Published in: European Radiology 6/2015

01-06-2015 | Pediatric

Low-dose biplanar radiography can be used in children and adolescents to accurately assess femoral and tibial torsion and greatly reduce irradiation

Authors: Olivier Meyrignac, Ramiro Moreno, Christiane Baunin, Julie Vial, Franck Accadbled, Agnes Sommet, Jérome Sales de Gauzy, Nicolas Sans

Published in: European Radiology | Issue 6/2015

Login to get access

Abstract

Purpose

To evaluate in children the agreement between femoral and tibial torsion measurements obtained with low-dose biplanar radiography (LDBR) and CT, and to study dose reduction ratio between these two techniques both in vitro and in vivo.

Materials and methods

Thirty children with lower limb torsion abnormalities were included in a prospective study. Biplanar radiographs and CTs were performed for measurements of lower limb torsion on each patient. Values were compared using Bland-Altman plots. Interreader and intrareader agreements were evaluated by intraclass correlation coefficients. Comparative dosimetric study was performed using an ionization chamber in a tissue-equivalent phantom, and with thermoluminescent dosimeters in 5 patients.

Results

Average differences between CT and LDBR measurements were –0.1° ±1.1 for femoral torsion and –0.7° ±1.4 for tibial torsion. Interreader agreement for LDBR measurements was very good for both femoral torsion (FT) (0.81) and tibial torsion (TT) (0.87). Intrareader agreement was excellent for FT (0.97) and TT (0.89). The ratio between CT scan dose and LDBR dose was 22 in vitro (absorbed dose) and 32 in vivo (skin dose).

Conclusion

Lower limb torsion measurements obtained with LDBR are comparable to CT measurements in children and adolescents, with a considerably reduced radiation dose.

Key points

LDBR and CT lower-limb torsion measurements are comparable in children and adolescents.
LDBR considerably reduced radiation dose necessary for lower-limb torsion measurements.
LDBR can be used for evaluation of lower limb-torsion in orthopaediatric patients.
Literature
1.
go back to reference Masson (2007) Anomalies rotationnelles des membres inférieurs chez l'enfant. EMC Appar Locomoteur 15:1–12 Masson (2007) Anomalies rotationnelles des membres inférieurs chez l'enfant. EMC Appar Locomoteur 15:1–12
2.
go back to reference Kim HY, Lee SK, Lee NK, Choy WS (2012) An anatomical measurement of medial femoral torsion. J Pediatr Orthop B 21:552–557CrossRefPubMed Kim HY, Lee SK, Lee NK, Choy WS (2012) An anatomical measurement of medial femoral torsion. J Pediatr Orthop B 21:552–557CrossRefPubMed
3.
go back to reference Hernandez RJ, Tachdjian MO, Poznanski AK, Dias LS (1981) CT determination of femoral torsion. AJR Am J Roentgenol 137:97–101CrossRefPubMed Hernandez RJ, Tachdjian MO, Poznanski AK, Dias LS (1981) CT determination of femoral torsion. AJR Am J Roentgenol 137:97–101CrossRefPubMed
4.
go back to reference Goutallier D, Van Driessche S, Manicom O et al (2006) Influence of lower-limb torsion on long-term outcomes of tibial valgus osteotomy for medial compartment knee osteoarthritis. J Bone Joint Surg Am 88:2439–2447CrossRefPubMed Goutallier D, Van Driessche S, Manicom O et al (2006) Influence of lower-limb torsion on long-term outcomes of tibial valgus osteotomy for medial compartment knee osteoarthritis. J Bone Joint Surg Am 88:2439–2447CrossRefPubMed
6.
go back to reference Dubousset J, Charpak G, Dorion I et al (2005) Une nouvelle imagerie ostéo-articulaire basse dose en position debout : le système EOS. Radioprotection 40:245–255CrossRef Dubousset J, Charpak G, Dorion I et al (2005) Une nouvelle imagerie ostéo-articulaire basse dose en position debout : le système EOS. Radioprotection 40:245–255CrossRef
7.
go back to reference Kalifa G, Charpak Y, Maccia C et al (1998) Evaluation of a new low-dose digital x-ray device: first dosimetric and clinical results in children. Pediatr Radiol 28:557–561CrossRefPubMed Kalifa G, Charpak Y, Maccia C et al (1998) Evaluation of a new low-dose digital x-ray device: first dosimetric and clinical results in children. Pediatr Radiol 28:557–561CrossRefPubMed
8.
go back to reference Mathews JD, Forsythe AV, Brady Z et al (2013) Cancer risk in 680,000 people exposed to computed tomography scans in childhood or adolescence: data linkage study of 11 million Australians. BMJ 346:f2360CrossRefPubMedCentralPubMed Mathews JD, Forsythe AV, Brady Z et al (2013) Cancer risk in 680,000 people exposed to computed tomography scans in childhood or adolescence: data linkage study of 11 million Australians. BMJ 346:f2360CrossRefPubMedCentralPubMed
9.
go back to reference Buck FM, Guggenberger R, Koch PP, Pfirrmann CWA (2012) Femoral and tibial torsion measurements with 3D models based on low-dose biplanar radiographs in comparison with standard CT measurements. Am J Roentgenol 199:W607–W612CrossRef Buck FM, Guggenberger R, Koch PP, Pfirrmann CWA (2012) Femoral and tibial torsion measurements with 3D models based on low-dose biplanar radiographs in comparison with standard CT measurements. Am J Roentgenol 199:W607–W612CrossRef
10.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310CrossRefPubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310CrossRefPubMed
11.
go back to reference Altman DG, Bland JM (1983) JSTOR: Journal of the Royal Statistical Society. Series D (The Statistician), Vol. 32, No. 3 (Sep., 1983), pp. 307-317. The statistician Altman DG, Bland JM (1983) JSTOR: Journal of the Royal Statistical Society. Series D (The Statistician), Vol. 32, No. 3 (Sep., 1983), pp. 307-317. The statistician
12.
go back to reference Rosskopf AB, Ramseier LE, Sutter R et al (2014) Femoral and tibial torsion measurement in children and adolescents: comparison of 3D models based on low-dose biplanar radiography and low-dose CT. Am J Roentgenol 202:W285–W291CrossRef Rosskopf AB, Ramseier LE, Sutter R et al (2014) Femoral and tibial torsion measurement in children and adolescents: comparison of 3D models based on low-dose biplanar radiography and low-dose CT. Am J Roentgenol 202:W285–W291CrossRef
13.
go back to reference Dubousset J, Charpak G, Skalli W, et al. (2007) [EOS stereo-radiography system: whole-body simultaneous anteroposterior and lateral radiographs with very low radiation dose]. In: Rev Chir Orthop Reparatrice Appar Mot. pp 141–143 Dubousset J, Charpak G, Skalli W, et al. (2007) [EOS stereo-radiography system: whole-body simultaneous anteroposterior and lateral radiographs with very low radiation dose]. In: Rev Chir Orthop Reparatrice Appar Mot. pp 141–143
14.
go back to reference Illés T, Tunyogi-Csapó M, Somoskeöy S (2011) Breakthrough in three-dimensional scoliosis diagnosis: significance of horizontal plane view and vertebra vectors. Eur Spine J 20:135–143CrossRefPubMedCentralPubMed Illés T, Tunyogi-Csapó M, Somoskeöy S (2011) Breakthrough in three-dimensional scoliosis diagnosis: significance of horizontal plane view and vertebra vectors. Eur Spine J 20:135–143CrossRefPubMedCentralPubMed
15.
go back to reference Deschênes S, Charron G, Beaudoin G et al (2010) Diagnostic imaging of spinal deformities: reducing patients radiation dose with a new slot-scanning X-ray imager. Spine 35:989–994CrossRefPubMed Deschênes S, Charron G, Beaudoin G et al (2010) Diagnostic imaging of spinal deformities: reducing patients radiation dose with a new slot-scanning X-ray imager. Spine 35:989–994CrossRefPubMed
16.
17.
18.
go back to reference Sugier A, Lecomte JF, Nenot JC (2007) Les recommandations 2007 de la Commission internationale de protection radiologique. Revue générale nucléaire Sugier A, Lecomte JF, Nenot JC (2007) Les recommandations 2007 de la Commission internationale de protection radiologique. Revue générale nucléaire
20.
go back to reference Delin C, Silvera S, Bassinet C et al (2014) Ionizing radiation doses during lower limb torsion and anteversion measurements by EOS stereoradiography and computed tomography. Eur J Radiol 83:371–377CrossRefPubMed Delin C, Silvera S, Bassinet C et al (2014) Ionizing radiation doses during lower limb torsion and anteversion measurements by EOS stereoradiography and computed tomography. Eur J Radiol 83:371–377CrossRefPubMed
21.
go back to reference Faria R, McKenna C, Wade R et al (2013) The EOS 2D/3D X-ray imaging system: a cost-effectiveness analysis quantifying the health benefits from reduced radiation exposure. Eur J Radiol 82:e342–e349CrossRefPubMed Faria R, McKenna C, Wade R et al (2013) The EOS 2D/3D X-ray imaging system: a cost-effectiveness analysis quantifying the health benefits from reduced radiation exposure. Eur J Radiol 82:e342–e349CrossRefPubMed
22.
go back to reference Staheli LT (1993) Rotational problems in children. J Bone Joint Surg Am 75:939–949 Staheli LT (1993) Rotational problems in children. J Bone Joint Surg Am 75:939–949
Metadata
Title
Low-dose biplanar radiography can be used in children and adolescents to accurately assess femoral and tibial torsion and greatly reduce irradiation
Authors
Olivier Meyrignac
Ramiro Moreno
Christiane Baunin
Julie Vial
Franck Accadbled
Agnes Sommet
Jérome Sales de Gauzy
Nicolas Sans
Publication date
01-06-2015
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 6/2015
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-014-3560-8

Other articles of this Issue 6/2015

European Radiology 6/2015 Go to the issue