Skip to main content
Top
Published in: Oral Radiology 1/2021

01-01-2021 | Original Article

Measuring the dose–width product and proposing the local diagnostic reference level in panoramic dental radiography: a multi-center study from Iran

Authors: Akbar Aliasgharzadeh, Davood Khezerloo, Bagher Farhood, Mehran Mohseni, Habiballah Moradi, Tohid Mortezazadeh

Published in: Oral Radiology | Issue 1/2021

Login to get access

Abstract

Objective

Although radiation exposure associated with dental radiography is relatively low, patient exposure must be kept practically low. Therefore, it is necessary for each country to establish its own diagnostic reference levels (DRLs) suitable for its equipment and practice. In the present study, dose-width product (DWP) values for panoramic dental radiography were measured and a local DRL was established.

Methods

Five panoramic devices from five radiology clinics of Kashan, Iran were selected to measure the DWP values of panoramic dental radiography. To investigate the DWP values, the parameters of each patient’s exposure (e.g., tube voltage, tube current, and exposure time) at these five radiology clinics were extracted. Then, the dose value received by each patient was measured based on a CT pencil chamber. Finally, the overall median DWP values for the patients with small, medium, and large sizes were obtained, and these values were considered as the local DRLs for panoramic dental radiography.

Results

A total of 99 adult patients were included in the present study. The findings demonstrated that the median and third-quartile DWP values for these five radiology clinics ranged from 42.3 to 94.3 and 49.7 to 142.8 mGy mm, respectively. The local DRL values, which were established as the overall median DWP values, were 43.4, 52.0, and 80.3 mGy  mm for the adults with small, medium, and large sizes, respectively.

Conclusion

The local DRL proposed in this study for the adult with standard/medium size was lower than those proposed by other reports and seemed acceptable for panoramic radiography in Kashan, Iran.
Literature
1.
go back to reference Granlund C, Thilander-Klang A, Ylhan B, Lofthag-Hansen S, Ekestubbe A. Absorbed organ and effective doses from digital intra-oral and panoramic radiography applying the ICRP 103 recommendations for effective dose estimations. Br J Radiol. 2016;89:20151052.CrossRef Granlund C, Thilander-Klang A, Ylhan B, Lofthag-Hansen S, Ekestubbe A. Absorbed organ and effective doses from digital intra-oral and panoramic radiography applying the ICRP 103 recommendations for effective dose estimations. Br J Radiol. 2016;89:20151052.CrossRef
2.
go back to reference Granlund CM, Lith A, Molander B, Grondahl K, Hansen K, Ekestubbe A. Frequency of errors and pathology in panoramic images of young orthodontic patients. Eur J Orthod. 2012;34:452–7.CrossRef Granlund CM, Lith A, Molander B, Grondahl K, Hansen K, Ekestubbe A. Frequency of errors and pathology in panoramic images of young orthodontic patients. Eur J Orthod. 2012;34:452–7.CrossRef
3.
go back to reference Maloney PL, Lincoln RE, Coyne CP. A protocol for the management of compound mandibular fractures based on the time from injury to treatment. J Oral Maxillofac Surg. 2001;59:879–40 (discussion 452-7).CrossRef Maloney PL, Lincoln RE, Coyne CP. A protocol for the management of compound mandibular fractures based on the time from injury to treatment. J Oral Maxillofac Surg. 2001;59:879–40 (discussion 452-7).CrossRef
4.
go back to reference Okano T, Sur J. Radiation dose and protection in dentistry. Jpn Dent Sci Rev. 2010;46:112–21.CrossRef Okano T, Sur J. Radiation dose and protection in dentistry. Jpn Dent Sci Rev. 2010;46:112–21.CrossRef
5.
go back to reference Lee JS, Kim YH, Yoon SJ, Kang BC. Reference dose levels for dental panoramic radiography in Gwangju. South Korea Radiat Prot Dosim. 2010;142:184–90.CrossRef Lee JS, Kim YH, Yoon SJ, Kang BC. Reference dose levels for dental panoramic radiography in Gwangju. South Korea Radiat Prot Dosim. 2010;142:184–90.CrossRef
6.
go back to reference Izawa M, Harata Y, Shiba N, Koizumi N, Ozawa T, Takahashi N, et al. Establishment of local diagnostic reference levels for quality control in intraoral radiography. Oral Radiol. 2017;33:38–44.CrossRef Izawa M, Harata Y, Shiba N, Koizumi N, Ozawa T, Takahashi N, et al. Establishment of local diagnostic reference levels for quality control in intraoral radiography. Oral Radiol. 2017;33:38–44.CrossRef
7.
go back to reference Protection BNR. Guidance notes for dental practitioners on the safe use of X-ray equipment. Documents of the NRPB. Chilton: NRPB; 2001. Protection BNR. Guidance notes for dental practitioners on the safe use of X-ray equipment. Documents of the NRPB. Chilton: NRPB; 2001.
8.
go back to reference International Commission on Radiological Protection. The 2007 recommendations of the International Commission on Radiological Protection ICRP Publication 103. Ann ICRP. 2007;37:2–4. International Commission on Radiological Protection. The 2007 recommendations of the International Commission on Radiological Protection ICRP Publication 103. Ann ICRP. 2007;37:2–4.
9.
go back to reference Isoardi P, Ropolo R. Measurement of dose–width product in panoramic dental radiology. Br J Radiol. 2003;76:129–31.CrossRef Isoardi P, Ropolo R. Measurement of dose–width product in panoramic dental radiology. Br J Radiol. 2003;76:129–31.CrossRef
10.
go back to reference International Commission on Radiological Protection. ICRP 1990 recommendations of the ICRP Ann ICRP Publication 60. Amsterdam: Elsevier; 1990. International Commission on Radiological Protection. ICRP 1990 recommendations of the ICRP Ann ICRP Publication 60. Amsterdam: Elsevier; 1990.
11.
go back to reference European Commission. Guidance on diagnostic reference levels (DRLs) for medical exposures. Radiation Protection 109. Luxembourg. 1999. European Commission. Guidance on diagnostic reference levels (DRLs) for medical exposures. Radiation Protection 109. Luxembourg. 1999.
12.
go back to reference Vañó E, Miller DL, Martin CJ, Rehani MM, Kang K, Rosenstein M, Ortiz-López P, Mattsson S, Padovani R, Rogers A. ICRP PUBLICATION 135: diagnostic reference levels in medical imaging. Ann ICRP. 2017;46:1–44.CrossRef Vañó E, Miller DL, Martin CJ, Rehani MM, Kang K, Rosenstein M, Ortiz-López P, Mattsson S, Padovani R, Rogers A. ICRP PUBLICATION 135: diagnostic reference levels in medical imaging. Ann ICRP. 2017;46:1–44.CrossRef
13.
go back to reference Napier I. Radiology: Reference doses for dental radiography. Br Dent J. 1999;186:392–6.CrossRef Napier I. Radiology: Reference doses for dental radiography. Br Dent J. 1999;186:392–6.CrossRef
14.
go back to reference Helmrot E, Alm CG. Measurement of radiation dose in dental radiology. Radiat Prot Dosim. 2005;114:168–71.CrossRef Helmrot E, Alm CG. Measurement of radiation dose in dental radiology. Radiat Prot Dosim. 2005;114:168–71.CrossRef
15.
go back to reference Poppe B, Looe H, Pfaffenberger A, Chofor N, Eenboom F, Sering M, et al. Dose-area product measurements in panoramic dental radiology. Radiat Prot Dosim. 2006;123:131–4.CrossRef Poppe B, Looe H, Pfaffenberger A, Chofor N, Eenboom F, Sering M, et al. Dose-area product measurements in panoramic dental radiology. Radiat Prot Dosim. 2006;123:131–4.CrossRef
16.
go back to reference Poppe B, Looe H, Pfaffenberger A, Eenboom F, Chofor N, Sering M, et al. Radiation exposure and dose evaluation in intraoral dental radiology. Radiat Prot Dosimetry. 2006;123:262–7.CrossRef Poppe B, Looe H, Pfaffenberger A, Eenboom F, Chofor N, Sering M, et al. Radiation exposure and dose evaluation in intraoral dental radiology. Radiat Prot Dosimetry. 2006;123:262–7.CrossRef
17.
go back to reference Doyle P, Martin C, Robertson J. Techniques for measurement of dose width product in panoramic dental radiography. Br J Radiol. 2006;79:142–7.CrossRef Doyle P, Martin C, Robertson J. Techniques for measurement of dose width product in panoramic dental radiography. Br J Radiol. 2006;79:142–7.CrossRef
18.
go back to reference Perisinakis K, Damilakis J, Neratzoulakis J, Gourtsoyiannis N. Determination of dose-area product from panoramic radiography using a pencil ionization chamber: Normalized data for the estimation of patient effective and organ doses. Med Phys. 2004;31:708–14.CrossRef Perisinakis K, Damilakis J, Neratzoulakis J, Gourtsoyiannis N. Determination of dose-area product from panoramic radiography using a pencil ionization chamber: Normalized data for the estimation of patient effective and organ doses. Med Phys. 2004;31:708–14.CrossRef
19.
go back to reference Williams J, Montgomery A. Measurement of dose in panoramic dental radiology. Br J Radiol. 2000;73:1002–6.CrossRef Williams J, Montgomery A. Measurement of dose in panoramic dental radiology. Br J Radiol. 2000;73:1002–6.CrossRef
20.
go back to reference Lubis LE, Bayuadi I, Bayhaqi YA, Ardiansyah F, Setiadi AR, Sugandi RD, et al. Radiation dose from dental radiography in indonesia: a five-year survey. Radiat Prot Dosimetry. 2019;183:342–7.CrossRef Lubis LE, Bayuadi I, Bayhaqi YA, Ardiansyah F, Setiadi AR, Sugandi RD, et al. Radiation dose from dental radiography in indonesia: a five-year survey. Radiat Prot Dosimetry. 2019;183:342–7.CrossRef
21.
go back to reference Niemann I, Van Zyl T, Van Staden A, Stofile C, Acho S. Assessment of dose-width products of pre-programmed exposure technique parameters in panoramic dental radiology: a comparison of methodologies. S Afr Dent J. 2015;70:242–5. Niemann I, Van Zyl T, Van Staden A, Stofile C, Acho S. Assessment of dose-width products of pre-programmed exposure technique parameters in panoramic dental radiology: a comparison of methodologies. S Afr Dent J. 2015;70:242–5.
22.
go back to reference National Radiological Protection Board. Guidelines on patient dose to promote the optimisation of protection for diagnostic medical exposures. Documents of the NRPB 1999;4(1). National Radiological Protection Board. Guidelines on patient dose to promote the optimisation of protection for diagnostic medical exposures. Documents of the NRPB 1999;4(1).
23.
go back to reference Kim YH, Lee JS, Yoon SJ, Kang BC. Reference dose levels for dental panoramic radiography in Anyang City. Korean J Oral Maxillofac Radiol. 2009;39:199–203. Kim YH, Lee JS, Yoon SJ, Kang BC. Reference dose levels for dental panoramic radiography in Anyang City. Korean J Oral Maxillofac Radiol. 2009;39:199–203.
24.
go back to reference Walker C, van der Putten W. Patient dosimetry and a novel approach to establishing Diagnostic Reference Levels in dental radiology. Phys Med. 2012;28:7–12.CrossRef Walker C, van der Putten W. Patient dosimetry and a novel approach to establishing Diagnostic Reference Levels in dental radiology. Phys Med. 2012;28:7–12.CrossRef
25.
go back to reference Sans Merce M, Damet J, Becker M. Comparative organ dose levels for dentomaxillofacial examinations performed with computed tomography, cone beam CT and panoramic radiographs. Radioprotection. 2018;53:287–91.CrossRef Sans Merce M, Damet J, Becker M. Comparative organ dose levels for dentomaxillofacial examinations performed with computed tomography, cone beam CT and panoramic radiographs. Radioprotection. 2018;53:287–91.CrossRef
26.
go back to reference Brenner DJ, Doll R, Goodhead DT, Hall EJ, Land CE, Little JB, et al. Cancer risks attributable to low doses of ionizing radiation: assessing what we really know. Proc Natl Acad Sci. 2003;100:13761–6.CrossRef Brenner DJ, Doll R, Goodhead DT, Hall EJ, Land CE, Little JB, et al. Cancer risks attributable to low doses of ionizing radiation: assessing what we really know. Proc Natl Acad Sci. 2003;100:13761–6.CrossRef
27.
go back to reference Bekas M, Pachocki K. The dose received by patients during dental X-ray examination and the technical condition of radiological equipment. Med Pr. 2013;64:755–9.PubMed Bekas M, Pachocki K. The dose received by patients during dental X-ray examination and the technical condition of radiological equipment. Med Pr. 2013;64:755–9.PubMed
28.
go back to reference Mihaela H, Maria M, Benjamin S, Ruben P, Caroline OA, Oana A, et al. Irradiation provided by dental radiological procedures in a pediatric population. Eur J Radiol. 2018;103:112–7.CrossRef Mihaela H, Maria M, Benjamin S, Ruben P, Caroline OA, Oana A, et al. Irradiation provided by dental radiological procedures in a pediatric population. Eur J Radiol. 2018;103:112–7.CrossRef
29.
go back to reference Wochos JF, Detorie NA, Cameron JR. Patients exposure from diagnostic X-rays: an analysis of 1972–1975 NEXT data. Health phys. 1979;36:127–34.CrossRef Wochos JF, Detorie NA, Cameron JR. Patients exposure from diagnostic X-rays: an analysis of 1972–1975 NEXT data. Health phys. 1979;36:127–34.CrossRef
30.
go back to reference Hart D, Hillier M, Wall B. National reference doses for common radiographic, fluoroscopic and dental X-ray examinations in the UK. Br J Radiol. 2009;82:1–12.CrossRef Hart D, Hillier M, Wall B. National reference doses for common radiographic, fluoroscopic and dental X-ray examinations in the UK. Br J Radiol. 2009;82:1–12.CrossRef
31.
go back to reference Shrimpton P, Wall B. An evaluation of the Diamentor transmission ionisation chamber in indicating exposure-area product (R cm2) during diagnostic radiological examinations. Phys Med Biol. 1982;27:871–8.CrossRef Shrimpton P, Wall B. An evaluation of the Diamentor transmission ionisation chamber in indicating exposure-area product (R cm2) during diagnostic radiological examinations. Phys Med Biol. 1982;27:871–8.CrossRef
32.
go back to reference Lee JS, Kang BC, Yoon SJ. The survey of the surface doses of the dental x-ray machines. Korean J Oral Maxillofac Radiol. 2005;35:87–90. Lee JS, Kang BC, Yoon SJ. The survey of the surface doses of the dental x-ray machines. Korean J Oral Maxillofac Radiol. 2005;35:87–90.
33.
go back to reference ICRP. Diagnostic reference levels in medical imaging: review and additional advice. Ann ICRP. 2001;31:33–52. ICRP. Diagnostic reference levels in medical imaging: review and additional advice. Ann ICRP. 2001;31:33–52.
Metadata
Title
Measuring the dose–width product and proposing the local diagnostic reference level in panoramic dental radiography: a multi-center study from Iran
Authors
Akbar Aliasgharzadeh
Davood Khezerloo
Bagher Farhood
Mehran Mohseni
Habiballah Moradi
Tohid Mortezazadeh
Publication date
01-01-2021
Publisher
Springer Singapore
Published in
Oral Radiology / Issue 1/2021
Print ISSN: 0911-6028
Electronic ISSN: 1613-9674
DOI
https://doi.org/10.1007/s11282-020-00428-z