Modular titanium alloy neck adapter failures in hip replacement - failure mode analysis and influence of implant material

Authors: Thomas M Grupp, Thomas Weik, Wilhelm Bloemer, Hanns-Peter Knaebel

Published in: BMC Musculoskeletal Disorders | Issue 1/2010

Login to get access

Abstract

Background

Modular neck adapters for hip arthroplasty stems allow the surgeon to modify CCD angle, offset and femoral anteversion intraoperatively. Fretting or crevice corrosion may lead to failure of such a modular device due to high loads or surface contamination inside the modular coupling. Unfortunately we have experienced such a failure of implants and now report our clinical experience with the failures in order to advance orthopaedic material research and joint replacement surgery.
The failed neck adapters were implanted between August 2004 and November 2006 a total of about 5000 devices. After this period, the titanium neck adapters were replaced by adapters out of cobalt-chromium. Until the end of 2008 in total 1.4% (n = 68) of the implanted titanium alloy neck adapters failed with an average time of 2.0 years (0.7 to 4.0 years) postoperatively. All, but one, patients were male, their average age being 57.4 years (36 to 75 years) and the average weight 102.3 kg (75 to 130 kg). The failures of neck adapters were divided into 66% with small CCD of 130° and 60% with head lengths of L or larger. Assuming an average time to failure of 2.8 years, the cumulative failure rate was calculated with 2.4%.

Methods

A series of adapter failures of titanium alloy modular neck adapters in combination with a titanium alloy modular short hip stem was investigated. For patients having received this particular implant combination risk factors were identified which were associated with the occurence of implant failure. A Kaplan-Meier survival-failure-analysis was conducted. The retrieved implants were analysed using microscopic and chemical methods. Modes of failure were simulated in biomechanical tests. Comparative tests included modular neck adapters made of titanium alloy and cobalt chrome alloy material.

Results

Retrieval examinations and biomechanical simulation revealed that primary micromotions initiated fretting within the modular tapered neck connection. A continuous abrasion and repassivation process with a subsequent cold welding at the titanium alloy modular interface. Surface layers of 10 - 30 μm titanium oxide were observed. Surface cracks caused by fretting or fretting corrosion finally lead to fatigue fracture of the titanium alloy modular neck adapters. Neck adapters made of cobalt chrome alloy show significantly reduced micromotions especially in case of contaminated cone connection. With a cobalt-chromium neck the micromotions can be reduced by a factor of 3 compared to the titanium neck. The incidence of fretting corrosion was also substantially lower with the cobalt-chromium neck configuration.

Conclusions

Failure of modular titanium alloy neck adapters can be initiated by surface micromotions due to surface contamination or highly loaded implant components. In the present study, the patients at risk were men with an average weight over 100 kg. Modular cobalt chrome neck adapters provide higher safety compared to titanium alloy material.
Appendix
Available only for authorised users
Literature
1.
go back to reference Toni A, Paderni S, Sudanese A, Guerra E, Traina F, Antonietti B, Giunti A: Anatomic cementless total hip arthroplasty with ceramic bearings and modular necks: 3 to 5 years follow-up. Hip International. 2001, 11 (1): 1-17. Toni A, Paderni S, Sudanese A, Guerra E, Traina F, Antonietti B, Giunti A: Anatomic cementless total hip arthroplasty with ceramic bearings and modular necks: 3 to 5 years follow-up. Hip International. 2001, 11 (1): 1-17.
2.
go back to reference Traina F, Baleani M, Viceconti M, Toni A: Modular Neck Primary Prosthesis: Experimental and Clinical Outcomes. Scientific Exhibit at the 71st AAOS Annual Meeting, San Francisco. 2004 Traina F, Baleani M, Viceconti M, Toni A: Modular Neck Primary Prosthesis: Experimental and Clinical Outcomes. Scientific Exhibit at the 71st AAOS Annual Meeting, San Francisco. 2004
3.
go back to reference Toni A, Sudanese A, Paderni S, Guerra E, Bianchi G, Antonietti B, Giunti A: Artroprotesi d'anca non cementata con collo modulare - Cementless hip arthroplasty with a modular neck. Chir Organi Mov. 2001, LXXXVI: 73-85. Toni A, Sudanese A, Paderni S, Guerra E, Bianchi G, Antonietti B, Giunti A: Artroprotesi d'anca non cementata con collo modulare - Cementless hip arthroplasty with a modular neck. Chir Organi Mov. 2001, LXXXVI: 73-85.
4.
go back to reference Braun A, Lazovic D, Zigan R: Modular short stem prosthesis- Implant positioning and the impact of navigation. Orthopedics Suppl. 2007, 30 (Suppl 10): 144-147. Braun A, Lazovic D, Zigan R: Modular short stem prosthesis- Implant positioning and the impact of navigation. Orthopedics Suppl. 2007, 30 (Suppl 10): 144-147.
5.
go back to reference Keggi KJ: My experience with proximal modular stems. Symposium "Cutting-edge developments on proximal modularity in THA" Annual AAHKS Meeting. 2008, 16-22. Keggi KJ: My experience with proximal modular stems. Symposium "Cutting-edge developments on proximal modularity in THA" Annual AAHKS Meeting. 2008, 16-22.
6.
go back to reference Keggi KJ, Kennon RE, Keggi JM: Hip reconstruction - primary total hip arthroplasty with modular stems. Current Orthopaedic Practice. 2008, 19 (2): 124-126. 10.1097/BCO.0b013e3282f53eca.CrossRef Keggi KJ, Kennon RE, Keggi JM: Hip reconstruction - primary total hip arthroplasty with modular stems. Current Orthopaedic Practice. 2008, 19 (2): 124-126. 10.1097/BCO.0b013e3282f53eca.CrossRef
7.
go back to reference Cameron HU, Leslie CJ, McTighe T: Design considerations and results for a modular neck in cemented THA. Abstract, Australian Orthopaedic Association 67th Annual Scientific Meeting. 2007 Cameron HU, Leslie CJ, McTighe T: Design considerations and results for a modular neck in cemented THA. Abstract, Australian Orthopaedic Association 67th Annual Scientific Meeting. 2007
8.
go back to reference Hermle T, Zeller R, Grupp TM, Blömer W: Metha short stem hip prosthesis - Examination of the modular cone connection. 2007, Springer Medizin Verlag, Heidelberg, implant, 12-17. 1/07 Hermle T, Zeller R, Grupp TM, Blömer W: Metha short stem hip prosthesis - Examination of the modular cone connection. 2007, Springer Medizin Verlag, Heidelberg, implant, 12-17. 1/07
9.
go back to reference Bobyn JD, Tanzer M, Krygier JJ, Dujovne AR, Brooks CE: Concerns with modularity in total hip arthroplasty. Clin Orth Rel Res. 1994, 27-38. 298 Bobyn JD, Tanzer M, Krygier JJ, Dujovne AR, Brooks CE: Concerns with modularity in total hip arthroplasty. Clin Orth Rel Res. 1994, 27-38. 298
10.
go back to reference Brown SA, Flemming CA, Kawalec JS, Placko HE, Vassaux C, Merritt K, Payer JH, Kraay MJ: Fretting corrosion accelerates crevice corrosion of modular hip tapers. J Appl Biomater. 1995, 6 (1): 19-26. 10.1002/jab.770060104.CrossRefPubMed Brown SA, Flemming CA, Kawalec JS, Placko HE, Vassaux C, Merritt K, Payer JH, Kraay MJ: Fretting corrosion accelerates crevice corrosion of modular hip tapers. J Appl Biomater. 1995, 6 (1): 19-26. 10.1002/jab.770060104.CrossRefPubMed
11.
go back to reference Lieberman JR, Rimnac CM, Garvin KL, Klein RW, Salvati EA: An analysis of the head-neck taper interface in retrieved hip prostheses. Clin Orth Rel Res. 1994, 162-167. 300 Lieberman JR, Rimnac CM, Garvin KL, Klein RW, Salvati EA: An analysis of the head-neck taper interface in retrieved hip prostheses. Clin Orth Rel Res. 1994, 162-167. 300
12.
go back to reference Gilbert JL, Buckley CA, Jacobs JJ: In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy coupling. Journal of Biomedical Materials Research. 1993, 27: 1533-1544.PubMed Gilbert JL, Buckley CA, Jacobs JJ: In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy coupling. Journal of Biomedical Materials Research. 1993, 27: 1533-1544.PubMed
13.
go back to reference Goldberg JR, Gilbert JL, Jacobs JJ, Bauer TW, Paprosky W, Leurgans S: A multicenter retrieval study of the taper interfaces of modular hip prostheses. Clin Orth Rel Res. 2002, 149-161. 10.1097/00003086-200208000-00018. 401 Goldberg JR, Gilbert JL, Jacobs JJ, Bauer TW, Paprosky W, Leurgans S: A multicenter retrieval study of the taper interfaces of modular hip prostheses. Clin Orth Rel Res. 2002, 149-161. 10.1097/00003086-200208000-00018. 401
14.
go back to reference Windler M: Korrosionsverhalten von modularen Verbindungen bei Hüftprothesen. Dissertation Eidgenössisch TechnischeHochschule Zürich No. 15284. 2003 Windler M: Korrosionsverhalten von modularen Verbindungen bei Hüftprothesen. Dissertation Eidgenössisch TechnischeHochschule Zürich No. 15284. 2003
15.
go back to reference Collier JP, Surprenant VA, Jensen RE, Mayor MB: Corrosion at the interface of cobalt-alloy heads on titanium-alloy stems. Clin Orth Rel Res. 1991, 305-312. 271 Collier JP, Surprenant VA, Jensen RE, Mayor MB: Corrosion at the interface of cobalt-alloy heads on titanium-alloy stems. Clin Orth Rel Res. 1991, 305-312. 271
16.
go back to reference McKellop H, Sarmiento A, Brien W, Park SH: Interface corrosion of a modular head total hip prosthesis. J Arthroplasty. 1992, 7 (3): 291-294. 10.1016/0883-5403(92)90051-Q.CrossRefPubMed McKellop H, Sarmiento A, Brien W, Park SH: Interface corrosion of a modular head total hip prosthesis. J Arthroplasty. 1992, 7 (3): 291-294. 10.1016/0883-5403(92)90051-Q.CrossRefPubMed
17.
go back to reference Jacobs JJ, Gilbert JL, Urban RM: Current concepts review - Corrosion of metal orthopaedic implants. J Bone Joint Surg [Am]. 1998, 80-A: 268-282. Jacobs JJ, Gilbert JL, Urban RM: Current concepts review - Corrosion of metal orthopaedic implants. J Bone Joint Surg [Am]. 1998, 80-A: 268-282.
18.
go back to reference Hoenig E, Morlock MM: "Bestimmung der Relativbewegung an der Konusverbindung der modularen Metha-Kurzschaftprothese". Study to determine the relative motion between the neck adapter and the stem Biomechanics Section - Hamburg University of Technology, Aesculap test report No. V1039. 2008 Hoenig E, Morlock MM: "Bestimmung der Relativbewegung an der Konusverbindung der modularen Metha-Kurzschaftprothese". Study to determine the relative motion between the neck adapter and the stem Biomechanics Section - Hamburg University of Technology, Aesculap test report No. V1039. 2008
19.
go back to reference Hoenig E, Morlock MM: Estimation of the relative motion in the taper lock interface of a modular hip prosthesis. Abstract Orthopedic Research Society, Las Vegas, Nevada. 2009 Hoenig E, Morlock MM: Estimation of the relative motion in the taper lock interface of a modular hip prosthesis. Abstract Orthopedic Research Society, Las Vegas, Nevada. 2009
20.
go back to reference Kaesche H: Die Korrosion der Metalle - Physikalisch-chemische Prinzipien und aktuelle Probleme. 1979, Springer-Verlag Berlin Heidelberg New York, 274-276.CrossRef Kaesche H: Die Korrosion der Metalle - Physikalisch-chemische Prinzipien und aktuelle Probleme. 1979, Springer-Verlag Berlin Heidelberg New York, 274-276.CrossRef
21.
go back to reference Gilbert JL, Jacobs JJ: The mechanical and electrochemical processes associated with taper fretting crevice corrosion - a review. Modularity of orthopedic implants, ASTM STP 1301. Edited by: Parr JE, Mayor MB. 1997, J ASTM International, 45-59. full_text. Gilbert JL, Jacobs JJ: The mechanical and electrochemical processes associated with taper fretting crevice corrosion - a review. Modularity of orthopedic implants, ASTM STP 1301. Edited by: Parr JE, Mayor MB. 1997, J ASTM International, 45-59. full_text.
22.
go back to reference Urban RM, Gilbert JL, Jacobs JJ: Corrosion of modular titanium alloy stems in cementless hip replacement. 2005, J ASTM International, 2 (10): 1-10. Urban RM, Gilbert JL, Jacobs JJ: Corrosion of modular titanium alloy stems in cementless hip replacement. 2005, J ASTM International, 2 (10): 1-10.
23.
go back to reference Rodrigues DC, Urban RM, Jacobs JJ, Gilbert JL: In vivo severe corrosion and hydrogen embrittlement of retrieved modular body titanium alloy hip-implants. J Biomed Mater Res Part B Appl Biomater. 2009, 88B (1): 206-219. 10.1002/jbm.b.31171.CrossRef Rodrigues DC, Urban RM, Jacobs JJ, Gilbert JL: In vivo severe corrosion and hydrogen embrittlement of retrieved modular body titanium alloy hip-implants. J Biomed Mater Res Part B Appl Biomater. 2009, 88B (1): 206-219. 10.1002/jbm.b.31171.CrossRef
24.
go back to reference Collier JP, Surprenant VA, Jensen RE, Mayor MB, Surprenant HP: Corrosion between the components of modular femoral hip prostheses. J Bone Joint Surg [Br]. 1992, 74-B: 511-517. Collier JP, Surprenant VA, Jensen RE, Mayor MB, Surprenant HP: Corrosion between the components of modular femoral hip prostheses. J Bone Joint Surg [Br]. 1992, 74-B: 511-517.
25.
go back to reference Cook SD, Barrack RL, Baffes GC, Clemow AJT, Serekian P, Dong N, Kester MA: Wear and corrosion of modular interfaces in total hip replacements. Clin Orth Rel Res. 1994, 80-88. 298 Cook SD, Barrack RL, Baffes GC, Clemow AJT, Serekian P, Dong N, Kester MA: Wear and corrosion of modular interfaces in total hip replacements. Clin Orth Rel Res. 1994, 80-88. 298
26.
go back to reference Goldberg JR, Buckley CA, Jacobs JJ, Gilbert JL: Corrosion testing of modular hip implants. Modularity of orthopedic implants, ASTM STP 1301. Edited by: Parr JE, Mayor MB. 1997, J ASTM International, 157-176. full_text. Goldberg JR, Buckley CA, Jacobs JJ, Gilbert JL: Corrosion testing of modular hip implants. Modularity of orthopedic implants, ASTM STP 1301. Edited by: Parr JE, Mayor MB. 1997, J ASTM International, 157-176. full_text.
27.
go back to reference Goldberg JR, Gilbert JL: In vitro corrosion testing of modular hip tapers. J Biomed Mater Res Part B: Appl Biomater. 2003, 64B: 78-93. 10.1002/jbm.b.10526.CrossRef Goldberg JR, Gilbert JL: In vitro corrosion testing of modular hip tapers. J Biomed Mater Res Part B: Appl Biomater. 2003, 64B: 78-93. 10.1002/jbm.b.10526.CrossRef
28.
go back to reference Duisabeau L, Combrade P, Forest B: Environmental effect on fretting of metallic materials for orthopaedic implants. Wear. 2004, 256: 805-816. 10.1016/S0043-1648(03)00522-2.CrossRef Duisabeau L, Combrade P, Forest B: Environmental effect on fretting of metallic materials for orthopaedic implants. Wear. 2004, 256: 805-816. 10.1016/S0043-1648(03)00522-2.CrossRef
29.
go back to reference Hoeppner DW, Gates FL: Fretting Fatigue Consideration in Engineering Design, Wear. 1981, 70: 155-164. Hoeppner DW, Gates FL: Fretting Fatigue Consideration in Engineering Design, Wear. 1981, 70: 155-164.
30.
go back to reference Viceconti M, Baleani M, Squarzoni S, Toni A: Fretting wear in modular neck hip prosthesis. Journal of Biomedical Materials Research. 1997, 35: 207-216. 10.1002/(SICI)1097-4636(199705)35:2<207::AID-JBM9>3.0.CO;2-R.CrossRefPubMed Viceconti M, Baleani M, Squarzoni S, Toni A: Fretting wear in modular neck hip prosthesis. Journal of Biomedical Materials Research. 1997, 35: 207-216. 10.1002/(SICI)1097-4636(199705)35:2<207::AID-JBM9>3.0.CO;2-R.CrossRefPubMed
31.
go back to reference Viceconti M, Ruggeri O, Toni A, Giunti A: Design-related fretting wear in modular neck hip prosthesis. Journal of Biomedical Materials Research. 1996, 30: 181-186. 10.1002/(SICI)1097-4636(199602)30:2<181::AID-JBM7>3.0.CO;2-N.CrossRefPubMed Viceconti M, Ruggeri O, Toni A, Giunti A: Design-related fretting wear in modular neck hip prosthesis. Journal of Biomedical Materials Research. 1996, 30: 181-186. 10.1002/(SICI)1097-4636(199602)30:2<181::AID-JBM7>3.0.CO;2-N.CrossRefPubMed
32.
go back to reference Hoeppner DW, Chandrasekaran V: Fretting in orthopaedic implants: a review, Wear. 1994, 173: 189-197. Hoeppner DW, Chandrasekaran V: Fretting in orthopaedic implants: a review, Wear. 1994, 173: 189-197.
33.
go back to reference Martinez SA, Sathish S, Mall S, Blodgett MP: Evolution of fretting fatigue damage and relaxation of residual stress in shot-peened Ti6Al4V. Metallurgical and Materials Trans A. 2005, 36A: 3385-3396. 10.1007/s11661-005-0012-8.CrossRef Martinez SA, Sathish S, Mall S, Blodgett MP: Evolution of fretting fatigue damage and relaxation of residual stress in shot-peened Ti6Al4V. Metallurgical and Materials Trans A. 2005, 36A: 3385-3396. 10.1007/s11661-005-0012-8.CrossRef
34.
go back to reference Forschungskuratorium Maschinenbau EV: Reibkorrosion - Vorhaben Nr. 6. Maschinenbau-Verlag GmbH, 6000 Frankfurt/M, Forschungsheft. 1976, 110-113. 56 Forschungskuratorium Maschinenbau EV: Reibkorrosion - Vorhaben Nr. 6. Maschinenbau-Verlag GmbH, 6000 Frankfurt/M, Forschungsheft. 1976, 110-113. 56
Metadata
Title
Modular titanium alloy neck adapter failures in hip replacement - failure mode analysis and influence of implant material
Authors
Thomas M Grupp
Thomas Weik
Wilhelm Bloemer
Hanns-Peter Knaebel
Publication date
01-12-2010
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2010
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/1471-2474-11-3

Other articles of this Issue 1/2010

BMC Musculoskeletal Disorders 1/2010 Go to the issue