01-11-2016 | Knee Arthroplasty
Mobile-bearing TKA improved the anteroposterior joint stability in mid-flexion range comparing to fixed-bearing TKA
Authors:
Yukihide Minoda, Mitsuhiko Ikebuchi, Shigekazu Mizokawa, Yoichi Ohta, Hiroaki Nakamura
Published in:
Archives of Orthopaedic and Trauma Surgery
|
Issue 11/2016
Login to get access
Abstract
Introduction
Proper anteroposterior (AP) joint displacement is an important indicator of good clinical outcome following total knee arthroplasty (TKA). We hypothesized that a newly introduced mobile-bearing posterior stabilized (PS) prosthesis reduces the AP joint displacement. The aim of this study is to compare the AP joint displacement between a newly introduced mobile-bearing PS TKA in one knee and a conventional fixed-bearing PS TKA in other knee.
Materials and methods
82 knees in 41 patients who had bilateral TKAs were investigated. All the patients received a conventional fixed-bearing PS prosthesis in one knee and a highly congruent mobile-bearing PS prosthesis in the other knee. AP joint displacement was measured using the KT-2000 arthrometer, at 30° and 75° in flexion, at average of 3.3 years after the operation.
Results
AP joint displacements at 30° in flexion were 6 ± 3 mm in the knees with the mobile-bearing PS prosthesis and 9 ± 4 mm in the knee with fixed-bearing PS prosthesis (p < 0.001). AP joint displacements at 75° in flexion were 4 ± 2 mm in the knees with the mobile-bearing PS prosthesis and 6 ± 3 mm in the knee with fixed-bearing PS prosthesis (p < 0.001).
Conclusions
This study suggested that the design of the prosthesis can improve the AP joint stability in mid-flexion range.