, vol. 8(1):147–148)
40.
go back to reference Lango T, Vijayan S, Rethy A, Vapenstad C, Solberg OV, Marvik R, Johnsen G, Hernes TN (2012) Navigated laparoscopic ultrasound in abdominal soft tissue surgery: technological overview and perspectives. Int J Comput Assist Radiol Surg 7(4):585–599PubMedCrossRef Lango T, Vijayan S, Rethy A, Vapenstad C, Solberg OV, Marvik R, Johnsen G, Hernes TN (2012) Navigated laparoscopic ultrasound in abdominal soft tissue surgery: technological overview and perspectives. Int J Comput Assist Radiol Surg 7(4):585–599PubMedCrossRef
41.
go back to reference Lugez E, Pichora DR, Akl SG, Ellis RE (2013) Accuracy of electromagnetic tracking in an image-guided surgery suite. Bone Jt J Orthop Proc Suppl 95(Supp 28):25 Lugez E, Pichora DR, Akl SG, Ellis RE (2013) Accuracy of electromagnetic tracking in an image-guided surgery suite. Bone Jt J Orthop Proc Suppl 95(Supp 28):25
42.
go back to reference Jain A, Kon R, Zhou Y, Fichtinger G (2005) C-arm calibration-is it really necessary? In: Medical image computing and computer-assisted intervention (MICCAI), pp 639–646 Jain A, Kon R, Zhou Y, Fichtinger G (2005) C-arm calibration-is it really necessary? In: Medical image computing and computer-assisted intervention (MICCAI), pp 639–646
43.
go back to reference Wan G, Wei Z, Gardi L, Downey DB, Fenster A (2005) Brachytherapy needle deflection evaluation and correction. Med Phys 32:902PubMedCrossRef Wan G, Wei Z, Gardi L, Downey DB, Fenster A (2005) Brachytherapy needle deflection evaluation and correction. Med Phys 32:902PubMedCrossRef
Metadata
Title
Needle deflection estimation: prostate brachytherapy phantom experiments
Authors
Hossein Sadjadi
Keyvan Hashtrudi-Zaad
Gabor Fichtinger
Publication date
01-11-2014
Publisher
Springer Berlin Heidelberg
Published in
International Journal of Computer Assisted Radiology and Surgery / Issue 6/2014
Print ISSN: 1861-6410
Electronic ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-014-0985-0

Other articles of this Issue 6/2014

International Journal of Computer Assisted Radiology and Surgery 6/2014 Go to the issue