Magnetic resonance of the heart in a muscular dystrophy patient with an MR conditional ICD: Assessment of safety, diagnostic value and technical limitations

Authors: Anca Florian, Anna Ludwig, Sabine Rösch, Udo Sechtem, Ali Yilmaz

Published in: Journal of Cardiovascular Magnetic Resonance | Issue 1/2013

Login to get access

Abstract

Cardiovascular magnetic resonance (CMR) studies in patients with pacemakers or implantable cardioverter/defibrillators (ICD) are increasingly required in daily clinical practice. Therefore, in the last years the manufacturers developed not only MR-conditional pacemakers, but also MR-conditional ICDs. However, the clinical experience regarding the feasibility and limitations of MR studies of the heart in patients with ICDs is still limited. In particular, there are hardly any CMR studies in the same patients performed prior to and post ICD implantation allowing a one-to-one comparison of the obtained CMR images. This is the first presentation of a CMR study in a patient with the world’s first and so far only MR-conditional ICD. In our case, a major problem related to the presence of the MR conditional ICD was an image artifact caused by the device’s generator which hampered the visualization of the midventricular and apical anterior and antero-lateral segments in all sequences performed. Considering previous studies, right chest implantation of the ICD could probably have helped in this setting and may be preferred in future ICD implantations. Our case report nicely illustrates the real clinical need for specially designed implantable devices that ensure safe and high-quality imaging in patients in whom serial CMR is required.
Appendix
Available only for authorised users
Literature
1.
go back to reference Verhaert D, Richards K, Rafael-Fortney JA, Raman SV: Cardiac involvement in patients with muscular dystrophies: magnetic resonance imaging phenotype and genotypic considerations. Circ Cardiovasc Imaging. 2011, 4 (1): 67-76. 10.1161/CIRCIMAGING.110.960740.PubMedCentralCrossRefPubMed Verhaert D, Richards K, Rafael-Fortney JA, Raman SV: Cardiac involvement in patients with muscular dystrophies: magnetic resonance imaging phenotype and genotypic considerations. Circ Cardiovasc Imaging. 2011, 4 (1): 67-76. 10.1161/CIRCIMAGING.110.960740.PubMedCentralCrossRefPubMed
2.
go back to reference Shinbane JS, Colletti PM, Shellock FG: Magnetic resonance imaging in patients with cardiac pacemakers: era of "MR Conditional" designs. J Cardiovasc Magn Reson. 2011, 13: 63-10.1186/1532-429X-13-63.PubMedCentralCrossRefPubMed Shinbane JS, Colletti PM, Shellock FG: Magnetic resonance imaging in patients with cardiac pacemakers: era of "MR Conditional" designs. J Cardiovasc Magn Reson. 2011, 13: 63-10.1186/1532-429X-13-63.PubMedCentralCrossRefPubMed
3.
go back to reference Nazarian S, Hansford R, Roguin A, Goldsher D, Zviman MM, Lardo AC, Caffo BS, Frick KD, Kraut MA, Kamel IR, Calkins H, Berger RD, Bluemke DA, Halperin HR: A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices. Ann Intern Med. 2011, 155 (7): 415-424. 10.7326/0003-4819-155-7-201110040-00004.PubMedCentralCrossRefPubMed Nazarian S, Hansford R, Roguin A, Goldsher D, Zviman MM, Lardo AC, Caffo BS, Frick KD, Kraut MA, Kamel IR, Calkins H, Berger RD, Bluemke DA, Halperin HR: A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices. Ann Intern Med. 2011, 155 (7): 415-424. 10.7326/0003-4819-155-7-201110040-00004.PubMedCentralCrossRefPubMed
4.
go back to reference Junttila MJ, Fishman JE, Lopera GA, Pattany PM, Velazquez DL, Williams AR, Trachtenberg BH, Sanina C, Mather J, Hare JM: Safety of serial MRI in patients with implantable cardioverter defibrillators. Heart. 2011, 97 (22): 1852-1856. 10.1136/heartjnl-2011-300153.PubMedCentralCrossRefPubMed Junttila MJ, Fishman JE, Lopera GA, Pattany PM, Velazquez DL, Williams AR, Trachtenberg BH, Sanina C, Mather J, Hare JM: Safety of serial MRI in patients with implantable cardioverter defibrillators. Heart. 2011, 97 (22): 1852-1856. 10.1136/heartjnl-2011-300153.PubMedCentralCrossRefPubMed
5.
go back to reference Quarta G, Holdright DR, Plant GT, Harkness A, Hausenloy D, Hyare H, Moon JC: Cardiovascular magnetic resonance in cardiac sarcoidosis with MR conditional pacemaker in situ. J Cardiovasc Magn Reson. 2011, 13: 26-10.1186/1532-429X-13-26.PubMedCentralCrossRefPubMed Quarta G, Holdright DR, Plant GT, Harkness A, Hausenloy D, Hyare H, Moon JC: Cardiovascular magnetic resonance in cardiac sarcoidosis with MR conditional pacemaker in situ. J Cardiovasc Magn Reson. 2011, 13: 26-10.1186/1532-429X-13-26.PubMedCentralCrossRefPubMed
6.
go back to reference Irnich W, Bartsch C, Vock J: Investigation of pacemaker position, lead configuration, and sensitivity setting in pacemakers of 579 deceased patients. Europace. 2011, 13 (1): 96-101. 10.1093/europace/euq405.CrossRefPubMed Irnich W, Bartsch C, Vock J: Investigation of pacemaker position, lead configuration, and sensitivity setting in pacemakers of 579 deceased patients. Europace. 2011, 13 (1): 96-101. 10.1093/europace/euq405.CrossRefPubMed
Metadata
Title
Magnetic resonance of the heart in a muscular dystrophy patient with an MR conditional ICD: Assessment of safety, diagnostic value and technical limitations
Authors
Anca Florian
Anna Ludwig
Sabine Rösch
Udo Sechtem
Ali Yilmaz
Publication date
01-12-2013
Publisher
BioMed Central
Published in
Journal of Cardiovascular Magnetic Resonance / Issue 1/2013
Electronic ISSN: 1532-429X
DOI
https://doi.org/10.1186/1532-429X-15-49

Other articles of this Issue 1/2013

Journal of Cardiovascular Magnetic Resonance 1/2013 Go to the issue