NEC Density and Liver ROI S/N Ratio for Image Quality Control of Whole-Body FDG-PET Scans: Comparison with Visual Assessment

Authors: Tetsuro Mizuta, Michio Senda, Terue Okamura, Keishi Kitamura, Yuichi Inaoka, Munehiro Takahashi, Keiichi Matsumoto, Makoto Abe, Yoshihiro Shimonishi, Susumu Shiomi

Published in: Molecular Imaging and Biology | Issue 6/2009

Login to get access

Abstract

Purpose

Patient noise equivalent count (NEC), NEC density, and liver region of interest (ROI) S/N have been proposed as physical indicators of image quality, but have not been thoroughly compared with visual assessments. In this study, those indicators were contrasted with blind visual evaluations for whole-body fluorodeoxyglucose–positron emission tomography (FDG-PET) images acquired under a variety of scanning conditions and body weights.

Methods

Images were acquired on 15 normal subjects using a SET-3000B/L PET scanner with a continuous bed motion. Body weight ranged from 50.2 to 95.7 kg, with injected activity ranging from 71 to 333 MBq (1.40 to 3.67 MBq/kg) and a scan duration from 10 to 30 min. Patient NEC (PNEC; counts/cm) was calculated as the NEC rate divided by bed speed. NEC density (counts/cm3) was defined as the PNEC divided by the cross-sectional area derived from transmission data. Both PNEC and NEC density were averaged from neck to abdomen. Liver S/N was obtained as the pixel mean/SD within the ROI. Blind reviews by 18 professionals were used to visually evaluate image quality.

Results

Average visual score correlated with liver S/N, PNEC, and NEC density, with a rank correlation coefficient of 0.81, 0.86, and 0.91, respectively (each p < 0.0003). The “acceptable” quality roughly corresponded to a liver S/N of 10, PNEC of 380 kcounts/cm, and NEC density of 550 counts/cm3 or more.

Conclusions

NEC density, representing count statistics per body volume, reflects the visual image quality assessment and may be utilized for quality control of whole-body FDG-PET images together with the liver ROI S/N ratio.
Literature
1.
go back to reference Eberl S, Fulham MJ, Meikle SR, Watson CC (2004) Optimization of injected dose and scanning protocol for 3D whole-body PETCT studies. J Nucl Med 45:426P Eberl S, Fulham MJ, Meikle SR, Watson CC (2004) Optimization of injected dose and scanning protocol for 3D whole-body PETCT studies. J Nucl Med 45:426P
2.
go back to reference Kinahan PE, Surti S, Badawi RD, Fakhri G El, Karp JS, Lewellen TK (2004) Effect of patient thickness on noise equivalent count rates for 2D and fully-3D whole-body PET imaging. J Nucl Med 45(5). Proceedings of the SNM 51 Annual Meeting No. 284 Kinahan PE, Surti S, Badawi RD, Fakhri G El, Karp JS, Lewellen TK (2004) Effect of patient thickness on noise equivalent count rates for 2D and fully-3D whole-body PET imaging. J Nucl Med 45(5). Proceedings of the SNM 51 Annual Meeting No. 284
3.
go back to reference Watson CC, Casey ME, Bendriem B, Carney JP, Townsend DW, Eberl S (2005) Optimizing injected dose in clinical PET by accurately modeling the counting-rate response functions specific to individual patient scans. J Nucl Med 46(11):1825–1834PubMed Watson CC, Casey ME, Bendriem B, Carney JP, Townsend DW, Eberl S (2005) Optimizing injected dose in clinical PET by accurately modeling the counting-rate response functions specific to individual patient scans. J Nucl Med 46(11):1825–1834PubMed
4.
go back to reference deKemp A, Caldwell CB, Famcombe TH, McKee BT, Wassenaar RW, Wells RG et al (2006) PET imaging standards and quality assurance for the multi-center trials of the Ontario Clinical Oncology Group. J Nucl Med 47(1):365P deKemp A, Caldwell CB, Famcombe TH, McKee BT, Wassenaar RW, Wells RG et al (2006) PET imaging standards and quality assurance for the multi-center trials of the Ontario Clinical Oncology Group. J Nucl Med 47(1):365P
5.
go back to reference Matsumoto K, Kitamura K, Mizuta T, Tanaka K, Yamamoto S, Sakamoto S et al (2006) Performance characteristics of a new 3-dimensional continuous emission and spiral transmission high-sensitivity and high-resolution PET camera evaluated with the NEMA NU2-2001 Standard. J Nucl Med 47:83–90PubMed Matsumoto K, Kitamura K, Mizuta T, Tanaka K, Yamamoto S, Sakamoto S et al (2006) Performance characteristics of a new 3-dimensional continuous emission and spiral transmission high-sensitivity and high-resolution PET camera evaluated with the NEMA NU2-2001 Standard. J Nucl Med 47:83–90PubMed
6.
go back to reference Kitamura K, Takahashi S, Tanaka A, Ishikawa A, Mizuta T, Tanaka K et al (2004) 3D continuous emission and spiral transmission scanning for high-throughput whole-body PET. IEEE NSS and MIC Conference Record Kitamura K, Takahashi S, Tanaka A, Ishikawa A, Mizuta T, Tanaka K et al (2004) 3D continuous emission and spiral transmission scanning for high-throughput whole-body PET. IEEE NSS and MIC Conference Record
7.
go back to reference Tanaka E, Amo Y (1998) A Fourier rebinning algorithm incorporating spectral transfer efficiency for 3D PET. Phys Med Biol 43:739–746CrossRefPubMed Tanaka E, Amo Y (1998) A Fourier rebinning algorithm incorporating spectral transfer efficiency for 3D PET. Phys Med Biol 43:739–746CrossRefPubMed
8.
go back to reference Kitamura K, Mizuta T, Tanaka K, Amano M (2003) A weighted Fourier rebinning method with high-pass filtering for continuous 3D scanning with a large aperture PET scanner. J Nucl Med 44(5):271P Kitamura K, Mizuta T, Tanaka K, Amano M (2003) A weighted Fourier rebinning method with high-pass filtering for continuous 3D scanning with a large aperture PET scanner. J Nucl Med 44(5):271P
9.
go back to reference Tanaka E, Kudo H (2003) Subset-dependent relaxation in block-iterative algorithms for image reconstruction in emission tomography. Phys Med Biol 48:1405–1422CrossRefPubMed Tanaka E, Kudo H (2003) Subset-dependent relaxation in block-iterative algorithms for image reconstruction in emission tomography. Phys Med Biol 48:1405–1422CrossRefPubMed
10.
go back to reference National Electrical Manufacturers Association (2001) NEMA Standards Publication NU 2-2001. Performance measurements of positron emission tomographs. National Electrical Manufacturers Association, Rosslyn, VA National Electrical Manufacturers Association (2001) NEMA Standards Publication NU 2-2001. Performance measurements of positron emission tomographs. National Electrical Manufacturers Association, Rosslyn, VA
11.
go back to reference Bendriem B, Townsend DW (1998) The theory and practice of 3D PET. Kluwer, The Netherlands Bendriem B, Townsend DW (1998) The theory and practice of 3D PET. Kluwer, The Netherlands
12.
go back to reference Cherry SR, Dahlbom M (2006) PET: physics, instrumentation, and scanners. Springer, New York Cherry SR, Dahlbom M (2006) PET: physics, instrumentation, and scanners. Springer, New York
13.
go back to reference Badawi RD, Marsden PK, Cronin BF, Sutcliffe JL, Maisey MN (1996) Optimization of noise-equivalent count rates in 3D PET. Phys Med Biol 41:1755–1776CrossRefPubMed Badawi RD, Marsden PK, Cronin BF, Sutcliffe JL, Maisey MN (1996) Optimization of noise-equivalent count rates in 3D PET. Phys Med Biol 41:1755–1776CrossRefPubMed
14.
go back to reference Juweid ME, Stroobants S, Hoekstra OS, Mottaghy FM, Dietlein M, Guermazi A et al (2007) Use of positron emission tomography for response assessment of lymphoma: consensus of the imaging subcommittee of international harmonization project in lymphoma. J Clin Oncol 25(5):571–578CrossRefPubMed Juweid ME, Stroobants S, Hoekstra OS, Mottaghy FM, Dietlein M, Guermazi A et al (2007) Use of positron emission tomography for response assessment of lymphoma: consensus of the imaging subcommittee of international harmonization project in lymphoma. J Clin Oncol 25(5):571–578CrossRefPubMed
15.
go back to reference Cheson BD, Pfistner B, Juweid ME, Gascoyne RD, Specht L, Horning SJ et al (2007) Revised response criteria for malignant lymphoma. J Clin Oncol 25(5):579–586CrossRefPubMed Cheson BD, Pfistner B, Juweid ME, Gascoyne RD, Specht L, Horning SJ et al (2007) Revised response criteria for malignant lymphoma. J Clin Oncol 25(5):579–586CrossRefPubMed
Metadata
Title
NEC Density and Liver ROI S/N Ratio for Image Quality Control of Whole-Body FDG-PET Scans: Comparison with Visual Assessment
Authors
Tetsuro Mizuta
Michio Senda
Terue Okamura
Keishi Kitamura
Yuichi Inaoka
Munehiro Takahashi
Keiichi Matsumoto
Makoto Abe
Yoshihiro Shimonishi
Susumu Shiomi
Publication date
01-11-2009
Publisher
Springer-Verlag
Published in
Molecular Imaging and Biology / Issue 6/2009
Print ISSN: 1536-1632
Electronic ISSN: 1860-2002
DOI
https://doi.org/10.1007/s11307-009-0214-3

Other articles of this Issue 6/2009

Molecular Imaging and Biology 6/2009 Go to the issue

Letter to the Editor

Letter to the Editor

Response to Letter to the Editor

Letter to the Editor