Measurement error of mean sac diameter and crown-rump length among pregnant women at Mulago hospital, Uganda

Authors: Sam Ali, Rosemary Kusaba Byanyima, Sam Ononge, Jerry Ictho, Jean Nyamwiza, Emmanuel Lako Ernesto Loro, John Mukisa, Angella Musewa, Annet Nalutaaya, Ronald Ssenyonga, Ismael Kawooya, Benjamin Temper, Achilles Katamba, Joan Kalyango, Charles Karamagi

Published in: BMC Pregnancy and Childbirth | Issue 1/2018

Login to get access

Abstract

Background

Ultrasonography is essential in the prenatal diagnosis and care for the pregnant mothers. However, the measurements obtained often contain a small percentage of unavoidable error that may have serious clinical implications if substantial. We therefore evaluated the level of intra and inter-observer error in measuring mean sac diameter (MSD) and crown-rump length (CRL) in women between 6 and 10 weeks’ gestation at Mulago hospital.

Methods

This was a cross-sectional study conducted from January to March 2016. We enrolled 56 women with an intrauterine single viable embryo. The women were scanned using a transvaginal (TVS) technique by two observers who were blinded of each other’s measurements. Each observer measured the CRL twice and the MSD once for each woman. Intra-class correlation coefficients (ICCs), 95% limits of agreement (LOA) and technical error of measurement (TEM) were used for analysis.

Results

Intra-observer ICCs for CRL measurements were 0.995 and 0.993 while inter-observer ICCs were 0.988 for CRL and 0.955 for MSD measurements. Intra-observer 95% LOA for CRL were ± 2.04 mm and ± 1.66 mm. Inter-observer LOA were ± 2.35 mm for CRL and ± 4.87 mm for MSD. The intra-observer relative TEM for CRL were 4.62% and 3.70% whereas inter-observer relative TEM were 5.88% and 5.93% for CRL and MSD respectively.

Conclusions

Intra- and inter-observer error of CRL and MSD measurements among pregnant women at Mulago hospital were acceptable. This implies that at Mulago hospital, the error in pregnancy dating is within acceptable margins of ±3 days in first trimester, and the CRL and MSD cut offs of ≥7 mm and ≥ 25 mm respectively are fit for diagnosis of miscarriage on TVS. These findings should be extrapolated to the whole country with caution. Sonographers can achieve acceptable and comparable diagnostic accuracy levels of MSD and CLR measurements with proper training and adherence to practice guidelines.
Literature
1.
go back to reference McNay MB, Fleming JE: Forty years of obstetric ultrasound 1957–1997: from A-scope to three dimensions. Ultrasound Med Bio 1999, 25(1):3–56. McNay MB, Fleming JE: Forty years of obstetric ultrasound 1957–1997: from A-scope to three dimensions. Ultrasound Med Bio 1999, 25(1):3–56.
2.
go back to reference Alan B, Goya C, Tunc S, Teke M, Hattapoglu S. Assessment of placental stiffness using acoustic radiation force impulse Elastography in pregnant women with fetal anomalies. Korean J Radiol. 2016;17(2):218–23.CrossRefPubMedPubMedCentral Alan B, Goya C, Tunc S, Teke M, Hattapoglu S. Assessment of placental stiffness using acoustic radiation force impulse Elastography in pregnant women with fetal anomalies. Korean J Radiol. 2016;17(2):218–23.CrossRefPubMedPubMedCentral
3.
go back to reference Padula F, Laganà A, Vitale S, Mangiafico L, D’Emidio L, Cignini P, Giorlandino M, Gulino F, Capriglione S, Giorlandino C. Ultrasonographic evaluation of placental cord insertion at different gestational ages in low-risk singleton pregnancies: a predictive algorithm. Facts Views Vis ObGyn. 2016;8(1):3.PubMedPubMedCentral Padula F, Laganà A, Vitale S, Mangiafico L, D’Emidio L, Cignini P, Giorlandino M, Gulino F, Capriglione S, Giorlandino C. Ultrasonographic evaluation of placental cord insertion at different gestational ages in low-risk singleton pregnancies: a predictive algorithm. Facts Views Vis ObGyn. 2016;8(1):3.PubMedPubMedCentral
4.
go back to reference Andrietti S, Carlucci S, Wright A, Wright D, Nicolaides KH. Repeat measurements of uterine artery pulsatility index, mean arterial pressure and serum placental growth factor at 12, 22 and 32 weeks in prediction of pre-eclampsia. Ultrasound Obstet Gynecol. 2017;50(2):221–7. Andrietti S, Carlucci S, Wright A, Wright D, Nicolaides KH. Repeat measurements of uterine artery pulsatility index, mean arterial pressure and serum placental growth factor at 12, 22 and 32 weeks in prediction of pre-eclampsia. Ultrasound Obstet Gynecol. 2017;50(2):221–7.
5.
go back to reference WHO. WHO manual of diagnostic ultrasound, vol. 2. 2nd ed. Geneva: Switzerland World Health Organization: World Health Organization & World Federation for Ultrasound in Medicine and Biology; 2013. WHO. WHO manual of diagnostic ultrasound, vol. 2. 2nd ed. Geneva: Switzerland World Health Organization: World Health Organization & World Federation for Ultrasound in Medicine and Biology; 2013.
6.
go back to reference Rumack CM: Diagnostic ultrasound, vol. Vol. 1: Elsevier/Mosby; 2011. Rumack CM: Diagnostic ultrasound, vol. Vol. 1: Elsevier/Mosby; 2011.
7.
go back to reference Chudleigh Trish, Thilaganathan B: obstetric ultrasound: how, why and when, third edn: Churchill Livingstone; 2004. Chudleigh Trish, Thilaganathan B: obstetric ultrasound: how, why and when, third edn: Churchill Livingstone; 2004.
8.
go back to reference Padula F, Capriglione S, Magliarditi M, De Sole R, Nuara R, Santonocito VC, Teodoro MC, Giorlandino C. Goal-directed junior ultrasound training in quantitative measurement of crown-rump length and fetal nuchal translucency: evaluation of a specific training program in a specialized center for prenatal diagnosis. Eur J Obstet Gynecol Reprod Biol. 2015;186:112–3.CrossRefPubMed Padula F, Capriglione S, Magliarditi M, De Sole R, Nuara R, Santonocito VC, Teodoro MC, Giorlandino C. Goal-directed junior ultrasound training in quantitative measurement of crown-rump length and fetal nuchal translucency: evaluation of a specific training program in a specialized center for prenatal diagnosis. Eur J Obstet Gynecol Reprod Biol. 2015;186:112–3.CrossRefPubMed
9.
go back to reference Harris EF, Smith RN. Accounting for measurement error: a critical but often overlooked process. Arch Oral Biol. 2009;54(Suppl 1):S107–17.CrossRefPubMed Harris EF, Smith RN. Accounting for measurement error: a critical but often overlooked process. Arch Oral Biol. 2009;54(Suppl 1):S107–17.CrossRefPubMed
10.
go back to reference Perini TA, Oliveira GL, Ornellas JD, Oliveira FP. Technical error of measurement in anthropometry. Rev Bras Med Esporte. 2005;11(1):81–5.CrossRef Perini TA, Oliveira GL, Ornellas JD, Oliveira FP. Technical error of measurement in anthropometry. Rev Bras Med Esporte. 2005;11(1):81–5.CrossRef
11.
go back to reference Pexsters A, Luts J, Van Schoubroeck D, Bottomley C, Van Calster B, Van Huffel S, Abdallah Y, D'Hooghe T, Lees C, Timmerman D, et al. Clinical implications of intra- and interobserver reproducibility of transvaginal sonographic measurement of gestational sac and crown-rump length at 6-9 weeks' gestation. Ultrasound Obstet Gynecol. 2011;38(5):510–5.CrossRefPubMed Pexsters A, Luts J, Van Schoubroeck D, Bottomley C, Van Calster B, Van Huffel S, Abdallah Y, D'Hooghe T, Lees C, Timmerman D, et al. Clinical implications of intra- and interobserver reproducibility of transvaginal sonographic measurement of gestational sac and crown-rump length at 6-9 weeks' gestation. Ultrasound Obstet Gynecol. 2011;38(5):510–5.CrossRefPubMed
12.
go back to reference Bickhaus J, Perry E, Schust DJ. Re-examining sonographic cut-off values for diagnosing early pregnancy loss. Gynecol Obstet (Sunnyvale, Calif). 2013;3(1):141. Bickhaus J, Perry E, Schust DJ. Re-examining sonographic cut-off values for diagnosing early pregnancy loss. Gynecol Obstet (Sunnyvale, Calif). 2013;3(1):141.
13.
go back to reference Lubinga SJ, Levine GA, Jenny AM, Ngonzi J, Mukasa-Kivunike P, Stergachis A, Babigumira JB. Health-related quality of life and social support among women treated for abortion complications in western Uganda. Health Qual Life Outcomes. 2013;11:118.CrossRefPubMedPubMedCentral Lubinga SJ, Levine GA, Jenny AM, Ngonzi J, Mukasa-Kivunike P, Stergachis A, Babigumira JB. Health-related quality of life and social support among women treated for abortion complications in western Uganda. Health Qual Life Outcomes. 2013;11:118.CrossRefPubMedPubMedCentral
14.
go back to reference Padula F, Gulino FA, Capriglione S, Giorlandino M, Cignini P, Mastrandrea ML, D'Emidio L, Giorlandino C. What is the rate of incomplete fetal anatomic surveys during a second-trimester scan? J Ultrasound Med. 2015;34(12):2187–91.CrossRefPubMed Padula F, Gulino FA, Capriglione S, Giorlandino M, Cignini P, Mastrandrea ML, D'Emidio L, Giorlandino C. What is the rate of incomplete fetal anatomic surveys during a second-trimester scan? J Ultrasound Med. 2015;34(12):2187–91.CrossRefPubMed
15.
go back to reference MoH: Health sector development plan 2015-16_2019–20. In. Ministry of Health, Uganda; 2015. MoH: Health sector development plan 2015-16_2019–20. In. Ministry of Health, Uganda; 2015.
16.
go back to reference WHO: Medical devices: managing the mismatch: an outcome of the priority medical devices project: World Health Organization; 2010. WHO: Medical devices: managing the mismatch: an outcome of the priority medical devices project: World Health Organization; 2010.
17.
go back to reference Kawooya MG. Training for rural radiology and imaging in sub-saharan Africa: addressing the mismatch between services and population. J Clin Imaging Sci. 2012;2:37.CrossRefPubMedPubMedCentral Kawooya MG. Training for rural radiology and imaging in sub-saharan Africa: addressing the mismatch between services and population. J Clin Imaging Sci. 2012;2:37.CrossRefPubMedPubMedCentral
18.
go back to reference Maru DS-R, Schwarz R, Andrews J, Basu S, Sharma A, Moore C. Turning a blind eye: the mobilization of radiology services in resource-poor regions. Glob Health. 2010;6(1):1.CrossRef Maru DS-R, Schwarz R, Andrews J, Basu S, Sharma A, Moore C. Turning a blind eye: the mobilization of radiology services in resource-poor regions. Glob Health. 2010;6(1):1.CrossRef
19.
go back to reference Sarris I, Ioannou C, Chamberlain P, Ohuma E, Roseman F, Hoch L, Altman DG, Papageorghiou AT. Intra- and interobserver variability in fetal ultrasound measurements. Ultrasound Obstet Gynecol. 2012;39(3):266–73.CrossRefPubMed Sarris I, Ioannou C, Chamberlain P, Ohuma E, Roseman F, Hoch L, Altman DG, Papageorghiou AT. Intra- and interobserver variability in fetal ultrasound measurements. Ultrasound Obstet Gynecol. 2012;39(3):266–73.CrossRefPubMed
20.
go back to reference MoH: Standards on Diagnostic Imaging and Therapeutic Radiology in Uganda In.: Ministry of Health 2012: 146. MoH: Standards on Diagnostic Imaging and Therapeutic Radiology in Uganda In.: Ministry of Health 2012: 146.
21.
go back to reference Souka AP, Pilalis A, Papastefanou I, Salamalekis G, Kassanos D. Reproducibility study of crown-rump length and biparietal diameter measurements in the first trimester. Prenat Diagn. 2012;32(12):1158–65.CrossRefPubMed Souka AP, Pilalis A, Papastefanou I, Salamalekis G, Kassanos D. Reproducibility study of crown-rump length and biparietal diameter measurements in the first trimester. Prenat Diagn. 2012;32(12):1158–65.CrossRefPubMed
22.
go back to reference Lane BF, Wong-You-Cheong JJ, Javitt MC, Glanc P, Brown DL, Dubinsky T, Harisinghani MG, Harris RD, Khati NJ, Mitchell DG, et al. ACR appropriateness criteria® first trimester bleeding. Ultrasound Quarterly. 2013;29(2):91–6.CrossRefPubMed Lane BF, Wong-You-Cheong JJ, Javitt MC, Glanc P, Brown DL, Dubinsky T, Harisinghani MG, Harris RD, Khati NJ, Mitchell DG, et al. ACR appropriateness criteria® first trimester bleeding. Ultrasound Quarterly. 2013;29(2):91–6.CrossRefPubMed
23.
go back to reference RCOG: Clinical practice guideline; management of early pregnancy miscarriage. In: Royal College of Obstetricians and Gynaecologists. Edited by Farah Nadine, Nadine Andrea Nugent, Anglim M; 2014. RCOG: Clinical practice guideline; management of early pregnancy miscarriage. In: Royal College of Obstetricians and Gynaecologists. Edited by Farah Nadine, Nadine Andrea Nugent, Anglim M; 2014.
24.
go back to reference McAlinden C, Khadka J, Pesudovs K. Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology. Ophthalmic Physiol Opt. 2011;31(4):330–8.CrossRefPubMed McAlinden C, Khadka J, Pesudovs K. Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology. Ophthalmic Physiol Opt. 2011;31(4):330–8.CrossRefPubMed
25.
go back to reference Weir JP: Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res 2005, 19(1):231–240. Weir JP: Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res 2005, 19(1):231–240.
26.
go back to reference Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979;86(2):420.CrossRefPubMed Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979;86(2):420.CrossRefPubMed
27.
go back to reference Bland JM, Altman DG. Applying the right statistics: analyses of measurement studies. Ultrasound Obstet Gynecol. 2003;22(1):85–93.CrossRefPubMed Bland JM, Altman DG. Applying the right statistics: analyses of measurement studies. Ultrasound Obstet Gynecol. 2003;22(1):85–93.CrossRefPubMed
28.
go back to reference Bland JM, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327(8476):307–10.CrossRef Bland JM, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327(8476):307–10.CrossRef
29.
go back to reference Verburg BO, Mulder PG, Hofman A, Jaddoe VW, Witteman JC, Steegers EA. Intra- and interobserver reproducibility study of early fetal growth parameters. Prenat Diagn. 2008;28(4):323–31.CrossRefPubMed Verburg BO, Mulder PG, Hofman A, Jaddoe VW, Witteman JC, Steegers EA. Intra- and interobserver reproducibility study of early fetal growth parameters. Prenat Diagn. 2008;28(4):323–31.CrossRefPubMed
30.
go back to reference Verwoerd-Dikkeboom CM, Koning AH, Hop WC, Rousian M, Van Der Spek PJ, Exalto N, Steegers EA. Reliability of three-dimensional sonographic measurements in early pregnancy using virtual reality. Ultrasound Obstet Gynecol. 2008;32(7):910–6.CrossRefPubMed Verwoerd-Dikkeboom CM, Koning AH, Hop WC, Rousian M, Van Der Spek PJ, Exalto N, Steegers EA. Reliability of three-dimensional sonographic measurements in early pregnancy using virtual reality. Ultrasound Obstet Gynecol. 2008;32(7):910–6.CrossRefPubMed
31.
go back to reference Kouchi M, Mochimaru M, Tsuzuki K, Yokoi T. Interobserver errors in anthropometry. J Hum Ergol. 1999;28(1/2):15–24. Kouchi M, Mochimaru M, Tsuzuki K, Yokoi T. Interobserver errors in anthropometry. J Hum Ergol. 1999;28(1/2):15–24.
32.
go back to reference Preisler J, Kopeika J, Ismail L, Vathanan V, Farren J, Abdallah Y, Battacharjee P, Van Holsbeke C, Bottomley C, Gould D. Defining safe criteria to diagnose miscarriage: prospective observational multicentre study. BMJ. 2015;351:h4579.CrossRefPubMedPubMedCentral Preisler J, Kopeika J, Ismail L, Vathanan V, Farren J, Abdallah Y, Battacharjee P, Van Holsbeke C, Bottomley C, Gould D. Defining safe criteria to diagnose miscarriage: prospective observational multicentre study. BMJ. 2015;351:h4579.CrossRefPubMedPubMedCentral
Metadata
Title
Measurement error of mean sac diameter and crown-rump length among pregnant women at Mulago hospital, Uganda
Authors
Sam Ali
Rosemary Kusaba Byanyima
Sam Ononge
Jerry Ictho
Jean Nyamwiza
Emmanuel Lako Ernesto Loro
John Mukisa
Angella Musewa
Annet Nalutaaya
Ronald Ssenyonga
Ismael Kawooya
Benjamin Temper
Achilles Katamba
Joan Kalyango
Charles Karamagi
Publication date
01-12-2018
Publisher
BioMed Central
Published in
BMC Pregnancy and Childbirth / Issue 1/2018
Electronic ISSN: 1471-2393
DOI
https://doi.org/10.1186/s12884-018-1769-2

Other articles of this Issue 1/2018

BMC Pregnancy and Childbirth 1/2018 Go to the issue