Assessment of magnetic field interactions and heating for cerebral aneurysm flow diverters during 7T MRI

Authors: Suguru Igarashi, Kohei Chida, Tsuyoshi Matsuda, Kota Takeda, Susumu Hayakawa, Shota Tsutsui, Shunrou Fujiwara, Yoshitaka Kubo, Makoto Sasaki, Kuniaki Ogasawara

Published in: Neuroradiology | Issue 12/2023

Login to get access

Abstract

Flow diverters (FDs) are utilized for a wide range of aneurysms, but show safety issues such as adverse interactions with static magnetic fields (displacement force and torque) and radiofrequency-induced heating during magnetic resonance imaging (MRI). The present study aimed to assess these adverse interactions in a 7-tesla (7T) static magnetic field and radiofrequency-induced heating during a 7T MRI for two types of FD. Displacement force and magnetically induced torque were assessed using the deflection angle method and low friction surface method, respectively. To assess heating, each FD was set in a phantom filled with gelled-saline mixed with polyacrylic acid and underwent a 7T MRI using a three-dimensional fast spin echo method. Displacement force and magnetically induced torque in the 7T static magnetic field were undetectable, and radiofrequency-induced heating during 7T MRI remained ≤ 0.6 °C for both types of FD, suggesting that magnetic field interactions and heating on FDs during a 7T MRI are acceptable from a safety perspective.
Appendix
Available only for authorised users
Literature
4.
go back to reference American Society for Testing and Materials (ASTM) International: F2182-19e2. Standard test method for measurement of radio frequency induced heating on or near passive implants during magnetic resonance imaging. https://www.astm.org/f2182-19e02.html. Accessed 30 Apr 2022 American Society for Testing and Materials (ASTM) International: F2182-19e2. Standard test method for measurement of radio frequency induced heating on or near passive implants during magnetic resonance imaging. https://​www.​astm.​org/​f2182-19e02.​html. Accessed 30 Apr 2022
Metadata
Title
Assessment of magnetic field interactions and heating for cerebral aneurysm flow diverters during 7T MRI
Authors
Suguru Igarashi
Kohei Chida
Tsuyoshi Matsuda
Kota Takeda
Susumu Hayakawa
Shota Tsutsui
Shunrou Fujiwara
Yoshitaka Kubo
Makoto Sasaki
Kuniaki Ogasawara
Publication date
13-09-2023
Publisher
Springer Berlin Heidelberg
Published in
Neuroradiology / Issue 12/2023
Print ISSN: 0028-3940
Electronic ISSN: 1432-1920
DOI
https://doi.org/10.1007/s00234-023-03208-8

Other articles of this Issue 12/2023

Neuroradiology 12/2023 Go to the issue