Introduction

Mesh-related adhesions are a significant clinical problem following intraperitoneal mesh placement. In this study, we evaluated adhesion formation to three relatively new meshes for intraperitoneal use.

Methods

Three new meshes for intraperitoneal use (Omyra® mesh, Physiomesh®, and Hi-Tex Endo-IP®) were implanted intraperitoneally in rats and compared with a polypropylene control mesh (Parietene®) after 7 or 90 days. Adhesion formation, incorporation (tensile strength), shrinkage, and foreign body reaction were scored.

Results

Hi-Tex Endo-IP and Physiomesh® showed significantly less adhesion formation when compared to Parietene at both time points (p < 0.05). Shrinkage was highest in Omyra mesh after 90 days, which was significantly more compared to Parietene® (p < 0.001). Physiomesh® only showed a significant reduction in craniocaudal mesh length, compared to Parietene and Hi-Tex Endo-IP (p < 0.05). After 90 days, Hi-Tex Endo-IP® showed significantly higher and Physiomesh® significantly lower incorporation strengths compared to all other groups (p < 0.05). Microscopic evaluation revealed massive foreign body reaction to Hi-Tex Endo-IP®, leading to an extensive and thick collagenous scar adherent to the abdominal wall. Fractioning of the Physiomesh® coating over time led to an increase in interfilamentary granuloma formation, leading to scar plate formation, but with only minimal to no abdominal wall adherence. Both Parietene® and Omyra® showed a mild foreign body response.

Conclusion

Although clear distinctions can be made between meshes and some meshes excel, none of the meshes are superior in all aspects required for effective and safe incisional hernia repair.
Literature
1.
go back to reference Kingsnorth A, LeBlanc K (2003) Hernias: inguinal and incisional. Lancet 8(362):1561–1571CrossRef Kingsnorth A, LeBlanc K (2003) Hernias: inguinal and incisional. Lancet 8(362):1561–1571CrossRef
2.
go back to reference Poulose BK, Shelton J, Phillips S, Moore D, Nealon W, Penson D et al (2012) Epidemiology and cost of ventral hernia repair: making the case for hernia research. Hernia 16:179–183PubMedCrossRef Poulose BK, Shelton J, Phillips S, Moore D, Nealon W, Penson D et al (2012) Epidemiology and cost of ventral hernia repair: making the case for hernia research. Hernia 16:179–183PubMedCrossRef
3.
go back to reference Burger JWA, Luijendijk RW, Hop WCJ, Halm JA, Verdaasdonk EGG, Jeekel J (2004) Long-term follow-up of a randomized controlled trial of suture versus mesh repair of incisional hernia. Ann Surg 240:578–583 discussion 583–585PubMedCentralPubMed Burger JWA, Luijendijk RW, Hop WCJ, Halm JA, Verdaasdonk EGG, Jeekel J (2004) Long-term follow-up of a randomized controlled trial of suture versus mesh repair of incisional hernia. Ann Surg 240:578–583 discussion 583–585PubMedCentralPubMed
4.
go back to reference Dumanian GA (2012) Discussion: adhesions and meshes: synthetic versus bioprosthetic. Plast Reconstr Surg 130:214S–215SPubMedCrossRef Dumanian GA (2012) Discussion: adhesions and meshes: synthetic versus bioprosthetic. Plast Reconstr Surg 130:214S–215SPubMedCrossRef
5.
go back to reference Helgstrand F, Rosenberg J, Kehlet H, Jorgensen LN, Bisgaard T (2013) Nationwide prospective study of outcomes after elective incisional hernia repair. J Am Coll Surg 216:217–228PubMedCrossRef Helgstrand F, Rosenberg J, Kehlet H, Jorgensen LN, Bisgaard T (2013) Nationwide prospective study of outcomes after elective incisional hernia repair. J Am Coll Surg 216:217–228PubMedCrossRef
6.
go back to reference Huang C-C, Lien H-H, Huang C-S (2013) Long-term follow-up of laparoscopic incisional and ventral hernia repairs. J Laparoendosc Adv Surg Tech A 23:199–203PubMedCrossRef Huang C-C, Lien H-H, Huang C-S (2013) Long-term follow-up of laparoscopic incisional and ventral hernia repairs. J Laparoendosc Adv Surg Tech A 23:199–203PubMedCrossRef
7.
go back to reference Jenkins ED, Yom V, Melman L, Brunt LM, Eagon JC, Frisella MM et al (2010) Prospective evaluation of adhesion characteristics to intraperitoneal mesh and adhesiolysis-related complications during laparoscopic re-exploration after prior ventral hernia repair. Surg Endosc 24:3002–3007PubMedCrossRef Jenkins ED, Yom V, Melman L, Brunt LM, Eagon JC, Frisella MM et al (2010) Prospective evaluation of adhesion characteristics to intraperitoneal mesh and adhesiolysis-related complications during laparoscopic re-exploration after prior ventral hernia repair. Surg Endosc 24:3002–3007PubMedCrossRef
8.
go back to reference Demir U, Mihmanli M, Coskun H, Dilege E, Kalyoncu A, Altinli E et al (2005) Comparison of prosthetic materials in incisional hernia repair. Surg Today 35:223–227PubMedCrossRef Demir U, Mihmanli M, Coskun H, Dilege E, Kalyoncu A, Altinli E et al (2005) Comparison of prosthetic materials in incisional hernia repair. Surg Today 35:223–227PubMedCrossRef
9.
go back to reference Turza KC, Butler CE (2012) Adhesions and meshes: synthetic versus bioprosthetic. Plast Reconstr Surg 130:206S–213SPubMedCrossRef Turza KC, Butler CE (2012) Adhesions and meshes: synthetic versus bioprosthetic. Plast Reconstr Surg 130:206S–213SPubMedCrossRef
10.
go back to reference Ellis H, Moran BJ, Thompson JN, Parker MC, Wilson MS, Menzies D et al (1999) Adhesion-related hospital readmissions after abdominal and pelvic surgery: a retrospective cohort study. Lancet 1(353):1476–1480CrossRef Ellis H, Moran BJ, Thompson JN, Parker MC, Wilson MS, Menzies D et al (1999) Adhesion-related hospital readmissions after abdominal and pelvic surgery: a retrospective cohort study. Lancet 1(353):1476–1480CrossRef
11.
go back to reference Van Der Krabben AA, Dijkstra FR, Nieuwenhuijzen M, Reijnen MMPJ, Schaapveld M, van Goor H (2000) Morbidity and mortality of inadvertent enterotomy during adhesiotomy. Br J Surg 87:467–471CrossRef Van Der Krabben AA, Dijkstra FR, Nieuwenhuijzen M, Reijnen MMPJ, Schaapveld M, van Goor H (2000) Morbidity and mortality of inadvertent enterotomy during adhesiotomy. Br J Surg 87:467–471CrossRef
12.
go back to reference Schreinemacher MHF, van Barneveld KWY, Dikmans REG, Gijbels MJJ, Greve JWM, Bouvy ND (2013) Coated meshes for hernia repair provide comparable intraperitoneal adhesion prevention. Surg Endosc 27:4202–4209PubMedCrossRef Schreinemacher MHF, van Barneveld KWY, Dikmans REG, Gijbels MJJ, Greve JWM, Bouvy ND (2013) Coated meshes for hernia repair provide comparable intraperitoneal adhesion prevention. Surg Endosc 27:4202–4209PubMedCrossRef
13.
go back to reference Schreinemacher MHF, Emans PJ, Gijbels MJJ, Greve JWM, Beets GL, Bouvy ND (2009) Degradation of mesh coatings and intraperitoneal adhesion formation in an experimental model. Br J Surg 96:305–313PubMedCrossRef Schreinemacher MHF, Emans PJ, Gijbels MJJ, Greve JWM, Beets GL, Bouvy ND (2009) Degradation of mesh coatings and intraperitoneal adhesion formation in an experimental model. Br J Surg 96:305–313PubMedCrossRef
14.
go back to reference Conze J, Rosch R, Klinge U, Weiss C, Anurov M, Titkowa S et al (2004) Polypropylene in the intra-abdominal position: influence of pore size and surface area. Hernia 8:365–372PubMedCrossRef Conze J, Rosch R, Klinge U, Weiss C, Anurov M, Titkowa S et al (2004) Polypropylene in the intra-abdominal position: influence of pore size and surface area. Hernia 8:365–372PubMedCrossRef
15.
go back to reference Novitsky YW, Harrell AG, Cristiano JA, Paton BL, Norton HJ, Peindl RD et al (2007) Comparative evaluation of adhesion formation, strength of ingrowth, and textile properties of prosthetic meshes after long-term intra-abdominal implantation in a rabbit. J Surg Res 140:6–11PubMedCrossRef Novitsky YW, Harrell AG, Cristiano JA, Paton BL, Norton HJ, Peindl RD et al (2007) Comparative evaluation of adhesion formation, strength of ingrowth, and textile properties of prosthetic meshes after long-term intra-abdominal implantation in a rabbit. J Surg Res 140:6–11PubMedCrossRef
16.
go back to reference Sanders DL, Kingsnorth AN (2012) Prosthetic mesh materials used in hernia surgery. Expert Rev Med Devices 9:159–179PubMedCrossRef Sanders DL, Kingsnorth AN (2012) Prosthetic mesh materials used in hernia surgery. Expert Rev Med Devices 9:159–179PubMedCrossRef
19.
go back to reference Greca F, Paula J, Biondo-Simões M, Costa F, Silva A, Time S et al (2001) The influence of differing pore sizes on the biocompatibility of two polypropylene meshes in the repair of abdominal defects. Hernia 5:59–64PubMedCrossRef Greca F, Paula J, Biondo-Simões M, Costa F, Silva A, Time S et al (2001) The influence of differing pore sizes on the biocompatibility of two polypropylene meshes in the repair of abdominal defects. Hernia 5:59–64PubMedCrossRef
20.
go back to reference Cobb WS, Kercher KW, Heniford BT (2005) The argument for lightweight polypropylene mesh in hernia repair. Surg Innov 12:63–69PubMedCrossRef Cobb WS, Kercher KW, Heniford BT (2005) The argument for lightweight polypropylene mesh in hernia repair. Surg Innov 12:63–69PubMedCrossRef
21.
go back to reference Sanders DL, Waydia S (2013) A systematic review of randomised control trials assessing mesh fixation in open inguinal hernia repair. Hernia 18:165–176 Sanders DL, Waydia S (2013) A systematic review of randomised control trials assessing mesh fixation in open inguinal hernia repair. Hernia 18:165–176
22.
go back to reference Hollinsky C, Kolbe T, Walter I, Joachim A, Sandberg S, Koch T et al (2010) Tensile strength and adhesion formation of mesh fixation systems used in laparoscopic incisional hernia repair. Surg Endosc Interv Tech 24:1318–1324CrossRef Hollinsky C, Kolbe T, Walter I, Joachim A, Sandberg S, Koch T et al (2010) Tensile strength and adhesion formation of mesh fixation systems used in laparoscopic incisional hernia repair. Surg Endosc Interv Tech 24:1318–1324CrossRef
23.
go back to reference Beldi G, Wagner M, Bruegger LE, Kurmann A, Candinas D (2011) Mesh shrinkage and pain in laparoscopic ventral hernia repair: a randomized clinical trial comparing suture versus tack mesh fixation. Surg Endosc 25:749–755PubMedCrossRef Beldi G, Wagner M, Bruegger LE, Kurmann A, Candinas D (2011) Mesh shrinkage and pain in laparoscopic ventral hernia repair: a randomized clinical trial comparing suture versus tack mesh fixation. Surg Endosc 25:749–755PubMedCrossRef
24.
go back to reference Zogbi L, Trindade EN, Trindade MRM (2013) Comparative study of shrinkage, inflammatory response and fibroplasia in heavyweight and lightweight meshes. Hernia 17:765–772 Zogbi L, Trindade EN, Trindade MRM (2013) Comparative study of shrinkage, inflammatory response and fibroplasia in heavyweight and lightweight meshes. Hernia 17:765–772
25.
go back to reference Eriksen JR, Gogenur I, Rosenberg J (2007) Choice of mesh for laparoscopic ventral hernia repair. Hernia 11:481–492PubMedCrossRef Eriksen JR, Gogenur I, Rosenberg J (2007) Choice of mesh for laparoscopic ventral hernia repair. Hernia 11:481–492PubMedCrossRef
26.
go back to reference Fortelny RH, Petter-Puchner AH, Glaser KS, Offner F, Benesch T, Rohr M (2010) Adverse effects of polyvinylidene fluoride‐coated polypropylene mesh used for laparoscopic intraperitoneal onlay repair of incisional hernia. Br J Surg 97:1140–1145PubMedCrossRef Fortelny RH, Petter-Puchner AH, Glaser KS, Offner F, Benesch T, Rohr M (2010) Adverse effects of polyvinylidene fluoride‐coated polypropylene mesh used for laparoscopic intraperitoneal onlay repair of incisional hernia. Br J Surg 97:1140–1145PubMedCrossRef
27.
go back to reference Schreinemacher MHF, Henatsch D, van Barneveld K, Bouvy N (2010) The need for standardised animal models and scoring systems in assessing mesh biocompatibility. Hernia 14:335–336PubMedCentralPubMedCrossRef Schreinemacher MHF, Henatsch D, van Barneveld K, Bouvy N (2010) The need for standardised animal models and scoring systems in assessing mesh biocompatibility. Hernia 14:335–336PubMedCentralPubMedCrossRef
28.
go back to reference Dubay DA, Wang X, Adamson B, Kuzon WM Jr, Dennis RG, Franz MG (2005) Progressive fascial wound failure impairs subsequent abdominal wall repairs: a new animal model of incisional hernia formation. Surgery 137:463–471PubMedCrossRef Dubay DA, Wang X, Adamson B, Kuzon WM Jr, Dennis RG, Franz MG (2005) Progressive fascial wound failure impairs subsequent abdominal wall repairs: a new animal model of incisional hernia formation. Surgery 137:463–471PubMedCrossRef
29.
go back to reference Bittner R, Bingener-Casey J, Dietz U, Fabian M, Ferzli GS, Fortelny RH et al (2014) Guidelines for laparoscopic treatment of ventral and incisional abdominal wall hernias (International Endohernia Society (IEHS)—part 1. Surg Endosc 28:2–29PubMedCentralPubMedCrossRef Bittner R, Bingener-Casey J, Dietz U, Fabian M, Ferzli GS, Fortelny RH et al (2014) Guidelines for laparoscopic treatment of ventral and incisional abdominal wall hernias (International Endohernia Society (IEHS)—part 1. Surg Endosc 28:2–29PubMedCentralPubMedCrossRef
Metadata
Title
Long-term evaluation of adhesion formation and foreign body response to three new meshes
Authors
R. R. M. Vogels
K. W. Y. van Barneveld
J. W. A. M. Bosmans
G. Beets
M. J. J. Gijbels
M. H. F. Schreinemacher
N. D. Bouvy
Publication date
01-08-2015
Publisher
Springer US
Published in
Surgical Endoscopy / Issue 8/2015
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-014-3936-5

Other articles of this Issue 8/2015

Surgical Endoscopy 8/2015 Go to the issue