01-06-2020 | Magnetic Resonance Imaging | Research Article

A hybrid (iron–fat–water) phantom for liver iron overload quantification in the presence of contaminating fat using magnetic resonance imaging

Authors: Nazanin Mobini, Malakeh Malekzadeh, Hamidreza Haghighatkhah, Hamidreza Saligheh Rad

Published in: Magnetic Resonance Materials in Physics, Biology and Medicine | Issue 3/2020

Login to get access

Abstract

Objective

Assessment of iron content in the liver is crucial for diagnosis/treatment of iron-overload diseases. Nonetheless, T2*-based methods become challenging when fat and iron are simultaneously present. This study proposes a phantom design concomitantly containing various concentrations of iron and fat suitable for devising accurate simultaneous T2* and fat quantification technique.

Materials and methods

A 46-vial iron–fat–water phantom with various iron concentrations covering clinically relevant T2* relaxation time values, from healthy to severely overloaded liver and wide fat percentages ranges from 0 to 100% was prepared. The phantom was constructed using insoluble iron (II, III) oxide powder containing microscale particles. T2*-weighted imaging using multi-gradient-echo (mGRE) sequence, and chemical shift imaging spin-echo (CSI-SE) Magnetic Resonance Spectroscopy (MRS) data were considered for the analysis. T2* relaxation times and fat fractions were extracted from the MR signals to explore the effects of fat and iron overload.

Results

Size distribution of iron oxide particles for Magnetite fits with a lognormal function with a mean size of about 1.17 µm. Comparison of FF color maps, estimated from bi- and mono-exponential model indicated that single-T2* fitting model resulted in lower NRMSD. Therefore, T2* values from the mono-exponential signal equation were used and expressed the relationship between relaxation time value across all iron (Fe) and fat concentration as \({\text{Fe}} = - 28.02 + \frac{302.84}{{T2^{*} }} - 0.045\,{\text{FF}}\), with R-squared = 0.89.

Discussion

The proposed phantom design with microsphere iron particles closely simulated the single-T2* behavior of fatty iron-overloaded liver in vivo.
Literature
1.
go back to reference Gupta CP (2014) Role of iron (Fe) in body. IOSR J Appl Chem (IOSR-JAC) 7:38–46 Gupta CP (2014) Role of iron (Fe) in body. IOSR J Appl Chem (IOSR-JAC) 7:38–46
2.
go back to reference Andrews NC (1999) Disorders of iron metabolism. N Engl J Med 341(26):1986–1995PubMed Andrews NC (1999) Disorders of iron metabolism. N Engl J Med 341(26):1986–1995PubMed
3.
go back to reference Brittenham GM, Badman DG (2003) Noninvasive measurement of iron: report of an NIDDK workshop. Blood 101(1):15–19PubMed Brittenham GM, Badman DG (2003) Noninvasive measurement of iron: report of an NIDDK workshop. Blood 101(1):15–19PubMed
4.
go back to reference Batts KP (2007) Iron overload syndromes and the liver. Mod Pathol 20(1s):S31PubMed Batts KP (2007) Iron overload syndromes and the liver. Mod Pathol 20(1s):S31PubMed
5.
go back to reference Yokoo T, Yuan Q, Sénégas J, Wiethoff AJ, Pedrosa I (2015) Quantitative R2* MRI of the liver with Rician noise models for evaluation of hepatic iron overload: Simulation, phantom, and early clinical experience. J Magn Reson Imaging 42(6):1544–1559PubMed Yokoo T, Yuan Q, Sénégas J, Wiethoff AJ, Pedrosa I (2015) Quantitative R2* MRI of the liver with Rician noise models for evaluation of hepatic iron overload: Simulation, phantom, and early clinical experience. J Magn Reson Imaging 42(6):1544–1559PubMed
6.
go back to reference Horng DE, Hernando D, Reeder SB (2017) Quantification of liver fat in the presence of iron overload. J Magn Reson Imaging 45(2):428–439PubMed Horng DE, Hernando D, Reeder SB (2017) Quantification of liver fat in the presence of iron overload. J Magn Reson Imaging 45(2):428–439PubMed
7.
go back to reference Yokoo T, Browning JD (2014) Fat and iron quantification in the liver: past, present, and future. Top Magn Reson Imaging 23(2):73–94PubMed Yokoo T, Browning JD (2014) Fat and iron quantification in the liver: past, present, and future. Top Magn Reson Imaging 23(2):73–94PubMed
8.
go back to reference Kühn J-P, Meffert P, Heske C, Kromrey M-L, Schmidt CO, Mensel B, Völzke H, Lerch MM, Hernando D, Mayerle J (2017) Prevalence of fatty liver disease and hepatic iron overload in a Northeastern German population by using quantitative MR imaging. Radiology 28(3):706–716 Kühn J-P, Meffert P, Heske C, Kromrey M-L, Schmidt CO, Mensel B, Völzke H, Lerch MM, Hernando D, Mayerle J (2017) Prevalence of fatty liver disease and hepatic iron overload in a Northeastern German population by using quantitative MR imaging. Radiology 28(3):706–716
10.
go back to reference Lidbury JA (2017) Getting the most out of liver biopsy. Vet Clin Small Animal Pract 47(3):569–583 Lidbury JA (2017) Getting the most out of liver biopsy. Vet Clin Small Animal Pract 47(3):569–583
11.
go back to reference Ratziu V, Charlotte F, Heurtier A, Gombert S, Giral P, Bruckert E, Grimaldi A, Capron F, Poynard T, Group LS (2005) Sampling variability of liver biopsy in nonalcoholic fatty liver disease. Gastroenterology 128(7):1898–1906PubMed Ratziu V, Charlotte F, Heurtier A, Gombert S, Giral P, Bruckert E, Grimaldi A, Capron F, Poynard T, Group LS (2005) Sampling variability of liver biopsy in nonalcoholic fatty liver disease. Gastroenterology 128(7):1898–1906PubMed
12.
go back to reference Li TQ, Aisen AM, Hindmarsh T (2004) Assessment of hepatic iron content using magnetic resonance imaging. Acta Radiol 45(2):119–129PubMed Li TQ, Aisen AM, Hindmarsh T (2004) Assessment of hepatic iron content using magnetic resonance imaging. Acta Radiol 45(2):119–129PubMed
13.
go back to reference Sirlin CB, Reeder SB (2010) Magnetic resonance imaging quantification of liver iron. Magn Reson Imaging Clin N Am 18(3):359–381PubMedPubMedCentral Sirlin CB, Reeder SB (2010) Magnetic resonance imaging quantification of liver iron. Magn Reson Imaging Clin N Am 18(3):359–381PubMedPubMedCentral
14.
go back to reference Bowen CV, Zhang X, Saab G, Gareau PJ, Rutt BK (2002) Application of the static dephasing regime theory to superparamagnetic iron-oxide loaded cells. Magn Reson Med 48(1):52–61PubMed Bowen CV, Zhang X, Saab G, Gareau PJ, Rutt BK (2002) Application of the static dephasing regime theory to superparamagnetic iron-oxide loaded cells. Magn Reson Med 48(1):52–61PubMed
15.
go back to reference Yu H, Shimakawa A, McKenzie CA, Brodsky E, Brittain JH, Reeder SB (2008) Multiecho water-fat separation and simultaneous R 2* estimation with multifrequency fat spectrum modeling. Magn Reson Med 60(5):1122–1134PubMedPubMedCentral Yu H, Shimakawa A, McKenzie CA, Brodsky E, Brittain JH, Reeder SB (2008) Multiecho water-fat separation and simultaneous R 2* estimation with multifrequency fat spectrum modeling. Magn Reson Med 60(5):1122–1134PubMedPubMedCentral
16.
go back to reference Clark PR, Chua-anusorn W, St. Pierre TG (2003) Bi-exponential proton transverse relaxation rate (R 2) image analysis using RF field intensity-weighted spin density projection: potential for R 2 measurement of iron-loaded liver. Magn Reson Imaging 21(5):519–530PubMed Clark PR, Chua-anusorn W, St. Pierre TG (2003) Bi-exponential proton transverse relaxation rate (R 2) image analysis using RF field intensity-weighted spin density projection: potential for R 2 measurement of iron-loaded liver. Magn Reson Imaging 21(5):519–530PubMed
17.
go back to reference Bernard CP, Liney GP, Manton DJ, Turnbull LW, Langton CM (2008) Comparison of fat quantification methods: a phantom study at 30T. J Magn Reson imaging 27(1):192–197PubMed Bernard CP, Liney GP, Manton DJ, Turnbull LW, Langton CM (2008) Comparison of fat quantification methods: a phantom study at 30T. J Magn Reson imaging 27(1):192–197PubMed
18.
go back to reference Hines CD, Yu H, Shimakawa A, McKenzie CA, Brittain JH, Reeder SB (2009) T1 independent, T2* corrected MRI with accurate spectral modeling for quantification of fat: validation in a fat-water-SPIO phantom. J Magn Reson Imaging JMRI 30(5):1215–1222PubMed Hines CD, Yu H, Shimakawa A, McKenzie CA, Brittain JH, Reeder SB (2009) T1 independent, T2* corrected MRI with accurate spectral modeling for quantification of fat: validation in a fat-water-SPIO phantom. J Magn Reson Imaging JMRI 30(5):1215–1222PubMed
19.
go back to reference Reeder SB, Hernando D, Sharma S (2016) Phantom for iron and fat quantification magnetic resonance imaging. United States Patent Reeder SB, Hernando D, Sharma S (2016) Phantom for iron and fat quantification magnetic resonance imaging. United States Patent
20.
go back to reference McMullan D (1995) Scanning electron microscopy 1928–1965. Scanning 17(3):175–185 McMullan D (1995) Scanning electron microscopy 1928–1965. Scanning 17(3):175–185
21.
go back to reference Reinoso RF, Telfe BA, Rowland M (1997) Tissue water content in rats measured by desiccation. J Pharmacol Toxicol Methods 38(2):87–92PubMed Reinoso RF, Telfe BA, Rowland M (1997) Tissue water content in rats measured by desiccation. J Pharmacol Toxicol Methods 38(2):87–92PubMed
22.
go back to reference Hernando D, Cook RJ, Diamond C, Reeder SB (2013) Magnetic susceptibility as a B0 field strength independent MRI biomarker of liver iron overload. Magn Reson Med 70(3):648–656PubMedPubMedCentral Hernando D, Cook RJ, Diamond C, Reeder SB (2013) Magnetic susceptibility as a B0 field strength independent MRI biomarker of liver iron overload. Magn Reson Med 70(3):648–656PubMedPubMedCentral
23.
go back to reference Fukuzawa K, Hayashi T, Takahashi J, Yoshihara C, Tano M, Ji Kotoku, Saitoh S (2017) Evaluation of six-point modified dixon and magnetic resonance spectroscopy for fat quantification: a fat-water-iron phantom study. Radiol Phys Technol 10(3):349–358PubMed Fukuzawa K, Hayashi T, Takahashi J, Yoshihara C, Tano M, Ji Kotoku, Saitoh S (2017) Evaluation of six-point modified dixon and magnetic resonance spectroscopy for fat quantification: a fat-water-iron phantom study. Radiol Phys Technol 10(3):349–358PubMed
24.
go back to reference Bydder M, Hamilton G, de Rochefort L, Desai A, Heba ER, Loomba R, Schwimmer JB, Szeverenyi NM, Sirlin CB (2018) Sources of systematic error in proton density fat fraction (PDFF) quantification in the liver evaluated from magnitude images with different numbers of echoes. NMR Biomed 31(1):e3843 Bydder M, Hamilton G, de Rochefort L, Desai A, Heba ER, Loomba R, Schwimmer JB, Szeverenyi NM, Sirlin CB (2018) Sources of systematic error in proton density fat fraction (PDFF) quantification in the liver evaluated from magnitude images with different numbers of echoes. NMR Biomed 31(1):e3843
25.
go back to reference Storey P, Thompson AA, Carqueville CL, Wood JC, de Freitas RA, Rigsby CK (2007) R2* imaging of transfusional iron burden at 3 T and comparison with 1.5 T. J Magn Reson Imaging 25(3):540–547PubMedPubMedCentral Storey P, Thompson AA, Carqueville CL, Wood JC, de Freitas RA, Rigsby CK (2007) R2* imaging of transfusional iron burden at 3 T and comparison with 1.5 T. J Magn Reson Imaging 25(3):540–547PubMedPubMedCentral
26.
go back to reference Meloni A, Positano V, Keilberg P, De Marchi D, Pepe P, Zuccarelli A, Campisi S, Romeo MA, Casini T, Bitti PP, Gerardi C, Lai ME, Piraino B, Giuffrida G, Secchi G, Midiri M, Lombardi M, Pepe A (2012) Feasibility, reproducibility, and reliability for the T* 2 iron evaluation at 3 T in comparison with 1.5 T. Magn Reson Med 68(2):543–551PubMed Meloni A, Positano V, Keilberg P, De Marchi D, Pepe P, Zuccarelli A, Campisi S, Romeo MA, Casini T, Bitti PP, Gerardi C, Lai ME, Piraino B, Giuffrida G, Secchi G, Midiri M, Lombardi M, Pepe A (2012) Feasibility, reproducibility, and reliability for the T* 2 iron evaluation at 3 T in comparison with 1.5 T. Magn Reson Med 68(2):543–551PubMed
27.
go back to reference Yamamura J, Keller S, Grosse R, Schoennagel B, Nielsen P, Wang ZJ, Graessner J, Kooijman H, Adam G, Fischer R (2016) Iron measurements by quantitative MRI-R2* at 3.0 and 1.5 T. In: ISMRM Yamamura J, Keller S, Grosse R, Schoennagel B, Nielsen P, Wang ZJ, Graessner J, Kooijman H, Adam G, Fischer R (2016) Iron measurements by quantitative MRI-R2* at 3.0 and 1.5 T. In: ISMRM
Metadata
Title
A hybrid (iron–fat–water) phantom for liver iron overload quantification in the presence of contaminating fat using magnetic resonance imaging
Authors
Nazanin Mobini
Malakeh Malekzadeh
Hamidreza Haghighatkhah
Hamidreza Saligheh Rad
Publication date
01-06-2020
Publisher
Springer International Publishing
Published in
Magnetic Resonance Materials in Physics, Biology and Medicine / Issue 3/2020
Print ISSN: 0968-5243
Electronic ISSN: 1352-8661
DOI
https://doi.org/10.1007/s10334-019-00795-7

Other articles of this Issue 3/2020

Magnetic Resonance Materials in Physics, Biology and Medicine 3/2020 Go to the issue