Modes of failure of a threaded acetabular cup

A radiographic study with EBRA of 42 revised cups

Authors: Thomas Ilchmann, Severin Neher, Franz Maurer, Kuno Weise

Published in: International Orthopaedics | Issue 2/2007

Login to get access

Abstract

A radiographic study of a singular type of threaded cup, revised due to aseptic loosening, was performed with Einzel-Bild-Roentgen-Analyse (EBRA) to detect eventual risk factors and patterns of loosening. Fifty-three cups of 50 patients were revised. Forty-two cups could be measured with EBRA. No cup had obvious osteolysis, 33 cups demonstrated early migration, and all cups had migrated at the time of revision. Twenty-eight of the cups had a change of inclination and 21 of anteversion, respectively. The mean migration was 1.9 mm in the medial and 7.9 mm in the cranial direction; the mean wear rate was 0.2 mm/year. Cups with early migration had a higher migration rate. All male patients had early migration; medially placed cups had less migration than the other cups. Wear was not significantly affected by the migration of the implant. No cup had a complete radiolucent line, and the only radiographic sign for loosening was the change of position of the cup. Medial placement showed less migration in case of loosening. Regular radiographic follow-up is recommended for the examined implant, and the cup should not be used in the future.
Literature
1.
go back to reference Biedermann R, Tonin A, Krismer M, Rachbauer F, Eibl G, Stockl B (2005) Reducing the risk of dislocation after total hip arthroplasty: the effect of orientation of the acetabular component. J Bone Joint Surg Br 87:762–769PubMedCrossRef Biedermann R, Tonin A, Krismer M, Rachbauer F, Eibl G, Stockl B (2005) Reducing the risk of dislocation after total hip arthroplasty: the effect of orientation of the acetabular component. J Bone Joint Surg Br 87:762–769PubMedCrossRef
2.
go back to reference Bruijn JD, Seelen JL, Feenstra RM, Hansen BE, Bernoski FP (1995) Failure of the Mecring screw-ring acetabular component in total hip arthroplasty. A three- to seven-year follow-up study. J Bone Joint Surg Am 77:760–766PubMed Bruijn JD, Seelen JL, Feenstra RM, Hansen BE, Bernoski FP (1995) Failure of the Mecring screw-ring acetabular component in total hip arthroplasty. A three- to seven-year follow-up study. J Bone Joint Surg Am 77:760–766PubMed
3.
go back to reference Delaunay CP, Kapandji AI (1998) Survivorship of rough surfaced threaded acetabular cups. 382 consecutive primary Zweymuller cups followed for 0.2–12 years. Acta Orthop Scand 69:379–383PubMed Delaunay CP, Kapandji AI (1998) Survivorship of rough surfaced threaded acetabular cups. 382 consecutive primary Zweymuller cups followed for 0.2–12 years. Acta Orthop Scand 69:379–383PubMed
4.
go back to reference Garcia-Cimbrelo E, Munuera L (1992) Early and late loosening of the acetabular cup after low-friction arthroplasty. J Bone Joint Surg Am 74:1119–1129PubMed Garcia-Cimbrelo E, Munuera L (1992) Early and late loosening of the acetabular cup after low-friction arthroplasty. J Bone Joint Surg Am 74:1119–1129PubMed
5.
go back to reference Grubl A, Chiari C, Gruber M, Kaider A, Gottsauner-Wolf F (2002) Cementless total hip arthroplasty with a tapered, rectangular titanium stem and a threaded cup: a minimum ten-year follow-up. J Bone Joint Surg Am 84:425–431PubMed Grubl A, Chiari C, Gruber M, Kaider A, Gottsauner-Wolf F (2002) Cementless total hip arthroplasty with a tapered, rectangular titanium stem and a threaded cup: a minimum ten-year follow-up. J Bone Joint Surg Am 84:425–431PubMed
6.
go back to reference Havelin LI, Vollset SE, Engesaeter LB (1995) Revision for aseptic loosening of uncemented cups in 4,352 primary total hip prostheses. A report from the Norwegian Arthroplasty Register. Acta Orthop Scand 66:494–500PubMed Havelin LI, Vollset SE, Engesaeter LB (1995) Revision for aseptic loosening of uncemented cups in 4,352 primary total hip prostheses. A report from the Norwegian Arthroplasty Register. Acta Orthop Scand 66:494–500PubMed
7.
go back to reference Huiskes R (1987) Finite element analysis of acetabular reconstruction. Non-cemented threaded cups. Acta Orthop Scand 58:620–625PubMed Huiskes R (1987) Finite element analysis of acetabular reconstruction. Non-cemented threaded cups. Acta Orthop Scand 58:620–625PubMed
8.
go back to reference Ilchmann T (1997) Radiographic assessment of cup migration and wear after hip replacement. Acta Orthop Scand Suppl 276:1–26PubMed Ilchmann T (1997) Radiographic assessment of cup migration and wear after hip replacement. Acta Orthop Scand Suppl 276:1–26PubMed
9.
go back to reference Ilchmann T, Markovic L, Joshi A, Hardinge K, Murphy J, Wingstrand H (1998) Migration and wear of long-term successful Charnley total hip replacements. J Bone Joint Surg Br 80:377–381PubMedCrossRef Ilchmann T, Markovic L, Joshi A, Hardinge K, Murphy J, Wingstrand H (1998) Migration and wear of long-term successful Charnley total hip replacements. J Bone Joint Surg Br 80:377–381PubMedCrossRef
10.
go back to reference Johnsson R, Franzen H, Nilsson LT (1994) Combined survivorship and multivariate analyses of revisions in 799 hip prostheses. A 10- to 20-year review of mechanical loosening. J Bone Joint Surg Br 76:439–443 Johnsson R, Franzen H, Nilsson LT (1994) Combined survivorship and multivariate analyses of revisions in 799 hip prostheses. A 10- to 20-year review of mechanical loosening. J Bone Joint Surg Br 76:439–443
11.
go back to reference Kesteris U, Hardinge K, Ilchmann T, Wingstrand H (2003) Polyethylene wear in prosthetic hips with loose components. J Arthroplasty 18:10–15PubMedCrossRef Kesteris U, Hardinge K, Ilchmann T, Wingstrand H (2003) Polyethylene wear in prosthetic hips with loose components. J Arthroplasty 18:10–15PubMedCrossRef
12.
go back to reference Kobayashi S, Takaoka K, Saito N, Hisa K (1997) Factors affecting aseptic failure of fixation after primary Charnley total hip arthroplasty. Multivariate survival analysis. J Bone Joint Surg Am 79:1618–1627PubMed Kobayashi S, Takaoka K, Saito N, Hisa K (1997) Factors affecting aseptic failure of fixation after primary Charnley total hip arthroplasty. Multivariate survival analysis. J Bone Joint Surg Am 79:1618–1627PubMed
13.
go back to reference Kody MH, Kabo JM, Markolf KL, Dorey FJ, Amstutz HC (1990) Strength of initial mechanical fixation of screw ring acetabular components. Clin Orthop:146–153 Kody MH, Kabo JM, Markolf KL, Dorey FJ, Amstutz HC (1990) Strength of initial mechanical fixation of screw ring acetabular components. Clin Orthop:146–153
14.
go back to reference Kordelle J, Starker M (2000) [Migration analysis of cemented Muller polyethylene acetabular cups versus cement-free Zweymuller screw-attached acetabular cups]. Z Orthop Ihre Grenzgeb 138:46–51PubMedCrossRef Kordelle J, Starker M (2000) [Migration analysis of cemented Muller polyethylene acetabular cups versus cement-free Zweymuller screw-attached acetabular cups]. Z Orthop Ihre Grenzgeb 138:46–51PubMedCrossRef
15.
go back to reference Krismer M, Bauer R, Tschupik J, Mayrhofer P (1995) EBRA: a method to measure migration of acetabular components. J Biomech 28:1225–1236PubMedCrossRef Krismer M, Bauer R, Tschupik J, Mayrhofer P (1995) EBRA: a method to measure migration of acetabular components. J Biomech 28:1225–1236PubMedCrossRef
16.
go back to reference Krismer M, Biedermann R, Stockl B, Fischer M, Bauer R, Haid C (1999) The prediction of failure of the stem in THR by measurement of early migration using EBRA-FCA. Einzel-Bild-Roentgen-Analyse-femoral component analysis. J Bone Joint Surg Br 81:273–280PubMedCrossRef Krismer M, Biedermann R, Stockl B, Fischer M, Bauer R, Haid C (1999) The prediction of failure of the stem in THR by measurement of early migration using EBRA-FCA. Einzel-Bild-Roentgen-Analyse-femoral component analysis. J Bone Joint Surg Br 81:273–280PubMedCrossRef
17.
go back to reference Krismer M, Stockl B, Fischer M, Bauer R, Mayrhofer P, Ogon M (1996) Early migration predicts late aseptic failure of hip sockets. J Bone Joint Surg Br 78:422–426PubMed Krismer M, Stockl B, Fischer M, Bauer R, Mayrhofer P, Ogon M (1996) Early migration predicts late aseptic failure of hip sockets. J Bone Joint Surg Br 78:422–426PubMed
18.
go back to reference Malchau H, Herberts P, Eisler T, Garellick G, Soderman P (2002) The Swedish Total Hip Replacement Register. J Bone Joint Surg Am 84 Suppl 2:2–20PubMed Malchau H, Herberts P, Eisler T, Garellick G, Soderman P (2002) The Swedish Total Hip Replacement Register. J Bone Joint Surg Am 84 Suppl 2:2–20PubMed
19.
go back to reference Mjoberg B (1997) The theory of early loosening of hip prostheses. Orthopedics 20:1169–1175PubMed Mjoberg B (1997) The theory of early loosening of hip prostheses. Orthopedics 20:1169–1175PubMed
20.
go back to reference Pupparo F, Engh CA (1991) Comparison of porous-threaded and smooth-threaded acetabular components of identical design. Two- to four-year results. Clin Orthop:201–206 Pupparo F, Engh CA (1991) Comparison of porous-threaded and smooth-threaded acetabular components of identical design. Two- to four-year results. Clin Orthop:201–206
21.
go back to reference Ranawat CS, Dorr LD, Inglis AE (1980) Total hip arthroplasty in protrusio acetabuli of rheumatoid arthritis. J Bone Joint Surg Am 62:1059–1065PubMed Ranawat CS, Dorr LD, Inglis AE (1980) Total hip arthroplasty in protrusio acetabuli of rheumatoid arthritis. J Bone Joint Surg Am 62:1059–1065PubMed
22.
go back to reference Schimmel JW, Huiskes R (1988) Primary fit of the Lord cementless total hip. A geometric study in cadavers. Acta Orthop Scand 59:638–642PubMedCrossRef Schimmel JW, Huiskes R (1988) Primary fit of the Lord cementless total hip. A geometric study in cadavers. Acta Orthop Scand 59:638–642PubMedCrossRef
23.
go back to reference Snorrason F, Karrholm J (1990) Primary migration of fully-threaded acetabular prostheses. A roentgen stereophotogrammetric analysis. J Bone Joint Surg Br 72:647–652PubMed Snorrason F, Karrholm J (1990) Primary migration of fully-threaded acetabular prostheses. A roentgen stereophotogrammetric analysis. J Bone Joint Surg Br 72:647–652PubMed
24.
go back to reference Tallroth K, Slatis P, Ylinen P, Paavolainen P, Paavilainen T (1993) Loosening of threaded acetabular components. Radiographic manifestations. J Arthroplasty 8:581–584PubMedCrossRef Tallroth K, Slatis P, Ylinen P, Paavolainen P, Paavilainen T (1993) Loosening of threaded acetabular components. Radiographic manifestations. J Arthroplasty 8:581–584PubMedCrossRef
25.
go back to reference Volkmann R, Schneider MA, Bretschneider C, Weise K (1999) More failures of uncemented acetabular screw-rings than of cemented polyethylene cups in total hip arthroplasties. Int Orthop 23:138–139PubMedCrossRef Volkmann R, Schneider MA, Bretschneider C, Weise K (1999) More failures of uncemented acetabular screw-rings than of cemented polyethylene cups in total hip arthroplasties. Int Orthop 23:138–139PubMedCrossRef
Metadata
Title
Modes of failure of a threaded acetabular cup
A radiographic study with EBRA of 42 revised cups
Authors
Thomas Ilchmann
Severin Neher
Franz Maurer
Kuno Weise
Publication date
01-04-2007
Publisher
Springer-Verlag
Published in
International Orthopaedics / Issue 2/2007
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-006-0155-2

Other articles of this Issue 2/2007

International Orthopaedics 2/2007 Go to the issue