Published in: International Orthopaedics 3/2014

01-03-2014 | Original Paper

Metal-on-metal hip prostheses: Correlation between debris in the synovial fluid and levels of cobalt and chromium ions in the bloodstream

Authors: Dalila De Pasquale, Susanna Stea, Stefano Squarzoni, Barbara Bordini, Marilina Amabile, Simona Catalani, Pietro Apostoli, Aldo Toni

Published in: International Orthopaedics | Issue 3/2014

Login to get access

Abstract

Purpose

Hip prostheses with metal-on-metal (MoM) coupling can release cobalt-chromium particles and ions. The aim of this work is to verify the correlation between particles in the synovial fluid and circulating ions.

Methods

Forty patients were enrolled; particles from synovial fluid were analysed by SEM–EDX (Scanning Electron Microscopy-Energy Dispersion X-rays analysis) and levels of circulating Co and Cr were assayed by ICP-MS (inductively-coupled plasma mass spectrometry).

Results

In 16 cases we did not find any particles in the synovial fluid and the Co level in whole blood was 0.05–4.42 ppb; in seven with few particles the blood level was 2.2–15.6 ppb; in six cases with several particles the level was 5.0–54.3 ppb; finally, in 11 cases we isolated not only Co-Cr particles, but also Cr particles with low or absent Co and in these patients the circulating level of Co was 23.8–109.6 ppb. Co in serum and Cr level both whole blood and serum have shown a similar trend to Co; the correlation between all these values and the corresponding particles is statistically significant in all cases.

Conclusion

Co and Cr both in serum and whole blood represents a systemic representation of the particle release at local level and can therefore be used to confirm a diagnosis and monitor the wear process of MoM articular prostheses.
Literature
6.
7.
8.
go back to reference De Smet K, De Haan R, Calistri A, Campbell PA, Ebramzadeh E, Pattyn C, Gill HS (2008) Metal ion measurement as a diagnostic tool to identify problems with metal-on-metal hip resurfacing. J Bone Joint Surg Am 90(Suppl 4):202–208. doi:10.2106/JBJS.H.00672 PubMedCrossRef De Smet K, De Haan R, Calistri A, Campbell PA, Ebramzadeh E, Pattyn C, Gill HS (2008) Metal ion measurement as a diagnostic tool to identify problems with metal-on-metal hip resurfacing. J Bone Joint Surg Am 90(Suppl 4):202–208. doi:10.​2106/​JBJS.​H.​00672 PubMedCrossRef
9.
go back to reference Natu S, Sidaginamale RP, Gandhi J, Langton DJ, Nargol AV (2012) Adverse reactions to metal debris: histopathological features of periprosthetic soft tissue reactions seen in association with failed metal on metal hip arthroplasties. J Clin Pathol 65(5):409–418. doi:10.1136/jclinpath-2011-200398 PubMedCrossRef Natu S, Sidaginamale RP, Gandhi J, Langton DJ, Nargol AV (2012) Adverse reactions to metal debris: histopathological features of periprosthetic soft tissue reactions seen in association with failed metal on metal hip arthroplasties. J Clin Pathol 65(5):409–418. doi:10.​1136/​jclinpath-2011-200398 PubMedCrossRef
10.
go back to reference Anissian L, Stark A, Dahlstrand H, Granberg B, Good V, Bucht E (2002) Cobalt ions influence proliferation and function of human osteoblast-like cells. Acta Orthop Scand 73(3):369–374PubMedCrossRef Anissian L, Stark A, Dahlstrand H, Granberg B, Good V, Bucht E (2002) Cobalt ions influence proliferation and function of human osteoblast-like cells. Acta Orthop Scand 73(3):369–374PubMedCrossRef
11.
go back to reference Back DL, Young DA, Shimmin AJ (2005) How do serum cobalt and chromium levels change after metal-on-metal hip resurfacing? Clin Orthop Relat Res 438:177–181PubMedCrossRef Back DL, Young DA, Shimmin AJ (2005) How do serum cobalt and chromium levels change after metal-on-metal hip resurfacing? Clin Orthop Relat Res 438:177–181PubMedCrossRef
12.
go back to reference Apostoli P, Catalani S, Zaghini A, Mariotti A, Poliani PL, Vielmi V, Semeraro F, Duse S, Porzionato A, Macchi V, Padovani A, Rizzetti MC, De Caro R (2012) High doses of cobalt induce optic and auditory neuropathy. Exp Toxicol Pathol 65(6):719–727. doi:10.1016/j.etp.2012.09.006 PubMed Apostoli P, Catalani S, Zaghini A, Mariotti A, Poliani PL, Vielmi V, Semeraro F, Duse S, Porzionato A, Macchi V, Padovani A, Rizzetti MC, De Caro R (2012) High doses of cobalt induce optic and auditory neuropathy. Exp Toxicol Pathol 65(6):719–727. doi:10.​1016/​j.​etp.​2012.​09.​006 PubMed
14.
go back to reference Gunther KP, Schmitt J, Campbell P, Delaunay CP, Drexler H, Ettema HB, Garcia-Cimbrelo E, Hannemann F, Hartmann A, Huberti H, Knahr K, Kunze J, Langton DJ, Lauer W, Learmonth I, Lohmann CH, Lutzner J, Morlock M, Seidler A, Wimmer MA, Zagra L (2012) Consensus statement "Current evidence on the management of metal-on-metal bearings" - April 16. Hip Int 23(1):2–5. doi:10.5301/HIP.2013.10754 CrossRef Gunther KP, Schmitt J, Campbell P, Delaunay CP, Drexler H, Ettema HB, Garcia-Cimbrelo E, Hannemann F, Hartmann A, Huberti H, Knahr K, Kunze J, Langton DJ, Lauer W, Learmonth I, Lohmann CH, Lutzner J, Morlock M, Seidler A, Wimmer MA, Zagra L (2012) Consensus statement "Current evidence on the management of metal-on-metal bearings" - April 16. Hip Int 23(1):2–5. doi:10.​5301/​HIP.​2013.​10754 CrossRef
17.
go back to reference Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV (2010) Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br 92(1):38–46. doi:10.1302/0301-620X.92B1.22770 PubMedCrossRef Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV (2010) Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br 92(1):38–46. doi:10.​1302/​0301-620X.​92B1.​22770 PubMedCrossRef
18.
go back to reference Beraudi A, Montesi M, Stea S, Sudanese A, Apostoli P, Toni A (2013) Detection of cobalt in synovial fluid from metal-on-metal hip prosthesis: correlation with the ionic haematic level. Biomarkers, Early Online 1–7. doi:10.3109/1354750X.2013.846413 Beraudi A, Montesi M, Stea S, Sudanese A, Apostoli P, Toni A (2013) Detection of cobalt in synovial fluid from metal-on-metal hip prosthesis: correlation with the ionic haematic level. Biomarkers, Early Online 1–7. doi:10.​3109/​1354750X.​2013.​846413
19.
go back to reference Toni A, Traina F, Stea S, Sudanese A, Visentin M, Bordini B, Squarzoni S (2006) Early diagnosis of ceramic liner fracture. Guidelines based on a twelve-year clinical experience. J Bone Joint Surg Am 88(Suppl 4):55–63PubMedCrossRef Toni A, Traina F, Stea S, Sudanese A, Visentin M, Bordini B, Squarzoni S (2006) Early diagnosis of ceramic liner fracture. Guidelines based on a twelve-year clinical experience. J Bone Joint Surg Am 88(Suppl 4):55–63PubMedCrossRef
20.
go back to reference Visentin M, Stea S, Squarzoni S, Antonietti B, Reggiani M, Toni A (2004) A new method for isolation of polyethylene wear debris from tissue and synovial fluid. Biomaterials 25(24):5531–5537PubMedCrossRef Visentin M, Stea S, Squarzoni S, Antonietti B, Reggiani M, Toni A (2004) A new method for isolation of polyethylene wear debris from tissue and synovial fluid. Biomaterials 25(24):5531–5537PubMedCrossRef
21.
go back to reference Stea S, Traina F, Beraudi A, Montesi M, Bordini B, Squarzoni S, Sudanese A, Toni A (2012) Synovial fluid microanalysis allows early diagnosis of ceramic hip prosthesis damage. J Orthop Res 30(8):1312–1320. doi:10.1002/jor.22077 PubMedCrossRef Stea S, Traina F, Beraudi A, Montesi M, Bordini B, Squarzoni S, Sudanese A, Toni A (2012) Synovial fluid microanalysis allows early diagnosis of ceramic hip prosthesis damage. J Orthop Res 30(8):1312–1320. doi:10.​1002/​jor.​22077 PubMedCrossRef
22.
23.
go back to reference Singh G, Meyer H, Ruetschi M, Chamaon K, Feuerstein B, Lohmann CH (2013) Large-diameter metal-on-metal total hip arthroplasties: a page in orthopedic history? J Biomed Mater Res A 101(11):3320–3326. doi:10.1002/jbm.a.34619 Singh G, Meyer H, Ruetschi M, Chamaon K, Feuerstein B, Lohmann CH (2013) Large-diameter metal-on-metal total hip arthroplasties: a page in orthopedic history? J Biomed Mater Res A 101(11):3320–3326. doi:10.​1002/​jbm.​a.​34619
24.
go back to reference Billi F, Benya P, Kavanaugh A, Adams J, McKellop H, Ebramzadeh E (2012) The John Charnley Award: an accurate and extremely sensitive method to separate, display, and characterize wear debris: part 2: metal and ceramic particles. Clin Orthop Relat Res 470(2):339–350. doi:10.1007/s11999-011-2058-9 PubMedCentralPubMedCrossRef Billi F, Benya P, Kavanaugh A, Adams J, McKellop H, Ebramzadeh E (2012) The John Charnley Award: an accurate and extremely sensitive method to separate, display, and characterize wear debris: part 2: metal and ceramic particles. Clin Orthop Relat Res 470(2):339–350. doi:10.​1007/​s11999-011-2058-9 PubMedCentralPubMedCrossRef
25.
go back to reference Doorn PF, Campbell PA, Worrall J, Benya PD, McKellop HA, Amstutz HC (1998) Metal wear particle characterization from metal on metal total hip replacements: transmission electron microscopy study of periprosthetic tissues and isolated particles. J Biomed Mater Res 42(1):103–111PubMedCrossRef Doorn PF, Campbell PA, Worrall J, Benya PD, McKellop HA, Amstutz HC (1998) Metal wear particle characterization from metal on metal total hip replacements: transmission electron microscopy study of periprosthetic tissues and isolated particles. J Biomed Mater Res 42(1):103–111PubMedCrossRef
26.
go back to reference Milosev I, Remskar M (2009) In vivo production of nanosized metal wear debris formed by tribochemical reaction as confirmed by high-resolution TEM and XPS analyses. J Biomed Mater Res A 91(4):1100–1110. doi:10.1002/jbm.a.32301 PubMedCrossRef Milosev I, Remskar M (2009) In vivo production of nanosized metal wear debris formed by tribochemical reaction as confirmed by high-resolution TEM and XPS analyses. J Biomed Mater Res A 91(4):1100–1110. doi:10.​1002/​jbm.​a.​32301 PubMedCrossRef
27.
go back to reference Bolland BJ, Culliford DJ, Langton DJ, Millington JP, Arden NK, Latham JM (2011) High failure rates with a large-diameter hybrid metal-on-metal total hip replacement: clinical, radiological and retrieval analysis. J Bone Joint Surg Br 93(5):608–615. doi:10.1302/0301-620X.93B5.26309 PubMedCrossRef Bolland BJ, Culliford DJ, Langton DJ, Millington JP, Arden NK, Latham JM (2011) High failure rates with a large-diameter hybrid metal-on-metal total hip replacement: clinical, radiological and retrieval analysis. J Bone Joint Surg Br 93(5):608–615. doi:10.​1302/​0301-620X.​93B5.​26309 PubMedCrossRef
28.
go back to reference Malviya A, Ramaskandhan JR, Bowman R, Hashmi M, Holland JP, Kometa S, Lingard E (2011) What advantage is there to be gained using large modular metal-on-metal bearings in routine primary hip replacement? A preliminary report of a prospective randomised controlled trial. J Bone Joint Surg Br 93(12):1602–1609. doi:10.1302/0301-620X.93B12.27533 PubMedCrossRef Malviya A, Ramaskandhan JR, Bowman R, Hashmi M, Holland JP, Kometa S, Lingard E (2011) What advantage is there to be gained using large modular metal-on-metal bearings in routine primary hip replacement? A preliminary report of a prospective randomised controlled trial. J Bone Joint Surg Br 93(12):1602–1609. doi:10.​1302/​0301-620X.​93B12.​27533 PubMedCrossRef
30.
go back to reference Maurer-Ertl W, Friesenbichler J, Sadoghi P, Pechmann M, Trennheuser M, Leithner A (2012) Metal ion levels in large-diameter total hip and resurfacing hip arthroplasty--preliminary results of a prospective five year study after two years of follow-up. BMC Musculoskelet Disord 13:56. doi:10.1186/1471-2474-13-56 PubMedCentralPubMedCrossRef Maurer-Ertl W, Friesenbichler J, Sadoghi P, Pechmann M, Trennheuser M, Leithner A (2012) Metal ion levels in large-diameter total hip and resurfacing hip arthroplasty--preliminary results of a prospective five year study after two years of follow-up. BMC Musculoskelet Disord 13:56. doi:10.​1186/​1471-2474-13-56 PubMedCentralPubMedCrossRef
31.
go back to reference Daniel J, Ziaee H, Salama A, Pradhan C, McMinn DJ (2006) The effect of the diameter of metal-on-metal bearings on systemic exposure to cobalt and chromium. J Bone Joint Surg Br 88(4):443–448PubMedCrossRef Daniel J, Ziaee H, Salama A, Pradhan C, McMinn DJ (2006) The effect of the diameter of metal-on-metal bearings on systemic exposure to cobalt and chromium. J Bone Joint Surg Br 88(4):443–448PubMedCrossRef
32.
Metadata
Title
Metal-on-metal hip prostheses: Correlation between debris in the synovial fluid and levels of cobalt and chromium ions in the bloodstream
Authors
Dalila De Pasquale
Susanna Stea
Stefano Squarzoni
Barbara Bordini
Marilina Amabile
Simona Catalani
Pietro Apostoli
Aldo Toni
Publication date
01-03-2014
Publisher
Springer Berlin Heidelberg
Published in
International Orthopaedics / Issue 3/2014
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-013-2137-5

Other articles of this Issue 3/2014

International Orthopaedics 3/2014 Go to the issue